Owner’s Guide
An Essential Part of Your Infection Control System
Congratulations on your purchase of a BioSonic® Ultrasonic Cleaner- the finest, most efficient and most popular ultrasonic cleaning system in dentistry. Developed specifically for the dental industry, BioSonic units feature unique product design to optimize time and safety in the office or laboratory.
BioSonic is a total ultrasonic cleaning system, with a full line of cleaning solutions and unit accessories which provide you with the flexibility to efficiently clean instruments, prostheses and other appliances. What is Ultrasonic Cleaning?
Ultrasonic Cleaning is a process created by high frequency soundwaves. The soundwaves, enhanced by specially formulated cleaning solutions, create high energy cavitation. During cavitation, millions of tiny bubbles form and then collapse or “implode,” releasing enormous amounts of energy and shock waves which scour the surface of instruments, appliances and other devices placed in the cleaning solution. This powerful scouring action reaches into minute crevices, which manual brushing cannot reach. The combination of energy and specially formulated solutions make ultrasonic cleaning the most effective method for removing gross and microscopic debris. I. Unpacking 1. Remove your BioSonic unit from the shipping container and inspect the unit for any possible damage which may have occurred during shipping.
2. Check the serial number of the unit (found on either the under side or rear of the unit) with the number on the shipping carton. If they are not identical, contact your dealer.
3. Fill out the warranty card and mail to Coltène/Whaledent Inc. within 10 days of receipt.
4. Select your desired location for the cleaner. It should be in close proximity to a sink or to waste lines to facilitate draining. The hose supplied with the unit will allow you to place the unit on either side of the sink or connect directly to the waste line.
5. Connect one end of the hose to the drain valve (Fig. 1) and place the other end in the sink. If desired, have your plumber connect the hose directly to the waste line with the appropriate fittings. II. Setup Instructions
1. CONNECT THE DISPLAY. Attach the connector on the front of the display module to the connector on the back top right-hand corner of the unit. Using the supplied hex L-key tool, install two screws (included) to secure the Liquid Crystal Display (LCD).
2. PLUG IN THE UNIT. Attach the UC125H’s electrical cord at the lower back of the unit and plug into a three-wire grounded AC outlet (in accordance with local regulations).
3. SELECT A LANGUAGE. Using the arrow keys on the LCD display, select the language code for your appropriate region. If you are unsure which code relates to your office, see the front page of the Owner’s Guide and choose the language you use in your office. Use the arrow keys to highlight the appropriate code. (The code will flash when it is selected.) Press ENTER to accept.
4. SET THE CLOCK. Use the arrow keys to set the time on the UC125H. Four numbers will appear on the screen. Use the “+” arrow key to increase the digit; use the “-” arrow key to decrease the digit. Press ENTER to accept.
NOTE: Users who select English must choose AM or PM and press ENTER to continue.
5. SET THE DATE. Use the arrow keys to set the date on the UC125H. “01/01/00” will appear on the screen. Each number will flash when it is selected. Use the “+” arrow key to increase the digit; use the “-” arrow key to decrease the digit. Press ENTER to accept. • US/Date Format (MM/DD/YY)
• International Date Format (DD/MM/YY)
6. SETUP COMPLETE. The display will now show the message “Setup Complete” in the language of your choice. Simply fill the tank of the UC125H with BioSonic Solution and water. You are ready to go! Enjoy your UC125H! The default screen will now display on the LCD screen, indicating that your unit is properly programmed.
NOTE: If a mistake is made while programming the LCD, a user may repeat this process by pressing and holding the “-” arrow key for 5 seconds.
MEMORY FEATURE: In the event of a power failure, your UC125H is equipped with a back-up power supply that will retain your programming information. This supply will be capable of storing user settings for a minimum of 12 days before it is again necessary to reprogram the unit. III. Filling Instructions
1. Insure that the blue drain handle (Fig. 2) is fully rotated clockwise to the “Closed” position. Fill the unit to the fill line indicator (Fig. 3).
2. Pour the required amount of ultrasonic cleaning solution (either BioSonic General Purpose UC30, BioSonic Germicidal Cleaner UC38/UC42 or BioSonic Enzymatic Ultrasonic Cleaner UC32) and water into the tank. Press the solution reset key.
3. Change the solution in the tank daily, or sooner if it appears soiled or discolored. When beakers are used, change the solution in the beaker after each use.
4. Drain unit by fully rotating the blue drain handle counter-clockwise to the “Open” position and allow solution to empty into sink or waste system. Rinse tank and fully rotate the blue drain handle clockwise to the “Closed” position. NOTE: Never operate the unit without solution. Operating a dry tank will void the warranty. Do not place objects directly on bottom of tank. Always use accessory baskets, cassette racks or beakers.
BioSonic® UC125H
4
IV. Degassing Instructions
Degassing the solution eliminates large air bubbles that interfere with the cleaning process. Anytime a new solution is mixed in the holding tank, the Degas key should be utilized. Degassing is necessary each time new solution is added to the tank. 1. PRESS THE DEGAS KEY. Press the Degas key from the default screen. To change the displayed degas time press ENTER. Use the arrow keys to select a degas time between 5 to 15 minutes. Now press the ENTER key to store your selected degas time in memory for easy recall.
2. Press STOP/START to begin degassing the solution.
3. WAIT. The UC125H will begin to cycle on and off during the degassing process. The timer will count down from the user-selected degas time. The words “Degas On” will be displayed.
4. DEGAS COMPLETE! Once the timer reaches zero, the words “Degas Complete” will be displayed. Your solution is now ready for use. Pressing any key will take the user back to the default screen. V. Operation Instructions
ACTIVATING THE UC125H. This section explains how to clean instruments with your BioSonic UC125H.
1. Press the STOP/START key from the default screen.
2. SELECT A LOAD. Three load icons will appear representing instrument baskets, beakers, and cassettes. To cycle through these icons:
a. PRESS THE ENTER KEY. Choose the appropriate icon for your particular load (the selected icon will flash).
b. USE THE ARROW KEYS to increase or decrease the amount of cleaning time between 5 and 60 minutes.
c. PRESS THE ENTER KEY to store the newly programmed default times. d. PRESS THE STOP/START KEY to begin cleaning.
3. WAIT. The UC125H will begin the cleaning process. The UC125H will continue to clean as the timer counts down the remaining minutes from the user-selected cleaning time. Once the timer reaches zero, “Cycle Complete” will display on the LCD until any button is pressed. The screen will then return to the default screen.
NOTE: The cleaning cycle can be stopped at any time by pressing the STOP/START button. The unit will then return to the user selected default value.
C-POWER: C-Power is an exclusive feature of the UC125H that will automatically adjust the power in the tank based on the load requirements. This function provides improved stability against varying solutions, solution level and instrument loads.
The BioSonic Ultrasonic cleaner should always be covered when operating to trap aerosols in the unit. The BioSonic UC125H is equipped with a “hinged” cover for easy one hand opening. The cover also conveniently lifts off for cleaning in spaces with low cabinets above the unit. VI. Heater Operation Instructions
ACTIVATING THE HEAT FEATURE. The UC125H model is equipped with a heater to rapidly increase the temperature of the solution in the tank.
NOTE: The temperature in the tank is displayed in the lower right-hand corner of the LCD screen. 1. PRESS THE HEAT KEY. Pressing the HEAT key to activate this feature will bring up a new screen on the display. The first time this feature is activated, the user must choose to use Celsius or Fahrenheit. Using the Arrow Keys, select “C” for Celsius or “F” for Fahrenheit and press ENTER.
2. SET THE DESIRED TEMPERATURE. Use the Arrow Keys to increase or decrease the tank temperature between 25°C (75°F) up to 60°C (140°F) in 5° degree increments. Press START to accept this temperature. The unit will now begin to heat.
3. WAIT. The UC125H will now display the default screen and begin to heat to the desired temperature. The Heat Toggle will now display “ON.” Once the desired temperature is reached, the Heat Toggle will display “OK.” Current temperature information will always be displayed in the lower right-hand corner of the LCD display.
CAUTION: The Heat feature of the UC125H is a pre-heat temperature only. Through prolonged use, the natural cavitation process will continue to heat the solution beyond the desired temperature. When this happens, the unit is still functioning properly.
NOTE: Once the unit is programmed to use either Celsius or Fahrenheit, it will not prompt you to do this again. If a mistake is made or the user desires to change from one to the other option, after entering heat mode, press and hold the HEAT key for three seconds and then repeat steps 1 through 3. VII. Solution Reset Operation Instructions
USING THE SOLUTION TIMER. The UC125H is equipped with a Solution Reset button. This feature tracks the amount of time your solution has been in use. The solution life is displayed in hours and minutes immediately above the SOLUTION RESET key. To utilize this feature:
1. PRESS THE SOLUTION RESET KEY once after every change of your ultrasonic cleaning solution. It is recommended to change your solution every 8 hours at a minimum, or whenever the solution becomes visibly cloudy or soiled. Coltène/Whaledent offers a variety of BioSonic ultrasonic cleaning solutions to meet your cleaning needs.
2. VERIFY. The solution timer should now display zeroes in the lower-left corner of the LCD display. VIII. Cleaning Methods A. In Tank
Basket Cleaning: Place parts to be cleaned into the basket (Fig. 5, 6, 7). Then place the basket directly into the main tank containing BioSonic General Purpose, Germicidal, or Enzymatic Cleaner. Clean according
5
Owner’s Guide
to instructions on bottle cleaner label.
B. In Beaker
For specialized item cleaning (e.g., prostheses): Place item into a beaker (Fig. 11), add desired cleaning solution (e.g., BioSonic Cement Remover or BioSonic Ammoniated General Purpose) according to the products directions. Suspend the beaker in the tank by the use of the beaker holder and place the beaker positioning ring around the beaker so that the bottom 2/3 will be suspended in the main bath.
C. Rack Cleaning
Place instruments to be cleaned in cassette. Then place cassette in wire rack and place the rack directly in the tank containing BioSonic General Purpose, Germicidal, or Enzymatic Cleaner. Clean according to instructions on bottle cleaner label (Fig. 8 & 9).
D. Instrument Draining
The UC125H is equipped with a bezel surrounding the tank. The bezel drains condensation back into the tank keeping the work area around the unit dry. The bezel also acts as a draining stand for the basket. Simply position the corner of the basket on two of the draining tabs (Fig. 4) at opposite corners of the tank and allow the solution to drain off instruments.
IX. Accessory Use Directions
A. Full Size FingerGuard™ BioSonic UC151XD Instrument Basket cleans large volumes of instruments and increases instrument processing efficiency. Use Directions:
Place instruments across ridges in bottom of basket. Place basket in tank and clean for desired time. After cleaning rinse basket and instruments thoroughly under running water.
B. Three-Quarter FingerGuard: The UC152XD Basket provides you with the versatility of cleaning a basket load of instruments and beaker cleaning for prostheses (or other small items) simultaneously.
Use Directions:
Place Basket into UC125H (Fig. 5). Place instruments across ridges in bottom of basket. Place basket in tank and clean for desired time.
C. Beaker Holder
Place three UC153 Beaker Holders on the bezel. Place positioning rings around the beaker so that the bottom 2/3 of the beaker is suspended in the main bath. Clean for desired time.
X. Environmental Conditions • Intended for indoor use.
• Maximum altitude 2000 m.
• Temperature range 5°C to 40°C (41°F to 104°F).
• Maximum relative humidity 80% for temperatures up to 31°C (88°F), decreasing linearly to 50% relative humidity at 40°C (104°F).
• Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage.
• Transient over voltages category II. • Pollution degree 2.
• Class I equipment. XI. Warnings
• Do not operate the unit without sufficient solution in the tank.
• Do not use water alone as the carrier bath. Water is not a satisfactory coupling agent unless a wetting agent is used. For best results use BioSonic General Purpose, BioSonic Germicidal Cleaner or BioSonic Enzymatic Ultrasonic Cleaning solutions.
• Do not place objects directly on bottom of the tank because they may block cavitation and cause buildup of heat in the transducer causing premature failure. Use only BioSonic UC125H accessory baskets, cassette racks or beakers to hold items.
• Do not use flammable liquids with BioSonic units.
• Do not restrict airflow. Provide adequate ventilation by maintaining at least 2 inches clearance at the back.
• Do not use acids in the stainless
steel tank.
• Do not subject unit to shock or impact.
• Do not immerse the unit in water. • Keep unit covered to reduce airborne contaminants.
• This product is intended for use only by trained healthcare professionals.
• Use this product in accordance with the instructions in the owner’s guide.
• Unspecified or improper use of this product may impair safety protection.
• Do not place fingers or hands in the ultrasonic tank while in operation.
• Before cleaning items of considerable value ultrasonically, consult the item’s manufacturer for cleaning precautions.
• Use only the supplied power cord or equivalent.
• The power off selection from the keypad does not turn off main power to the unit.
• Operation for long periods of time will cause the cleaning solution to get very hot. To avoid scalding, do not place hands in the solution or spill any onto you. Be careful when removing an item from the tank as it may also be hot.
• Do not disassemble the unit. There are no user serviceable parts inside. XII. Order Information Cat.No
Description
Unit and accessories: UC125H
Qty
100 Unit (100V, 50/60 Hz) 1 115 Unit (115V, 50/60 Hz) 1 230CE Unit (230V, 50/60 Hz) 1 230UK Unit (230V, 50/60 Hz) 1 240 Unit (240V, 50/60 Hz) 1 UC155XD Accessory Kit 3/4 FingerGuard Basket, (2) Beakers, (2) Covers, (2) Positioning Rings and (2) Beaker Holders UC151XD Full Size FingerGuard Basket 1 UC152XD 3/4 FingerGuard Basket 1 UC153 One Beaker Holder 1 UC350 Replacement Drain Filters 3 UC360 Replacement Drain Hose 1 UC53 600 ml Beaker with Cover and Positioning Ring 1
BioSonic® UC125H
6
Tank Solutions: UC30 UC31 UC32 UC34 UC38 UC39 UC42
General Purpose Cleaning Solution Concentrate in MeterDose® Dispensing Bottle (16 oz.) 473 ml General Purpose Cleaner + Super Rust Inhibitor 20 Single Dose Packets 15 ml Enzymatic Ultrasonic Cleaner Concentrate in MeterDose Dispensing Bottle (8 oz.) 236 ml Plaster & Stone Remover Powder Concentrate in Jar (1.85 lbs.) 840 g Germicidal Ultrasonic Cleaning Concentrate in MeterDose Dispensing Bottle (8 oz.) 236 ml Cement Remover Concentrate in MeterDose Dispensing Bottle (16 oz.) 473 ml Germicidal Ultrasonic Cleaning Concentrate in MeterDose Dispensing Bottle (16 oz.) 473 ml
XIII. Maintenance A. Cleaning
All components should be cleaned each day by wiping with a soft cloth moistened with a commercially available mineral deposit remover followed by cleaning with isopropyl alcohol or a mild disinfectant.
CAUTION: Do not use petroleum based solvents, iodophors or phenolic based products. (Iodophors and phenolics can stain the surface of the unit.) Clean up all liquid spills immediately. Rust deposits on tank can be cleaned/removed with a commercially available stainless steel cleaner.
B. Inspecting the Power Cord Periodically check the power cord for damage and see that the connectors at both ends of the cord are fully seated.
4. Examine the fuse (Fig. 13). If blown, replace with a fuse that matches the unit’s voltage and current (see section XIV). If fuse appears to be in good condition return the unit to your dealer or Coltène/Whaledent.
5. Plug the power cord into the power cord receptacle on the rear of the unit. Then plug the power cord into the wall receptacle. XIV. Unit Specifications Transducers: 2 UC125H unit
100 [100V 50/60 Hz, 337W (max.)] 115 [115V 50/60 Hz, 337W (max.)]
230CE [230V 50/60 Hz, 337W (max.)]
230UK [230V 50/60 Hz, 337W (max.)]
240 [240V 50/60 Hz, 337W (max.)] Operating Frequency:
Tank Capacity:
1.25 gallon or 4.75 L
Overall Dimensions:
16.5 x 10.5 x 13.8 in. 42 X 27 X 35 cm
Use Capacity:
Tank Dimensions: Fuse Type:
Time Lag:
100-115V Heated units 230-240V Heated units
2. The fuse compartment is located directly next to the power cord receptacle. To access, take a small, flat-headed screwdriver to the tab on the fuse holder. Apply slight pressure to release the fuse holder (Fig. 12).
3. Gently remove the fuse from the fuse holder.
1 gallon or 3.78 L
11.25 x 5.62 x 5 in. 28.6 x 14.3 x 12.7 cm
5 x 20 mm Cartridge
IEC 60127-2 Sheet 3
EMC: Class A
4 Amp 2.5 Amp
IEC 60417-5041 Caution, Hot Surface
C. Changing the Fuse
1. Remove the power cord from the wall outlet first and then from the power cord receptacle on the rear of the unit.
52 kHz
!
ISO 3864, No B.3. Caution: Refer to accompanying documents ISO 3864, No B.3.6 Caution: Risk of electric shock
XV. Warranty Information
BioSonic Ultrasonic Cleaners are carefully manufactured to meet stringent quality assurance requirements. BioSonic products are manufactured from new parts or new and serviceable used parts. Regardless, our warranty terms apply. This product has been developed specifically for use in dentistry and is intended to be operated only by qualified dental professionals in accordance with the instructions contained in this guide. However, notwithstanding anything contained herein to the contrary, the user shall at all times be solely responsible for determining the suitability of the product for the intended purpose and the method of its use. Any guidance on application technology offered by or on behalf of the manufacturer, whether written, verbal or by demonstration, shall not relieve the dental professional of his/her obligation to control the product and to make all professional judgments regarding its use. BioSonic Ultrasonic Cleaners are warranted in accordance with the terms of a written Certificate of Limited Warranty accompanying each product. Except for the warranties specifically set forth in the Certificate of Limited Warranty, Coltène/Whaledent Inc. provides no warranties or guarantees of any kind covering the product, expressed or implied, including, without limitation, any warranties as to merchantability or fitness for a particular purpose. The purchaser/user is referred to the Certificate of Limited Warranty for all of the terms, conditions and limitations of the warranty covering this product. This Section XV of the user manual is not intended to in any way modify or add to the warranty provided in the Certificate of Limited Warranty. Any claim for damage or breakage to the product in transit should be made to the carrier promptly upon discovery. C/W does not warrant the product against shipping damage.
BioSonic® UC125H
56
ÂÚ›ÙˆÛË ‰È·ÎÔ‹˜ Ú‡̷ÙÔ˜, ÙÔ UC125H ‰È·ı¤ÙÂÈ ¤Ó· ÂʉÚÈÎfi Û‡ÛÙËÌ· ÙÚÔÊÔ‰ÔÛ›·˜, Ì ÙÔ ÔÔ›Ô ‰È·ÙËÚÔ‡ÓÙ·È ÔÈ ÏËÚÔÊÔڛ˜ ÚÔÁÚ·ÌÌ·ÙÈÛÌÔ‡. ∞˘Ùfi ÙÔ Û‡ÛÙËÌ· ÙÚÔÊÔ‰ÔÛ›·˜ ¤¯ÂÈ ÙËÓ ÈηÓfiÙËÙ· Ó· ·ÔıË·ÂÈ ÙȘ Ú˘ıÌ›ÛÂȘ ÙÔ˘ ¯Ú‹ÛÙË ÁÈ· ÂÚ›Ô˘ 12 Ë̤Ú˜ ÚÈÓ ¯ÚÂÈ·ÛÙ› Ó· Í·Ó·ÚÔÁÚ·ÌÌ·Ù›ÛÂÙ ÙË ÌÔÓ¿‰·. III. √‰ËÁ›Â˜ Ï‹ÚˆÛ˘ 1. µÂ‚·Èˆı›Ù fiÙÈ ¤¯ÂÙ ÂÚÈÛÙÚ¤„ÂÈ ÙËÓ ÌÏ ‚·Ï‚›‰· ·ÔÛÙÚ¿ÁÁÈÛ˘ Û¯. 2) ÚÔ˜ Ù· ‰ÂÍÈ¿ ̤¯ÚÈ ÙÔ (Û Ù¤ÚÌ· ÛÙË ı¤ÛË “∫ÏÂÈÛÙ‹”. °ÂÌ›ÛÙ ÙË ÌÔÓ¿‰·, ̤¯ÚÈ ÙËÓ ¤Ó‰ÂÈÍË Û¯. 3). ÁÚ·ÌÌ‹˜ Ï‹ÚˆÛ˘ (Û 2. °ÂÌ›ÛÙ ÙÔÓ Î¿‰Ô Ì ÙËÓ ··ÈÙÔ‡ÌÂÓË ÔÛfiÙËÙ· ηı·ÚÈÛÙÈÎÔ‡ ‰È·Ï‡Ì·ÙÔ˜ ˘ÂÚ‹¯ˆÓ (BioSonic General Purpose UC30, BioSonic Germicidal Cleaner UC38 ‹ BioSonic Enzymatic Ultrasonic Cleaner UC32) Î·È ÓÂÚfi. ¶π∂™∆∂ ∆√ ∫√Àª¶π ª∏¢∂¡π™ª√À ¢π∞§Àª∞∆√™. 3. ¶Ú¤ÂÈ Ó· ·ÓÙÈηıÈÛٿ٠ÙÔ ‰È¿Ï˘Ì· ÛÙÔÓ Î¿‰Ô ηıËÌÂÚÈÓ¿ ‹ Û˘¯ÓfiÙÂÚ·, Â¿Ó Ê·›ÓÂÙ·È fiÙÈ ¤¯ÂÈ ‚ÚÔÌ›ÛÂÈ ‹ ·Ô¯ÚˆÌ·ÙÈÛÙ›. ŸÙ·Ó ¯ÚËÛÈÌÔÔț٠ٷ ÁÈ¿ÏÈÓ· ‰Ô¯Â›·, ·ÓÙÈηıÈÛٿ٠ÙÔ ‰È¿Ï˘Ì· ÛÙÔ ÁÈ¿ÏÈÓÔ ‰Ô¯Â›Ô ÌÂÙ¿ ·fi οı ¯Ú‹ÛË. 4. ∞‰ÂÈ¿ÛÙ ÙË ÌÔÓ¿‰· ÂÚÈÛÙÚ¤ÊÔÓÙ·˜ ÙËÓ ÌÏ ‚·Ï‚›‰· ·ÔÛÙÚ¿ÁÁÈÛ˘ ÚÔ˜ Ù· ·ÚÈÛÙÂÚ¿ ̤¯ÚÈ ÙÔ Ù¤ÚÌ· ÛÙË ı¤ÛË “∞ÓÔȯً” Î·È ‰ÈÔ¯ÂÙ‡ÛÙ ÙÔ ‰È¿Ï˘Ì· Û ¤Ó· ÓÂÚÔ¯‡ÙË ‹ ÛÙÔ Û‡ÛÙËÌ· ·Ô¯¤Ù¢Û˘. •ÂχÓÙ ÙÔ ‰Ô¯Â›Ô Î·È ÂÚÈÛÙÚ¤„Ù ÙËÓ ÌÏ ‚·Ï‚›‰· ·ÔÛÙÚ¿ÁÁÈÛ˘ ÚÔ˜ Ù· ‰ÂÍÈ¿ ̤¯ÚÈ ÙÔ Ù¤ÚÌ· ÛÙË ı¤ÛË “∫ÏÂÈÛÙ‹”. ™∏ª∂πø™∏: ªË ı¤ÙÂÙ ÔÙ¤ ÙË ÌÔÓ¿‰· Û ÏÂÈÙÔ˘ÚÁ›· ¯ˆÚ›˜ ‰È¿Ï˘Ì·. ™Â ÂÚ›ÙˆÛË ¯Ú‹Û˘ Ù˘ ÌÔÓ¿‰·˜ Ì ¿‰ÂÈÔ Î¿‰Ô, ·‡ÂÈ Ó· ÈÛ¯‡ÂÈ Ë ÂÁÁ‡ËÛË. ªËÓ ÙÔÔıÂÙ›Ù ·ÓÙÈΛÌÂÓ· ·Â˘ı›·˜ ÛÙÔÓ ¿ÙÔ ÙÔ˘ ‰Ô¯Â›Ô˘. ÃÚËÛÈÌÔÔț٠¿ÓÙÔÙ ٷ ηϿıÈ·, Ù· Ï·›ÛÈ· ηÛÂÙÒÓ ‹ Ù· ÁÈ¿ÏÈÓ· ‰Ô¯Â›·. IV. √‰ËÁ›Â˜ ÂÍ·¤ÚˆÛ˘ ªÂ ÙËÓ ÂÍ·¤ÚˆÛË ÙÔ˘ ‰È·Ï‡Ì·ÙÔ˜ ·ÔÌ·ÎÚ‡ÓÔÓÙ·È ÔÈ ÌÂÁ¿Ï˜ Ê˘Û·Ï›‰Â˜
·¤Ú· Ô˘ ÂÌÔ‰›˙Ô˘Ó ÙÔÓ ·ÔÙÂÏÂÛÌ·ÙÈÎfi ηı·ÚÈÛÌfi. ∫¿ı ÊÔÚ¿ Ô˘ ÚÔÛÙ›ıÂÙ·È Ó¤· ÔÛfiÙËÙ· ‰È·Ï‡Ì·ÙÔ˜ ÛÙÔÓ Î¿‰Ô Ù˘ Û˘Û΢‹˜, Ú¤ÂÈ Ó· ¯ÚËÛÈÌÔÔÈÂ›Ù·È ÙÔ ÎÔ˘Ì› ÂÍ·¤ÚˆÛ˘. ∂Í·¤ÚˆÛË ··ÈÙÂ›Ù·È Î¿ı ÊÔÚ¿ Ô˘ ÚÔÛÙ›ıÂÙ·È Ó¤· ÔÛfiÙËÙ· ‰È·Ï‡Ì·ÙÔ˜ ÛÙÔÓ Î¿‰Ô. 1. ¶π∂™∆∂ ∆√ ∫√Àª¶π ∂•∞∂ƒø™∏™. ¶È¤ÛÙ ÙÔ ÎÔ˘Ì› ÂÍ·¤ÚˆÛ˘ ÛÙË ‚·ÛÈ΋ ÔıfiÓË. °È· Ó· ·ÏÏ¿ÍÂÙ ÙÔ ¯ÚfiÓÔ ÂÍ·¤ÚˆÛ˘ Ô˘ ÂÌÊ·Ó›˙ÂÙ·È, ȤÛÙ ÙÔ ENTER. ÃÚËÛÈÌÔÔÈ‹ÛÙ ٷ ÎÔ˘ÌÈ¿ ‚¤ÏÔ˘˜, ÁÈ· Ó· ÂÈϤÍÂÙ ÌÈ· ÙÈÌ‹ ¯ÚfiÓÔ˘ ÂÍ·¤ÚˆÛ˘ ·Ó¿ÌÂÛ· Û 5 Î·È 15 ÏÂÙ¿. ∆ÒÚ· ȤÛÙ ÙÔ ÎÔ˘Ì› ENTER, ÁÈ· Ó· ·ÔıË·ÛÂÙ ÙÔ ¯ÚfiÓÔ ÂÍ·¤ÚˆÛ˘ Ô˘ ÂÈϤͷÙ ÛÙË ÌÓ‹ÌË, ¤ÙÛÈ ÒÛÙ ӷ ÌÔÚ›Ù ӷ ÙÔÓ Î·Ï¤ÛÂÙ ‡ÎÔÏ·. 2. ¶È¤ÛÙ ÙÔ ÎÔ˘Ì› STOP/START, ÁÈ· Ó· ÍÂÎÈÓ‹ÛÂÙ ÙËÓ ÂÍ·¤ÚˆÛË ÙÔ˘ ‰È·Ï‡Ì·ÙÔ˜. 3. ¶∂ƒπª∂¡∂∆∂. ∆Ô UC125H ı· ·Ú¯›ÛÂÈ Ó· ÂÓÂÚÁÔÔÈÂ›Ù·È Î·È Ó· ·ÂÓÂÚÁÔÔÈÂ›Ù·È ÂÓ·ÏÏ¿Í Î·Ù¿ ÙË ‰È·‰Èηۛ· ÂÍ·¤ÚˆÛ˘. √ ¯ÚÔÓԉȷÎfiÙ˘ ı· Ú·ÁÌ·ÙÔÔÈ‹ÛÂÈ ·ÓÙ›ÛÙÚÔÊË Ì¤ÙÚËÛË ÙÔ˘ ¯ÚfiÓÔ˘ ÂÍ·¤ÚˆÛ˘ Ô˘ ¤¯ÂÈ ÂÈϯı› ·fi ÙÔ ¯Ú‹ÛÙË. ∏ ÊÚ¿ÛË “∂Í·¤ÚˆÛË ÂÓÂÚÁÔÔÈË̤ÓË” ÂÌÊ·Ó›˙ÂÙ·È. 4. ∏ ∂•∞∂ƒø™∏ √§√∫§∏ƒø£∏∫∂! ªfiÏȘ Ô ¯ÚÔÓԉȷÎfiÙ˘ ÊÙ¿ÛÂÈ ÛÙÔ Ìˉ¤Ó, ÂÌÊ·Ó›˙ÂÙ·È Ë ÊÚ¿ÛË “∏ ÂÍ·¤ÚˆÛË ÔÏÔÎÏËÚÒıËΔ. ∆Ô ‰È¿Ï˘Ì¿ Û·˜ Â›Ó·È ÙÒÚ· ¤ÙÔÈÌÔ ÁÈ· ¯Ú‹ÛË. ªÂ ÙÔ ¿ÙËÌ· ÔÔÈÔ˘‰‹ÔÙ ÎÔ˘ÌÈÔ‡, Ô ¯Ú‹ÛÙ˘ ı· ÂÈÛÙÚ¤„ÂÈ ÛÙË ‚·ÛÈ΋ ÔıfiÓË. V. √‰ËÁ›Â˜ ¯Ú‹Û˘ ∂¡∂ƒ°√¶√π∏™∏ ∆√À UC125H. ™Â ·˘Ù‹Ó ÙËÓ ÂÓfiÙËÙ· ÂÂÍËÁÂ›Ù·È Ô ÙÚfiÔ˜ ηı·ÚÈÛÌÔ‡ ÙˆÓ ÂÚÁ·Ï›ˆÓ Ì ÙÔ BioSonic UC125H. 1. ¶È¤ÛÙ ÙÔ ÎÔ˘Ì› STOP/START ÛÙË ‚·ÛÈ΋ ÔıfiÓË. 2. ∂¶π§∂•∆∂ º√ƒ∆π√. £· ÂÌÊ·ÓÈÛÙÔ‡Ó ÙÚ›· ÂÈÎÔÓ›‰È· ÊÔÚÙ›Ô˘, Ù· ÔÔ›· ·ÓÙÈÛÙÔÈ¯Ô‡Ó Û ηϿıÈ·, ÁÈ¿ÏÈÓ· ‰Ô¯Â›· Î·È Î·Û¤Ù˜ ÂÚÁ·Ï›ˆÓ. °È· Ó· ÂÈϤÍÂÙ ÂÓ·ÏÏ¿Í ·˘Ù¿ Ù· ÂÈÎÔÓ›‰È·:
a. ¶π∂™∆∂ ∆√ ∫√Àª¶π ENTER. ∂ÈϤÍÙ ÙÔ ·Ó¿ÏÔÁÔ ÂÈÎÔÓ›‰ÈÔ ÁÈ· ÙÔ ÂοÛÙÔÙ ÊÔÚÙ›Ô (ÙÔ ÂÈÏÂÁ̤ÓÔ ÂÈÎÔÓ›‰ÈÔ ·Ó·‚ÔÛ‚‹ÓÂÈ). b. Ã∏™πª√¶√π∏™∆∂ ∆∞ ∫√Àª¶π∞ µ∂§√À™, ÁÈ· Ó· ·˘Í‹ÛÂÙ ‹ Ó· ÌÂÈÒÛÂÙ ÙË ‰È¿ÚÎÂÈ· ÙÔ˘ ¯ÚfiÓÔ˘ ηı·ÚÈÛÌÔ‡ ÌÂٷ͇ 5 Î·È 60 ÏÂÙÒÓ. c. ¶π∂™∆∂ ∆√ ∫√Àª¶π ENTER, ÁÈ· Ó· ·ÔıË·ÛÂÙ ÙÔ˘˜ Ó¤Ô˘˜ ÚÔÁÚ·ÌÌ·ÙÈṲ̂ÓÔ˘˜ ¯ÚfiÓÔ˘˜. d. ¶π∂™∆∂ ∆√ ∫√Àª¶π STOP/START, ÁÈ· Ó· ·Ú¯›ÛÂÙ ÙÔÓ Î·ı·ÚÈÛÌfi. 3. ¶∂ƒπª∂¡∂∆∂. ∆Ô UC125H ı· ·Ú¯›ÛÂÈ ÙË ‰È·‰Èηۛ· ηı·ÚÈÛÌÔ‡. ∆Ô UC125H ı· Û˘Ó¯›ÛÂÈ Ó· ηı·Ú›˙ÂÈ Î·ıÒ˜ Ô ¯ÚfiÓÔ˜ ÌÂÙÚ¿ÂÈ Û ·ÓÙ›ÛÙÚÔÊË Ì¤ÙÚËÛË Ù· ÏÂÙ¿ Ô˘ ˘ÔÏ›ÔÓÙ·È ·fi ÙÔ ¯ÚfiÓÔ Î·ı·ÚÈÛÌÔ‡ Ô˘ ¤¯ÂÈ ÂÈϤÍÂÈ Ô ¯Ú‹ÛÙ˘. ªfiÏȘ Ô ¯ÚÔÓԉȷÎfiÙ˘ ÊÙ¿ÛÂÈ ÛÙÔ Ìˉ¤Ó, ı· ÂÌÊ·ÓÈÛÙ› ÙÔ Ì‹Ó˘Ì· “√ ·ÎÏÔ˜ ÔÏÔÎÏËÚÒıËΔ ÛÙËÓ ÔıfiÓË LCD, ̤¯ÚÈ Ó· ·Ù‹ÛÂÙ οÔÈÔ ¿ÏÏÔ ÎÔ˘Ì›. ™ÙË Û˘Ó¤¯ÂÈ· ÛÙËÓ ÔıfiÓË ı· ÂÌÊ·ÓÈÛÙ› Ë ‚·ÛÈ΋ ÔıfiÓË. ™ËÌ›ˆÛË: √ ·ÎÏÔ˜ ηı·ÚÈÛÌÔ‡ ÌÔÚ› Ó· ‰È·ÎÔ› ·Ó¿ ¿Û· ÛÙÈÁÌ‹ Ì ÙÔ ¿ÙËÌ· ÙÔ˘ ÎÔ˘ÌÈÔ‡ STOP/START. ™ÙË Û˘Ó¤¯ÂÈ· Ë ÌÔÓ¿‰· ı· ÂÈÛÙÚ¤„ÂÈ ÛÙËÓ ÙÈÌ‹ Ô˘ ¤¯ÂÈ ÂÈϤÍÂÈ ˆ˜ ÚÔÂÈÏÔÁ‹ Ô ¯Ú‹ÛÙ˘. C-POWER: ∆Ô C-Power Â›Ó·È ÌÈ· ·ÔÎÏÂÈÛÙÈ΋ ÏÂÈÙÔ˘ÚÁ›· ÙÔ˘ UC125H, Ë ÔÔ›· ÚÔÛ·ÚÌfi˙ÂÈ ·˘ÙfiÌ·Ù· ÙËÓ ÈÛ¯‡ ÛÙÔÓ Î¿‰Ô ‚¿ÛÂÈ ÙˆÓ ··ÈÙ‹ÛÂˆÓ ÊÔÚÙ›Ô˘. ∞˘Ù‹ Ë ÏÂÈÙÔ˘ÚÁ›· ·Ú¤¯ÂÈ ‚ÂÏÙȈ̤ÓË ÛÙ·ıÂÚfiÙËÙ· Û ۯ¤ÛË Ì ‰È·ÊÔÚÂÙÈο ‰È·Ï‡Ì·Ù·, ‰È·ÊÔÚÂÙÈΤ˜ ÛÙ¿ı̘ ‰È·Ï‡Ì·ÙÔ˜ Î·È ‰È·ÊÔÚÂÙÈο ÊÔÚÙ›· ÂÚÁ·Ï›ˆÓ. ∆Ô ™‡ÛÙËÌ· ηı·ÚÈÛÌÔ‡ ˘ÂÚ‹¯ˆÓ BioSonic Ú¤ÂÈ Ó· Â›Ó·È ¿ÓÙ· Î·Ï˘Ì̤ÓÔ Î·Ù¿ ÙË ÏÂÈÙÔ˘ÚÁ›· ÙÔ˘, ÒÛÙ ӷ ÌË ‰È·Ê‡ÁÔ˘Ó Ù· ·ÂÚÔχ̷ٷ ·fi ÙË ÌÔÓ¿‰·. ∆Ô BioSonic UC125H Â›Ó·È ÂÍÔÏÈṲ̂ÓÔ Ì ¤Ó· Î¿Ï˘ÌÌ· Ì “¿ÚıÚˆÛË” ÁÈ· ‡ÎÔÏÔ ¿ÓÔÈÁÌ· Ì ÙÔ ¤Ó· ¯¤ÚÈ. ∆Ô Î¿Ï˘ÌÌ· ·Ó˘„ÒÓÂÙ·È Â›Û˘ ‡ÎÔÏ· ÁÈ· ÙÔÓ Î·ı·ÚÈÛÌfi Û ¯ÒÚÔ˘˜ fiÔ˘ ˘¿Ú¯Ô˘Ó ¯·ÌËÏ¿ ÓÙÔ˘Ï¿È· ¿Óˆ ·fi ÙË ÌÔÓ¿‰·.
BioSonic® UC125H
58
·ÔÙÂÏÂÛÌ·ÙÈÎfiÙËÙ· ÂÂÍÂÚÁ·Û›·˜ ÙˆÓ ÂÚÁ·Ï›ˆÓ. √‰ËÁ›Â˜ ¯Ú‹Û˘: ∆ÔÔıÂÙ‹ÛÙ ٷ ÂÚÁ·Ï›· οıÂÙ· ÚÔ˜ ÙȘ Ó¢ÚÒÛÂȘ Ô˘ ˘¿Ú¯Ô˘Ó ÛÙÔÓ ¿ÙÔ ÙÔ˘ ηϷıÈÔ‡. ∆ÔÔıÂÙ‹ÛÙ ÙÔ Î·Ï¿ıÈ ÛÙÔÓ Î¿‰Ô Î·È Î·ı·Ú›ÛÙ ÁÈ· fiÛÔ ¯ÚfiÓÔ ı¤ÏÂÙÂ. ∞ÊÔ‡ ÔÏÔÎÏËÚˆı› Ô Î·ı·ÚÈÛÌfi˜, ÍÂχÓÂÙ ηϿ ÙÔ Î·Ï¿ıÈ Î·È Ù· ÂÚÁ·Ï›· Ì ÙÚ¯ԇÌÂÓÔ ÓÂÚfi. B. FingerGuardì ÙÚÈÒÓ ÙÂÙ¿ÚÙˆÓ: ∆Ô ∫·Ï¿ıÈ UC152XD Û·˜ ·Ú¤¯ÂÈ ÙË ‰˘Ó·ÙfiÙËÙ· ηı·ÚÈÛÌÔ‡ ÂÓfi˜ ηϷıÈÔ‡ Ì ÂÚÁ·Ï›· Î·È Ù·˘Ùfi¯ÚÔÓÔ˘ ηı·ÚÈÛÌÔ‡ ÚÔı¤ÛÂˆÓ (‹ ¿ÏÏˆÓ ÌÈÎÚÒÓ ·ÓÙÈÎÂÈ̤ӈÓ) Û ÁÈ¿ÏÈÓÔ ‰Ô¯Â›Ô. √‰ËÁ›Â˜ ¯Ú‹Û˘: ∆ÔÔıÂÙ‹ÛÙ ÙÔ ∫·Ï¿ıÈ ÛÙË Û¯. 5). ÌÔÓ¿‰· UC125H (Û ∆ÔÔıÂÙ‹ÛÙ ٷ ÂÚÁ·Ï›· οıÂÙ· ÚÔ˜ ÙȘ Ó¢ÚÒÛÂȘ Ô˘ ˘¿Ú¯Ô˘Ó ÛÙÔÓ ¿ÙÔ ÙÔ˘ ηϷıÈÔ‡. ∆ÔÔıÂÙ‹ÛÙ ÙÔ Î·Ï¿ıÈ ÛÙÔÓ Î¿‰Ô Î·È Î·ı·Ú›ÛÙ ÁÈ· fiÛÔ ¯ÚfiÓÔ ı¤ÏÂÙ C. ∂Í¿ÚÙËÌ· ÛÙ‹ÚÈ͢ ÁÈ¿ÏÈÓÔ˘ ‰Ô¯Â›Ô˘: ∆ÔÔıÂÙ‹ÛÙ ÙÚ›· ∂Í·ÚÙ‹Ì·Ù· ÛÙ‹ÚÈ͢ ÁÈ¿ÏÈÓˆÓ ‰Ô¯Â›ˆÓ UC153 ÛÙÔ ¯Â›ÏÔ˜ ·ÔÛÙÚ¿ÁÁÈÛ˘. ∆ÔÔıÂÙ‹ÛÙ ‰·ÎÙ˘Ï›Ô˘˜ ÙÔÔı¤ÙËÛ˘ Á‡Úˆ ·fi ÙÔ ÁÈ¿ÏÈÓÔ ‰Ô¯Â›Ô ÒÛÙ ÙÔ 2/3 ÙÔ˘˜ Ó· Â›Ó·È ‚˘ıÈṲ̂ӷ ̤۷ ÛÙÔ Î˘Ú›ˆ˜ ÏÔ˘ÙÚfi. ∫·ı·Ú›ÛÙ ÁÈ· fiÛÔ ¯ÚfiÓÔ ı¤ÏÂÙÂ. X. ™˘Óı‹Î˜ ÂÚÈ‚¿ÏÏÔÓÙÔ˜ • ¶ÚÔÔÚ›˙ÂÙ·È ÁÈ· ¯Ú‹ÛË Û ÂÛˆÙÂÚÈÎfi ¯ÒÚÔ. • ª¤ÁÈÛÙÔ ˘„fiÌÂÙÚÔ 2.000 m. • ∂‡ÚÔ˜ ıÂÚÌÔÎÚ·Û›·˜ 5 ÆC ¤ˆ˜ 40 ÆC. • ª¤ÁÈÛÙË Û¯ÂÙÈ΋ ˘ÁÚ·Û›· 80% ÁÈ· ıÂÚÌÔÎڷۛ˜ ¤ˆ˜ 31 ÆC, ÌÂÈÒÓÂÙ·È ÁÚ·ÌÌÈο Û ۯÂÙÈ΋ ˘ÁÚ·Û›· 50% ÛÙÔ˘˜ 40 ÆC. • √È ·˘ÍÔÌÂÈÒÛÂȘ Ù˘ Ù¿Û˘ ÙÔ˘ Ú‡̷ÙÔ˜ ·fi ÙÔ ‰›ÎÙ˘Ô ‰ÂÓ Ú¤ÂÈ Ó· ˘ÂÚ‚·›ÓÔ˘Ó ± 10% Ù˘ ÔÓÔÌ·ÛÙÈ΋˜ Ù¿Û˘. • ªÂÙ·‚·ÙÈ΋ ˘¤ÚÙ·ÛË II.
• µ·ıÌfi˜ Ú‡·ÓÛ˘ 2.
• ∂ÍÔÏÈÛÌfi˜ ηÙËÁÔÚ›·˜ I. XI. ¶ÚÔÂȉÔÔÈ‹ÛÂȘ
• ªË ı¤ÙÂÙ Û ÏÂÈÙÔ˘ÚÁ›· ÙË ÌÔÓ¿‰· Â¿Ó ‰ÂÓ ˘¿Ú¯ÂÈ Â·Ú΋˜ ÔÛfiÙËÙ· ‰È·Ï‡Ì·ÙÔ˜ ÛÙÔ ‰Ô¯Â›Ô.
• ªË ¯ÚËÛÈÌÔÔț٠ÛΤÙÔ ÓÂÚfi ˆ˜ Î·Ù·Ï˘ÙÈÎfi ÏÔ˘ÙÚfi. ∆Ô ÓÂÚfi ‰ÂÓ ·ÔÙÂÏ› ÈηÓÔÔÈËÙÈÎfi ·Ú¿ÁÔÓÙ· Û‡˙¢Í˘ ÂÎÙfi˜ Â¿Ó ¯ÚËÛÈÌÔÔÈÂ›Ù·È ·Ú¿ÁÔÓÙ·˜ ‰È·‚ÚÔ¯‹˜. °È· ηχÙÂÚ· ·ÔÙÂϤÛÌ·Ù·, ¯ÚËÛÈÌÔÔț٠¢È¿Ï˘Ì· ÁÂÓÈ΋˜ ¯Ú‹Û˘ BioSonic, ªÈÎÚÔ‚ÈÔÎÙfiÓÔ ‰È¿Ï˘Ì· BioSonic ‹ ∂Ó˙˘Ì·ÙÈÎfi ‰È¿Ï˘Ì· ÁÈ· ηı·ÚÈÛÌfi ˘ÂÚ‹¯ˆÓ BioSonic. • ªËÓ ÙÔÔıÂÙ›Ù ·ÓÙÈΛÌÂÓ· ·Â˘ı›·˜ ÛÙÔÓ ¿ÙÔ ÙÔ˘ ‰Ô¯Â›Ô˘, ‰ÈfiÙÈ ·˘Ù¿ ÌÔÚ› Ó· ·ÚÂÌÔ‰›ÛÔ˘Ó ÙË ÛËÏ·›ˆÛË Î·È Ó· ÚÔηϤÛÔ˘Ó ·‡ÍËÛË Ù˘ ıÂÚÌfiÙËÙ·˜ ÛÙÔ ÌÂÙ·ÙÚÔ¤· Ì ·ÔÙ¤ÏÂÛÌ· ÚfiˆÚË ˙ËÌÈ¿. ÃÚËÛÈÌÔÔț٠ÌfiÓÔ Î·Ï¿ıÈ·, Ï·›ÛÈ· ηÛÂÙÒÓ ‹ ÂÍ·ÚÙ‹Ì·Ù· ÛÙ‹ÚÈ͢ ÁÈ¿ÏÈÓˆÓ ‰Ô¯Â›ˆÓ BioSonic UC125H Ô˘ ·ÁÔÚ¿˙ÂÙ ˆ˜ ·ÍÂÛÔ˘¿Ú, ÁÈ· ÙË Û˘ÁÎÚ¿ÙËÛË ÙˆÓ ·ÓÙÈÎÂÈ̤ӈÓ. • ªË ¯ÚËÛÈÌÔÔț٠‡ÊÏÂÎÙ· ˘ÁÚ¿ Ì ÙȘ ÌÔÓ¿‰Â˜ BioSonic • ªËÓ ÂÚÈÔÚ›˙ÂÙ ÙË ÚÔ‹ ÙÔ˘ ·¤Ú·. ¶·Ú¤¯ÂÙ ·Ú΋ ·ÂÚÈÛÌfi ‰È·ÙËÚÒÓÙ·˜ ÙÔ˘Ï¿¯ÈÛÙÔÓ 5 ÂηÙÔÛÙ¿ ÂχıÂÚÔ ¯ÒÚÔ ÛÙÔ ›Ûˆ ̤ÚÔ˜. • ªË ¯ÚËÛÈÌÔÔț٠Ôͤ· ÛÙÔ ‰Ô¯Â›Ô Ô˘ Â›Ó·È Î·Ù·Û΢·Ṳ̂ÓÔ ·fi ·ÓÔÍ›‰ˆÙÔ ¯¿Ï˘‚·. • ªËÓ ˘Ô‚¿ÏÏÂÙ ÙË ÌÔÓ¿‰· Û ÎÚÔ‡ÛÂȘ ‹ ÚfiÛÎÚÔ˘ÛË. • ªË ‚˘ı›˙ÂÙ ÙË ÌÔÓ¿‰· Û ÓÂÚfi. • ¢È·ÙËÚ›Ù ÙË ÌÔÓ¿‰· Ì ÙÔ Î¿Ï˘ÌÌ· ÎÏÂÈÛÙfi ÒÛÙ ӷ ÂÚÈÔÚ›˙ÔÓÙ·È ÔÈ ÂÓ·¤ÚÈÔÈ Ú‡ÔÈ. • ∞˘Ùfi ÙÔ ÚÔ˚fiÓ ÚÔÔÚ›˙ÂÙ·È ÁÈ· ¯Ú‹ÛË ÌfiÓÔ ·fi ÂÎ·È‰Â˘Ì¤ÓÔ˘˜ ·ÁÁÂÏ̷ٛ˜ ÙÔ˘ ÙÔ̤· Ù˘ È·ÙÚÈ΋˜ ÂÚ›ı·Ï„˘. • ÃÚËÛÈÌÔÔț٠·˘Ùfi ÙÔ ÚÔ˚fiÓ Û‡Ìʈӷ Ì ÙȘ Ô‰ËÁ›Â˜ ¯Ú‹Û˘. • ∆˘¯fiÓ ¯Ú‹ÛË Ô˘ ‰ÂÓ ÂÚÈÁÚ¿ÊÂÙ·È ÛÙȘ Ô‰ËÁ›Â˜ ¯Ú‹Û˘ ‹ ·Î·Ù¿ÏÏËÏË ¯Ú‹ÛË ·˘ÙÔ‡ ÙÔ˘ ÚÔ˚fiÓÙÔ˜ ÂÓ‰¤¯ÂÙ·È Ó· ¤¯ÂÈ ÂÈÙÒÛÂȘ ÁÈ· ÙËÓ ·ÛÊ¿ÏÂÈ·. • ªË ‚¿˙ÂÙ ٷ ‰¿¯Ù˘Ï· ‹ Ù· ¯¤ÚÈ· Û·˜ ̤۷ ÛÙÔ ¢Ô¯Â›Ô ˘ÂÚ‹¯ˆÓ ηٿ ÙË ÏÂÈÙÔ˘ÚÁ›·. • ¶ÚÈÓ Î·ı·Ú›ÛÂÙ Ì ˘ÂÚ‹¯Ô˘˜ ·ÓÙÈΛÌÂÓ· ÌÂÁ¿Ï˘ ·Í›·˜, Û˘Ì‚Ô˘Ï¢ı›Ù ÙÔÓ Î·Ù·Û΢·ÛÙ‹ ÙÔ˘ ·ÓÙ›ÛÙÔÈ¯Ô˘ ·ÓÙÈÎÂÈ̤ÓÔ˘ ÁÈ· ÏËÚÔÊÔڛ˜ Û¯ÂÙÈο Ì ÙȘ ÚÔÊ˘Ï¿ÍÂȘ Ô˘ Ú¤ÂÈ Ó· Ï¿‚ÂÙÂ
ηٿ ÙÔÓ Î·ı·ÚÈÛÌfi. • ÃÚËÛÈÌÔÔț٠ÌfiÓÔ ÙÔ ·Ú¯fiÌÂÓÔ Î·ÏÒ‰ÈÔ ÙÚÔÊÔ‰ÔÛ›·˜ ‹ ÈÛfiÙÈÌÔ. • ∏ ÂÈÏÔÁ‹ ·ÂÓÂÚÁÔÔ›ËÛ˘ ·fi ÙÔ ÏËÎÙÚÔÏfiÁÈÔ ‰ÂÓ ÎÏ›ÓÂÈ ÙËÓ ÙÚÔÊÔ‰ÔÛ›· Ù¿Û˘ ‰ÈÎÙ‡Ô˘ ÛÙË ÌÔÓ¿‰·. • ∏ ÏÂÈÙÔ˘ÚÁ›· ÁÈ· ÌÂÁ¿Ï· ¯ÚÔÓÈο ‰È·ÛÙ‹Ì·Ù· ı· ÚÔηϤÛÂÈ ˘ÂÚı¤ÚÌ·ÓÛË ÙÔ˘ ηı·ÚÈÛÙÈÎÔ‡ ‰È·Ï‡Ì·ÙÔ˜. °È· Ó· ·ÔʇÁÂÙ ÙÔÓ Î›Ó‰˘ÓÔ ÂÁη˘Ì¿ÙˆÓ, ÌË ‚¿˙ÂÙ ٷ ¯¤ÚÈ· Û·˜ ÛÙÔ ‰È¿Ï˘Ì· Î·È ÚÔÛ¤¯ÂÙ ӷ ÌË ¯˘ı› ‰È¿Ï˘Ì· ¿ӈ Û·˜. ¶ÚÔÛ¤¯ÂÙ ȉȷ›ÙÂÚ· ηٿ ÙËÓ ·Ê·›ÚÂÛË ÔÔÈÔ˘‰‹ÔÙ ÂÍ·ÚÙ‹Ì·ÙÔ˜ ·fi ÙÔÓ Î¿‰Ô, ηıÒ˜ ÌÔÚ› Î·È ·˘Ùfi Ó· Â›Ó·È Î·˘Ùfi. • ªËÓ ·ÔÛ˘Ó·ÚÌÔÏÔÁ›Ù ÙË ÌÔÓ¿‰·. ¢ÂÓ ÂÚȤ¯ÂÈ Î·Ó¤Ó· ÂÍ¿ÚÙËÌ· Ô˘ Ó· ÌÔÚ› Ó· ‰Â¯ı› Ù¯ÓÈ΋ ˘ÔÛÙ‹ÚÈÍË ·fi ÙÔ ¯Ú‹ÛÙË. XII. ¶ÏËÚÔÊÔڛ˜ ·Ú·ÁÁÂÏ›·˜ ∫ˆ‰.
¶ÂÚÈÁÚ·Ê‹
¶ÔÛ.
ªÔÓ¿‰· Î·È ·ÍÂÛÔ˘¿Ú: UC125H
100 ªÔÓ¿‰· (100V, 50/60 Hz) 115 ªÔÓ¿‰· (115V, 50/60 Hz) 230CE ªÔÓ¿‰· (230V, 50/60 Hz) 230UK ªÔÓ¿‰· (230V, 50/60 Hz) 240 ªÔÓ¿‰· (240V, 50/60 Hz) UC155XD ™ÂÙ ·ÍÂÛÔ˘¿Ú ∫·Ï¿ıÈ 3/4 FingerGuard, (2) °È¿ÏÈÓ· ‰Ô¯Â›·, (2) ∫·Ï‡ÌÌ·Ù·, (2) ¢·ÎÙ‡ÏÈÔÈ ∆ÔÔı¤ÙËÛ˘ Î·È (2) ∂Í·ÚÙ‹Ì·Ù· ÛÙ‹ÚÈ͢ ‰Ô¯Â›ˆÓ ˙¤Ûˆ˜
UC151XD ∫·Ï¿ıÈ ÌÂÁ¿ÏÔ˘ ÌÂÁ¤ıÔ˘˜ FingerGuard UC152XD ∫·Ï¿ıÈ 3/4 FingerGuard UC153 ŒÓ· ÂÍ¿ÚÙËÌ· ÛÙ‹ÚÈ͢ ÁÈ¿ÏÈÓÔ˘ ‰Ô¯Â›Ô˘ UC350 ∞ÓÙ·ÏÏ·ÎÙÈο Ê›ÏÙÚ· ·ÔÛÙÚ¿ÁÁÈÛ˘ UC360 ∞ÓÙ·ÏÏ·ÎÙÈÎfi˜ ۈϋӷ˜ ·ÔÛÙÚ¿ÁÁÈÛ˘ UC53
1 1 1 1 1
1 1 1 3 1
600 ml °È¿ÏÈÓÔ ‰Ô¯Â›Ô ÌÂ Î¿Ï˘ÌÌ· Î·È ‰·ÎÙ‡ÏÈÔ ÙÔÔı¤ÙËÛ˘ 1
¢È·Ï‡Ì·Ù· ΢ڛˆ˜ ‰Ô¯Â›Ô˘: UC30
UC31
UC32
™˘Ì‡Îӈ̷ ηı·ÚÈÛÙÈÎÔ‡ ‰È·Ï‡Ì·ÙÔ˜ ÁÂÓÈ΋˜ ¯Ú‹Û˘ Û ¢ÔÛÔÌÂÙÚÈ΋ ÊÈ¿ÏË MeterDose® (16 oz.) 473 ml General Purpose Cleaner + Super Rust Inhibitor 20 Single Dose Packets 15 ml ∂Ó˙˘Ì·ÙÈÎfi ‰È¿Ï˘Ì· ηı·ÚÈÛÌÔ‡ ˘ÂÚ‹¯ˆÓ Û ¢ÔÛÔÌÂÙÚÈ΋ ÊÈ¿ÏË
√‰ËÁ›Â˜ ¯Ú‹Û˘
MeterDose (8 oz.)
59
236 ml
UC34
™˘Ì‡Îӈ̷ ÛÎfiÓ˘ ·ÔÌ¿ÎÚ˘ÓÛ˘ Á‡„Ô˘ & ¤ÙÚ·˜ Û ‚¿˙Ô (1,85 lbs.) 840 g
UC38
ªÈÎÚÔ‚ÈÔÎÙfiÓÔ Î·ı·ÚÈÛÙÈÎfi Û˘Ì‡Îӈ̷ ˘ÂÚ‹¯ˆÓ Û ¢ÔÛÔÌÂÙÚÈ΋ ÊÈ¿ÏË MeterDose (8oz.) 236 ml
UC39
UC42
™˘Ì‡Îӈ̷ ·ÔÌ¿ÎÚ˘ÓÛ˘ ÎÔÓ›·˜ Û ‰ÔÛÔÌÂÙÚÈ΋ ÊÈ¿ÏË MeterDose (16 oz.) 473 ml ªÈÎÚÔ‚ÈÔÎÙfiÓÔ Î·ı·ÚÈÛÙÈÎfi Û˘Ì‡Îӈ̷ ˘ÂÚ‹¯ˆÓ Û ‰ÔÛÔÌÂÙÚÈ΋ ÊÈ¿ÏË MeterDose (16oz.) 473 ml
XIII. ™˘ÓÙ‹ÚËÛË A. ∫·ı·ÚÈÛÌfi˜ ¶Ú¤ÂÈ Ó· ηı·Ú›˙ÂÙ ηıËÌÂÚÈÓ¿ fiÏ· Ù· ÂÍ·ÚÙ‹Ì·Ù·, ÛÎÔ˘›˙ÔÓÙ¿˜ Ù· Ì ¤Ó· Ì·Ï·Îfi ·Ó› ÌÔ˘ÛÎÂ̤ÓÔ Ì ÚÔ˚fiÓ ·ÔÌ¿ÎÚ˘ÓÛ˘ ·Ï¿ÙˆÓ Î·È ÛÙË Û˘Ó¤¯ÂÈ· ηı·Ú›˙ÔÓÙ·˜ Ì ÈÛÔÚÔ˘ÏÈ΋ ·ÏÎÔfiÏË ‹ ÌË
·ÛÊ¿ÏÂÈ·˜. ∞Û΋ÛÙ ÂÏ·ÊÚÈ¿ ›ÂÛË ÁÈ· Ó· ÂÏ¢ıÂÚÒÛÂÙ ÙË ‚¿ÛË Ù˘ ·ÛÊ¿ÏÂÈ·˜ (Û¯. 12).
4. ∂ÍÂÙ¿ÛÙ ÙËÓ ·ÛÊ¿ÏÂÈ·. (Û¯. 13) ∂¿Ó Â›Ó·È Î·Ì¤ÓË, ·ÓÙÈηٷÛÙ‹ÛÙ ÙËÓ Ì ÌÈ· ·ÛÊ¿ÏÂÈ· Ô˘ Û˘ÌʈÓ› Ì ÙËÓ Ù¿ÛË Î·È ÙÔ Ú‡̷ Ù˘ ÌÔÓ¿‰·˜ (‚Ï. ÂÓfiÙËÙ· XIV). ∂¿Ó Ë ·ÛÊ¿ÏÂÈ· Ê·›ÓÂÙ·È Ó· ÌËÓ Â›Ó·È Î·Ì¤ÓË Î·È Ë ÌÔÓ¿‰· ‰ÂÓ ·Ó¿‚ÂÈ fiÙ·Ó Á˘ÚÓ¿Ù ÙÔ ÁÂÓÈÎfi ‰È·ÎfiÙË ÏÂÈÙÔ˘ÚÁ›·˜ on/off ÛÙË ı¤ÛË “on”, ÂÈÛÙÚ¤„Ù ÙË ÌÔÓ¿‰· ÛÙËÓ ·ÓÙÈÚÔۈ›· ‹ ÛÙËÓ Coltène/Whaledent Inc. 5. ™˘Ó‰¤ÛÙ ÙÔ Î·ÏÒ‰ÈÔ ÙÚÔÊÔ‰ÔÛ›·˜ ÛÙËÓ ·ÓÙ›ÛÙÔÈ¯Ë ˘Ô‰Ô¯‹ Ô˘ ‚Ú›ÛÎÂÙ·È ÛÙËÓ ›Ûˆ ÏÂ˘Ú¿ Ù˘ ÌÔÓ¿‰·˜. ™ÙË Û˘Ó¤¯ÂÈ·, Û˘Ó‰¤ÛÙ ÙÔ Î·ÏÒ‰ÈÔ ÙÚÔÊÔ‰ÔÛ›·˜ ÛÙËÓ Ú›˙· ÙÔ˘ ÙÔ›¯Ô˘. XIV. ¶ÚԉȷÁڷʤ˜ ÌÔÓ¿‰·˜ ªÂÙ·ÙÚÔ›˜ :
2 ÌÂÙ·ÙÚÔ›˜
UC125H ªÔÓ¿‰·
B. ∂ÈıÂÒÚËÛË ÙÔ˘ ηψ‰›Ô˘ ÙÚÔÊÔ‰ÔÛ›·˜
ÈÚËÙÈÎfiÙËÙ· ο‰Ô˘: 1,25 Á·ÏfiÓÈ· ‹ 4,75 Ï›ÙÚ·
C. ∞ÓÙÈηٿÛÙ·ÛË Ù˘ ·ÛÊ¿ÏÂÈ·˜ 1. µÁ¿ÏÙ ÙÔ Î·ÏÒ‰ÈÔ ÙÚÔÊÔ‰ÔÛ›·˜ ÚÒÙ· ·fi ÙËÓ Ú›˙· ÙÔ˘ ÙÔ›¯Ô˘ Î·È ‡ÛÙÂÚ· ·fi ÙËÓ ˘Ô‰Ô¯‹ Ô˘ ‚Ú›ÛÎÂÙ·È ÛÙËÓ ›Ûˆ ÏÂ˘Ú¿ Ù˘ ÌÔÓ¿‰·˜. 2. ∏ ·ÛÊ·ÏÂÈÔı‹ÎË ‚Ú›ÛÎÂÙ·È ·ÎÚÈ‚Ò˜ ‰›Ï· ÛÙËÓ ˘Ô‰Ô¯‹ ÙÔ˘ ηψ‰›Ô˘ ÙÚÔÊÔ‰ÔÛ›·˜. °È· Ó· ·ÔÎÙ‹ÛÂÙ ÚfiÛ‚·ÛË Û ·˘Ù‹Ó, ÙÔÔıÂÙ‹ÛÙ ¤Ó· ÌÈÎÚfi ›ÛÈÔ Î·ÙÛ·‚›‰È ÛÙÔ ÁψÛÛ›‰È Ô˘ ˘¿Ú¯ÂÈ ÛÙË ‚¿ÛË Ù˘
IEC 60417-5041 ¶ÚÔÛÔ¯‹, ∫·˘Ù‹ ∂ÈÊ¿ÓÂÈ·
3. µÁ¿ÏÙ ÚÔÛÂÎÙÈο ÙËÓ ·ÛÊ¿ÏÂÈ· ·fi ÙË ‚¿ÛË Ù˘.
‰È·‚ÚˆÙÈÎfi ·ÔÏ˘Ì·ÓÙÈÎfi. ¶ÚÔÛÔ¯‹: ªË ¯ÚËÛÈÌÔÔț٠‰È·Ï˘ÙÈο Ì ‚¿ÛË ÙÔ ÂÙڤϷÈÔ, Ȉ‰ÈˆÌ¤Ó· ÚÔ˚fiÓÙ· ‹ ÚÔ˚fiÓÙ· Ì ‚¿ÛË ÙË Ê·ÈÓfiÏË. (∆· Ȉ‰ÈˆÌ¤Ó· Î·È Ù· Ê·ÈÓÔÏÈο ÚÔ˚fiÓÙ· ÂÓ‰¤¯ÂÙ·È Ó· ·Ê‹ÛÔ˘Ó ÏÂΤ‰Â˜ ÛÙËÓ ÂÈÊ¿ÓÂÈ· Ù˘ ÌÔÓ¿‰·˜.) ∫·ı·Ú›˙ÂÙ ·Ì¤Ûˆ˜ fiϘ ÙȘ ÈÙÛÈÏȤ˜ ˘ÁÚÒÓ. ∆· ˘ÔÏ›ÌÌ·Ù· ÛÎÔ˘ÚÈ¿˜ ÛÙÔ ‰Ô¯Â›Ô Ú¤ÂÈ Ó· ηı·Ú›˙ÔÓÙ·È/·ÔÌ·ÎÚ‡ÓÔÓÙ·È Ì ¤Ó· ÚÔ˚fiÓ Î·ı·ÚÈÛÌÔ‡ ·ÓÔÍ›‰ˆÙÔ˘ ¯¿Ï˘‚· ÙÔ˘ ÂÌÔÚ›Ô˘.
∂ϤÁ¯ÂÙ ηٿ ‰È·ÛÙ‹Ì·Ù· ÙÔ Î·ÏÒ‰ÈÔ ÙÚÔÊÔ‰ÔÛ›·˜ ÁÈ· Ù˘¯fiÓ ˙ËÌȤ˜ Î·È ÊÚÔÓÙ›˙ÂÙ ӷ ÂÊ·ÚÌfi˙Ô˘Ó Î·Ï¿ Ù· ÊȘ Î·È ÛÙ· ‰‡Ô ¿ÎÚ· ÙÔ˘ ηψ‰›Ô˘.
EMC: ∫·ÙËÁÔÚ›· A
100 [100V 50/60 Hz, 337W ( ̤Á.)] 115 [115V 50/60 Hz, 337W ( ̤Á.)]
230CE [230V 50/60 Hz, 337W ( ̤Á.)] 230UK [230V 50/60 Hz, 337W ( ̤Á.)] 240 [240V 50/60 Hz, 337W ( ̤Á.)] ™˘¯ÓfiÙËÙ· ÏÂÈÙÔ˘ÚÁ›·˜:
52 kHz
ÈÚËÙÈÎfiÙËÙ· ¯Ú‹Û˘:
1 Á·ÏfiÓÈ ‹ 3,78 Ï›ÙÚ·
™˘ÓÔÏÈΤ˜ ‰È·ÛÙ¿ÛÂȘ:
16,5 x 10,5 x 13.8 ›ÓÙÛ˜ 42 x 27 x 35 ÂÎ.
¢È·ÛÙ¿ÛÂȘ ο‰Ô˘:
11,2 x 5,62 x 5 ›ÓÙÛ˜ 28,6 x 14,3 x 12,7 ÂÎ.
∆‡Ô˜ ·ÛÊ¿ÏÂÈ·˜:
5 x 20 mm ·ÌԇϷ,
‚Ú·‰Â›·˜ ً͈˜:
IEC 60127-2 º‡ÏÏÔ 3
£ÂÚÌ·ÈÓfiÌÂÓ˜ ÌÔÓ¿‰Â˜ 100-115V £ÂÚÌ·ÈÓfiÌÂÓ˜ ÌÔÓ¿‰Â˜ 230-240V
4 Amp 2,5 Amp
!
ISO 3864, No B.3. ¶ÚÔÛÔ¯‹: (·Ó·ÙÚ¤ÍÙ ÛÙ· Û˘Óԉ¢ÙÈο ¤ÓÙ˘·)
ISO 3864, No B.3.6 ¶ÚÔÛÔ¯‹: ∫›Ó‰˘ÓÔ˜ ËÏÂÎÙÚÔÏËÍ›·˜
XV. ¶§∏ƒ√º√ƒπ∂™ ∂°°À∏™∏™ √È Û˘Û΢¤˜ ηı·ÚÈÛÌÔ‡ BioSonic ηٷÛ΢¿˙ÔÓÙ·È ‚¿ÛÂÈ ¿ÚÙÈˆÓ ÚÔÙ‡ˆÓ, ¤¯Ô˘Ó ۯ‰ȷÛÙ› ÂȉÈο ÁÈ· ¯Ú‹ÛË ÛÙËÓ Ô‰ÔÓÙÈ·ÙÚÈ΋ Î·È ÚÔÔÚ›˙ÔÓÙ·È ÁÈ· ¯ÂÈÚÈÛÌfi ÌfiÓÔ ·fi ·ÁÁÂÏ̷ٛ˜ Ù˘ Ô‰ÔÓÙÈ·ÙÚÈ΋˜, Û‡Ìʈӷ Ì ÙȘ ·ÚÔ‡Û˜ Ô‰ËÁ›Â˜ ¯Ú‹Û˘. øÛÙfiÛÔ, ·ÓÂÍ¿ÚÙËÙ· ·fi ÙȘ fiϘ ÙȘ ·ÚÔ‡Û˜ Ô‰ËÁ›Â˜, Ô ¯Ú‹ÛÙ˘ ·Ú·Ì¤ÓÂÈ ·Ó¿ ¿Û· ÛÙÈÁÌ‹ Ô ÌfiÓÔ˜ ˘Â‡ı˘ÓÔ˜ ÁÈ· ÙÔÓ ÚÔÛ‰ÈÔÚÈÛÌfi Ù˘ ηٷÏÏËÏfiÙËÙ·˜ ÙÔ˘ ÚÔ˚fiÓÙÔ˜ ÁÈ· Û˘ÁÎÂÎÚÈ̤ÓÔ ÛÎÔfi Î·È ÁÈ· ÙË Ì¤ıÔ‰Ô ¯Ú‹Û˘ ÙÔ˘. √ÔÈ·‰‹ÔÙ ԉËÁ›· Û¯ÂÙÈο Ì ÙËÓ Ù¯ÓÔÏÔÁ›· ÂÊ·ÚÌÔÁÒÓ, Ë ÔÔ›· ·Ú¤¯ÂÙ·È ·fi ÙÔÓ Î·Ù·Û΢·ÛÙ‹ ‹ ÁÈ· ÏÔÁ·ÚÈ·ÛÌfi ·˘ÙÔ‡, ÁÚ·Ù‹, ÚÔÊÔÚÈ΋ ‹ Ì ›‰ÂÈÍË, ‰ÂÓ ··ÏÏ¿ÛÛÂÈ ÙÔÓ Â·ÁÁÂÏÌ·Ù›· Ù˘ Ô‰ÔÓÙÈ·ÙÚÈ΋˜ ·fi ÙËÓ ˘Ô¯Ú¤ˆÛ‹ ÙÔ˘/Ù˘ Ó· ÂϤÁ¯ÂÈ ÙÔ ÚÔ˚fiÓ Î·È Ó· Ú·ÁÌ·ÙÔÔÈ› fiϘ ÙȘ ·ÁÁÂÏÌ·ÙÈΤ˜ ·ÍÈÔÏÔÁ‹ÛÂȘ Û¯ÂÙÈο Ì ÙË ¯Ú‹ÛË ÙÔ˘. ∆· ÚÔ˚fiÓÙ· Ù˘ Coltène/Whaledent ηٷÛ΢¿˙ÔÓÙ·È ·fi ηÈÓÔ‡ÚÁÈ· ÂÍ·ÚÙ‹Ì·Ù· ‹ ·fi ηÈÓÔ‡ÚÁÈ· Î·È ÂÈÛ΢·Ṳ̂ӷ ÌÂÙ·¯ÂÈÚÈṲ̂ӷ ÂÍ·ÚÙ‹Ì·Ù·. ∞ÓÂÍ·Úًو˜ ÂÍ·ÚÙËÌ¿ÙˆÓ, Ë ÂÁÁ‡ËÛË ÈÛ¯‡ÂÈ. ∆· ÚÔ˚fiÓÙ· ηχÙÔÓÙ·È ·fi ÂÁÁ‡ËÛË Û‡Ìʈӷ Ì ÙÔ˘˜ fiÚÔ˘˜ Ô˘ ÂÚÈÏ·Ì‚¿ÓÔÓÙ·È ÛÙÔ ¤ÁÁÚ·ÊÔ ¶ÈÛÙÔÔÈËÙÈÎfi ¶ÂÚÈÔÚÈṲ̂Ó˘ ∂ÁÁ‡ËÛ˘. ∏ Coltène/Whaledent Inc. ‰ÂÓ ·Ú¤¯ÂÈ Î·Ì›· ÂÁÁ‡ËÛË ÁÈ· Î¿Ï˘„Ë ·˘ÙÔ‡ ÙÔ˘ ÚÔ˚fiÓÙÔ˜, ÚËÙ‹ ‹ ÛȈËÚ‹, Û˘ÌÂÚÈÏ·Ì‚·ÓÔ̤ӈÓ, ¯ˆÚ›˜ ÂÚÈÔÚÈÛÌfi, ÔÔȈӉ‹ÔÙ ÂÁÁ˘‹ÛˆÓ
√‰ËÁ›Â˜ ¯Ú‹Û˘
61
Owner’s Guide
63
8. Hu-Friedy IMS * Instrument Cassette
7.
®
9. Hu-Friedy IMS * Cassette Rack ®
11.
12.
10.
13. * IMS is a registered trademark of Hu-Friedy Inc.