Instructions for use/Technical description Spine instruments Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapImplantsystems.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-866-229-3002. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Wirbelsäuleninstrumentarium Mode d’emploi/Description technique Instrumentation pour la colonne vertébrale Instrucciones de manejo/Descripción técnica Instrumental para la columna vertebral Istruzioni per l’uso/Descrizione tecnica Strumentario per interventi sulla colonna vertebrale Instruções de utilização/Descrição técnica Instrumentário para a coluna vertebral Gebruiksaanwijzing/Technische beschrijving Wervelkolominstrumentarium Brugsanvisning/Teknisk beskrivelse Rygsøjleinstrumenter Bruksanvisning/Teknisk beskrivning Instrumentuppsättning för ryggrad Käyttöohje/Tekninen kuvaus Selkärankainstrumentit Lietošanas instrukcijas/tehniskais apraksts Mugurkaula instrumenti Naudojimo instrukcija/techninis aprašas Prietaisų rinkinys stuburui Инструкция по примению/Техническое описание Инструментарий для позвоночника Návod k použití/Technický popis Sada páteřních nástrojů Instrukcja użytkowania/Opis techniczny Instrumentarium do zabiegów kręgosłupowych Návod na použitie/Technický opis Inštrumentárium pre chrbticu Használati útmutató/Műszaki leírás Gerincsebészeti műszerek Navodila za uporabo/Tehnični opis Hrbtenični instrumentarij Upute za uporabu/Tehnički opis Instrument za kralježnicu Manual de utilizare/Descriere tehnică Instrumente spinale Упътване за употреба/Техническо описание Инструментариум за гръбначен стълб Kullanım Kılavuzu/Teknik açiklama Omurga aletleri Οδηγίες χρήσης/Τεχνική περιγραφή Εργαλεία σπονδυλικής στήλης
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap® – a B. Braun brand TA014810
0482
2020-09
V6
Change No. 63431
en
3.2 ®
Aesculap Spine instruments 1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply to instruments having lumens or gouges, instruments that can be easily disassembled and instruments that cannot be taken apart, made entirely of robust materials (e.g. stainless steel, PEEK, silicone). Note The applicable CE mark for the product can be found on the label or packaging of the product. ► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at
eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result.
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
3.3 Reusable products ■ No maximum number of applications and processing cycles has been specified for the product. ■ The service life of the product is limited by damage, normal wear, the type and duration of usage, as well as the handling, storage and transportation of the product.
CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
■ Careful visual and functional testing prior to the next use is the best way to identify a malfunctioning product. 3.4
Preparations at the place of use
2.
Clinical use
2.1
Areas of use and limitations of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
2.1.1
Intended use
3.5
The spinal instrument set is used for applications as part of procedures such as implantation of implants for the stabilization of the spine or for the fusion ofvertebral bodies.
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Contraindications
Preparing for cleaning
► Disassemble product before cleaning.
3.6
Cleaning/Disinfection
3.6.1
Product-specific safety information on the reprocessing method
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning agents and disinfectants in accordance with the manufacturer's instructions. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
No known contraindications.
2.2
Safety information
► Do not use oxidizing chemicals (e.g.H2O2) on PVD coated products, as the former can cause bleaching or loss of
2.2.1
Clinical user
► For wet disposal, use suitable cleaning agents/disinfectants. To prevent foam from forming and reduced effec-
layers.
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user.
tiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly under running water ► Put jaw protecting cap on the product.
Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located.
cloth or medical compressed air
Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
3.6.2
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning using ultrasound and immersion disinfection
■ Suitable cleaning brush ■ 20 ml disposable syringe ■ Drying phase: Use a lint-free
Chapter Manual cleaning/disinfection and subsection:
Manual pre-cleaning with ultrasound and a brush followed by mechanical alkaline cleaning and thermal disinfection
is suitable for cleaning (avoid rinsing blind spots).
■ Connect components with
General safety information
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
with ultrasound and brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
cleaning purposes. placing the product on the screen basket.
3.7 3.1
■ Chapter Manual pre-cleaning
■ Open the links and joints before
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
Validated reprocessing procedure
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
■ Keep working ends open for
Application
3.
ultrasound and immersion disinfection
lumens and channels directly to the rinsing port of the injector carriage.
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
■ Suitable cleaning brush ■ 20 ml disposable syringe ■ Place the product on a tray that
■ Chapter Manual cleaning with
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
3.7.1
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
3.8.2
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
I
Ultrasonic cleaning
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Phase
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
RT (cold)
1
RT
-
IV
V
Final rinse
Drying
-
-
FD-W
-
-
-
D–W: FD–W:
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH ~ 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
D–W: FD–W:
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.8
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.8.1
Manual pre-cleaning with ultrasound and brush
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
RT (cold)
1
Rinsing
-
D–W
Inspection
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.9.1
Visual inspection
► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check products having long, narrow geometries (in particular rotating instruments) for deformations. ► Check the cutting edges for continuity, sharpness, nicks and other damage. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.9.2
Functional test
CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check rotating products (e.g. reusable drills and cutters) for bending and deformation. To do this, roll product,
Mechanical cleaning/disinfection with manual pre-cleaning
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
II
3.9
for example, on a flat surface. ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.10 Packaging ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.11 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets).
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh
► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.12 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4. ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. TA014810