Instructions for use/Technical description Bulldog/Vessel clamps - cannot be disassembled or pivoted Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Bulldog-/Gefäßklemmen - nicht demontierbar, nicht schwenkbar Mode d’emploi/Description technique Pinces bulldog/hémostatiques - non démontables, non pivotantes Instrucciones de manejo/Descripción técnica Pinzas de Dieffenbach/pinzas hemostáticas - no desmontables, no giratorias Istruzioni per l’uso/Descrizione tecnica Pinze Bulldog/vascolari - non smontabili, non girevoli Instruções de utilização/Descrição técnica Pinças hemostáticas/Clamps Bulldog - não desmontáveis, não articuláveis Gebruiksaanwijzing/Technische beschrijving Bulldog-/vaatklemmen - niet demonteerbaar, niet zwenkbaar Brugsanvisning/Teknisk beskrivelse Bulldog-/karklemmer - kan ikke afmonteres eller drejes Bruksanvisning/Teknisk beskrivning Bulldog-/kärlklämmor - ej demonterbara, ej svängbara Käyttöohje/Tekninen kuvaus Bulldog-/verisuonipihdit - ei purettavat, ei kääntyvät Lietošanas instrukcijas/tehniskais apraksts Bulldog/asinsvadu skavas - nav demontējamas, nav grozāmas Naudojimo instrukcija/techninis aprašas Neišmontuojami, nepasukami bulldog tipo spaustukai/hemostatai Инструкция по примению/Техническое описание Зажимы для сосудов Bulldog - не демонтируемые, не поворотные Návod k použití/Technický popis Cévní svorky/svorky Bulldog – nedemontovatelné, neotočné Instrukcja użytkowania/Opis techniczny Zaciski szczękowe/naczyniowe – bez możliwości demontażu i obracania Návod na použitie/Technický opis Cievne buldogové svorky – bez možnosti demontáže a otočenia Használati útmutató/Műszaki leírás Bulldog- /érfogók - nem szétszerelhető, nem forgatható Navodila za uporabo/Tehnični opis Bulldog- /prijemke se ne dajo demontirati, negibljive Upute za uporabu/Tehnički opis Buldog hvataljke/vaskularne hvataljke - bez mogućnosti demontiranja i zakretanja Manual de utilizare/Descriere tehnică Cleme Bulldog/vasculare - nedemontabile, non-pivotante Упътване за употреба/Техническо описание Булдог/съдови клампи – неразглобяеми, без шарнири Kullanım Kılavuzu/Teknik açiklama Buldog/damar pensleri - sökülemez, döndürülemez Οδηγίες χρήσης/Τεχνική περιγραφή Κλιπ τύπου Bulldog και κλιπ απολίνωσης αγγείων, μη αποσυναρμολογούμενα, μη περιστρεφόμενα
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap® – a B. Braun brand TA013100
2020-10
V6
Change No. 61645
3.2
en ®
Aesculap Bulldog/Vessel clamps - cannot be disassembled or pivoted 1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply to the products listed below: Bulldog and vessel clamps, cannot be disassembled or pivoted ► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a potential hazard. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2.
Clinical use
2.1
Areas of use and limitations of use
Intended use Bulldog clamps and vascular clips are used for temporary occlusion of blood vessels in the course of surgical operations.
Indications Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
Absolute contraindications No known absolute contraindications.
Relative contraindications The following conditions, individual or combined, can lead to delayed healing or compromise the success of the operation: ■ Medical or surgical conditions (e.g. comorbidities) which could hinder the success of the operation. In the presence of relative contraindications, the user decides individually regarding the use of the product.
2.2
Safety information
Clinical user General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user.
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
3.4
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Cleaning/Disinfection
Product-specific safety information on the reprocessing method Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning agents and disinfectants according to the manufacturer's instructions, – which are approved for plastics and high-grade steel. – and that do not attack softeners (e.g., in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C. ► If the microsurgical products can be securely fixed in machines or storage devices in such a way that they will
be cleaned thoroughly, clean and disinfect them mechanically.
Validated cleaning and disinfection procedure Validated procedure
Specific requirements
Reference
Manual cleaning with ultrasound and immersion disinfection
■ Suitable cleaning brush ■ Keep working ends open for
Chapter Manual cleaning/disinfection and subsection:
cleaning.
Sterility The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
■ In the ultrasonic bath: For products that cannot be placed on the basket with opened jaws, use wire or similar means to keep the jaws open. Make sure that the jaw part is opened sufficiently wide to have the wall surfaces accessible for the cleaning media.
■ Drying phase: Use a lint-free cloth or medical compressed air Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
Validated reprocessing procedure General safety information
with ultrasound and brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
made of metal and be round if at all possible. Plastic materials or elastic tubes (silicone or PVC) are not permitted (not all areas will be reached by water jets).
■ Position the wire or similar in the area of the olive or in the area without teeth.
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
■ Chapter Manual pre-cleaning
■ Wire or equipment must be
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
Note If there is no final sterilization, then a virucidal disinfectant must be used.
■ Keep working ends open for
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
placed on the basket with opened jaws, use wire or similar means to keep the jaws open. Make sure that the jaw part is opened sufficiently wide to have the wall surfaces accessible for the cleaning media.
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this.
is suitable for cleaning (avoid rinsing blind spots).
■ For products that cannot be
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
3.1
■ Place the product on a tray that
cleaning.
Application
3.
ultrasound and immersion disinfection
with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
■ Chapter Manual cleaning with
■ When cleaning instruments
3.6
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Manual cleaning with ultrasound and immersion disinfection
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
I
Step
Ultrasonic cleaning
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Phase
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
RT (cold)
1
RT
-
IV
V
Final rinse
Drying
-
-
FD-W
-
-
-
D–W: FD–W:
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH ~ 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
D–W: FD–W:
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.7
Mechanical cleaning/disinfection with manual pre-cleaning
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.8
Inspection
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
Functional test CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.9
Packaging
► Store products with ratchet locks fully opened or locked no further than in the first notch. ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.10 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
Manual pre-cleaning with ultrasound and brush
opening any valves and faucets). ► Validated sterilization process
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
– Steam sterilization with the fractionated vacuum process – Steam sterilizer in accordance with DIN EN 285 and validated in accordance with DIN EN ISO 17665 – Sterilization with the fractionated vacuum process at 134 °C, holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
II
Rinsing
RT (cold)
1
-
D–W
-
3.11 Storage
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA013100