Ball trocar
40 Pages
Preview
Page 1
Aesculap®
Aesculap Endoscopic Technology
Instructions for use/Technical description Ball trocar Gebrauchsanweisung/Technische Beschreibung Kugeltrokar Mode d’emploi/Description technique Trocart à bille Instrucciones de manejo/Descripción técnica Trocar de rótula Istruzioni per l’uso/Descrizione tecnica Trocar a sfera Instruções de utilização/Descrição técnica Trocarte esférico Gebruiksaanwijzing/Technische beschrijving Kogeltrocart Bruksanvisning/Teknisk beskrivning Kultrokar Инструкция по примению/Техническое описание Троакар с шарниром Návod k použití/Technický popis Kulový trokar Instrukcja użytkowania/Opis techniczny Trokar kulkowy Návod na použivanie/Technický opis Guľový trokár Kullanım Kılavuzu/Teknik açiklama Küresel trokar
1
2
3 4
5
7 6 13 12 Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com
8
9 10
Aesculap – a B. Braun company TA-Nr. 010940
03/13
V6
Änd.-Nr. 46884
11 - DIR 93/42/EEC Technical alterations reserved
► Remove sealing cap for Luer Lock connector 5. ► Fully unscrew fixing screw 10 and turn trocar shaft 11 upside down.
Clamping element 8 will drop out of trocar shaft 11.
Aesculap® Ball trocar
Assembling Leaks caused by damaged or defective sealing elements (sealing cap, silicone flap valve, sealing cap for Luer Lock connector, code ring)! ► Replace damaged/defective sealing elements.
Legend 1 Sealing cap 2 Trocar body 3 Silicone flap valve 4 Ring (with or without Luer Lock insufflation connector) 5 Sealing cap for Luer Lock connector 6 Trocar holder 7 Straining screw 8 Clamping element 9 Code ring 10 Fixing screw 11 Trocar shaft (with ball) 12 Obturator 13 Ball trocar
Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
CAUTION ► Slide clamping element 8 into trocar shaft 11. When doing this make certain that the nose of trocar shaft 11
fits in the groove of clamping element 8. ► Insert and slightly tighten fixing screw 10. ► Mount code ring 9. ► Mount sealing cap for Luer Lock connector 5. ► Slide ring 4 onto trocar shaft 11. ► Press sealing cap 1 on trocar body 2. ► Insert silicone flap valve 3 in trocar body 2. Make certain that silicone flap valve 3 is pushed down to its position
below the thread. ► Screw trocar body 2 onto trocar shaft 11. ► Insert the ball of trocar shaft 11 in trocar holder 6 so that it engages in its proper position. ► Insert straining screw 7 of trocar holder 6 into the tension ring. ► Slightly tighten straining screw 7 until the intended friction is set.
Validated reprocessing procedure General safety instructions
Latex-free
Applicable to ► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
at www.extranet.bbraun.com Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Intended use The ball trocar is used in laparoscopic surgery to insert, position and manipulate instruments and endoscopes in the operating field. Note The ball trocar may only be used in conjunction with the silicone lip valve 3.
Available sizes See Endoscopy online catalog www.endoscopy-catalog.com
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts.
Safe operation Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product. WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING ► Insert ball trocar 13 through the abdominal wall into the abdomen, using an obturator 12. ► Push trocar shaft 11 through the abdominal wall as far as possible down to the ball. ► Mount trocar holder 6 on the ball of trocar shaft 11 so that it engages in its proper position. ► Fasten the trocar holder 6 on Unitrac holding arm RT040R or flexible holding arm FF169R, see instructions for
use of Unitrac holding arm or the flexible holding arm. ► Tighten the straining screw 7 until the ball trocar has the desired amount of stiffness in its ball joint. Instruments or endoscopes will now be held in place and can be moved into various positions without disengaging the holding arm. ► To fix an instrument or endoscope inside the ball trocar itself, tighten the fixing screw 10. ► Connect the insufflation tube to ring 4 directly or via a metal or plastic stop-cock. ► To avoid damaging the silicone lip valve 3, sharp-edged or hooked instruments should be removed from the ball trocar with a slight twist.
Disassembling ► Unscrew the straining screw 7 on trocar holder 6. ► Remove straining screw 7 from the tension ring. ► Press on the tension ring of trocar holder 6 and extract the ball of trocar shaft 11 from the tension ring of trocar
holder 6. ► Unscrew trocar body 2 from trocar shaft 11. ► Remove silicone flap valve 3 from trocar body 2. ► Remove sealing cap 1 from trocar body 2. ► Remove ring 4 from trocar shaft 11. ► Remove code ring 9.
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Preparations at the place of use ► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Preparation before cleaning ► Dismantle the product prior to cleaning, see Disassembling. ► Disassemble the product prior to cleaning.
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for high-grade steel, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 94 °C.
Validated cleaning and disinfection procedure
Manual pre-cleaning with a brush
Validated procedure
Specific requirements
Reference
Phase
Step
t [min]
Conc. [%]
Water quality
Chemical
■ 20 ml disposable syringe ■ When cleaning products with
T [°C/°F]
Manual cleaning with immersion disinfection
Chapter Manual cleaning/disinfection and sub-chapter:
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
movable hinges, ensure that these are in an open position and, if applicable, move the joint while cleaning.
■ Chapter Manual cleaning with immersion disinfection
■ Drying phase: Use a lint-free
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed
cloth or medical compressed air. Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ 20 ml disposable syringe ■ Place the product in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Connect components with lumens and channels directly to the special injector carriage attachment.
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
■ Chapter Manual pre-cleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
■ Store any products with hinges or joints on the tray in such a way that the joints are kept open.
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Manual cleaning/disinfection Risk to patients! ► The product must only be cleaned mechanically!
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound
DANGER
Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
– pH = 13 – <5 % anionic surfactant
Manual cleaning with immersion disinfection Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe.
CAUTION
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational
and undamaged. ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service.
Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully.
► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened.
► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
Note The product can be sterilized either in disassembled or in assembled condition.
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
Mechanical cleaning/disinfection with manual pre-cleaning Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Packaging
Steam sterilization
Note Sterilize the trocar obturator separately (i.e. not inserted into trocar or valve unit). ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Validated sterilization process
– Disassemble the product – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6, Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Accessories/Spare parts See Endoscopy online catalog www.endoscopy-catalog.com
Disposal ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
Distributor in the US/Contact in Canada for product information and complaints 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 010940