Instructions for use/Technical description MACSTL torque wrench FW331R Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapImplantsystems.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-866-229-3002. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung MACSTL Drehmomentschlüssel FW331R Mode d’emploi/Description technique Clé dynamométrique MACSTL FW331R Instrucciones de manejo/Descripción técnica Llave dinamométrica MACSTL FW331R Istruzioni per l’uso/Descrizione tecnica Chiave dinamometrica MACSTL FW331R Instruções de utilização/Descrição técnica Chave dinamométrica MACSTL FW331R Gebruiksaanwijzing/Technische beschrijving MACSTL momentsleutel FW331R Brugsanvisning/Teknisk beskrivelse MACSTL-momentnøgle FW331R Bruksanvisning/Teknisk beskrivning MACSTL momentnyckel FW331R Käyttöohje/Tekninen kuvaus MACSTL -momenttiavain FW331R Lietošanas instrukcijas/tehniskais apraksts MACSTL griezes momenta uzgriežņu atslēga FW331R Naudojimo instrukcija/techninis aprašas MACSTL dinamometrinis veržliaraktis FW331R Инструкция по примению/Техническое описание Динамометрический ключ MACSTL FW331R Návod k použití/Technický popis Momentový klíč MACSTL FW331R Instrukcja użytkowania/Opis techniczny Klucz dynamometryczny MACSTL FW331R Návod na použitie/Technický opis MACSTL momentový kľúč FW331R Használati útmutató/Műszaki leírás MACSTL nyomatékkulcs FW331R Navodila za uporabo/Tehnični opis Navorni ključ MACSTL FW331R Upute za uporabu/Tehnički opis Momentni ključ MACSTL FW331R Manual de utilizare/Descriere tehnică Cheie dinamometrică MACSTL FW331R Упътване за употреба/Техническо описание Динамометричен гаечен ключ MACSTL FW331R Kullanım Kılavuzu/Teknik açiklama MACSTL tork anahtarı FW331R Οδηγίες χρήσης/Τεχνική περιγραφή Δυναμομετρικό κλειδί σύσφιξης MACSTL FW331R
1 2
3 4
5 7
8
6
10
9
11
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap® – a B. Braun brand TA010383
2020-07
V6
Change No. 62719
en
2.2.3 ®
Aesculap MACSTL torque wrench FW331R Legend 1 Jaw part SW 13 mm 2 Wrench 3 Release lever 4 Casing 5 Grip 6 Calibration point 7 Jaw part SW 11 mm 8 Insertion instrument handle 9 Socket wrench for locknuts 10 Ratchet handle for insertion instrument 11 Screwdriver for clamping screws
1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
2.3
Designation
FW331R
MACSTL torque wrench
Application
The wrench 2 is equipped with two different jaw parts: ■ Jaw part SW 13 mm 1 for fixing locknuts. ■ Jaw part SW 11 mm 7 for fixing clamping screws. WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product. WARNING Risk of injury/correction loss by using an unchecked break-away torque wrench! ► Prior to each use, check whether the next service date has been reached. ► Do not continue to use torque wrenches with an expired service date. ► Send torque wrenches with an expired service date to technical service for checking; see the chapter on Maintenance. ► When applying the torque wrench, be sure that it is always oriented with the marking (arrow) facing upwards.
2.3.1
Fixing locknuts
► Apply jaw part SW 13 mm 1 onto the outer hexagon of the socket wrench 9 for locknuts. ► When tightening the locknut, exert counterforce on the handle 8 of the insertion instrument to prevent the
torque from being transferred to the patient and implant.
These instructions for use apply for the following products: Art. no.
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3.2
Fixing clamping screws
► Apply jaw part SW 11 mm 7 onto the outer hexagon of the screwdriver for clamping screws 11. ► Activate the ratchet on the ratchet handle 10 of the insertion instrument by rotating it to the right, see instruc-
tions for use for the screwdriver for clamping screws (TA010385). ► To ensure the torque is transferred: Do not exert counterforce on the insertion instrument ratchet handle 10
Article-specific instructions for use as well as information on material compatibility and lifetime are available online at eIFU.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows:
when tightening clamping screws.
3.
Validated reprocessing procedure
3.1
General safety instructions
WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result.
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
2.
Clinical use
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
2.1
Areas of use and limitations of use
2.1.1
Intended use
The MACSTL torque wrench is used for fixing MACSTL locknuts and clamping screws with a defined torque. ■ 15±1.5 Nm for locknuts ■ 10±1.5 Nm for clamping screws
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Contraindications
No known contraindications.
2.2
Safety information
2.2.1
Clinical user
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Product
Product-specific safety information WARNING Risk of injury caused by incorrect operation of the product! ► Attend appropriate product training before using the product. ► For information about product training, please contact your national B. Braun/Aesculap agency. The application of the MACSTL spinal system requires precise knowledge about spine stabilization and the biomechanical situation at the spine. Operative use of the MACSTL spinal system is described in detail in the relevant O.R. manual. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Prior to first application, thoroughly read the O.R. manual for the MACSTL spinal system. ► Inform patients of the advantages and disadvantages of using the MACSTL spinal system. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts.
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
3.4
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Disassemble the product prior to cleaning, see Disassembly.
3.6
Disassembly
Note The grip 5 must be removed for cleaning, disinfection and sterilization. If only the wrench part 2 needs changing, the grip 5 only needs to be unscrewed. ► Unscrew and detach grip 5 in a counterclockwise direction.
This releases the ratchet mechanism on the wrench 2. ► Apply slight pressure on the protruding end of the wrench 2.
The wrench 2 disengages downwards. ► Open the release lever 3. To do this, rotate the release lever 90°. ► Pull the release lever 3 out to the stop. ► Pull the wrench 2 forward, away from the ratchet mechanism and out of the casing 4.
3.7
Cleaning/Disinfection
3.9
3.7.1
Product-specific safety information on the reprocessing method
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Following the manufacturer's instructions, use cleaning and disinfecting agents – that are approved for (e.g. aluminum, plastics, high-grade steel), – that do not attack softeners (e.g., in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C. For products with plasma layers (e.g. Noir instruments), the layer is attacked or removed when special cleaning procedures with the addition of oxidizing chemicals (e.g. hydrogen peroxide H2O2) are used. ► Do not use oxidizing chemicals for cleaning.
Mechanical cleaning/disinfection with manual pre-cleaning
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.9.1
Manual pre-cleaning with ultrasound and brush
Phase
Step
D [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
► Use suitable cleaning/disinfecting agents if the product is disposed of in a wet condition. To prevent foaming and
degradation of the efficacy of the process chemicals: prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water
3.7.2
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with ultrasound and immersion disinfection
■ Suitable cleaning brush ■ Disposable syringe 20 ml ■ When cleaning instruments
Chapter Manual cleaning/disinfection and subsection:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
■ Drying phase: Use a lint-free cloth or medical compressed air Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
■ Suitable cleaning brush ■ Disposable syringe 20 ml ■ Place the product on a tray that
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
is suitable for cleaning (avoid rinsing blind spots).
with ultrasound and brush
■ Place the product on the tray
■ Chapter Manual pre-cleaning ■ Chapter Mechanical alkaline
with all product links and joints open.
3.8
cleaning and thermal disinfecting
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.9.2
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Manual cleaning/disinfection
Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
3.8.1
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
– pH = 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD–W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
IV
Final rinse
RT (cold)
1
-
FD-W
-
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline
V
Drying
RT
-
-
-
-
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
D–W: FD–W:
3.10 Inspection ► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.10.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the products with long, slim shapes (in particular rotating instruments) for deformities. ► Check the product for damage to the spiral element. ► Check the cutting edges for continuity, sharpness, nicks and other damage. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.10.2 Functional test CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.11 Assembly ► Insert wrench 2 in casing 4 in such a way that the release lever 3 engages in the axial hole in the wrench 2. ► Push the release lever 3 in open position into the casing 4 to the stop. ► Close the release lever 3. To do this, rotate the release lever 90°. ► Turn the wrench 2 so that it engages in the casing 4. ► Unscrew the grip 5 by rotating it clockwise to the stop.
3.12 Packaging ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.13 Steam sterilization Note The product may only be sterilized in disassembled condition. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Disassemble the product – Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.14 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. WARNING Risk of injury due to sharp-edged and/or pointed products! ► When disposing of or recycling the product, ensure that the packaging prevents injury by the product. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA010383