Instructions for use/Technical description Bent cup insertion instrument with thread Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapimplantsystemsifus.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-866-229-3002. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Gebogenes Pfanneneinsetzinstrument mit Gewinde Mode d’emploi/Description technique Instrument d’insertion coudé pour cotyle avec filetage Instrucciones de manejo/Descripción técnica Instrumento curvo para copas acetabulares con rosca Istruzioni per l’uso/Descrizione tecnica Strumento inseritore arcuato per cotili con filettatura Instruções de utilização/Descrição técnica Instrumento introdutor de soquete curvo com rosca Gebruiksaanwijzing/Technische beschrijving Gebogen kominbrenginstrument met schroefdraad Brugsanvisning/Teknisk beskrivelse Bøjet indsætningsinstrument til skål med gevind Bruksanvisning/Teknisk beskrivning Böjt insättningsinstrument för ledpanna med gänga Käyttöohje/Tekninen kuvaus Kierteellä varustettu taivutettu nivelkuopan asetusinstrumentti Lietošanas instrukcijas/tehniskais apraksts Izliekts bļodiņas ievietošanas instruments ar vītni Naudojimo instrukcija/techninis aprašas Lenktas gūžduobės įstatymo instrumentas su sriegiu Инструкция по примению/Техническое описание Изогнутый импактор чашки, с резьбой Návod k použití/Technický popis Zahnutý nástroj k nasazování jamky se závitem Instrukcja użytkowania/Opis techniczny Wygięty panewkowy instrument aplikacyjny z gwintem Návod na použitie/Technický opis Ohnutý nástroj na zavádzanie kĺbovej jamky so závitom Használati útmutató/Műszaki leírás Menetes hajlított vápabehelyező eszköz Navodila za uporabo/Tehnični opis Upognjeni instrument za vstavljanje ponvice z navojem Upute za uporabu/Tehnički opis Savijeni instrument za umetanje čašice s navojem Manual de utilizare/Descriere tehnică Instrument de inserție pentru cupe, cotit, cu filet Упътване за употреба/Техническо описание Извит инструмент с резба за вмъкване на ацетабуларна чашка Kullanım Kılavuzu/Teknik açiklama Dişli ile eğimli uygulama aleti Οδηγίες χρήσης/Τεχνική περιγραφή Κυρτό εργαλείο εισαγωγής κυπελίου κοτύλης με σπείρωμα
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com AESCULAP® – a B. Braun brand TA013803
2021-02
Change No. 64200
NT411R / NT579R
NT412R
NT416R
ND476R
3 2
NT411R
A
B
NT579R
NT579R 5
C
4
D
ND476R
en
2.3 ®
AESCULAP Bent cup insertion instrument with thread Legend 1 2 3 4 5
Impact plate Universal thread Universal joint Hexagon for rotation lock Knock-out plate
1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply for the following products: Art. no.
Designation
NT411R
Bent cup insert insertion instrument with thread
NT579R
Bent revision cup insert insertion instrument with thread
► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at
eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows:
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product. WARNING Risk of injury and/or damage to the implant/insertion instrument from carrying out a position correction or if the connection is loose! ► Do not correct the position of an implant that has been firmly impacted in place. ► Carry out position corrections only if the implant has not been impacted in place. ► If the connection between the implant and insertion instrument becomes loose, re-tighten this connection immediately. CAUTION Breakage of components if the torque is too high! ► Use only the intended screwdriver to tighten and fasten the cup. ► Do not cause an increase in torque by turning the cup opposite the complete impactor.
2.3.1
Implantation
Note The explantation plate NT416R can be used for improved rotation control. The use of the explantation plate NT416R is only possible with cup insertion instrument NT411R. ► Place the insertion instrument centrally onto the adaption site of the cup; see fig. A. Ensure correct seating.
Note In the case of insertion instrument NT579R, the hexagon 4 enables rotationally stable seating; see figure C. ► Insert screwdriver NT412R into the opening of the insertion instrument. ► Connect screwdriver NT412R with the universal joint 3. ► Screw universal thread 2 into the cup by turning clockwise with screwdriver NT412R; this will clamp the inser-
tion instrument with the cup. ► Implant the implant device by applying strong hammer blows to the impact plate 1.
WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2.
Clinical use
2.1
Areas of use and limitations of use
2.1.1
Intended use
2.3.2
Explantation
Note Prior to explantation, the insertion instrument is clamped with the cup following the procedure described in the 'implantation' section, see Implantation. Cup insertion instrument NT411R ► Insert explantation plate NT416R into one of the holes under the impact plate 1, until the cone jams; see fig. B. ► Explant the hip cup by applying strong hammer blows to the explantation plate NT416R. Cup insertion instrument revision NT579R ► Explant the hip cup by applying strong hammer blows to the knock-out plate 5; see fig. D.
The bent insertion instrument is used to implant, and if necessary explant, Aesculap hip cups. The bent shape enables operations to be performed that are minimally invasive and gentle on the soft tissue. As a result of the hexagon on NT579R, a rotationally stable accommodation of the implant is possible.
3.
Validated reprocessing procedure
3.1
General safety information
2.1.2
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. The indications conform to those of the implant system used.
2.1.3
Absolute contraindications
The absolute contraindications conform to those of the implant system used.
2.1.4
Relative contraindications
The relative contraindications conform to those of the implant system used. In the presence of relative contraindications, the user decides individually regarding the use of the product.
2.2
Safety information
2.2.1
Clinical user
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
3.7
Mechanical cleaning/disinfection with manual pre-cleaning
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
3.4
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Cleaning/Disinfection
3.5.1
Product-specific safety information on the reprocessing method
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning agents and disinfectants according to the manufacturer's instructions. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
3.7.1
Manual pre-cleaning with ultrasound and brush
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
► Use suitable cleaning/disinfecting agents if the product is disposed of in a wet condition. To prevent foaming and
degradation of the efficacy of the process chemicals: prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water.
3.5.2
Validated cleaning and disinfection procedure
Validated procedure Manual cleaning with ultrasound and immersion disinfection
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
Specific requirements ► Use a suitable cleaning brush. ► Use a 20 ml disposable syringe. ► Drying phase: use a lint-free cloth
Chapter Manual cleaning/disinfection and subsection:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
► Use a suitable cleaning brush. ► Use a 20 ml disposable syringe. ► Place the product on a tray that is
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
suitable for cleaning (avoid rinsing blind spots).
with ultrasound and brush
■ Chapter Manual pre-cleaning ■ Chapter Mechanical alkaline cleaning and thermal disinfecting
3.6
infection procedure.
Reference
or medical compressed air
Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh
Manual cleaning/disinfection
Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.7.2
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
3.6.1
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
RT (cold)
1
RT (cold)
5
– pH ~ 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11*
II
Intermediate rinse
III
Disinfection
-
2
D–W
D–W
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
-
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.8
Inspection
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
D–W: FD–W:
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.8.1
Visual inspection
► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps.
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.8.2
Functional test
► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.9
Packaging
► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.10 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.11 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Technical service
CAUTION Modifications to medical technology equipment result in loss of approval and can result in forfeiture of guarantee/warranty claims. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA013803