BBraun
YASARGIL aneurysm clip applier forceps
27 Pages
Preview
Page 1
Use with locking function
Aesculap® YASARGIL aneurysm clip applier forceps Legend 1 2 3 4 5 6 7
Inserting the aneurysm clip and engaging the lock ► Insert the aneurysm clip into jaws 1. ► Press lock push button 4 while compressing branches left/right 2, see Fig. A.
Lock 5 is engaged, see Fig. B. With lock 5 engaged, jaws 1are tensioned so that the aneurysm clip is lightly clamped in between jaws 1.
Releasing the lock
Jaw part Branches left/right Color-coding button Lock pushbutton Lock Spring Clip applier forceps
► Compress branches left/right 2 to the positive stop.
Jaws 1 of clip applier forceps 7 closes and the aneurysm clip opens. Spring 6 switches automatically to the unlocked position, see Fig. C.
Detach the clip applier forceps from the aneurysm clip ► Fully release branches left/right 2, see Fig. D.
Jaw piece 1 of clip applier forceps 7 opens and the aneurysm clip closes. ► Detach clip applier forceps 7 from the aneurysm clip.
Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
Validated reprocessing procedure General safety notes Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for reprocessing.
Intended use The YASARGIL aneurysm clip applier forceps are used for opening, closing and applying Aesculap YASARGIL aneurysm clips. Note In the following text, Aesculap YASARGIL aneurysm clips are referred to as aneurysm clips.
Indications Indications, see Intended use. Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications.
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD, or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note It should be noted that successful reprocessing of this medical device can only be guaranteed following prior validation of the reprocessing method. The operator/reprocessing technician is responsible for this. The specified chemistry was used for validation. Note For up-to-date information about reprocessing and material compatibility, see also the Aesculap Extranet at www.extranet.bbraun.com. The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
General information
Contraindications None known.
Risks and side effects As part of the legal duty to inform, the following typical risks and side effects associated with the use of surgical instruments are referred to. These are predominantly procedure-specific, non product-specific and not limited to unwanted damage to surrounding tissue resulting in e.g. bleeding, infection, material incompatibilities or instrument parts remaining unnoticed in the patient, etc.
Available sizes All clip applier forceps are marked according to size (Mini or Standard) and clip material (Phynox or titanium) so that their correct application with aneurysm clips of the respective size and material is ensured. For detailed information on the available sizes, see the YASARGIL aneurysm clip brochurefor clip applier forceps.
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Clean the new product mechanically after removing its transport packaging and prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts.
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive neutralizing agents or basic cleaners may result in a chemical attack and/or fading and the laser marking becoming unreadable either visually or by machine. Residues containing chlorine or chlorides, e.g., in surgical residues, medicines, saline solutions, and in the service water used for cleaning, disinfection, and sterilization, will cause corrosion damage (pitting, stress corrosion) and result damage to metallic products. These must be removed by rinsing thoroughly with demineralized water and then drying. Perform additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes in materials, e.g., fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging, or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surface and could cause corrosion ► For further detailed information on hygienically safe and material-preserving/value-preserving reprocessing, see www.a-k-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Disassembling the product before carrying out the reprocessing procedure ► Open up products with hinges.
Preparations at the place of use ► If applicable, rinse surfaces that are not accessible to visible inspection (preferably with demineralized water),
using a disposable syringe, for example. ► Remove any visible surgical residues as much as possible with a damp, lint-free cloth. ► Place the dry product in a sealed waste container and forward it on for cleaning and disinfection within 6 hours.
Safe operation Cleaning/disinfection Damage to, imprecise functioning and incorrect closing force of the aneurysm clips caused by using the wrong applier forceps! ► Use aneurysm clips only with the matching applier forceps; note the respective color of the color-coding button.
CAUTION
Product-specific safety notes on the reprocessing procedure Danger to the patient! ► Reprocess the product only with manual pre-cleaning followed by mechanical cleaning.
► Observe color of color-coding button 3.
DANGER Size
Titanium aneurysm clips
Phynox aneurysm clips
Standard
Blue
Black
Mini
Red
Gray
Risk to patient due to cross contamination! ► Do not clean contaminated products together with uncontaminated products in a tray. DANGER
Use clips and clip applier forceps only in the following combinations: ■ Phynox aneurysm clips with the clip applier forceps for Phynox aneurysm clips ■ Titanium aneurysm clips with the clip applier forceps for titanium aneurysm clips ■ Mini and Standard aneurysm clips must be applied with the clip applying forceps of the appropriate size (Mini or Standard, respectively). Locking function Note The clip applier forceps can be used with or without locking function.
Use without locking function ► Insert the aneurysm clip into jaw piece 1. ► Compress branches left/right 2.
The inserted aneurysm clip opens. ► Fully release branches left/right 2. The aneurysm clip closes. ► Detach clip applier forceps 7 from the aneurysm clip.
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for plastic materials, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 55 °C.
► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam forma-
tion and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water. ► Clean and disinfect microsurgical products mechanically if they can be placed securely in the machine or on the positioning aids.
Mechanical cleaning/disinfection with manual pre-cleaning Note The cleaning and disinfecting machine must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection machine used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with a brush
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6 , Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemicals
Prevacuum
270 °F/275 °F
4 min
20 min
I
Disinfectant cleaning
RT (cold)
>15
1
D–W
Dr. Weigert neodisher® SeptoClean*
II
Rinsing
RT (cold)
1
-
D–W
-
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc. WARNING for the US market If this device is/was used in a patient with, or suspected of having Creutzfeldt-Jakob Disease (CJD), the device cannot be reused and must be destroyed due to the inability to reprocess or sterilize to eliminate the risk of crosscontamination.
D–W: Drinking water RT: Room temperature *Prion-inactivating detergent (see Technical information Dr. Weigert neodisher® SeptoClean) Phase I ► Fully immerse the product in the cleaning/disinfecting solution for at least 15 minutes. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed. ► If applicable, brush non-visible surfaces for at least 1 minute with a suitable cleaning brush. ► Mobilize non-rigid components, such as set screws and hinges, during cleaning. ► Then flush these areas thoroughly at least five times with the cleaning disinfectant solution using a disposable syringe (20 ml). Phase II ► Rinse/flush the instrument thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws and hinges, during rinsing.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Mechanical alkaline cleaning and thermal disinfection Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemicals
I
Pre-rinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD–W
Dr. Weigert neodisher® SeptoClean 1 %* working solution
III
Intermediate rinse
>10/50
1
FD–W
-
IV
Thermal disinfection
90/194
5
FD–W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Prion-inactivating detergent (see Technical information Dr. Weigert neodisher® SeptoClean)
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Disposal ► Place the product in a tray that is suitable for cleaning (avoiding rinsing blind spots) ► Place instruments in the tray with their hinges open. ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
CAUTION
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service. ► Check for compatibility with associated products.
Packaging ► Appropriately protect products with fine working tips. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractional vacuum process at 134 °C, holding time 18 minutes for prion inactivation ► When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 012544
2020-03
V6
Änd.-Nr. 62653