BBraun
Vascular clamps including application in the central circulatorysystem
70 Pages
Preview
Page 1
AESCULAP® en USA
de fr es it pt nl da sv fi lv lt ru
cs pl sk hu sl hr ro bg tr el
Instructions for use/Technical description Vascular clamps including application in the central circulatory system Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusaifus.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Gefäßklemmen einschließlich der Anwendung im zentralen Kreislaufsystem Mode d’emploi/Description technique Pinces vasculaires, y compris l'application dans le système circulatoire central Instrucciones de manejo/Descripción técnica Abrazaderas vasculares incluyendo la aplicación en el sistema circulatorio central Istruzioni per l’uso/Descrizione tecnica Morsetti vascolari, inclusa l'applicazione nel sistema circolatorio centrale Instruções de utilização/Descrição técnica Pinças hemostáticas que podem ser usadas no sistema circulatório central Gebruiksaanwijzing/Technische beschrijving Vaatklemmen inclusief toepassing in de centrale bloedsomloop Brugsanvisning/Teknisk beskrivelse Vaskulære klemmer, herunder anvendelse i det centrale kredsløbssystem Bruksanvisning/Teknisk beskrivning Kärlklämmor inklusive applicering i centrala cirkulationssystemet Käyttöohje/Tekninen kuvaus Verisuonipuristimet sekä niiden käyttö keskusverenkiertojärjestelmässä Lietošanas instrukcijas/tehniskais apraksts Asinsvadu skavas lietošanai centrālajā asinsrites sistēmā Naudojimo instrukcija/techninis aprašas Kraujagyslių spaustuvai, kuriuos galima naudoti ir centrinei kraujotakos sistemai Инструкция по примению/Техническое описание Сосудистые зажимы с возможностью применения в центральной системе кровообращения Návod k použití/Technický popis Cévní svorky včetně použití v centrální oběhové soustavě Instrukcja użytkowania/Opis techniczny Zaciski naczyniowe, w tym ich zastosowanie w centralnym układzie krążenia Návod na použitie/Technický opis Cievne svorky vrátane aplikácie v centrálnom obehovom systéme Használati útmutató/Műszaki leírás Érfogók, beleértve a központi keringési rendszerben történő alkalmazásukat is Navodila za uporabo/Tehnični opis Žilne prijemalke, namenjene tudi za uporabo na osrednjih obtočilih Upute za uporabu/Tehnički opis Vaskularne stezaljke, uključujući primjenu u središnjem krvožilnom sustavu Manual de utilizare/Descriere tehnică Cleme vasculare, inclusiv pentru utilizarea în sistemul circulator central Упътване за употреба/Техническо описание Съдови клампи, включващи употреба в областта на централната кръвоносна система Kullanım Kılavuzu/Teknik açiklama Merkezi dolaşım sisteminde uygulamayı da kapsayan vasküler kelepçeler Οδηγίες χρήσης/Τεχνική περιγραφή Αγγειακοί σφιγκτήρες συμπεριλαμβανομένης της εφαρμογής στο κεντρικό κυκλοφορικό σύστημα
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com AESCULAP® – a B. Braun brand TA015901
2021-12
Change No. AE0061393
en AESCULAP® Vascular clamps including application in the central circulatory system 1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply for vascular clamps including application in the central circulatory system. ► For article-specific instructions for use as well as information on material compatibility and lifetime see B. Braun eIFU at eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2.
Clinical use
3.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating precleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
Note The short summary of safety and clinical performance of the product is available in the European Database for Medical Products (EUDAMED).
3.4
2.1
Areas of use and limitations of use
► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
2.1.1
Intended use
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for
example.
Vascular clamps are surgical instruments for the temporary clamping of tissues and blood vessels, including the heart and the vessels of the central circulatory system.
3.5
Cleaning/Disinfection
3.5.1
Product-specific safety information on the reprocessing method
2.1.2
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Following the manufacturer's instructions, use cleaning and disinfecting agents that are approved for stainless steel. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Contraindications
No known contraindications.
3.5.2
2.2
Safety information
Validated procedure
Specific requirements
Reference
2.2.1
Clinical user
Manual cleaning with immersion disinfection
■ Suitable cleaning brush ■ Keep working ends open for cleaning. ■ Place the product on the tray with all pro-
Chapter Manual cleaning/disinfection and subsection:
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
■ All except FC295R, FC296R and FC297R
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
3.
Validated reprocessing procedure
3.1
General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
duct links and joints open.
■ When cleaning instruments with movable
■ Chapter Manual cleaning with immersion disinfection
hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
■ Drying phase: Use a lint-free cloth or medical compressed air Manual cleaning with ultrasound and immersion disinfection Valid for:
■ FC295R ■ FC296R ■ FC297R
■ Suitable cleaning brush ■ Keep working ends open for cleaning. ■ Place the product on the tray with all product links and joints open.
■ When cleaning instruments with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
Chapter Manual cleaning/disinfection and subsection:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
■ Drying phase: Use a lint-free cloth or
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
Validated cleaning and disinfection procedure
medical compressed air Mechanical alkaline cleaning and thermal disinfection
■ Place the product on a tray that is suit-
■ All except FC295R,
■ Keep working ends open for cleaning. ■ Place the product on the tray with all pro-
FC296R and FC297R
able for cleaning (avoid rinsing blind spots).
duct links and joints open. Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection Valid for:
■ FC295R ■ FC296R ■ FC297R
■ Suitable cleaning brush ■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
■ Keep working ends open for cleaning. ■ Place the product on the tray with all product links and joints open.
Chapter Mechanical cleaning/disinfecting and subsection:
■ Chapter Mechanical alkaline cleaning and thermal disinfection Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subchapter:
■ Chapter Manual precleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
3.6
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of
the disinfecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
3.6.1
Manual cleaning with immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.7
3.7.1
D–W: Drinking water FD–W: Fully desalinated water (demineralized, microbiological, at least of drinking water quality) RT: Room temperature *Recommended: BBraun Stabimed fresh
Mechanical cleaning/disinfecting
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfection
► Process the implant in its system storage device. ► Place the implants on a tray that is suitable for cleaning (avoiding rinsing blind spots and damage).
Machine type: single-chamber cleaning/disinfecting machine without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemistry/Note
I
Pre-rinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and
disinfection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.6.2
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenolfree, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
RT (cold)
5
III
Disinfection
2
D–W
Aldehyde-free, phenolfree, and QUAT-free concentrate, pH ~ 9*
– pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
D–W: Drinking water FD–W: Fully desalinated water (demineralized, microbiological, at least of drinking water quality) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.8
Mechanical cleaning/disinfection with manual pre-cleaning
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.8.1
Manual pre-cleaning with a brush
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and
disinfection procedure. IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and
disinfection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully.
Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.8.2
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD–W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.9
Inspection
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.9.1
Visual inspection
► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts,
recessed areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see
Technical service.
3.9.2
Functional test
CAUTION Damage (metal cold welding / friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical
service.
3.10 Packaging ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during stor-
age.
3.11 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces
(e.g., by opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.12 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled
area.
4.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA015901
2021-12
Change No. AE0061393