Instructions for use/Technical description MIDCAB Retractor set Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusaifus.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung MIDCAB Retraktor-Set Mode d’emploi/Description technique Écarteur MIDCAB Instrucciones de manejo/Descripción técnica Kit de retractores MIDCAB Istruzioni per l’uso/Descrizione tecnica Set di retrattori MIDCAB Instruções de utilização/Descrição técnica Conjunto de retrator MIDCAB Gebruiksaanwijzing/Technische beschrijving MIDCAB-retractorset Brugsanvisning/Teknisk beskrivelse MIDCAB Retraktor-sæt Bruksanvisning/Teknisk beskrivning MIDCAB retraktor-set Käyttöohje/Tekninen kuvaus MIDCAB retraktorisetti Lietošanas instrukcijas/tehniskais apraksts MIDCAB retraktora komplekts Naudojimo instrukcija/techninis aprašas MIDCAB retraktorių rinkinys Инструкция по примению/Техническое описание Ретракционная система MIDCAB (для малоинвазивного прямого шунтирования коронарной артерии) Návod k použití/Technický popis Sada retraktoru MIDCAB Instrukcja użytkowania/Opis techniczny Zestaw retraktora MIDCAB Návod na použitie/Technický opis Súprava retraktora MIDCAB Használati útmutató/Műszaki leírás MIDCAB retraktor-készlet Navodila za uporabo/Tehnični opis MIDCAB Komplet retraktorja Upute za uporabu/Tehnički opis MIDCAB set retraktora Manual de utilizare/Descriere tehnică Set retractor MIDCAB Упътване за употреба/Техническо описание MIDCAB комплект ретрактори Kullanım Kılavuzu/Teknik açiklama MIDCAB retraktör seti Οδηγίες χρήσης/Τεχνική περιγραφή MIDCAB Σετ διαστολέα
11
D
6 5
12 C 4
B
1
3 A
1
2 2
B
3 9 8
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com
6
AESCULAP® – a B. Braun brand TA013912
2021-03
D
Change No. 62817
12 13
C
C 10
7
4
5
en ®
AESCULAP MIDCAB Retractor set Legend
Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Product-specific safety information
A Retractor 1 Movable Retractor Arm 2 Ratchet bar 3 Crank 4 Thread position B Double blade 5 Ball holder C Stabilizer 6 Handle 7 Suction foot 8 LUER cone 9 Suction hole 10 Shaft D Adapter 11 Large thumbscrew 12 Small thumbscrew 13 Ball joint
► Check compatibility with LUER-Konus before clinical use.
1.
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
2.2.3
2.3
Art. no.
Designation
FC520R
Retractor Double Joint Only
FC521R
MIDCAB adapter for FC522R
FC522R
MIDCAB Stabilizer
FC525R
MIDCAB lock complete
FC530R to FC534R
Double blade
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
3.
Validated reprocessing procedure
3.1
General safety information
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2 Note The applicable CE mark for the product can be found on the label or packaging of the product. ► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at
eifu.bbraun.com
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2.
Clinical use
Note The short summary of safety and clinical performance of the product is available in the European Database for Medical Products (EUDAMED).
2.1
Areas of use and limitations of use
2.1.1
Intended use
The MIDCAB Retractor is used in cardiac surgery to provide access to the heart through an intercostal space. The retractor arms are equipped with joints to better adjust them to the patients’ anatomy. Blades of five different lengths are available. A ball-type quick-fastener is used to attach the double blades to the retractor. A stabilizer can be mounted at any of the several threads that are distributed over the entire surface of the retractor. MIDCAB Stabilizer FC522R is used for the partial immobilization of the target region around the coronary artery.
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Absolute contraindications
Do not position MIDCAB Stabilizer FC522R directly over coronary arteries, over aneurysmatic tissue or over infected tissue. No absolute contraindications are known for the remaining products.
2.1.4
Application
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Scope
These instructions for use apply for the following products:
1.2
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
Relative contraindications
The following conditions, individual or combined, can lead to delayed healing or compromise the success of the operation: ■ Medical or surgical conditions (e.g. comorbidities) which could hinder the success of the operation. In the presence of relative contraindications, the user decides individually regarding the use of the product.
2.2
Safety information
2.2.1
Clinical user
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located.
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
3.4
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Disassemble the product prior to cleaning, see Disassembly. ► Open up product with hinges.
3.6
Disassembly
► Loosen the large thumb screw 11 and remove the adapter D with the mounted stabilizer C from the retractor A,
see Fig. 1. ► Slightly loosen the small thumb screw 12 and pull the stabilizer C from the ball joint 13, see Fig. 5. ► Press onto ball holder 5 and slide double blades B out of the retractor A, see Fig. 1. Use the forceps BV399R to change the double blades. ► Turn the crank 3 and remove the moving retractor arm 1 from the ratchet bar 2, see Fig. 2.
3.7
Cleaning/Disinfection
3.7.1
Product-specific safety information on the reprocessing method
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning agents and disinfectants in accordance with the manufacturer's instructions. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C. ► Use suitable cleaning/disinfecting agents if the product is disposed of in a wet condition. To prevent foaming and
degradation of the efficacy of the process chemicals: prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water.
3.7.2
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with immersion disinfection
■ Use a cleaning brush ■ Use a disposable syringe ■ When cleaning instruments
Chapter Manual cleaning/disinfection and subsection:
■ FC520R ■ FC522R ■ FC530R to FC534
■ Chapter Manual cleaning with immersion disinfection
with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
■ Drying phase: Use a lint-free
Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.8.2
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
cloth or medical compressed air Manual cleaning with ultrasound and immersion disinfection
■ FC521R
■ Use a cleaning brush ■ Use a disposable syringe ■ When cleaning instruments
Chapter Manual cleaning/disinfection and subsection:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
■ Drying phase: Use a lint-free cloth or medical compressed air Mechanical alkaline cleaning and thermal disinfection
■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
■ FC530R to FC534R
Chapter Mechanical cleaning/disinfection and subsection:
■ Chapter Mechanical alkaline
■ Connect components with
cleaning and thermal disinfection
■ Place the product on the tray
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
with all product links and joints open.
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
lumens and channels directly to the rinsing port of the injector carriage.
Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ FC520R ■ FC521R ■ FC522R
■ Use a cleaning brush ■ Use disposable syringe ■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
■ Connect components with
infection procedure. Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
■ Chapter Manual pre-cleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
lumens and channels directly to the rinsing port of the injector carriage.
Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully.
■ Place the product on the tray with all product links and joints open.
3.8
D–W: FD–W:
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
3.8.1
Manual cleaning with immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
IV
Final rinse
RT (cold)
1
-
FD-W
-
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
V
Drying
RT
-
-
-
-
3.9.1
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.9
Mechanical cleaning/disinfection
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened.
Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Pre-rinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.10 Mechanical cleaning/disinfection with manual pre-cleaning
3.12 Assembly
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Retractor assembly ► Push movable retractor arm 1 onto the ratchet bar 2 of the retractor A and insert the crank 3 into the large drilled hole, see Fig. 2. ► Set the desired spacing between the two retractor arms: Turn the crank 3, see Fig. 2. ► Insert suitable double blades into the ball holder 5 of the retractor A, see Fig. 1. The retractor is now ready for use.
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.10.1 Manual pre-cleaning with a brush Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Installation stabilizer If required, a suction foot 7 can be fitted to the Stabilizer C, with which the LUER-Konus 8 can be used to permanently or temporarily suction blood and/or other body fluids. ► Push the suction foot 7 all the way onto the stabilizer C, see Fig. 5. Installation of stabilizer at adapter ► Push the stabilizer C shaft 10 first into the ball joint 13 of the adapter D. ► Use the small thumb screw 12 to lightly attach the stabilizer C at the adapter D to prevent it from sliding out of the ball joint 13, see Fig. 5. Installation of adapter with stabilizer at retractor ► Use the large thumb screw 11 to position, orient and install the adapter D at the desired thread position 4. ► Grasp stabilizer C with one hand holding the handle 6. ► Use the other hand to loosen the small thumb screw 12. The stabilizer C is then articulated again. ► Place the stabilizer C in the desired position, orient it properly and fasten it using the small thumb screw 12, see Fig. 1.
3.13 Packaging ► Store products with ratchet locks fully opened or locked no further than in the first notch. ► Place the product in its dedicated holder or in a suitable tray. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.14 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
Machine type: single-chamber cleaning/disinfection device without ultrasound
– Steam sterilization in fractionated vacuum process – Steam sterilizer in accordance with DIN EN 285 and validated in accordance with DIN EN ISO 17665 – Sterilization in fractionated vacuum process at 134 °C, holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
3.15 Storage
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
3.10.2 Mechanical alkaline cleaning and thermal disinfecting
► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4. – pH ~ 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.11 Inspection ► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.11.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely scratched and fractured components. ► Check the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Technical service.
3.11.2 Functional test CAUTION Damage (metal cold welding / friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA013912