BBraun
HiLAN pneumatic trepan motor GA742R
160 Pages
Preview
Page 1
Aesculap® HiLAN HiLAN pneumatic trepan motor GA742R
® HiLAN pneumatic trepanAesculap motor GA742R HiLAN
Legend 1 HUDSON connector 2 Unlocking sleeve 3 HiLAN pneumatic trepan motor 4 Compressed-air connector 5 Fast-lock coupling (motor-side) 6 Unlocking sleeve (motor-side) 7 Tube coupling 8 Motor tube 9 Aesculap-Dräger plug 10 Unlocking sleeve 11 Fast-lock coupling (for foot control) 12 Unlocking sleeve (for foot control) 13 Tube coupling 14 Connecting tube 15 Aesculap-Dräger socket 16 Socket (at foot control) 17 Plug (at foot control) 18 Diffuser 19 DIN connector 20 Pressure reducer 21 DIN socket 22 Unlocking sleeve (at DIN connection) 23 Schrader socket
4.1.2 5. 5.1 5.2 5.3 5.3.1 5.3.2 5.3.3 5.4 5.5 6. 6.1 6.2 6.3 6.4 6.5 6.6 6.6.1
Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
2011-02
Notice indicating the next scheduled service (date) by international B. Braun/Aesculap agency, see Technical Service
Contents 1. 2. 3. 3.1 3.2 3.3 3.4 3.4.1 3.4.2 2
Applicable to Safe handling Product description Scope of supply Components required for operation Intended use Operating principle HiLAN pneumatic trepan motor Compressed-air tubes
3.4.3 3.4.4 4. 4.1 4.1.1
3 3 3 3 3 3 3 3 4
6.6.2 6.7 6.7.1 6.8 6.8.1 6.9 6.10 6.11 6.12 6.13 7. 8. 9. 10. 11. 12. 13.
Exit air recycling via the compressed-air tube Foot control Preparation Compressed-air supply Operation with central gas supply according to DIN 13260 and EN 737 Operation with compressed-gas cylinders according to DIN 4664 Working with the HiLAN pneumatic trepan motor System set-up Connecting the accessories Connecting the HiLAN pneumatic trepan motor to the compressed-air supply Connecting the HiLAN pneumatic trepan motor to the foot control Connecting the foot control to the compressed air supply Coupling a cranial perforator with HUDSON coupling to the HiLAN pneumatic trepan motor Function checks Safe operation Validated reprocessing procedure General safety instructions General information Dismantling prior to carrying out the reprocessing procedure Preparations at the place of use Preparation before cleaning Cleaning/disinfection Product-specific safety instructions for the reprocessing procedure Validated cleaning and disinfection procedure Manual cleaning/disinfecting Manual cleaning and wipe disinfecting Mechanical cleaning/disinfecting Mechanical alkaline cleaning and thermal disinfecting Inspection, maintenance and checks Packaging Steam sterilization Sterilization for the US market Storage Maintenance Troubleshooting list Technical Service Accessories/Spare parts Technical data Disposal Distributor in the US/Contact in Canada for product information and complaints
4 4 4 4 4 4 4 4 5 5 5 5 6 6 6 7 7 7 7 7 7 8 8 8 9 9 10 10 10 10 11 11 11 11 12 12 13 13 13 13
1.
Applicable to
Compressed-air tubes
► For item-specific instructions for use and information on material
compatibility, see also www.extranet.bbraun.com
2.
the
Aesculap
Extranet
Safe handling
CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Prior to use, check that the product is in good working order. ► To prevent damage caused by improper setup or operation, and in order not to compromise warranty and manufacturer liability: – Use the product only according to these instructions for use. – Follow the safety and maintenance instructions. – Only combine Aesculap products with each other. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Keep the instructions for use accessible for the user.
3.
Product description
3.1
Scope of supply
Art. no.
Aesculap-Dräger system GA464R
3m
GA466R
5m
GA194R
Extension tube 3 m
GA195R
Extension tube 5 m
DIN connection GA461R
3 m, with diffuser
GA463R
3 m, without diffuser
GA465R
5 m, without diffuser
GA087
Extension tube 5 m
Motor tubes GA513R
3m
GA523R
2.3 m
GA505R
5m
GA194R
Extension tube 3 m
GA195R
Extension tube 5 m
3.3 Art. no.
Designation
GA742R
HiLAN pneumatic trepan motor
GB600820
STERILIT adapter
3.2
Components required for operation
Art. no.
Designation
–
Compressed-air supply with a minimum of 6 bar (0.6 MPa)
e.g., GA099
For compressed air supply from compressed air cylinders: Pressure reducer
GA521
Foot control
–
HiLAN pneumatic trepan motor
Designation
at
Intended use
The HiLAN pneumatic trepan motor is used as a drive unit for trepanning tools in neurosurgery.
3.4
Operating principle
3.4.1
HiLAN pneumatic trepan motor
The HiLAN pneumatic trepan motor converts the energy induced by the compressed-air supply into rotary movement. Only cranial perforators with a HUDSON coupling can be coupled to this device. The HiLAN pneumatic trepan motor is controlled via the foot control. The rotary speed of the HiLAN pneumatic trepan motor can be continuously adjusted between "ZERO" and "MAXIMAL" by applying more or less pressure on the control pedal. Note The operator is responsible for the compatibility and proper, safe use of the tools used with this applied part.
3
Aesculap® HiLAN HiLAN pneumatic trepan motor GA742R
3.4.2
Compressed-air tubes
4.1.1
Operation with central gas supply according to DIN 13260 and EN 737
All compressed-air tubes have a fast-lock coupling. Consequently, the compressed air tube need not be disconnected from the compressed air supply when changing the motor.
The gas supply must provide clean, dry air and must be set as follows: Nominal operating pressure
8 bar
3.4.3
Maximum allowable operating pressure
10 bar
Minimum operating pressure
6 bar
Minimum gas-supply delivery
300 l/min
Exit air recycling via the compressed-air tube
■ Complete exit air recycling (Aesculap-Dräger system): The exit air is passed outside the operating room. Absence of air turbulence in the operating room: no effect on either ambient air temperature or hygiene in the operating room. ■ Partial exit-air recycling (DIN connection and others): The exit air emerges from the tube end, either directly or through a diffuser inside the operating room. 3.4.4
Foot control
The foot control GA521 is equipped with a forced ventilation system. As soon as the foot control pedal is released, the supply air is vented on the motor side. In this way the HiLAN pneumatic trepan motor is stopped more rapidly.
4.
Preparation
Non-compliance with the following instructions will preclude all responsibility and liability in this respect on the part of Aesculap. ► When setting up and operating the product, adhere to – national regulations for installation and operation, – national regulations on fire and explosion protection. ► Prior to using the HiLAN pneumatic trepan motor and its accessories, check the motor and its accessories for visible damage. ► Use the HiLAN pneumatic trepan motor and its accessories only if they are in perfect condition.
4.1
Operation with compressed-gas cylinders according to DIN 4664
The following conditions must be met: ■ Compressed air or nitrogen is used ■ An adequate stock of full gas cylinders is in place ■ Cylinder pressure reduced to operating pressure of 6 bar to 10 bar through a pressure reducer (e.g., GA099) ► Operating pressure must be read from the scale on the device and set with the T-screw.
5.
Working with the HiLAN pneumatic trepan motor
5.1
System set-up
WARNING
Compressed-air supply
DANGER
CAUTION
4
4.1.2
Danger of fire and explosion when operating devices that run on oxygen! ► Do not operate the HiLAN pneumatic trepan motor with oxygen under any circumstances. ► Operate the HiLAN pneumatic trepan motor with compressed air or nitrogen. Damage to the HiLAN pneumatic trepan motor caused by operating pressure in excess of 10 bar! ► Set an operating pressure of between 6 bar and 10 bar.
Risk of injury and material damage caused by inadvertent activation of the motor! ► To avoid injuries caused by inadvertent starting of the motor, never connect the compressed-air tube directly to a compressed-air source without the foot control installed between the supply and the motor.
WARNING
Risk of infection and contamination! The HiLAN pneumatic trepan motor and its accessories are supplied in unsterile condition. ► Carry out sterile processing of the HiLAN pneumatic trepan motor and its accessories prior to putting them into operation.
DANGER
Risk of injury caused by inadvertent activation of the motor! ► Disconnect the HiLAN pneumatic trepan motor from the motor tube before carrying out any tool change.
5.2
Connecting the accessories
Combinations of accessories that are not mentioned in the present instructions for use may only be employed if they are specifically intended for the respective application, and if they do not compromise the performance and safety characteristics of the products. ► Please address your B. Braun/Aesculap partner or Aesculap Technical Service with any inquiries in this respect; for a contact address, see Technical Service.
5.3
Connecting the HiLAN pneumatic trepan motor to the compressed-air supply
Long tubes may cause a backup of the exhaust air and may thus result in lower power of the HiLAN pneumatic trepan motor. ► Keep tube lines as short as possible. ► Make certain that the tube lines are not buckled or crushed. ► Use a suitable extension tube if necessary, see Preparation. Note The HiLAN pneumatic trepan motor and the motor tube are sterile. Sterile separation takes place at the motor tube. 5.3.1
Connecting the HiLAN pneumatic trepan motor to the foot control
Coupling: ► Pull back unlocking sleeve 6 on the motor tube 8 and connect motor
tube with fast-lock coupling 5 to the compressed-air connector 4 on the HiLAN pneumatic trepan motor 3. ► Release unlocking sleeve 6. ► Insert plug 9 of motor tube 8 in socket 16 of the foot control so that it fully engages. Uncoupling: ► Pull back unlocking sleeve 6 at motor tube 8. ► Pull off fast-lock coupling 5 from HiLAN pneumatic trepan motor 3. ► Pull back unlocking sleeve 10 on plug 9 of motor tube 8. ► Pull out plug 9 from socket 16 of the foot control.
5.3.2
Connecting the foot control to the compressed air supply
Aesculap-Dräger system Coupling: ► Pull back unlocking sleeve 12 on connecting tube 14 and connect tube with fast-lock coupling 11 to plug 17 on the foot control. ► Release unlocking sleeve 12. ► Insert plug 9 of connecting tube 14 in Aesculap-Dräger socket 15 so that it fully engages. Uncoupling: ► Pull back unlocking sleeve 12 at connecting tube 14. ► Pull off fast-lock coupling 11 from foot control plug 17. ► Pull back unlocking sleeve 10 on plug 9 of connecting tube 14. ► Pull out plug 9 from Aesculap-Dräger socket 15. DIN connection Coupling: ► Pull back unlocking sleeve 12 on connecting tube 14 and connect tube
with fast-lock coupling 11 to plug 17 on the foot control. ► Release unlocking sleeve 12. ► Insert DIN connector 19 of connecting tube 14 in DIN socket 21 so
that it fully engages. Uncoupling: ► Pull back unlocking sleeve 12 at connecting tube 14. ► Pull off fast-lock coupling 11 from foot control plug 17. ► Pull back unlocking sleeve 22 on DIN socket 21. ► Pull out DIN connector 19 from DIN socket 21.
5
Aesculap® HiLAN HiLAN pneumatic trepan motor GA742R
5.3.3
Coupling a cranial perforator with HUDSON coupling to the HiLAN pneumatic trepan motor
WARNING
CAUTION
Risk of injury and material damage due to inappropriate use of the cranial perforators! ► Follow the instructions for use of the cranial perforators. Damage to the HiLAN pneumatic trepan motor and/or the cranial perforators! ► Couple/uncouple cranial perforators only with the motor inactive.
5.4
Function checks
► Always perform a dry run of the pneumatic motor system (HiLAN pneu-
matic trepan motor, foot control, tubes) prior to use. ► Make certain the HiLAN pneumatic trepan motor and the foot control
function properly. ► Make certain that all tubes are properly connected and free of leaks. ► Check that the cutting edges of the skull trepan are not showing any
mechanical damage. Test run ► Ensure that the perforator A is seated securely in the tool coupling C. ► Briefly run the trepan motor at maximum speed. Check for proper and
reliable functioning.
WARNING
5.5
Risk of burns to the patient, caused by blunt perforators or lack/inadequacy of maintenance of handpieces! ► Prior to each use and before clamping in place, check the perforator for any wear or damage.
Safe operation
Fig. 1
Risk of injury caused by inadvertent activation of the HiLAN pneumatic trepan motor! ► Always secure the HiLAN pneumatic trepan motor against inadvertent activation when it is not in active use. ► After use, disconnect the motor tube from the HiLAN pneumatic trepan motor.
Legend A Cranial perforator (ready for operation) B Tool shaft C Tool coupling 2 Unlocking sleeve 3 HiLAN pneumatic trepan motor
DANGER
Coupling the cranial perforator
WARNING
Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product.
WARNING
Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control.
► Pull and hold back uncoupling sleeve 2. ► Insert tool shaft B into tool holder of the HiLAN pneumatic trepan
motor 3 as far as it will go. ► Release unlocking sleeve 2. ► Pull at cranial perforator A to check for secure seating. Uncoupling the cranial perforator ► Pull and hold back uncoupling sleeve 2. ► Remove cranial perforator A. ► Release unlocking sleeve 2.
6
► Press the foot control to set the speed of HiLAN pneumatic trepan
motor 3.
6.
Validated reprocessing procedure
6.1
General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation.
6.3
Note If there is no final sterilization, then a virucidal disinfectant must be used.
6.4
Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
6.2
General information
Dismantling prior to carrying out the reprocessing procedure
► Disassemble the product immediately after use, as described in the
respective instructions for use.
Preparations at the place of use
► Remove any visible surgical residues to the extent possible with a
6.5
Preparation before cleaning
► Carry out non-fixating/NaCl-free pre-cleaning immediately after use.
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating precleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary.
7
Aesculap® HiLAN HiLAN pneumatic trepan motor GA742R
6.6
Cleaning/disinfection
6.6.1
Product-specific safety instructions for the reprocessing procedure
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for high-grade steel, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 60 °C.
► Dry the product for at least 10 minutes at a maximum of 120 °C.
Note The indicated drying temperature is a guide temperature only. It must be checked taking into account the specific conditions (e.g. load) and adjusted if applicable. ► Do not clean motors/handpieces by ultrasonic treatment and do not
immerse them in any liquids. To avoid the risk of corrosion/malfunctioning, allow any fluid that has entered the product to drain out immediately.
6.6.2
Validated cleaning and disinfection procedure
Validated procedure
Special features
Manual cleaning and wipe disinfection
■ When cleaning instruments with movable hinges, ensure that these
■ HiLAN pneumatic trepan motor
Reference
Chapter Manual cleaning/disinfecting and subare in an open position and, if applicable, move the joint while clean- section: ing. ■ Chapter Manual cleaning and wipe disinfecting ■ Drying phase: Use a lint-free cloth or compressed air for medical purposes
■ Ensure that the product is positioned in such a way that water will not enter the product e.g. through coupling interfaces. (Immediately remove any fluid that entered the product inadvertently.) Mechanical alkaline cleaning and thermal disinfecting
■ HiLAN pneumatic trepan motor
8
■ Insert the product in its proper position in the Eccos holder GB609R.
Chapter Mechanical cleaning/disinfecting and subsection:
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
6.7
Manual cleaning/disinfecting
► Prior to manual disinfecting, allow water to drip off for a sufficient
length of time to prevent dilution of the disinfecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for
residues. ► Repeat the cleaning/disinfection process if necessary.
6.7.1
Manual cleaning and wipe disinfecting
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Cleaning
RT (cold)
-
-
D–W
-
II
Drying
RT
-
-
-
-
III
Wipe disinfection
-
>1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
IV
Final rinse
RT (cold)
0.5
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W: RT:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) Room temperature
Phase I ► Clean the product under running faucet water, using a suitable clean-
ing brush until all visible residues have been removed from the surfaces. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. Phase II ► Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air), see Validated cleaning and disinfection procedure. Phase III ► Wipe all surfaces of the product with a single-use disinfectant wipe.
Phase IV ► After the specified exposure time (at least 1 min), rinse the disinfected
surfaces under running FD water. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air), see Validated cleaning and disinfection procedure.
9
Aesculap® HiLAN HiLAN pneumatic trepan motor GA742R
6.8
Mechanical cleaning/disinfecting
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals. 6.8.1
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: Drinking water FD–W: Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline
► Check visible surfaces for residues after mechanical cleaning/disinfect-
ing.
6.9
► Follow the instructions for use for the applied packaging and storage
Inspection, maintenance and checks
► Allow the product to cool down to room temperature. ► Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged. ► Spray the HiLAN pneumatic trepan motor with Aesculap-STERILIT oil
spray GB600 with adapter GB600820 for approx. 1 second. ► Check the product for any damage, abnormal running noise, overheat-
ing or excessive vibration. ► Inspect tools for broken, damaged or blunt edges. ► Set aside the product if it is damaged.
10
6.10 Packaging systems (e.g. instructions for use TA009721 for Aesculap Eccos storage system). ► Insert the product in its proper position in the Eccos holder, or put it on a tray in such a way that the product is protected against damage. ► Pack trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging will prevent a recontamination of the product.
6.11 Steam sterilization
7.
Note The foot control must not be sterilized.
To ensure reliable operation, the product must be maintained as indicated on the maintenance label. For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
Note The product may only be sterilized when dismantled.
Maintenance
► Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and faucets). ► Validated sterilization process – Disassemble the instrument – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
6.12 Sterilization for the US market
■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.
■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6, Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
6.13 Storage ► Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
11
Aesculap® HiLAN HiLAN pneumatic trepan motor GA742R
8.
Troubleshooting list
Malfunction
Detection
Cause
Remedy
Motor not running
No compressed air
Gas cylinder empty Manometer set to "0"
Change cylinder
No compressed air at the wall outlet
Open valve
Motor tube/connecting tube not connected correctly
Fully connect the fast-lock coupling
Pedal movement too light or pedal not moving
Foot control defective
Have product repaired by the manufacturer
Insufficient air supply
Operating pressure too low (at least 6 bar)
Have compressed-air supply checked
Pressure reducer on gas cylinder defective
Have pressure reducer checked
Operating pressure setting at pressure reducer too low
Adjust to higher operating pressure setting
Connecting tube too long
Use shorter connecting tube
Connecting tube excessively buckled or crushed
Free the connecting tube
Exit-air ducting
Unblock diffuser/exit air channel
HiLAN pneumatic trepan motor defective
Have product repaired by the manufacturer
Insufficient power
Strong air resistance in connecting tube
Motor not running despite correct air supply HiLAN pneumatic trepan motor overheating HiLAN pneumatic trepan motor running jerkily
9.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product.
WARNING ► For service and repairs, please contact your national B. Braun/
Aesculapp agency. Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany
12
Phone: +49 (7461) 95 -1601 Fax: +49 (7461) 14 -939 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 USA Aesculap Repair Hotline Phone: +1 (800) 214 -3392 Fax: +1 (314) 895 -4420 Other service addresses can be obtained from the address indicated above.
10. Accessories/Spare parts
11. Technical data
Art. no.
Designation
Classification acc. to Directive 93/42/EEC
GA521
Foot control (accelerator pedal)
Art. no.
Designation
Class
GB459R
Eccos mounting set for holders
GA742R
HiLAN pneumatic trepan motor
IIa
GB564R
Eccos holder for motor tube
GB609R
Eccos holder for trepan motor
GB600
Aesculap STERILIT oil spray
GB600820
STERILIT adapter
JF213R
Tray (64 mm)
JK431
Container
Compressed-air tubes Art. no.
Designation
Nominal operating pressure
8 bar ± 2 bar
Max. operating pressure
10 bar
Air consumption
approx. 300 l/min
Max. permissible backup in the exit air
0.4 bar to 0.6 bar
Torque
>3 Nm
Weight
495 g
Size
∅ 39 mm x 130 mm
Aesculap-Dräger system GA464R
3m
GA466R
5m
GA194R
Extension tube 3 m
GA195R
Extension tube 5 m
DIN connection
12. Disposal Note The user institution is obliged to process the product before its disposal, see Validated reprocessing procedure. ► Detailed information concerning the disposal of the product is avail-
GA461R
3 m, with diffuser
GA463R
3 m, without diffuser
GA465R
5 m, without diffuser
GA087
Extension tube 5 m
Motor tubes GA513R
3m
GA523R
2.3 m
GA505R
5m
GA194R
Extension tube 3 m
GA195R
Extension tube 5 m
able through your national B. Braun/Aesculap agency, see Technical Service.
13. Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA
13