Instructions
2 Pages
Preview
Page 1
3,1 bar
2 - 3V 4V Unifix
Aquilon Do not sterilise fig. 1
A B C
Instruction 0120
REF 2100033-ENG/05.11
English Medical devices entirely made in Switzerland by Bien-Air Dental SA. Type Dental air micromotor, non-sterilisable. Description Aquilon with internal or external spray, without light as per ISO Standard 3964. Intended use Product intended for professional use only. Use in dentistry for prophylaxis, general dentistry and endodontic work.
Servicing Never disassemble the device. For any modification and repair, we recommend that you contact your regular supplier or Bien-Air Dental directly. BienAir Dental asks the user to have its dynamic instruments checked or inspected at least once a year.
Never immerse in disinfectant solutions.
Classification Class IIa in accordance with European Directive 93/42/EEC concerning medical devices.
Lubrication Lubrication at least twice a day. Only use the spray Lubrifluid. • Place the device in a cloth so as to catch the debris and foreign matter expelled by the atomizer jet. • Insert the nozzle into the back of the instrument sleeve. Spray for about 1 second fig. 5.
Dimensions Ø 23 x 65 mm Weight 90 g without the cable. Recommanded rotation speed 5,000 rpm up to 25,000 rpm; rotation in both directions, positions A, B and C fig. 2. Mechanical power 16 W Torque 25 mNm
Transport and storage conditions Temperature between -40°C (-40°F) and 70°C (158°F), relative humidity between 10% and 100%, atmospheric pressure 50 kPa to 106 kPa (7.3 to 15.3 psi). Information The technical specifications, illustrations and dimensions contained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equipment, without amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover. Other precautions for use
Noise level In accordance with ISO 11498, less than 70 dBA at 45 cm (17.72 in). Air consumption output 56 normliter/min on average. Required air pressure Adjust the required air pressure to 3.1 bar with the assistance of a Bien-Air Dental inspection pressure gauge - 3-hole version or 4-hole version, placed between hose and Unifix coupling, fig. 1. Changing the seals fig. 3 Maintenance Non-sterilisable Clean and lubricate the device before the treatment of each patient. Only use maintenance products and components from Bien-Air Dental. The use of other products and components can void the guarantee. REF 2100033-ENG/05.11
B
Disinfection Disinfect using a clean cloth soaked in a suitable product. Products containing acetone, chlorine and bleaches are not recommended as disinfectants.
Not suitable for ultrasonic bath.
Important Never mount an instrument on a rotating motor.
1 sec.
A
Cleaning Carefully clean all surfaces with a cloth impregnated with Spraynet, fig. 4.
Technical data and assembly
Coupling For 2-3- or 4-way coupling, fig. 1
fig. 3
fig. 2
The device must be used by a qualified person in accordance with the current legal provisions concerning industrial safety, health and accident prevention measures, and these working instructions. In accordance with these requirements, the operators: • must only use operating devices that are in perfect working order; in the event of irregular functioning, excessive vibration, abnormal heating or other signs indicating malfunction of the device, the work must be stopped immediately; in this case, contact a repair centre that is approved by Bien-Air Dental; • must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamination through the use of the product. The device is intended for medical treatment only; any use other than that for which this product is intended is unauthorised and may be dangerous. The medical device meets all the current legal requirements.
fig. 4
fig. 5
Recommendations It is essential to use dry, purified compressed air in order to ensure the long working life of the device. Maintain the quality of the air and the water by regular maintenance of the compressor and the filtration systems. The use of unfiltered hard water will lead to early blockage of the tubes, connectors and spray cones.
Claims under the terms of the guarantee will be considered only on presentation, together with the product, of the invoice or the consignment note, on which the date of purchase, the product reference and the Serial No. should be clearly indicated.
Rest the device on a suitable support to avoid risks of infection for yourself, the patient or third parties. Guarantee Terms of guarantee Bien-Air Dental grants the user a guarantee covering all functional defects, material or production faults. The device is covered by this guarantee for 12 months from the date of invoicing. Bien-Air Dental offers a 24-month guarantee for the glass-bar light conductors. In the event of justified claim, Bien-Air Dental or its authorised representative will fulfil the company’s obligations under this guarantee by repairing or replacing the product free of charge. Any other claims, of whatever nature, in particular in the form of a claim for damages and interest, are excluded. Bien-Air Dental shall not be held responsible for damage or injury and the consequences thereof, resulting from: • excessive wear and tear • improper use • non-observance of the instructions for installation, operation and maintenance • unusual chemical, electrical or electrolytic influences • poor connections, whether of the air, water or electricity supply. The guarantee does not cover flexible “fibre optic” type conductors, or any parts made of synthetic materials. The guarantee shall become null and void if the damage and its consequences are due to improper manipulation of the product, or modifications to the product carried out by persons not authorised by Bien-Air Dental. 1/2
Set supplied
Set 4V
Set 2-3V
REF
Legend
1700058-001
Aquilon 830 with internal spray, without light set with 2-3-hole Unifix-coupling Aquilon 830 with internal spray, without light set with 4-hole Unifix-coupling Aquilon 830 with internal spray, without light, with 2-3 and 4-hole Unifix rotating quick coupling Aquilon 830 with external spray, without light, with 2-3 and 4-hole Unifix rotating quick coupling Unifix 2-3-hole coupling Unifix 4-hole coupling Gauge for 3-hole attachment Gauge for 4-hole attachment Attachment seals for micromotors Spraynet, cleaning spray 500 ml, box of 6 cans Lubrifluid, lubricant 500 ml, box of 6 cans
1700211-001 1600070-001 1600362-001
REF 1700058-001
1600081-001 1600082-001 1600242-001 1600243-001 1300145-010 1600036-006 1600064-006
REF 1700211-001
List of Bien-Air Dental SA registered trade mark products ®:
Optional accessories
Aquilon® Bora® Boralina® ChiroPro® 3-hole
4-hole
REF 1600242-001
REF 1600243-001
Lubrifluid® Lubrimed® MX® PowerCare®
Prestilina® Spraynet®
In these instructions, “Device” corresponds to the product described in the heading “Type”. For example, turbine, contra-angle, handpiece, micromotor, tube, electronics, connectors, station etc.
REF 1600070-01
REF 1600362-001
Eolia® Gyro® Gyrolina® Isolite®
Symbols Manufacturer.
0120
CE Marking with number of the notified body.
Attention.
2-3V
4V
REF 1600081-001
REF 1600082-001
6x
6x
REF 1600036-006
REF 1600064-006
10x REF 1300145-010
Wear rubber gloves. 3-hole connection. 4-hole connection.
Bien-Air Dental SA Länggasse 60 Case postale 2500 Bienne 6, Switzerland Tel. +41 (0)32 344 64 64 Fax +41 (0)32 344 64 91 [email protected] Bien-Air Deutschland GmbH Jechtinger Strasse 11 79111 Freiburg, Deutschland Tel. +49 (0)761 45 57 40 Fax +49 (0)761 47 47 28 [email protected] Bien-Air España, SA Entença, 169 Bajos 08029 Barcelona, España Tel. +34 934 25 30 40 Fax +34 934 23 98 60 [email protected]
This product may be covered by one or more of the following patents: EP Europe: 745358 / 688539 / 948294 / 1145688 / 1563800 / 1563801 / 1675523 / 1753360 DE Germany: 29616023.7 DK Denmark: 9600315 FR France: 2722972 CH Switzerland: 693922 CN China: 100528099 / 100522100 / 100522099 / 100553584 JP Japan: 3892485 / 4298933 / 7000419 US United-States: 5453008 / 6033220 / 6319003 / 7214060 / 7448870 RU Russia: 2361540 / 2361541 / 2372046 REF 1600070-001 MOT AQUILON830 Spray interne / REF 1600362-001 MOT AQUILON830 Spray externe
REF 2100033-ENG/05.11
Bien-Air USA, Inc. Medical Technologies 5 Corporate Park Suite 160 Irvine, CA 92606 USA Phone 1-800-433-BIEN Phone 949-477-6050 Fax 949-477-6051 [email protected] Bien-Air France Sàrl 55-57, avenue Jean Lolive 93508 Pantin Cedex, France Tel. +33 (0)1 41 83 60 70 Fax +33 (0)1 48 96 07 40 [email protected]
Move in the direction indicated. Move fully to the stop, in the direction indicated.
Bien-Air UK Ltd Arundel House Unit 1 - Ground Floor Amberley Court, Whitworth Road Crawley, West Sussex, RH11 7XL, England Telephone +44 (0)1293 550200 Fax: +44 (0)1293 520481 [email protected] Bien-Air Asia Ltd. Nishi-Ikebukuro Daiichi-Seimei Bldg. 10F 2-40-12 Ikebukuro, Toshimaku Tokyo, 171-0014, Japan
Tel. +81 (3) 5954-7661 Fax +81 (3) 5954-7660 [email protected] Beijing Bien-Air Medical Instrument Technology Service Co. Ltd. Room 907, The Exchange Beijing, No 118 Jian Guo Lu Yi, Chao Yang District, Beijing 100022, China
Tel. +86 10 6567 0651 Fax +86 10 6567 8047 [email protected]
Bien-Air Italia s.r.l. Via Vaina 3 20122 Milano, Italia Tel. +39 (02) 58 32 12 51/52/54 Fax +39 (02) 58 32 12 53 [email protected] 2/2