Bien-Air Medical Technology
OPTIMA User Guide Sept 2019
User Guide
118 Pages
Preview
Page 1
OPTIMA INT Contents Description Identification Function Fields of use Assembly (see installation instructions 2100307)
Page 28 28 28 28
Meaning of functions Preset (SETUP) Starting Display Mode Program description Operative - Programs - Transmission ratio - Rotation speed of bur - Torque Program description Endo - Programs - Transmission ratio - Rotation speed of bur - Torque List of errors & Troubleshooting
Maintenance
29 30-31 32 32 32-33
34-35
36-37
38
Overhaul Application Information Cleaning Important
Other precautions for use and guarantee
39
Other precautions for use Terms of guarantee
27
ENG
Operation
OPTIMA INT Description OPTIMA INT (REF 1600715-001) is manufactured entirely in Switzerland by Bien-Air Dental.
Function
OPTIMA INT controls the Bien-Air Dental MX2 and MCX micromotor.
Note
The pre-defined torque and speed values in the 10 Endo programs and the 10 Operational programs are only intended as a guide. The operator must always ensure that the pre-defined parameters correspond to its planned treatment and to its instruments being used. Under no circumstances can Bien-Air Dental be held responsible for an incorrect choice of value.
Fields of use
OPTIMA INT is designed for work in general and restorative dentistry and endodentics.
Assembly
See installation instructions 2100307.
USA Caution
Federal law restricts this device to sale by or on order of a dentist.
Applied part (per IEC60601-1)
MCX and MX2 micromotor.
ENG
Identification
28
OPTIMA INT Operation Meaning of functions REV ‘Reverse rotation’ of motor function Bien-Air Dental OPTIMA INT
SAVE Save settings
ENG
ESC ‘Back’ function
SETUP Change configuration
REV
The ‘reverse rotation’ function can be selected directly from all programs in the mode Operative. If selected, an audible signal indicates reverse rotation. For Endo setting must be carried out in SETUP (Auto-forward/Auto-reverse).
SAVE
There are two ways of saving program settings: - Hold down button, values are memorised directly - Short press on the button and the following screen appears: Save -> values are memorised directly Save as (10 left) -> save values under a new name (up to 10 programs) Enter a new name: with the alphabet appears with move cursor left or right with «REV» delete letters with «SETUP» insert letters Rename -> change the name of a program (as above) Delete -> delete a program Progr. listing -> Change the order of programs with select a program with «OK» confirm (the selection flashes) with move the program to desired position
ESC
BACK function. «ESC» makes you leave current screen. If the display flashes when leaving, changes will NOT be recorded. The changes must always be confirmed using «SAVE», otherwise they will be lost.
29
OPTIMA INT Operation Preset
SETUP
ENG
OK
OK
Language
English
Light
Français
Endo
Deutsch
Beep
Italiano
Advanced
Español
Select language required and save with «SAVE». Return with «ESC»
Russkiy
ON/OFF: with
Language
Light
Light
Intensity
Endo
Delay
Beep
Endo Light
Advanced
Select value required and confirm with «OK». Return with «ESC»
and «SAVE»
Only display under Light ON: Intensity (1-16): precision setting with Delay (OFF and 1-30 sec.): with and «SAVE» Endo Light: AUTO/OFF with
and «SAVE»
OK OK
Language Light
Auto-forward
Endo
Time 1.7 s
Auto-forward: Auto-reverse:
Beep Advanced
Select value required and confirm with «OK». Return with «ESC» OK
30
and «SAVE»
Only display under Auto-forward: Time (0.5 - 3 sec.): with and «SAVE»
OK
Language
Operative rev
Light
Endo reverse
Endo
Error
Beep
Overload
Advanced
with
Operative rev (High/Low): with «SAVE» Endo reverse (High/Low): with Error (High/Low): with
Select value required and confirm with «OK». Return with «ESC»
Overload (High/Low): with
and
and «SAVE»
and «SAVE» and «SAVE»
OPTIMA INT Operation Preset
OK
OK
Contrast
Light
Pedal
Endo
Defined Speed
Beep
Defined Torque
Advanced
Defined ratio Info system Restore Select value required and confirm with «OK». Return with «ESC»
Contrast (-5/clearly to 5/dark): with and «SAVE» Pedal (Operative/Endo): ON/OFF or Progressive with
and «SAVE»
ENG
Language
Defined Speed (100 - 40’000 rpm): with and «SAVE» -> 4 more positions (EMPTY), for saving additional values: with and «SAVE» Defined Torque / Unit: Percent - mNm - Ncm (10 - 100% / 0.35 - 3.5 Ncm) Select with and «SAVE». Adjustable with and «SAVE» Defined ratio: Select with and «OK» -> Value settings (for ex.) 20 : 1 LEFT RIGHT Confirm with «SAVE» Info system: series number, electronic Restore: The device’s original parameters will be restored. All your settings will be lost! Continue: with «OK» Cancel: with «ESC»
31
OPTIMA INT Operation Starting Display
Bien-Air Dental OPTIMA INT
ENG
OPTIMA INT Mode OK
Select required usage and confirm with «OK».
Operative Endo
Programm description Operative
Operative Programms (example below for MX2) OK
OK
Operative Endo
OK
Crown preparation
Caries removal
Ratio 1:5
Dentine reduction
200’000 rpm
Enamel reduction
Torque 0.70 Ncm
Tooth sectioning
Direct selecting of required program with .
Porcelain adjust Amalgam removal Crown preparation Deep caries remov
Select required program and confirm with «OK».
Polishing Cavity preparation
Return with «ESC»
can save up to 10 additional programs (Process refer to «SAVE» button)
32
OPTIMA INT Operation Contenu Operative Transmission ratio (example below for MX2) OK
Crown preparation Ratio 1:5 200’000 rpm Torque 0.70 Ncm
Operative Endo
Ajustable with
.
OK
1:5 1:4 1:2 1:1 7:1 10 : 1 20 : 1 22 : 1 30 : 1 100 : 1 Other
Select required transmission ratio and confirm with «OK». Return with «ESC» Other: for ex. Value:
ENG
OK
20 : 1 LEFT
RIGHT
max. 128
max. 5
Save with «SAVE»
Operative Bur rotational speed depends on transmission ratio (example below for MX2) OK
OK
Crown preparation Ratio 1:5 200’000 rpm Torque 0.70 Ncm
Operative Endo
Ajustable with
.
OK
500 1’500 30’000 50’000 100’000 200’000
Select required speed range and confirm with «OK». Return with «ESC»
Operative Torque depends on transmission ratio (example below for MX2) OK
OK
Operative Endo
OK
Crown preparation
10% 0.07 Ncm
Ratio 1:5
20% 0.14 Ncm
200’000 rpm
30% 0.21 Ncm 40% 0.28 Ncm
Torque 0.70 Ncm Direct selecting of required torque with .
50% 0.35 Ncm 60% 0.42 Ncm 70% 0.49 Ncm 80% 0.56 Ncm 90% 0.63 Ncm 100% 0.70Ncm
Select required torque speed and confirm with «OK». Return with «ESC»
33
OPTIMA INT Operation Programm description Endo Endo Programms (example below for MX2) OK
OK
OK
Endo file 1
Endo File 1
Endo
Ratio 1:1
Endo File 2
250 rpm
Endo File 3
Torque 3.01 Ncm
Endo File 4
ENG
Operative
Endo File 5
Direct selecting of required program with .
Select required program and confirm with «OK».
Endo File 6 Endo File 7 Endo File 8 Endo File 9
Returm with «ESC»
Endo File 10
can save up to 10 additional programs (Process refer to «SAVE» button)
Endo Transmission ratio (example below for MX2) OK
OK
OK
Operative
Endo file 1
1:5
Endo
Ratio 1:1
1:4
250 rpm
1:2
Torque 3.01 Ncm
1:1
Direct selecting of required transmission ratio with .
7:1 10 : 1
Return with «ESC»
20 : 1
Other: for ex.
22 : 1 30 : 1
34
Select required transmission ratio and confirm with «OK».
20 : 1
100 : 1
Value: LEFT max. 128
Other
Save with «SAVE»
RIGHT max. 5
OPTIMA INT Operation Endo Bur rotational speed depends on transmission ratio (example below for MX2) OK
OK
Operative
Endo file 1
100
Endo
Ratio 1:1
300
250 rpm
6’000
Torque 3.01 Ncm
10’000
Select required speed range and confirm with «OK».
20’000 Ajustable with
.
Return with «ESC» 40’000
Endo Torque depends on transmission ratio (example below for MX2) OK
OK
OK
Operative
Endo file 1
10% 0.35 Ncm
Endo
Ratio 1:1
20% 0.7 Ncm
250 rpm
30% 1.05 Ncm
Torque 3.01 Ncm
40% 1.40 Ncm 50% 1.75 Ncm
Ajustable with
.
60% 2.10 Ncm 70% 2.45 Ncm
Select torque
80% 2.80 Ncm 90% 3.15 Ncm
and confirm with «OK».
100% 3.50 Ncm
Return with «ESC»
Display icons description Left scroll possible using left key
Hourglass = Wait
Right scroll possible using right key
Floppy disk = Value memorized
Up scroll possible using up key Down scroll possible using down key
Info = information or error displayed
35
ENG
OK
OPTIMA INT List of errors & troubleshooting Warning. DMCX/DMX2/DMX3 memory configured check. Message Root cause Action
The DMCX/DMX2/DMX3 is not correctly configured. Press key SAVE to setup the DMCX/DMX2/DMX3. DMCX/DMX2/DMX3 memory data must be updated. Press the SAVE key to reload.
ENG
Equipment initialisation failures. DMCX/DMX2/DMX3 connection / type / dip switches configuration / minimum compatible version / serial number check. Message
Motor drive error. Please contact your retailer. Key ESC -> restart Root causes DMCX/DMX2/DMX3 connection attempt failed. • DMCX/DMX2/DMX3 type is unexpected. • DMCX/DMX2/DMX3dip switches configuration is wrong. Troubleshooting Check DMCX/DMX2/DMX3 connection (red 8-pin connector). This connector might be “reverse” plugged: it is correctly plugged when wires are oriented outside of the DMCX/DMX2/DMX3 board. • Check DMCX/DMX2/DMX3 type (DMX2 Plus, DMX2 Pro,...). • Check dip switches configuration for requested serial communication type. Action Press ESC to try to restart. Contact retailer. Equipment operation failures. Following errors may occur at any time during equipment operation. Communication timeout. Message Root cause Action
Communication timeout DMCX/DMX2/DMX3 connection lost. Check DMCX/DMX2/DMX3 connection (red 8-pin connector). Contact retailer.
Motor cable fault. Message Root cause Action
Error(s): #1 motor cable fault DMCX/DMX2/DMX3 board short circuit detected on motor phases. Check DMCX/DMX2/DMX3 connection with MCX/MX2 motor. Contact retailer.
Motors is not connected. Message Root cause Action
36
Error(s): #2 motor is not connected DMCX/DMX2/DMX3 board connected motor phase(s) not detected. MCX/MX2 Motor might not be connected to DMCX/DMX2/DMX3 board. Check DMCX/DMX2/DMX3 connection with MCX/MX2 motor. Contact retailer.
OPTIMA INT List of errors & troubleshooting Communication timeout. Message Root cause Action
Error(s): #3 comm. timeout Communication failed with the DMCX/DMX2/DMX3 board. Check DMCX/DMX2/DMX3 connection (red 8-pin connector). Contact retailer.
DMCX/DMX2/DMX3 board over-temperature. Error(s): #4 temperature DMCX/DMX2/DMX3 board too hot (electronic device overheating). Wait until the DMCX/DMX2/DMX3 cools down. Contact retailer.
ENG
Message Root cause Action
DMCX/DMX2/DMX3 board under-voltage. Message Root cause Action
Error(s): #5 under voltage DMCX/DMX2/DMX3 board 32V power supply detected voltage too low. Contact retailer.
DMCX/DMX2/DMX3 board over-voltage. Message Root cause Action
Error(s): #6 over voltage DMCX/DMX2/DMX3 board 32V power supply detected voltage too high. Contact retailer.
DMCX/DMX2/DMX3 unknown failure. Message Root cause Action
Error(s): #0 unknown DMCX/DMX2/DMX3 unknown failure (unidentified cause). Contact Bien-Air Dental SA.
37
OPTIMA INT Maintenance Only use maintenance products and components from Bien-Air Dental. The use of other products and components can void the guarantee. Overhaul Never disassemble the device. For any modification and repair, we recommend that you contact your regular supplier or Bien-Air Dental directly. Bien-Air Dental asks the user to have its dynamic instruments checked or inspected at least once a year.
ENG
Application Product intended for professional use only. Information The technical specifications, illustrations and dimensions contained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equipment, without amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover. Cleaning • Use cleaner Spraynet, Bien-Air Dental. Disinfect with a clean damp cloth. • Do not exert pressure on the screen. • Products containing acetone, chlorine and bleach are not recommended as disinfectants. • Never immerse in solutions. • Not suitable for ultrasonic bath. Important to maintain the - MX2 Micromotor, see instructions REF 2100199 - MX2-400 hose, see instructions REF 2100223 - DMX2 Plus/DMX2 Pro, see instructions REF 2100202 - DMX3, see instructions REF 2100278 Information The disposal and/or recycling of materials must be performed in accordance with the legislation in force. This unit and its accessories must be recycled. Electrical and electronic equipment may contain dangerous substances which constitute health and environmental hazards. The user must return the unit to its dealer or establish direct contact with an approved body for treatment and recovery of this type of equipment (European Directive 2002/96/EC).
38
OPTIMA INT
Other precautions for use The device must be used by a qualified person in accordance with the current legal provisions concerning industrial safety, health and accident prevention measures, and these working instructions. In accordance with these requirements, the operators: • must only use operating devices that are in perfect working order; in the event of irregular functioning, excessive vibration, abnormal heating or other signs indicating malfunction of the device, the work must be stopped immediately; in this case, contact a repair centre that is approved by Bien-Air Dental; • must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamination through the use of the product. The device is intended for medical treatment only; any use other than that for which this product is intended is unauthorised and may be dangerous. The medical device meets all the current legal requirements. The device is not authorised for use in an explosive atmosphere (anaesthetic gas). Never immerse in disinfectant solutions. Guarantee Terms of guarantee Bien-Air Dental grants the user a guarantee covering all functional defects, material or production faults. The device is covered by this guarantee for 12 months from the date of invoicing. In the event of justified claim, Bien-Air Dental or its authorised representative will fulfil the company’s obligations under this guarantee by repairing or replacing the product free of charge. Any other claims, of whatever nature, in particular in the form of a claim for damages and interest, are excluded. Bien-Air Dental shall not be held responsible for damage or injury and the consequences thereof, resulting from: • excessive wear and tear • improper use • non-observance of the instructions for installation, operation and maintenance • unusual chemical, electrical or electrolytic influences • poor connections, whether of the air, water or electricity supply. The guarantee does not cover flexible “fibre optic” type conductors, or any parts made of synthetic materials. The guarantee shall become null and void if the damage and its consequences are due to improper manipulation of the product, or modifications to the product carried out by persons not authorised by Bien-Air Dental. Claims under the terms of the guarantee will be considered only on presentation, together with the product, of the invoice or the consignment note, on which the date of purchase, the product reference and the Serial No. should be clearly indicated.
39
ENG
Other precautions for use and guarantee
2100222
93 93 93 93
94 95-96 97 97 97-98
99-100
101-102
103
JPN
104
92
Bien-Air Dental OPTIMA INT
REV
JPN
SAVE
ESC
94
SETUP OK
OK
Language
English
Light
Français
Endo
Deutsch
Beep
Italiano
Advanced
Español Russkiy
Language
Light
Light
Intensity
Endo
Delay Endo Light
Beep Advanced
OK OK
Language Light
Auto-forward
Endo
Time 1.7 s
Beep Advanced
OK
Language
Operative rev
Light
Endo reverse
Endo
Error
Beep
Overload
JPN
OK
Advanced
95
OK
OK
Language
Contrast
Light
Pedal
Endo
Defined Speed
Beep
Defined Torque
Advanced
Defined ratio Info system
JPN
Restore
96
Bien-Air Dental OPTIMA INT
OPTIMA INT
OK
Operative Endo
֮ԱΊ0;ΉΉ OK
Operative Endo
OK
Crown preparation
Caries removal
Ratio 1:5
Dentine reduction
200’000 rpm
Enamel reduction
Torque 0.70 Ncm
Tooth sectioning Porcelain adjust Amalgam removal Crown preparation Deep caries remov Polishing Cavity preparation
JPN
OK
97
֮ԱΊ0;ΉΉ OK
OK
OK
Crown preparation Ratio 1:5 200’000 rpm Torque 0.70 Ncm
Operative Endo
1:5 1:4 1:2 1:1 7:1 10 : 1 20 : 1 22 : 1 30 : 1 100 : 1 Other
֮ԱΊ0;ΉΉ OK
OK
Crown preparation Ratio 1:5 200’000 rpm Torque 0.70 Ncm
Operative Endo
OK
500 1’500 30’000 50’000 100’000 200’000
֮ԱΊ0;ΉΉ OK
OK
Operative
Crown preparation
10% 0.07Ncm
Endo
Ratio 1:5
20% 0.14Ncm
200’000 rpm
30% 0.21Ncm 40% 0.28Ncm
Torque 0.70 Ncm
JPN
OK
50% 0.35Ncm 60% 0.42Ncm 70% 0.49Ncm 80% 0.56Ncm 90% 0.63Ncm 100% 0.70Ncm 98
֮ԱΊ0;ΉΉ OK
OK
OK
Operative
Endo file 1
Endo File 1
Endo
Ratio 1:1
Endo File 2
250 rpm
Endo File 3
Torque 3.01 Ncm
Endo File 4 Endo File 5 Endo File 6 Endo File 7 Endo File 8 Endo File 9 Endo File 10
֮ԱΊ0;ΉΉ OK
OK
Operative
Endo file 1
1:5
Endo
Ratio 1:1
1:4
250 rpm
1:2
Torque 3.01 Ncm
1:1 7:1 10 : 1 20 : 1 22 : 1 30 : 1
JPN
OK
8
100 : 1 Other
99