Instructions
2 Pages
Preview
Page 1
➇ ➈
Hoses MC2 & Isolite
fig. 1
fig. 2
Do not sterilise
Instruction
0120
REF 2100017-ENG/09.11
English Medical devices entirely made in Switzerland by Bien-Air Dental SA.
To keep the surface of the hose in good condition, it is advisable to periodically wipe the complete length of it with a cloth dusted with talcum powder. Never immerse in disinfectant solutions.
Type Electric micromotor hose with revolving connector for use with the MC2, the Isolite 100 and 300, with or without light. Quality hoses straight or coiled, in black, grey or white. Standard length 1.7 m or special lengths up to 3 m. Intended use Product intended for professional use only. Use in dentistry for prophylaxis, general dentistry and endodontic work. Technical data and assembly Classification Class IIa, «active medical devices», as per section III «Classification», Rule 2 of Annex IX to European Directive 93/42/EEC, namely: «Non-invasive devices intended for channelling liquids or gases for the purpose of administration into the body, which may be connected to an active medical device in Class IIa». Fitting Keep the initial alignment of the leads and conduits unchanged. Place the back-up ring ➇ in the locking area fig. 1. Tie the security string ➈ fig. 2. Description fig. 3 - 4 ➀ Ø 1.5/2.5 mm air spray (small white) with “A” or (blue) ➁ Ø 1.5/2.5 mm water spray (small white) with “W” or (green) ➂ Ø 1.5/2.5 mm motor cooling (small white) or (transparent) Ø 2.8/4.1 mm ➃ (+) red: motor ➄ black: motor ➅ (+) brown: bulb ➆ (0V) blue: bulb ➇ back-up ring ➈ security string (if existing) Accessories O-Ring joint REF 1300414-010. Changing the Oring joint fig. 5. The split ring REF 212.65.01 fig. 6 is required only for connection to an MC2 IR or GTAV motor fig. 7-8. Connection (without the split ring) to an Isolite electric motor fig. 9-10. The conical piece fig. 11 can be useful for connecting the cooling tube to the Unit tube ➆ fig. 12. Maintenance Non-sterilisable Cleaning Carefully clean all surfaces with a cloth impregnated with Spraynet, fig. 13. Disinfection Disinfect using a clean cloth soaked in a suitable product. Products containing acetone, chlorine and bleaches are not recommended as disinfectants. REF 2100017-ENG/09.11
Not suitable for ultrasonic bath. Transport Temperature between -40°C (-40°F) and 70°C (158°F), relative humidity between 10% and 100%, atmospheric pressure 50 kPa to 106 kPa (7.3 to 15.3 psi). Information The technical specifications, illustrations and dimensions contained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equipment, without amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover.
fig. 3
REF 1300414-010
fig. 4
fig. 5
REF 212.65.01-001
MC2 IR & MC2 GTAV REF 212.65.01-001
Other precautions for use The device must be used by a qualified person in accordance with the current legal provisions concerning industrial safety, health and accident prevention measures, and these working instructions. In accordance with these requirements, the operators: • must only use operating devices that are in perfect working order; in the event of irregular functioning, excessive vibration, abnormal heating or other signs indicating malfunction of the device, the work must be stopped immediately; in this case, contact a repair centre that is approved by Bien-Air Dental; • must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamination through the use of the product. The device is intended for medical treatment only; any use other than that for which this product is intended is unauthorised and may be dangerous. The medical device meets all the current legal requirements.
fig. 6
fig. 7
MC2 IR & MC2 GTAV
Isolite
fig. 8
fig. 9
Isolite
fig. 10
The device is not authorised for use in an explosive atmosphere (anaesthetic gas). Recommendations It is essential to use dry, purified compressed air in order to ensure the long working life of the device. Maintain the quality of the air and the water by regular maintenance of the compressor and the filtration systems. The use of unfiltered hard water will lead to early blockage of the tubes, connectors and spray cones.
fig. 11
A
B
Rest the device on a suitable support to avoid risks of infection for yourself, the patient or third parties. Only use maintenance products and components from Bien-Air Dental. The use of other products and components can void the guarantee. fig. 13
1/2
Guarantee Terms of guarantee Bien-Air Dental grants the user a guarantee covering all functional defects, material or production faults. The device is covered by this guarantee for 12 months from the date of invoicing. Bien-Air Dental offers a 24-month guarantee for the glass-bar light conductors. In the event of justified claim, Bien-Air Dental or its authorised representative will fulfil the company’s obligations under this guarantee by repairing or replacing the product free of charge. Any other claims, of whatever nature, in particular in the form of a claim for damages and interest, are excluded. Bien-Air Dental shall not be held responsible for damage or injury and the consequences thereof, resulting from: • excessive wear and tear • improper use
• non-observance of the instructions for installation, operation and maintenance • unusual chemical, electrical or electrolytic influences • poor connections, whether of the air, water or electricity supply.
REF
Legend
1600120-001
Grey hose with fixed connector, for micromotors MC2 and Isolite 1600297-001 Black hose with fixed connector, for micromotors MC2 and Isolite 1600298-001 Grey hose, revolving connector for micromotors Isolite 1600299-001 Black hose, revolving connector for micromotors Isolite 1600314-001 White hose with fixed connector, for micromotors MC2 and Isolite 1600315-001 Black coiled hose with fixed connector for micromotors MC2 and Isolite 1600316-001 White hose, revolving connector for micromotors Isolite 1300414-010 O-Ring joint 212.65.01-001 Split ring 1600036-006 Spraynet, cleaning spray 500 ml, box of 6 cans
The guarantee does not cover flexible “fibre optic” type conductors, or any parts made of synthetic materials. The guarantee shall become null and void if the damage and its consequences are due to improper manipulation of the product, or modifications to the product carried out by persons not authorised by Bien-Air Dental. Claims under the terms of the guarantee will be considered only on presentation, together with the product, of the invoice or the consignment note, on which the date of purchase, the product reference and the Serial No. should be clearly indicated.
List of Bien-Air Dental SA registered trade mark products ®: Aquilon® Bora® Boralina®
ChiroPro® Eolia® Gyro®
Gyrolina® Isolite® Lubrifluid®
Lubrimed® Prestilina® MX® Spraynet® PowerCare®
In these instructions, “Device” corresponds to the product described in the heading “Type”. For example, turbine, contra-angle, handpiece, micromotor, tube, electronics, connectors, station etc. Designation micromotor REF micromotor Designation hose Grey hose with fixed connector Black hose with fixed connector Grey hose, revolving connector Black hose, revolving connector White hose with fixed connector Black coiled hose with fixed connector White hose, revolving connector
REF hose 1600120-001 1600297-001 1600298-001 1600299-001 1600314-001 1600315-001 1600316-001
MC2 40 IR 1600073-001
MC2 40 GTAV 1600075-001
MC2 Isolite 300 1600078-001
MC2 Isolite 100 1600079-001
x x
(x) (x)
x x x x x x x
x x x x x x x
Symbols Manufacturer.
x x
(x) (x)
0120
CE Marking with number of the notified body.
Wear rubber gloves.
Move in the direction indicated.
Attention.
Move fully to the stop, in the direction indicated.
(X) Do not connect supply of water and air spray (risk to damage the micromotor)
This product may be covered by one or more of the following patents: EP Europe: 745358 / 688539 / 948294 / 1145688 / 1563800 / 1563801 / 1675523 / 1753360 DE Germany: 29616023.7 DK Denmark: 9600315 FR France: 2722972 CH Switzerland: 693922 CN China: 100528099 / 100522100 / 100522099 / 100553584 JP Japan: 3892485 / 4298933 / 7000419 US United-States: 5453008 / 6033220 / 6319003 / 7214060 / 7448870 RU Russia: 2361540 / 2361541 / 2372046 REF 1600120-001 HOSE ISOLITE/MC2 GREY / REF 1600297-001 HOSE ISOLITE/MC2 BLACK / REF 1600314-001 HOSE ISOLITE/MC2 WHITE / REF 1600315-001 HOSE ISOLITE/MC2 COILED BLACK / REF 1600298-001 HOSE ISOLITE/MC2 SWIVEL GREY / REF 1600299-001 HOSE ISOLITE/MC2 SWIVEL BLACK / REF 1600316-001 HOSE ISOLITE/MC2 SWIVEL WHITE
Set supplied
Optional accessories
Fixed connector for MC2 10x
REF 1600120-001 REF 1600297-001
REF 212.65.01-001
REF 1600314-001
REF 1600315-001
REF 1600298-001
REF 1600316-001
REF 2100017-ENG/09.11
Bien-Air Deutschland GmbH Jechtinger Strasse 11 79111 Freiburg, Deutschland Tel. +49 (0)761 45 57 40 Fax +49 (0)761 47 47 28 [email protected] Bien-Air España, SA Entença, 169 Bajos 08029 Barcelona, España Tel. +34 934 25 30 40 Fax +34 934 23 98 60 [email protected] Bien-Air USA, Inc. Medical Technologies 5 Corporate Park Suite 160 Irvine, CA 92606 USA Phone 1-800-433-BIEN Phone 949-477-6050 Fax 949-477-6051 [email protected] Bien-Air France Sàrl 55-57, avenue Jean Lolive 93508 Pantin Cedex, France Tel. +33 (0)1 41 83 60 70 Fax +33 (0)1 48 96 07 40 [email protected]
Rotating connector for MC2
REF 1600299-001
REF 1300414-010
Bien-Air Dental SA Länggasse 60 Case postale 2500 Bienne 6, Switzerland Tel. +41 (0)32 344 64 64 Fax +41 (0)32 344 64 91 [email protected]
6x
REF 1600036-006
Bien-Air UK Ltd Arundel House Unit 1 - Ground Floor Amberley Court, Whitworth Road Crawley, West Sussex, RH11 7XL, England Telephone +44 (0)1293 550200 Fax: +44 (0)1293 520481 [email protected] Bien-Air Asia Ltd. Nishi-Ikebukuro Daiichi-Seimei Bldg. 10F 2-40-12 Ikebukuro, Toshimaku Tokyo, 171-0014, Japan
Tel. +81 (3) 5954-7661 Fax +81 (3) 5954-7660 [email protected] Beijing Bien-Air Medical Instrument Technology Service Co. Ltd. Room 907, The Exchange Beijing, No 118 Jian Guo Lu Yi, Chao Yang District, Beijing 100022, China
Tel. +86 10 6567 0651 Fax +86 10 6567 8047 [email protected]
Bien-Air Italia s.r.l. Via Vaina 3 20122 Milano, Italia Tel. +39 (02) 58 32 12 51/52/54 Fax +39 (02) 58 32 12 53 [email protected] 2/2