Aesculap® 1
Aesculap Endoscopic Technology
Instructions for use/Technical description Monopolar 3.5-mm hook electrode Gebrauchsanweisung/Technische Beschreibung Monopolare 3,5-mm-Hakenelektrode Mode d’emploi/Description technique Electrode monopolaire à crochet 3,5 mm Instrucciones de manejo/Descripción técnica Electrodo monopolar de gancho de 3,5 mm Istruzioni per l’uso/Descrizione tecnica Elettrodo a uncino monopolare da 3,5 mm Instruções de utilização/Descrição técnica Eléctrodo de gancho monopolar de 3,5 mm Gebruiksaanwijzing/Technische beschrijving Monopolaire 3,5mm-haakelektrode Bruksanvisning/Teknisk beskrivning Monopolär 3,5 mm hakelektrod Инструкция по примению/Техническое описание Однополярный 3,5 мм электрод-крючок Návod k použití/Technický popis Monopolární 3,5mm elektroda ve tvaru háčku Instrukcja użytkowania/Opis techniczny Jednobiegunowa elektroda zagięta 3,5 mm Návod na použivanie/Technický opis Monopolárna 3,5 mm háková elektróda Kullanım Kılavuzu/Teknik açiklama Monopolar 3,5-mm-çengelli elektrot 사용 설명서 / 기술 설명 재사용가능손조절식전기수술기용전극 (RU Monopolar, 형명개별기재 )
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap – a B. Braun company TA-Nr. 013327
09/14
- DIR 93/42/EEC Technical alterations reserved
V6
Änd.-Nr. 50991
2
Fig. 1
3
Fig. 2
Disassembling ► Pull out hook electrode 1 from electrode handle 3.
Aesculap® Monopolar 3.5-mm hook electrode
Assembling ► Fully insert hook electrode 1 with hexagon 2 in electrode handle 3.
The anti-twist guard at hexagon 2 is now effective, see Fig. 2.
Legend 1 Hook electrode 2 Hexagon 3 Electrode handpiece
Validated reprocessing procedure
Symbols on product and packages
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
General safety instructions
Caution, general warning symbol Caution, see documentation supplied with the product
Date of manufacture
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation.
Intended use The monopolar 3.5-mm hook electrode is used in combination with an electrode handle in endoscopic procedures. It serves to prepare, dissect and coagulate tissue by means of monopolar HF current.
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician!
WARNING
Risk of injury caused by incorrect operation of the product! ► Attend appropriate product training before using the product. ► For information about product training, please contact your national B. Braun/ Aesculap agency.
► Ensure that the product and its accessories are operated and used only by persons with the requisite training,
knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial
sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts. ► Do not use products that are in open or damaged sterile packaging. ► To avoid damage to the working end: Carefully insert the product through the working channel (e.g. trocar).
Risk of injury from ignition or explosion of flammable gases! Sparks may occur when using the HF device as directed. ► Observe the safety guidelines in the instructions for use of the HF device. WARNING
Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Disassembling the product before carrying out the reprocessing procedure ► Disassemble the product immediately after use, as described in the respective instructions for use.
WARNING
Thermal injuries to patients/users due to insufficient insulation of leads in active accessories! ► Adjust the HF device to an appropriate setting to ensure that the peak output voltage does match or not exceed the accessory voltage rating specified for the product.
The manufacturer has tested the product and verified that its insulation can withstand 20 reprocessing cycles. In clinical practice, the service life will depend on the individual intraoperative usage and the hospital's specific reprocessing conditions. ► Adjust the HF power output to the intended surgical intervention. Take into account clinical experience or reference values. ► Select the lowest possible HF power output. ► Keep the product’s contact surfaces clean during surgery. Remove encrusted tissue residues or body fluids with a moistened swab. The plug end of the product is fitted with the following connector: Pin 4 mm with hexagon anti-twist guard Refer to our brochures to find a compatible cable. The accessory voltage rating of the product is 1 500 Vp. The accessory voltage rating must exceed or match the peak output voltage with which the product is operated in combination with a suitable HF device at an appropriate operating mode/setting (see IEC/DIN EN 60601-2-2). To avoid HF burns: ► Always keep the working end of the product in the user’s field of vision whenever the HF power is activated. ► Prior to activating the HF device, check that the working end of the product is not touching any electrically conductive accessories. ► Prior to each use, visually inspect the product for: damage or surface changes to the insulation. ► Never place the product on or next to the patient. ► When using accessories for endoscopy or laparoscopy, deactivate the automatic switch-on mode of the HF device. ► Follow the instructions for use of the HF device.
Safe operation
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Preparation before cleaning ► Dismantle the product prior to cleaning, see Disassembling.
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – be approved for plastic material and high-grade steel, ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 94 °C.
Immersion treatment in a 3 % H2O2 solution for approx. 5 minutes is a particularly effective and gentle method to dissolve encrustations from HF instruments. Subsequently, the debris can be removed by hand, with a medium-hard brush and or in an ultrasonic bath. This is followed by the conventional reprocessing steps. ► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam formation and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water.
Validated cleaning and disinfection procedure Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product.
WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING
Preparations at the place of use
Validated procedure
Specific requirements
Reference
Manual cleaning with immersion disinfection
■ Drying phase: Use a lint-free
Chapter Manual cleaning/disinfection and sub-chapter:
cloth or medical compressed air
■ Chapter Manual cleaning with immersion disinfection
Mechanical alkaline cleaning and thermal disinfection
■ Place the instrument in a tray that is suitable for cleaning (avoiding rinsing blind spots).
Chapter Mechanical cleaning/disinfecting and sub-chapter:
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
Manual cleaning/disinfection
Inspection, maintenance and checks
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational,
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Manual cleaning with immersion disinfection Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
and free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Technical Service. ► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
Packaging ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in sterile Aesculap containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization
IV
Final rinse
RT (cold)
1
-
FD-W
-
Note The product may only be sterilized when dismantled.
V
Drying
RT
-
-
-
-
► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
► Validated sterilization process
– Disassemble the instrument – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6 , Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Mechanical cleaning/disinfecting Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11*
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
Note Accessories and spare parts under http://www.endoscopy-catalog.com.
V
Drying
-
-
-
According to the program for cleaning and disinfection device
Disposal
Accessories/Spare parts
► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 013327