Bio-logic® NavPRO ONE ABR screening Quick start guide Doc. No.7-26-1388-EN/00 Part No.7-26-13808-EN
Natus Medical Denmark ApS. +45 45 75 55 55. [email protected] Natus Medical Incorporated. 1-800-289-2150. [email protected] www.otometrics.com
Bio-logic®
Quick Guide: Bio-logic® NavPRO ONE® ABR screening
Turning On/Off, and charging the device
On/off button on right side
Power supply connector
Start menu Note: this screen (test options) can vary depending on your set up
Battery indicator Battery status Current time Select patient
Customize
Turn off device
2
Additional information about options for current screen
Otometrics - Next generation Bio-logic products
Bio-logic®
Quick Guide: Bio-logic® NavPRO ONE® ABR screening
Auditory brainstem response (abr) screening Please note that single ear testing is shown for ABR screening. For binaural testing instructions, please refer to the ABR diagnostic Quick Start Guide
Connect the electrode cable to the white and the probe to the red connector.
Clean the patient’s skin and place the electrode pads. Connect the clips to the corresponding sites. White: high forehead Black: cheek Red: mastoid
3
Otometrics - Next generation Bio-logic products
Quick Guide: Bio-logic® NavPRO ONE® ABR screening
Bio-logic®
Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making sure that the ear tip fits tightly into the ear canal.
Select QuickABR button
Select the test ear by pressing either the Right or Left button
A calibration will be performed prior to measurement. A green light indicates successful calibration.
An impedance check will be performed prior to measurement. An impedance below 6 k ohm is recommended for best test results.
4
After successful ear probe calibration and impedance check, the test will start automatically. If yellow or red lights are displayed, check to make sure you are using the correct probe tip and ear tip size, that the eartip is inserted correctly, and that there is no excessive ambient noise.
Otometrics - Next generation Bio-logic products
Quick Guide: Bio-logic® NavPRO ONE® ABR screening
Bio-logic®
ABR sreening test screen Normative data for selected age: red – wave III, green – wave V
Swipe screen up and down to change display scale
Response detection bar Ongoing EEG – value should remain in the lower third
Impedance monitor – impedance should be below 6 kOhm for best results
Test progress Stop test
Pause/resume test
ABR sreening test results
Swipe screen up and down to change display scale
Overall screening test result
Test parameters and conditions Test other ear Print results to a label or to a pdf file
5
Otometrics - Next generation Bio-logic products
Bio-logic®
Quick Guide: Bio-logic® NavPRO ONE® ABR screening
Quick Guide Information Release date: 2016-12-22 Revision: 01 Valid from: Firmware Rev. 2.1, Mira PC Software Rev. 1.6.4 Applies to: Model: SOH100360, web descriptions SO H06, SOH07, SOH08 and SOH09 All mentioned items, products, brands and trademarks are registered or owned by the mentioned companies. All information, illustrations, and specifications provided within this manual are based on the latest product information available at the time of publication. PATH medical reserves the right to make changes at any time without notice. Please note that not all products displayed on our webpage or advertising materials are available in all territories. Please contact your local distributor or PATH medical for further information. The latest revision of the Quick Start Guide is available online at www.pathme.de/download. Errors and omissions excepted. Copyright Notice No part of this manual may be reproduced, translated, stored, or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written consent of PATH medical GmbH. Copyright © 2016 PATH medical GmbH PATH medical GmbH Landsberger Straße 65 82110 Germering Germany Tel: +49 89 800 76 502 Fax: +49 89 800 76 503 Internet: www.pathme.de