Instructions for use/Technical description S4® Element pedicle screw system instruments for minimally invasive surgery (FW753R, FW759R, FW760R, FW765R) Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapimplantsystemsifus.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-866-229-3002. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung S4® Element Pedikelschraubensystem–Instrumentarium für die minimal-invasive Chirurgie (FW753R, FW759R, FW760R, FW765R) Mode d’emploi/Description technique Instrument pour système de vis pédiculaires S4® Element pour chirurgie mini-invasive (FW753R, FW759R, FW760R, FW765R) Instrucciones de manejo/Descripción técnica Instrumental del sistema de tornillos pediculares S4® Element para la cirugía mínimamente invasiva (FW753R, FW759R, FW760R, FW765R) Istruzioni per l’uso/Descrizione tecnica S4® Element Strumentario per sistema di viti transpeduncolari per chirurgia mini-invasiva (FW753R, FW759R, FW760R, FW765R) Instruções de utilização/Descrição técnica Sistema de parafusos pediculares S4® Element para cirurgia minimamente invasiva (FW753R, FW759R, FW760R, FW765R) Gebruiksaanwijzing/Technische beschrijving S4® element pedikelschroefsysteem-instrumentarium voor minimaal invasieve chirurgie (FW753R, FW759R, FW760R, FW765R) Brugsanvisning/Teknisk beskrivelse S4® Element Pedikelskruesystem - instrument til minimal-invasiv kirurgi (FW753R, FW759R, FW760R, FW765R) Bruksanvisning/Teknisk beskrivning S4® Element instrumentuppsättning för pedikelskruvsystem för minimalinvasiv kirurgi (FW753R, FW759R, FW760R, FW765R) Käyttöohje/Tekninen kuvaus S4® Element mini-invasiivisen kirurgian pedikkeliruuvijärjestelmän välineistö (FW753R, FW759R, FW760R, FW765R) Kasutusjuhend/Tehniline kirjeldus S4® Minimaalselt invasiivseks kirurgiaks mõeldud elementidest koosnev jätkekruvide süsteemi instrumentide kogum (FW753R, FW759R, FW760R, FW765R) Lietošanas instrukcijas/tehniskais apraksts S4® Element Pedikulu skrūvju sistēma - instrumenti minimāli invazīvai ķirurģijai (FW753R, FW759R, FW760R, FW765R) Naudojimo instrukcija/techninis aprašas Pedikulinių sraigtų sistemos instrumentai S4® Element, skirti minimalios invazijos operacijoms (FW753R, FW759R, FW760R, FW765R) Инструкция по примению/Техническое описание Комплект инструментов транспедикулярной винтовой системы S4® Element для минимально инвазивной хирургии (FW753R, FW759R, FW760R, FW765R) Návod k použití/Technický popis S4® Element – instrumentárium pro systém pedikulárních šroubů pro minimálně invazivní chirurgii (FW753R, FW759R, FW760R, FW765R) Instrukcja użytkowania/Opis techniczny Element systemu instrumentów do śrub pedikularnych S4® do chirurgii minimalnie inwazyjnej (FW753R, FW759R, FW760R, FW765R) Návod na použitie/Technický opis S4® Element pedikulárny systém pre minimálnu invazívnu chirurgiu (FW753R, FW759R, FW760R, FW765R) Használati útmutató/Műszaki leírás S4® Element pediculuscsavar-rendszer – Eszköztár minimálinvazív sebészethez (FW753R, FW759R, FW760R, FW765R) Navodila za uporabo/Tehnični opis Sistem vijakov za pedikel S4® Element – instrumenti za minimalno invazivno kirurgijo (FW753R, FW759R, FW760R, FW765R) Upute za uporabu/Tehnički opis Sustav pedikularnog vijka za S4® Element sustav – instrumenti za minimalno invazivnu kirurgiju (FW753R, FW759R, FW760R, FW765R) Manual de utilizare/Descriere tehnică Instrumentar pentru sistemul de șuruburi pediculare S4® Element pentru chirurgia minim-invazivă (FW753R, FW759R, FW760R, FW765R) Упътване за употреба/Техническо описание S4® Елемент система от винтове за педикула–инструментариум за минимално-инвазивна хирургия (FW753R, FW759R, FW760R, FW765R) Kullanım Kılavuzu/Teknik açiklama Minimal invazif cerrahi için S4® element pedikül vida sistemi aletleri (FW753R, FW759R, FW760R, FW765R) Οδηγίες χρήσης/Τεχνική περιγραφή Σύστημα διαυχενικών βιδών S4® Element – Εργαλεία για ελάχιστα επεμβατική χειρουργική (MIS) (FW753R, FW759R, FW760R, FW765R)
1 1
2
A
3 4 3
B
4
5
7 8 5
6 6
7 8
C
11 9 9 11 10 10
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com AESCULAP® – a B. Braun brand TA013859
2021-09
Change No. AE0060979
11 8
D
en
2.3 ®
AESCULAP S4® Element pedicle screw system instruments for minimally invasive surgery (FW753R, FW759R, FW760R, FW765R) Legend 1 S4 Element MIS hand-held rod insertion instrument (FW760R) 2 Inner shaft (FW760R) 3 S4 Element MIS rod length measuring instrument (FW759R) 4 Slider (FW759R) 5 S4 Element MIS compression instrument (FW765R) 6 Toothed rack (FW765R) 7 Ratchet unit (FW765R) 8 Wing bolt (FW753R/FW765R) 9 S4 Element MIS connector ferrule (FW753R) 10 Frame (FW753R) 11 Ratchet unit (FW753R)
1.
About this document
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
3.
Validated reprocessing procedure
3.1
General safety information
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this.
Note General risk factors associated with surgical procedures are not described in these instructions for use.
Note If there is no final sterilization, then a virucidal disinfectant must be used.
1.1
Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
Scope
These instructions for use apply for the following products: Art. no.
Designation
3.2
FW753R
S4 Element MIS connector ferrule
FW759R
S4 Element MIS rod length measuring instrument
FW760R
S4 Element MIS hand-held rod insertion instrument
FW765R
S4 Element MIS compression instrument
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
Note The applicable CE mark for the product can be found on the label or packaging of the product. ► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at
eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result.
General information
CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
3.3
2.
Clinical use
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best option to determine if a product is no longer functional, see Inspection.
2.1
Areas of use and limitations of use
2.1.1
Intended use
The instruments FW760R, FW759R, FW765R and FW753R are used for inserting and handling implants of the S4 series pedicle screw system in patients with spinal fusion surgery.
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications.
3.4
Reusable products
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Do not disassemble anchoring screws/bolts that permanently connect the various components. ► Disassemble the product prior to cleaning, see Disassembly.
For indications, see Intended use.
3.6
Disassembly
2.1.3
3.6.1
S4 Element MIS hand-held rod insertion instrument(FW760R)
Contraindications
No known contraindications.
► To clean, remove the inner sheath 2 from the main unit, see Fig. A.
2.2
Safety information
3.6.2
2.2.1
Clinical user
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Product specific safety information
► To avoid damage to the working end: Carefully insert the product through the working channel (e.g. trocar). ► Read, observe and store the OR Manual S4 Element MIS surgical technique O82002.
S4 Element MIS hand-held rod insertion instrument FW760R Note The product consists of metal alloys which contain cobalt.
2.2.3
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
S4 Element MIS rod length measuring instrument (FW759R)
► Pull slider 4 to remove it from the main part of the instrument, see Fig. B.
3.6.3
S4 Element MIS compression instrument (FW765R)
► Remove the ratchet unit 7 from the toothed rack 6. ► Remove the wing bolt 8 from the ratchet unit 7, see Fig. C.
3.6.4
S4 Element MIS connector ferrule (FW753R)
► Remove the ratchet unit 11 from the frame 10. ► Remove the wing bolt 8 from the ratchet unit 11, see Fig. D.
3.7
Cleaning/Disinfection
3.7.1
Product-specific safety information on the reprocessing method
Hazardous to patients! S4 Element MIS rod length measuring instrument FW759R may not be processed manually. ► The S4 Element MIS rod length measuring instrument FW759R may only be processed mechanically. Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning agents and disinfectants in accordance with the manufacturer's instructions. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
3.7.2
3.8.2
Manual cleaning with ultrasound and immersion disinfection
Validated procedure
Validated cleaning and disinfection procedure Specific requirements
Reference
Phase
Step
t [min]
Conc. [%]
Water quality
Chemical
Manual cleaning using immersion disinfection
■ Cleaning brush: e.g. TA011944,
T [°C/°F]
Chapter Manual cleaning/disinfection and subsection:
■ FW765R ■ FW753R ■ FW759R
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
■ Disposable syringe 20ml ■ Clean products having movable
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
TE654202, GK469200
■ Chapter Manual cleaning with immersion disinfection
hinges in the open position or while moving the joints.
■ Drying phase: Use a lint-free cloth or medical compressed air Manual cleaning using ultrasound and immersion disinfection
■ Cleaning brush: e.g. TA011944,
■ FW760R
■ Disposable syringe 20ml ■ Clean products having movable
TE654202, GK469200
Chapter Manual cleaning/disinfection and sub-sections:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
hinges in the open position or while moving the joints.
D–W: FD–W:
■ Drying: Use lint free cloth or medical compressed air for drying
■ Place product on screen basket
Alkaline machine cleaning and thermal disinfection
suitable for cleaning (make sure all areas will be reached by water jets).
■ FW759R
Chapter Machine cleaning/disinfection and sub-sections:
■ Chapter Mechanical alkaline cleaning and thermal disinfection
■ Open the links and joints before placing the product on the screen basket. Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ FW765R ■ FW753R ■ FW760R
■ Cleaning brush: e.g. TA011944, TE654202, GK469200
■ Disposable syringe 20ml ■ Place the product on a tray that is suitable for cleaning (make sure all areas will be reached by water jets).
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
■ Chapter Manual pre-cleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
■ Open the links and joints before placing the product on the screen basket.
3.8
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
3.8.1
Manual cleaning with immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
3.9
Machine cleaning/disinfection
II
Intermediate rinse
RT (cold)
1
-
D–W
-
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
3.9.1
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Pre-rinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
■ 0.5 % working solution
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
– pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.10 Mechanical cleaning/disinfection with manual pre-cleaning Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.14 Steam sterilization
3.10.1 Manual pre-cleaning with a brush Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Rinsing
RT (cold)
1
Note FW760R: The product may only be sterilized in its disassembled state. Note FW759R/FW765R: The product can be sterilized both in its disassembled and assembled state. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). II
-
D–W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
► Validated sterilization process
– Disassemble the product if required – Steam sterilization in fractionated vacuum process – Steam sterilizer in accordance with DINEN 285 and validated in accordance with DIN EN ISO 17665 – Sterilization in the fractionated vacuum process at 134°C, holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.15 Storage
Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.10.2 Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
CAUTION Modifications to medical equipment may result in the voiding of any guarantee/warranty claims and any approvals. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5. – pH ~ 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.11 Inspection ► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.11.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.11.2 Functional test CAUTION The product may become damaged (metal pitting/fretting corrosion) if not sufficiently oiled! ► Oil any moving parts (e.g. joints, spool parts and threaded rods) prior to the functional test using oil suitable for the applied sterilization process (e.g. STERILIT® I oil spray JG600 for steam sterilization or STERILIT® Idrip-feed lubricator JG598). ► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check rotating products (e.g. reusable drills and cutters) for bending and deformation. To do this, roll product,
for example, on a flat surface. ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.12 Assembly Assemble the instruments as in disassembly but in reverse order, see Disassembly.
3.13 Packaging ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
Technical service
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. TA013859