Instructions for use/Technical description Tonsillotome handle and blades for tonsil knives Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Tonsillotomgriff und Aufsätze für Tonsillenmesser Mode d’emploi/Description technique Manche et lames pour couteau à amygdale Instrucciones de manejo/Descripción técnica Mango de amigdalótomo y hojas para bisturís para amígdalas Istruzioni per l’uso/Descrizione tecnica Impugnatura per tonsillotomo e lame per bisturi per tonsille Instruções de utilização/Descrição técnica Punho de tonsilótomo e lâminas para bisturi de amígdalas Gebruiksaanwijzing/Technische beschrijving Tonsillotoom handgreep en bladen voor tonsillenmessen Brugsanvisning/Teknisk beskrivelse Tonsillotom-håndtag og klinger til tonsilknive Bruksanvisning/Teknisk beskrivning Tonsillotomhandtag och blad för tonsillknivar Käyttöohje/Tekninen kuvaus Nielurisaveitsen kahva ja terät nielurisaveitsiin Lietošanas instrukcijas/tehniskais apraksts Tonsilotoma rokturis un mandeļu nažu asmeņi Naudojimo instrukcija/techninis aprašas Tonzilių šalinimo įrankio rankena ir ašmenys, skirti tonzilių peiliams Инструкция по примению/Техническое описание Рукоятка тонзиллотома и лезвия ножей для тонзиллэктомии Návod k použití/Technický popis Rukojeť tonsilotomu a čepele pro nože na mandle Instrukcja użytkowania/Opis techniczny Uchwyt tonsilotomu i ostrza do noży do migdałków Návod na použitie/Technický opis Rukoväť nástroja na tonzilotómiu a čepele nože na mandle Használati útmutató/Műszaki leírás Tonsillotom nyél és pengék mandulaműtétekhez használatos késekhez Navodila za uporabo/Tehnični opis Ročaj in rezila za tonzilotome Upute za uporabu/Tehnički opis Ručka za tonzilotom i oštrice za noževe za tonzile Manual de utilizare/Descriere tehnică Mâner de tonsilotom și lame pentru bisturie de tonsilectomie Упътване за употреба/Техническо описание Ръкохватка за тонзилотомия и ножове за сливици Kullanım Kılavuzu/Teknik açiklama Tonsil bıçaklar için tonsilotom tutamaçları ve bıçakları Οδηγίες χρήσης/Τεχνική περιγραφή Λαβή και λεπίδες αμυγδαλεκτομής για μαχαίρια αμυγδαλών
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap® – a B. Braun brand TA014577
2020-12
V6
Change No. 63998
7 8
9
9
A 3 10 2 1
3 D
B
C 4
6 7 8 1 2 9
D
E
9
en ®
Aesculap Tonsillotome handle and blades for tonsil knives Legend 1 Tonsil knife holder 2 Tonsil knife 3 Leaf spring 4 Lever 5 Thumb screw 6 Knurled head screw 7 Activation handle piece 8 Counter handle piece 9 Handle spring 10 Post of leaf spring
1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply for the following products: Art. no.
Designation
OM801R
Sluder-Ballenger blade for tonsil knives small
OM802R
Sluder-Ballenger blade for tonsil knives medium
OM803R
Sluder-Ballenger blade for tonsil knives large
OM806R
Sluder-Ballenger tonsillotome handle for OM801R to OM803R
Note The applicable CE mark for the product can be seen on the label or packaging of the product. ► For article-specific instructions for use as well as information on material compatibility and lifetime see B. Braun
eIFU at eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
3.
Validated reprocessing procedure
3.1
General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2
3.3
Clinical use
2.1
Areas of use and limitations of use
2.1.1
Intended use
Tonsillotomes are used in adenotomy and tonsillectomy. Blades for tonsil knives are applied to a tonsillotome handle.
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Contraindications
No known contraindications.
2.2
Safety information
2.2.1
Clinical user
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
3.4
2.
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to "AKI-Brochures", "Red brochure".
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Disassemble the product prior to cleaning, see Disassembly.
3.6
Disassembly
► Unscrew and remove thumb screw 5, see Fig. A. ► Remove tonsil knife 2 and tonsil knife holder 1 from the handle. ► Unscrew and remove knurled head screw 6 and separate the two handle pieces 7 and 8 from each other. ► Unlock leaf spring 3 from tonsil knife holder 1, see Fig. B. ► Turn leaf spring 3 until the post 10 is in line with the groove in tonsil knife 2, see Fig. C. Then, remove leaf
spring 3. ► Separate tonsil knife 2 from tonsil knife holder 1, see Fig. D.
3.7
Cleaning/Disinfection
3.7.1
Product-specific safety information on the reprocessing method
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Following the manufacturer's instructions, use cleaning and disinfecting agents that are approved for stainless steel. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C. ► Use suitable cleaning/disinfecting agents if the product is disposed of in a wet condition. To prevent foaming and
degradation of the efficacy of the process chemicals: prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water
3.7.2
3.9
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with immersion disinfection
■ Suitable cleaning brush ■ Disposable syringe 20 ml ■ Keep working ends open for
Chapter Manual cleaning/disinfection and subsection:
■ Chapter Manual cleaning with immersion disinfection
cleaning.
■ When cleaning instruments
■ Drying phase: Use a lint-free cloth or medical compressed air
■ Mount jaws protection on the product.
■ Place the product on a tray that
3.9.1
Mechanical alkaline cleaning and thermal disinfecting
Phase
Step
D [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH ~ 13 – <5 % anionic surfactant
Chapter Mechanical cleaning/disinfection and subsection:
■ Chapter Mechanical alkaline
■ working solution 0.5%
cleaning and thermal disinfecting
is suitable for cleaning (avoid rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
■ To flush the product: Use a flushing nozzle or flushing sleeve.
– pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
DW: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline
■ Keep working ends open for cleaning.
■ Place the product on the tray with all product links and joints open.
3.8
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals. Machine type: single-chamber cleaning/disinfecting machine without ultrasound
with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
Mechanical alkaline cleaning and thermal disinfection
Mechanical cleaning/disinfection
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.10 Inspection
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
3.8.1
Manual cleaning with immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.10.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. Water quality
Chemical
► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components.
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
► Check the product for missing or faded labels. ► Check the products with long, slim shapes (in particular rotating instruments) for deformities. ► Check the product for damage to the spiral element. ► Check the cutting edges for continuity, sharpness, nicks and other damage. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
D–W: Drinking water FD–W: Fully desalinated water (demineralized, microbiological, at least of drinking water quality) RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
3.10.2 Functional test CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check rotating products (e.g. reusable drills and cutters) for bends and deformities. To do this, roll the product
on an even surface.
Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe.
► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully.
► Screw thumb screw 5 into handle pieces. Do not fully tighten. ► Put tonsil knife 2 and tonsil knife holder 1 onto lever 4 of the handle piece 7. ► Press handle pieces 7 and 8 together and fixate tonsil knife 2 and tonsil knife holder 1 on handles.
Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.11 Assembly ► Slide tonsil knife 2 into the groove in tonsil knife holder 1, see Fig. D. ► Insert post 10 of leaf spring 3 through the groove in tonsil knife holder 1 and tonsil knife 2, see Fig. C. ► Turn leaf spring 3 until it locks into the recess in tonsil knife holder 1, see Fig. B. ► Put both handle pieces 7 and 8 together and completely screw knurled head screw 6 into handle pieces, see Fig.
E. Make sure, that both handle springs 9 are connected.
Note If tonsil knife 2 and tonsil knife holder 1 cannot be fixated, loosen thumb screw 5 and try again. ► Fully tighten thumb screw 5.
3.12 Packaging ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.13 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.14 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. WARNING Risk of injury due to sharp-edged and/or pointed products! ► When disposing of or recycling the product, ensure that the packaging prevents injury by the product. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA014577
2020-12