BBraun
Veress cannula
29 Pages
Preview
Page 1
Validated reprocessing procedure Aesculap® Veress cannula
General safety information
Legend
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD, or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
1 2 3 4 5 6 7 8
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for reprocessing.
Outer shaft Compression spring Inner shaft Spring cap LL-connector Lever Plug, maintenance free (must not be oiled) Marking on cock plug and outer shaft
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note It should be noted that successful reprocessing of this medical device can only be guaranteed following prior validation of the reprocessing method. The operator/reprocessor is responsible for this. The recommended chemical was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used.
Symbols on product and packages
Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at https://extranet.bbraun.com
Caution, general warning symbol Caution, see documentation supplied with the product
General notes
Applicable to ► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
at https://extranet.bbraun.com
Intended use The Veress cannula is used to puncture the abdominal wall at the start of a laparoscopic operation and to insufflate CO2 gas into the abdominal cavity.
Available sizes Art. no.
Designation
PG003
Veress insufflation needle
Length
Diameter
100 mm
2mm
PG008
120 mm
PG011
150 mm
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Carry out sterile processing of the product prior to every use. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts.
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore, no more than 6 hours should be left between use and sterile processing, pre-clean temperatures >45 °C liable to fusing should not be employed and disinfectants liable to fusing (aldehyde or alcohol-based) should not be used. Excessive neutralizing agents or basic cleaners may result in a chemical attack and/or fading and the laser marking becoming unreadable either visually or by machine in case of non stainless steel. Residues containing chlorine or chlorides, e.g., in surgical residues, medicines, saline solutions, and in the service water used for cleaning, disinfection, and sterilization, will cause corrosion damage (pitting, stress corrosion) and result in damage to the products in case of non stainless steel. These must be removed by rinsing thoroughly with demineralized water and then drying. Perform additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for reprocessing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes in materials, e.g., fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging, or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surface and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, section Publications, Red Brochure – Proper maintenance of instruments.
Disassembling the product before carrying out the reprocessing procedure ► Disassemble the product immediately after use, as described in the respective instructions for use.
Preparation at the place of use ► If applicable, rinse surfaces that are not accessible to visible inspection (preferably with demineralized water),
using a disposable syringe, for example. ► Remove any visible surgical residues as much as possible with a damp, lint-free cloth. ► Place the dry product in a sealed waste container and forward it on for cleaning and disinfection within 6 hours.
Preparation before cleaning ► Dismantle the product prior to cleaning, see Disassembling. ► Open taps (if present).
Cleaning/disinfection
Safe operation
Product-specific safety notes on the reprocessing procedure Risk of injury due to incorrect handling! The Veress cannula can damage organs and tissue. ► Perform the incision with the Veress cannula with care.
WARNING
CAUTION
CAUTION Risk of injury due to loss of gas! ► Remove the instrument only after the procedure is complete. ► Prior to puncturing the abdominal wall, make certain that the lever of the Veress cannula is closed.
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for use, for example, on aluminum, plastic materials, and high-grade steel, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 96 °C.
► Use the LL-connector 5 to connect the instrument with the gas line of the insufflator. ► Puncture the abdominal wall at the navel with the Veress cannula.
► Do not use oxidizing chemicals (e.g. H2O2), which could cause bleaching/layer loss of the product. ► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam forma-
When puncturing the abdominal wall, the inner shaft 3 slides backward into the outer shaft 1 and the sharp tip of the outer shaft 1 is exposed. When the Veress cannula has penetrated the abdominal cavity, the inner shaft 3 springs forward again via the compression spring 2 and covers the sharp tip of the outer shaft 1. ► Open lever 6.
tion and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water. ► Carry out ultrasound cleaning: – as an effective mechanical supplement to manual cleaning/disinfecting. – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/disinfecting. – as an integrated mechanical support measure for mechanical cleaning/disinfecting. – for additional cleaning of products with residues left after mechanical cleaning/disinfecting.
Disassembling ► Unscrew outer shaft 1 and remove from inner shaft 3, see Fig. A. ► Unscrew spring cap 4 and remove plug 7, see Fig. B.
Manual cleaning/disinfection ► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
Assembling Note Cock plug 7 and outer shaft 1 are each identified by a marking 8, see Fig. D. The marking indicates that only the matched cock plug may be used. ► Apply outer shaft 1 onto inner shaft 3 and screw tight, see Fig. A. ► Insert the plug 7 into the stop cock so that the pin on the plug 7 sits in the recess in the stop cock, see Fig. C. ► Screw together the plug 7 and spring cap 4.
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Inspection, maintenance and checks
Manual cleaning with ultrasound and immersion disinfection Phase
Step
D [°C/°F]
t [min]
Conc. [%]
Water quality
Chemistry
I
pre-cleaning
RT (cold)
>10
-
D–W
-
II
Ultrasound cleaning
40/104
15
0.8
D–W
Enzymatic cleaning agent1)
III
Intermediate rinse
RT
>20
-
Di-W
-
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service.
IV
Drying
50/122
>10
-
-
-
► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
V
Disinfection
RT (cold)
15
4
D–W
Mucocit-T, pH ~ 10,5 (diluted)
Packaging
D-W: Di–W: RT: 1):
Drinking water Deionized water Room temperature Validated with Johnson & Johnson Cidezyme/Enzol and Merz Hygiene GmbH Mucadont Zymaktiv
► Appropriately protect products with fine working tips. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization Phase I ► Place instrument in cold water: >10 min. ► Brush instrument under cold water until all visible contamination has been removed. ► Rinse out cavities, bores and threads using cold water with a cleaning gun: >20s at 3-5 bar. Phase II ► Clean the product in an ultrasound cleaning bath (frequency of 35 kHz) for at least 15 minutes. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Turn and move product repeatedly during cleaning in the ultrasound bath. Phase III ► Rinse out cavities, bores and threads using cold water with a cleaning gun: >20 s at 3 bar to 5 bar. ► Immerse instrument in deionized water and rinse out cavities multiple times with deionized water. ► Clean interior spaces of the product under water with a cleaning brush. Phase IV ► Dry product inside and out: >10 min at 50-100 °C and/or blow out with medical grade compressed air. Phase V ► Disinfect product with disinfectant according to manufacturer's usage instructions.
Note The product can be sterilized either in disassembled or in assembled condition. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Place product in sterilization machine so that no components touch and the steam can freely circulate. ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C, holding time 5 min to 30 min, 3 bar (44 PSI) ► When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6 , Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters*
Mechanical cleaning/disinfection with manual pre-cleaning Manual pre-cleaning with ultrasound and brush Phase
Step
D [°C/°F]
t [min]
Conc. [%]
Water quality
Chemistry
II
Ultrasound cleaning
40/104
15
0.8
D–W
Enzymatic cleaning agent1)
D-W: 1) :
Drinking water Validated with Johnson & Johnson Cidezyme/Enzol and Merz Hygiene GmbH Mucadont Zymaktiv
► Clean the product in an ultrasound cleaning bath (frequency of 35 kHz) for at least 15 minutes. Ensure that all
accessible surfaces are immersed and acoustic shadows are avoided. ► Turn and move product repeatedly during cleaning in the ultrasound bath.
Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service
Mechanical alkaline cleaning Machine type: single-chamber cleaning machine without ultrasound Phase
Step
D [°C/°F]
t [min]
Water quality
Chemistry
I
Pre-rinse
<25/77
1
D–W
-
II
Repeated prerinsing
<25/77
3
D–W
-
III
Cleaning
55/131
5
Di-W
Dr. Weigert neodisher 0.5 %, pH ~ 12.2 to 14 (diluted)
IV
Neutralize
55/131
3
Di-W
-
V
Rinsing
<25/77
2
Di-W
-
VI
Drying
90/194
15-25
–
–
Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
D-W: Di–W:
Drinking water Deionized water
► Place instruments in a strainer bowl on the MIC infeed carriage of the cleaning machine so that the cleaning
agent has access to all inner and outer surfaces. ► Connect rinsing opening (if present) to MIC infeed carriage. ► Remove product from the cleaning machine immediately after the end of the program. ► If necessary, additionally blow out product with medical grade compressed air until it is completely dry.
Thermal disinfecting
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Machine type: single-chamber disinfecting machine without ultrasound Phase
Step
D [°C/°F]
t [min]
Water quality
Chemistry
I
Disinfect
90/194
>5
Di-W/FD-W
-
Di–W: Deionized water FD–W: Fully desalinated water ► Place instruments in a strainer bowl on the MIC infeed carriage of the cleaning machine so that the cleaning
agent has access to all inner and outer surfaces. ► Connect rinsing opening (if present) to MIC infeed carriage. ► Remove product from the cleaning machine immediately after the end of the program. ► If necessary, additionally blow out product with medical grade compressed air until it is completely dry.
Accessories/Replacement Parts Art. no.
Designation
PG003200
Cock plug
PG003201
Spring cap
Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 008562
2018-01
V6
Änd.-Nr. 56432