Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
® HiLAN XS pneumatic motor Aesculap GA740R/GA741R HiLAN XS
Legend 1 HiLAN XS pneumatic motor 2 Fast-lock coupling (motor-side) 3 Unlocking sleeve (motor-side) 4 Motor tube 5 Plug (on motor side) 6 Unlocking sleeve (for foot control) 7 Foot control socket 8 Foot control plug 9 Connecting tube 10 Unlocking sleeve (for foot control) 11 Fast-lock coupling (for foot control) 12 Diffuser 13 DIN connector 14 Aesculap-Dräger socket 15 DIN socket 16 Hi-Line XS handpiece 17 Threaded ring 18 Safety stop 19 Tube connection 20 Handpiece coupling 21 Notches
3. 3.1 3.2 3.3
4. 4.1
5. 5.1
5.2 5.3 6. 6.1 6.2 6.3 6.4 6.5
Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
6.6 Rotate “right” (clockwise) to unlock the handpiece 6.7 Rotate “left” (counterclockwise) to lock the handpiece
I
HiLAN XS pneumatic motor “ready”
0
HiLAN XS pneumatic motor “off” Direction of motor rotation “right” Direction of motor rotation “left”
Contents 1. 2. 2
Applicable to Safe handling
3 3
6.8 6.9 6.10 6.11 6.12 7. 8. 9. 10. 11. 11.1 12. 13.
Product description Components required for operation Intended use Operating principle HiLAN XS pneumatic motor Compressed-air tubes Exit air recycling via the compressed air tube Foot control Preparation and setup Compressed air supply Operation with central gas supply according to DIN 13260 and EN 737 Operation with compressed gas cylinders according to DIN4664 Working with the HiLAN XS pneumatic motor System set-up Connecting the accessories Connecting the HiLAN XS pneumatic motor to the compressed-air supply Function checks Safe operation Validated reprocessing procedure General safety instructions General information Preparations at the place of use Preparation before cleaning Cleaning/disinfection Product-specific safety instructions for the reprocessing procedure Validated cleaning and disinfection procedure Manual cleaning/disinfecting Manual cleaning and wipe disinfecting Mechanical cleaning/disinfecting Mechanical alkaline cleaning and thermal disinfecting Inspection, maintenance and checks Packaging Steam sterilization Sterilization for the US market Storage Maintenance Troubleshooting list Technical Service Accessories/Spare parts Technical data Ambient conditions Disposal Distributor in the US/Contact in Canada for product information and complaints
3 3 4 4 4 4 4 4 4 4 4 4 5 5 5 5 6 6 7 7 7 8 8 8 8 9 10 10 11 11 12 12 12 12 12 12 13 14 14 15 15 15 15
1.
Applicable to
3.
Product description
3.1
Components required for operation
► For item-specific instructions for use and information on material
compatibility, see also www.extranet.bbraun.com
the
Aesculap
Extranet
at
When putting the HiLAN XS pneumatic motor into operation, the following components are required
2.
Safe handling
CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Prior to use, check that the product is in good working order. ► To prevent damage caused by improper setup or operation, and in order not to compromise warranty and manufacturer liability: – Use the product only according to these instructions for use. – Follow the safety and maintenance instructions. – Only combine Aesculap products with each other. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Keep the instructions for use accessible for the user. ► Only use Hi-Line XS handpieces and tools expressly approved for the HiLAN XS Power System.
Designation
Special features
Art. no.
Compressed air supply with a minimum of 6 bar (0.6 MPa)
When supplying compressed air with a compressed gas cylinder use a pressure reducer
Pressure reducer: e.g. GA099
Compressed-air tubes
Aesculap-Dräger system GA464 (3 m), GA466 (5 m) if needed, extension tube GA194 (3 m), GA195 (5 m)
Foot control
DIN coupling
GA461 (3 m, with diffuser) GA463 (3 m, without diffuser) GA465 (5 m, without diffuser) if needed, extension tube GA087 (5 m)
Motor tube
GA513 (3 m) GA523 (2.3 m) GA505 (5 m) if needed, extension tube GA194 (3 m), GA195 (5 m) GA521
HiLAN XS pneumatic motor Hi-Line XS handpiece
3
Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
3.2
Intended use
The HiLAN XS pneumatic motor GA740R/GA741R is used for driving Aesculap-Hi-Line XS handpieces in microsurgery, neurosurgery, spine surgery, ENT and maxillary surgery as well as in orthopedics.
3.3
Operating principle
3.3.1
HiLAN XS pneumatic motor
In the HiLAN XS pneumatic motor, the energy provided by the compressed air supply is converted into rotary movement. The HiLAN XS pneumatic motor is equipped with a safety stop, which prevents any unintended start-up of the HiLAN XS motor. 3.3.2
Preparation and setup
Non-compliance with the following instructions will preclude all responsibility and liability in this respect on the part of Aesculap. ► When setting up and operating the product, adhere to – the national regulations for installation and operation, and – national regulations on fire and explosion protection ► Prior to using the HiLAN XS pneumatic motor and its accessories, check to be sure they are not damaged. ► Use the HiLAN XS pneumatic motor and its accessories only if they are in perfect condition.
4.1
Compressed air supply
Compressed-air tubes
All compressed air tubes have a fast-lock coupling. Consequently, the compressed air tube need not be disconnected from the compressed air supply when changing the motor. 3.3.3
4.
DANGER
Exit air recycling via the compressed air tube
■ Complete exit air recycling (Aesculap-Dräger system): The exit air is passed outside the operating room. Absence of air turbulence in the operating room: no effect on either ambient air temperature or hygiene in the operating room. ■ Partial exit air recycling (DIN coupling and others): The exit air emerges from the tube end, either directly or through a diffuser inside the operating room. 3.3.4
CAUTION
4.1.1
Foot control
The foot control GA521 is equipped with a forced ventilation system. As soon as the foot control pedal is released, the air is forced out towards the motor. This allows the HiLAN XS pneumatic motor to stop more rapidly.
Danger of fire and explosion when operating devices that run on oxygen! ► Never operate the HiLAN XS pneumatic motor with oxygen! ► Operate the HiLAN XS pneumatic motor with compressed air or nitrogen. Damage to the HiLAN XS pneumatic motor caused by operating pressure in excess of 10 bar! ► Set an operating pressure of between 6 bar and 10 bar.
Operation with central gas supply according to DIN 13260 and EN 737
The gas supply must provide clean, dry air and must be set as follows: Nominal operating pressure
8 bar
Maximum allowable operating pressure
10 bar
Minimum operating pressure
6 bar
Minimum gas supply delivery
300 l/min
4.1.2
Operation with compressed gas cylinders according to DIN4664
The following conditions must be met: ■ Compressed air or nitrogen is used ■ An adequate stock of gas cylinders is in place ■ Cylinder pressure reduced to operating pressure of 6 bar to 10 bar through a pressure reducer (e.g. GA099) ► Operating pressure must be read from the scale on the device and set with the T-screw.
4
5.
5.1
Working with the HiLAN XS pneumatic motor System set-up
WARNING
WARNING
5.1.1
Injury to the operator and damage to the device components can result from accidentally putting the HiLAN XS pneumatic motor into operation! ► Use the safety stop to prevent any accidental activation of the HiLAN XS pneumatic motor. ► Only attach Hi-Line XS handpieces, tools and the compressed air supply when the HiLAN XS pneumatic motor is shut off. Risk of infection and contamination! The HiLAN XS pneumatic motor, its accessories and the Hi-Line XS handpieces are delivered unsterile! ► Sterilize the HiLAN XS pneumatic motor, its accessories and the Hi-Line XS handpieces before putting them into operation.
Connecting the accessories
Combinations of accessories that are not mentioned in the present instructions for use may only be employed if they are specifically intended for the respective application, and if they do not compromise the performance and safety characteristics of the products. ► Please address your B. Braun/Aesculap partner or Aesculap Technical Service with any inquiries in this respect; for a contact address, see Technical Service.
5.1.2
Connecting the HiLAN XS pneumatic motor to the compressed-air supply
Long tubes may cause a backup of the exhaust air and may thus result in lower power of the HiLAN XS motor. ► Keep tube connections as short as possible. ► Make certain that the tubes are not buckled or crushed. ► Use a suitable extension tube if necessary, see Components required for operation. Note Motors GA740R/GA741R and the motor tube are sterile. Sterile separation is carried out at the motor hose. Connecting the HiLAN XS pneumatic motor to the foot control Coupling: ► Pull back unlocking sleeve 3 on the motor tube 4 and attach motor tube with the fast-lock coupling 2 onto the tube connection 19 on the motor 1. ► Release unlocking sleeve 3. ► Insert plug 5 of the motor tube 4 into the socket 7 on the foot control, until it engages. Uncoupling: ► Pull back unlocking sleeve 3 on the motor tube 4. ► Remove fast-lock coupling 2 from the HiLAN XS pneumatic motor 1. ► Pull back unlocking sleeve 6 on the plug 5 of the motor tube 4. ► Pull out plug 5 from the socket 7 on the foot control. Note If coolant pump GA530 is employed for cooling/irrigating the tool blades, connect the coolant pump between the foot control and the HiLAN XS pneumatic motor, see instructions for use of the coolant pump TA022140.
Coupling a Hi-Line XS handpiece (e.g. GB740R-GB758R) ► Turn safety stop 18 on the HiLAN XS pneumatic motor 1 to “0”. ► Turn threaded ring 17 in the direction of the arrow in symbol
as far as it will go. ► Attach Hi-Line XS handpiece 16 onto the HiLAN XS pneumatic motor 1 in such a way that the pin on the Hi-Line XS engages in one of the notches 21 (The Hi-Line XS handpiece can be mounted in 15 different positions) ► Turn threaded ring 17 in the direction of the arrow in symbol as far as it will go. The Hi-Line XS handpiece is locked now. ► Make certain the threaded ring is fastened to its limit stop, because otherwise it may come loose due to vibrations during operation. Detaching the Hi-Line XS handpiece ► Turn safety stop 18 on the HiLAN XS pneumatic motor 1 to “0”. ► Turn threaded ring 17 in the direction of the arrow in symbol
as far as it will go. ► Pull off the Hi-Line XS handpiece from the HiLAN XS pneumatic motor.
5
Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
Connecting the foot control to the compressed air supply
5.3
Safe operation
Aesculap-Dräger system Coupling: ► Pull back unlocking sleeve 10 on the connecting tube 9 and attach tube with the fast-lock coupling 11 onto the plug 8 on the foot control. ► Release unlocking sleeve 10. ► Insert plug 5 of the connecting tube 9 into the Aesculap-Dräger socket 14 until it engages. Uncoupling: ► Pull back unlocking sleeve 10 on the connecting tube 9. ► Remove fast-lock coupling 11 from the plug 8 on the foot control. ► Pull back unlocking sleeve 6 on the plug 5 of the connecting tube 9. ► Pull out plug 5 from the Aesculap-Dräger socket 14.
WARNING
WARNING
WARNING
Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product. Danger of damage to patients’ hearing from sound of motor! ► When using the HiLAN XS pneumatic motor in otology, keep as much distance to the ear as possible. Heat necroses with GA741R with serrated tools in counterclockwise mode! ► Only run diamond tools in counterclockwise mode.
DIN coupling
► Turn safety stop 18 for GA740R on the HiLAN XS pneumatic motor 1
Coupling: ► Pull back unlocking sleeve 10 on the connecting tube 9 and attach tube with the fast-lock coupling 11 onto the plug 8 on the foot control. ► Release unlocking sleeve 10. ► Insert DIN connector 13 of the connecting tube 9 into DIN socket 15 until it engages. Uncoupling: ► Pull back unlocking sleeve 10 on the connecting tube 9. ► Remove fast-lock coupling 11 from the plug 8 on the foot control. ► Pull back unlocking sleeve on the DIN socket 15. ► Pull out DIN connector 13 from the DIN socket 15.
until the mark on the unlocking sleeve is aligned with the “I” mark on the motor. ► Turn safety stop 18 for GA741R to the intended direction of rotation: – Rotation “right”: Turn safety stop 18 clockwise to symbol . – Rotation “left”: Push safety stop 18 axially forward and then turn counterclockwise to symbol . ► Carefully press down the foot control pedal. The motor starts running. ► Adjusting speed: Press down on the pedal to a greater or lesser extent. ► To select maximum speed: Press down on the pedal as far as it will go. ► To switch off the HiLAN XS pneumatic motor, e.g. to change a tool or handpiece intraoperatively: Turn safety stop 18 until the mark on the unlocking sleeve is aligned with the “0” mark on the motor.
5.2
Function checks
► Always do a dry run of the pneumatic motor system (HiLAN XS pneu-
matic motor, foot control, tube) prior to use. ► Make certain that the HiLAN XS pneumatic motor and foot control are in good working order. ► Make certain that all tubes are properly connected and free of leaks.
6
6. 6.1
Validated reprocessing procedure General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
6.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating precleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
7
Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
6.3
Preparations at the place of use
► Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
6.4
Preparation before cleaning
► Carry out non-fixating/NaCl-free pre-cleaning immediately after use.
6.5
Cleaning/disinfection
6.5.1
Product-specific safety instructions for the reprocessing procedure
DANGER
CAUTION
CAUTION
Fire hazard! ► Do not use flammable or explosive cleaning or disinfecting solutions! ► Ensure that no fluids will penetrate the product. Damage to the product due to inappropriate cleaning/disinfecting agents! ► Only use cleaning/disinfecting agents approved for surface cleaning. Follow the manufacturer’s instructions for the respective cleaning/disinfecting agent. Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – be approved for plastic material and highgrade steel, – do not attack softeners (e.g. in silicone), – that do not cause stress cracking of plastics (e.g. PPSU). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 55 °C.
► Dry the product for at least 10 minutes at a maximum of 120 °C.
Note The indicated drying temperature is a guide temperature only. It must be checked taking into account the specific conditions (e.g. load) and adjusted if applicable.
8
► Do not clean motors/handpieces by ultrasonic treatment and do not
immerse them in any liquids. To avoid the risk of corrosion/malfunctioning, allow any fluid that has entered the product to drain out immediately. ► If necessary, apply ultrasonic cleaning to tools: – as an effective mechanical supplement to manual cleaning/disinfection – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/disinfection. – as an integrated mechanical support measure for mechanical cleaning/disinfection. – for additional cleaning of products with residues left after mechanical cleaning/disinfecting.
6.5.2
Validated cleaning and disinfection procedure
Validated procedure
Special features
Manual cleaning and wipe disinfection
■ When cleaning instruments with movable hinges, ensure that these
Reference
Chapter Manual cleaning/disinfecting and subare in an open position and, if applicable, move the joint while clean- section: ing. ■ Chapter Manual cleaning and wipe disinfecting ■ Drying phase: Use a lint-free cloth
■ Ensure that the product is positioned in such a way that water will not enter the product e.g. through coupling interfaces. (Immediately remove any fluid that entered the product inadvertently.) Mechanical alkaline cleaning and thermal disinfecting
■ Insert the product in its proper position in the Eccos holder. ■ Mechanical cleaning of the products is only permitted if a corresponding Eccos holder is used.
Chapter Mechanical cleaning/disinfecting and subsection:
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
9
Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
6.6
Manual cleaning/disinfecting
► Prior to manual disinfecting, allow water to drip off for a sufficient
length of time to prevent dilution of the disinfecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for
residues. ► Repeat the cleaning/disinfection process if necessary.
6.6.1
Manual cleaning and wipe disinfecting
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Cleaning
RT (cold)
-
-
D–W
-
II
Drying
RT
-
-
-
-
III
Wipe disinfection
-
>1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
IV
Final rinse
RT (cold)
0.5
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W: RT:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) Room temperature
Phase I ► Clean the product under running faucet water, using a suitable clean-
ing brush until all visible residues have been removed from the surfaces. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. Phase II ► Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air), see Validated cleaning and disinfection procedure. Phase III ► Wipe all surfaces of the product with a single-use disinfectant wipe.
Phase IV ► After the specified exposure time (at least 1 min), rinse the disinfected
surfaces under running FD water. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air), see Validated cleaning and disinfection procedure.
10
6.7
Mechanical cleaning/disinfecting
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals. 6.7.1
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: Drinking water FD–W: Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline
► Check visible surfaces for residues after mechanical cleaning/disinfect-
ing.
11
Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
6.8
Inspection, maintenance and checks
► Allow the product to cool down to room temperature. ► Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged. ► Spray through the Hi-Line motor/handpiece with Aesculap-STERILIT oil spray GB600 with adapter GB600820 until clear oil emerges from the product. ► Check the product for any damage, abnormal running noise, overheating or excessive vibration. ► Inspect tools for broken, damaged or blunt edges. ► Set aside the product if it is damaged. Note Lubricating the HiLAN XS pneumatic motors with the original Aesculap oil spray can lead to oil residues entering the compressed air tubes. These oil residues do not harm the materials of which the compressed air tubes are made.
6.9
Packaging
► Follow the instructions for use for the applied packaging and storage
systems (e.g. instructions for use TA009721 for Aesculap Eccos storage system). ► Insert the product in its proper position in the Eccos holder, or put it on a tray in such a way that the product is protected against damage. Ensure that all cutting edges are protected. ► Pack trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging will prevent a recontamination of the product.
6.10 Steam sterilization Note Regarding materials and temperatures, the HiLAN XS pneumatic motors have been designed for steam sterilization. Consequently, do not sterilize HiLAN XS pneumatic motors with hot air or ethylene oxide (EO). ► Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and faucets). ► Validated sterilization process – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
12
6.11 Sterilization for the US market
■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.
■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6, Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
6.12 Storage ► Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
7.
Maintenance
To ensure reliable operation, the product must be maintained at least once a year. For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
8.
Troubleshooting list
Malfunction
Detection
HiLAN XS pneumatic motor not running
Valves at the pressure gauge closed No compressed air
Open the valves at the pressure gauge
Gas cylinder empty Manometer set to "0"
Fill or change cylinder
No compressed air at the wall outlet
Open valve
Turn safety stop 18 to "0"
Turn safety stop 18 to “I”
Tubes not correctly connected
Fully connect the hose connector
Pedal can be easily pressed or can- Foot control defective not be pressed at all
Have product repaired by the manufacturer
Safety stop 18 not in position “I”
Turn safety stop 18 to “I”
Insufficient power
Operating pressure of the central compressed air supply too low (min. 6 bar)
Compressed air supply pressure too low
Remedy
Have compressed-air supply checked
Pressure reducer on gas cylinder is faulty
Have pressure reducer checked
Pressure on pressure reducer set too low
Set higher pressure
Tube too long
Insufficient cutting power of tool
Cause
High air resistance in the tube
Use shorter tubes
Tubes excessively buckled or squeezed
Do not buckle or squeeze tubes
Exit air does not flow freely
Unblock diffuser 12 or exit air channel
Motor does not run in spite of cor- HiLAN XS pneumatic motor faulty rect compressed air supply
Have product repaired by the manufacturer
HiLAN XS pneumatic motor heats up
Have product repaired by the manufacturer
HiLAN XS pneumatic motor runs jerkily
Have product repaired by the manufacturer
Serrated tool run in counterclockwise mode
HiLAN XS pneumatic motor running in counterclockwise mode
Use serrated tools only in clockwise mode
13
Aesculap® HiLAN XS HiLAN XS pneumatic motor GA740R/GA741R
9.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product.
WARNING ► For service and repairs, please contact your national B. Braun/
Aesculapp agency. Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95 -1601 Fax: +49 (7461) 14 -939 E-Mail: [email protected] Or in the US: Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 USA Aesculap Repair Hotline Phone: +1 (800) 214 -3392 Fax: +1 (314) 895 -4420 Other service addresses can be obtained from the address indicated above.
14
10. Accessories/Spare parts Art. no.
Designation
GB600
STERILIT Power Systems oil spray
GB600820
STERILIT Power Systems adapter
GA059
Special oil
GA530
HiLAN irrigation pump
GA087 (5 m)
Extension tube DIN coupling
GA194 (3 m) GA195 (5 m)
Extension tube Aesculap-Dräger system
GA461 (3 m) GA465 (5 m)
Connecting tube DIN coupling
GA464 (3 m) GA466 (5 m)
Connecting tube Aesculap-Dräger system
GA468 (5 m)
Connecting tube Aesculap-Schrader system
GA508 R
Angled tube adapter
GA513R (3 m) GA523R (2.3 m) GA505R (5 m) GA506R (5 m) GA507R (5 m)
Motor tube
GA521
Foot control
GB473
Eccos storage mounting set
GB564 R
Storage aid (for motor tube)
GB577R
Storage aid (for HiLAN XS pneumatic motor with angled tube adapter)
GB610R
Storage aid (for HiLAN XS pneumatic motor)
JF223R (64 mm high)
Tray
JK400, JK401
Containers
11. Technical data
12. Disposal
GA740R Operating pressure
8 ± 2 bar
Note The user institution is obliged to process the product before its disposal, see Validated reprocessing procedure.
Max. operating pressure
10 bar
► Detailed information concerning the disposal of the product is avail-
Max. motor speed
110 000 1/min
Air consumption
approx. 180 l/min
Max. permissible backup in the exhaust air
0.7 bar
Noise
<70 dB(A) at 1 m distance
Weight
approx. 64 g
Dimensions
∅ 16 mm (∅ 20.5 mm) x 65 mm
GA741R Operating pressure
8 ± 2 bar
Max. operating pressure
10 bar
Max. motor speed
110 000 1/min
Air consumption
approx. 180 l/min
Max. permissible backup in the exhaust air
0.7 bar
Noise
<70 dB(A) at 1 m distance
Weight
approx. 67 g
Dimensions
∅ 16 mm (∅ 20.5 mm) x 65 mm
able through your national B. Braun/Aesculap agency, see Technical Service.
13. Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA
11.1 Ambient conditions Operation
Storage and transport
Temperature
Relative humidity
Atmospheric pressure
15
- DIR 93/42/EEC
Technical alterations reserved