HANDBOOK
MU - I-MOVE –V5
Manufacturer: BMB MEDICAL BP 405 - Z.I. de Fetan F-01604 TREVOUX CEDEX Phone: +33(0)4 74 08 71 71 Fax: +33(0) 4 74 08 71 72 www.bmb-medical.com
The purpose of this manual is to familiarise you with the use, maintenance and safety aspects of your new medical appliance. We strongly advise you to read this document and follow its recommendations to be able to make the best use your appliance in complete safety.
Stick the label with the product serial number here
MU - I-MOVE –V5
Contents I. INTRODUCTION SPECIFICATIONS ... 4 1.1 – PRODUCT PRESENTATION ... 4 1.1.1 I-MOVE transfer chair ... 4 1.1.2 I-MOVE transfer chair for overweight patients ... 6 1.1.3 Motorised I-MOVE transfer chair ... 8 Overall view ... 9 1.2 – APPLICATION FIELD ... 10 1.3 – PURPOSE OF THE USER'S MANUAL ... 10 1.4 – STORAGE ... 10 1.5 – SYMBOLS AND STANDARDS USED ... 10 II. INSTRUCTIONS FOR USE ... 11 2.1 – BATTERY ... 11 2.2 – MOTORISED DRIVE ... 12 2.3 – BRAKING ... 13 2.4 – ARM RESTS ... 13 2.5 – FOOTREST... 13 2.6 – OPTIONS ... 14 2.6.1 – Leg rest ... 14 2.6.2 – Removable drip stand / Anti-theft bar... 14 III. CLEANING AND DISINFECTION ... 16 3.1 – INTRODUCTION ... 16 3.2 – INTERVALS ... 16 3.3 – RECOMMENDED EQUIPMENT AND PRODUCTS... 16 3.4 – CLEANING AND DISINFECTION ADVICE ... 17 IV. PRECAUTIONS ... 17 4.1 – PRECAUTIONS WHEN USING THIS APPLIANCE ... 17 4.2 – ABNORMAL USE ... 18 4.3 – TRANSPORT AND STORAGE ... 19 4.4 – WARRANTY... 19 4.5 – AFTER SALES SERVICE ... 19 4.6 – ENVIRONMENT... 20
The information in this document may be modified without notice and does not imply any responsibility of any sort of BMB MEDICAL. Any representation or copying, whether complete or partial, without authorisation from BMB MEDICAL is illegal. This illegal representation or copying, by any method at all, would constitute an infringement of rights. The illustrations in this document are not contractual.
MU - I-MOVE –V5
I.
INTRODUCTION SPECIFICATIONS
1.1 – PRODUCT PRESENTATION 1.1.1 I-MOVE transfer chair Technical characteristics REFERENCE
I-MOVE
DESCRIPTION
PATIENT TRANSFER CHAIR
SAFE WORKING LOAD
200 kg (maximum acceptable weight comprising the weight of the patient, the mattress and the accessories)
WEIGHT OVERALL DIMENSION CUSHION AND BACK SIZE BACK HEIGHT
EQUIPMENT
OPTIONS
31 kg 105 x 70 x 102 cm 46 x 46 cm 55 cm - Steel chassis with epoxy coating - 300 mm diameter rear wheels with a central brake (with a control accessible by the person accompanying the chair in the standing position, and a foot pedal). - 125 mm diameter pivoting front wheels - 5 cm thick foam upholstery - Seamless vinyl covering - Retractable and self-locking foot rest - Retractable polyurethane arms - Removable drip stand, adjustable in two parts, 2 hooks - Anti-theft bar fitted with 2 drip hooks - Removable effect basket including a cylinder holder - Bag hooks - Leg rest fitting - Adjustable leg rest - Coin handler - Wall plate
MU - I-MOVE –V5
Overall view
MU - I-MOVE –V5
1.1.2 I-MOVE transfer chair for overweight patients Technical characteristics REFERENCE
I-MOVE transfer chair for overweight patients
DESCRIPTION
OVERWEIGHT PATIENT TRANSFER CHAIR
SAFE WORKING LOAD
250 kg (shows the maximum acceptable weight comprising the weight of the patient, the mattress and the accessories)
WEIGHT OVERALL DIMENSION CUSHION AND BACK SIZE BACK HEIGHT
EQUIPMENT
OPTIONS
50 kg 105 x 90 x 102 cm 46 x 65 cm 55 cm - Steel chassis with epoxy coating - Twin 300 mm diameter rear wheels with a central brake (with a control accessible by the person accompanying the chair in the standing position, and a foot pedal). - Strengthened 125 mm diameter pivoting front wheels - 8 cm thick foam upholstery - PVC coated stockinette coating - Retractable and self-locking foot rest - Retractable polyurethane arms - Removable drip stand, adjustable in two parts, 2 hooks - Anti-theft bar fitted with 2 drip hooks - Removable personal effect basket including a cylinder holder - Bag hooks - Leg rest fitting - Adjustable leg rest - Coin handler - Wall plate
MU - I-MOVE –V5
Overall view
MU - I-MOVE –V5
1.1.3 Motorised I-MOVE transfer chair Technical characteristics REFERENCE
Motorised I-MOVE
DESCRIPTION
MOTORISED PATIENT TRANSFER CHAIR
SAFE WORKING LOAD
I-Move 200 kg
l-Move for overweight patients 250 kg
50 kg
65 kg
WEIGHT OVERALL DIMENSION
EQUIPMENT
OPTIONS
105 x 70 x 102 cm 105 x 90 x 102 cm - Steel chassis with epoxy coating - 300 mm diameter rear wheels with a central brake (with a control accessible by the person accompanying the chair in the standing position, and a foot pedal). - Strengthened 125 mm diameter pivoting front wheels - Motorised wheel with integral battery and charger - 8 cm thick foam upholstery - PVC coated seamless stockinette coating - Retractable foot rest - Retractable polyurethane arms - Fixed effects basket - Removable drip stand, adjustable in two parts, 2 hooks - Anti-theft bar fitted with 2 drip hooks - Bag hooks - Leg rest fitting - Adjustable leg rest - Coin handler - Wall plate
MU - I-MOVE –V5
Overall view
MU - I-MOVE –V5
1.2 – APPLICATION FIELD This chair is intended for the reception and transfer of patients in the sitting position. Its technical characteristics also allow it to be used outdoors.
1.3 – PURPOSE OF THE USER'S MANUAL Health professionals, this user's manual has been written for you to provide you with all the information necessary and all the precautions to be observed when using the chair independently and in complete safety. The instructions relating to maintenance and surveillance of the product ensure its service life. Consequently it is essential that all personnel have access to, and are informed of, the handling procedures and the risks connected with the use of the transfer chair. It should be used by a person who is qualified and aware of the risks inherent in incorrect use.
1.4 – STORAGE This user's manual should be kept close to the product in a location dedicated to this effect. It should be kept protected from all elements or substances detrimental to its legibility.
1.5 – SYMBOLS AND STANDARDS USED Type B appliance Indicates that failure to respect recommendations may cause danger for the patient or the user, or damage the appliance. Complies with the essential requirements of Directive 93/42/EC covering medical equipment.
IPx4
Protection against water splashing and projection.
Indicates the safe working load
I-MOVE and I-MOVE chairs for overweight patients are non-invasive and non-active Class I medical appliances. The I-MOVE EZ-GO transfer chair is a non-invasive and active Class I medical appliance.
MU - I-MOVE –V5
II.
INSTRUCTIONS FOR USE
2.1 – BATTERY The battery should be charged for 5 hours on a mains supply socket before being put into use for the first time. To be sure to maintain an acceptable level of charge, we recommend plugging motorised chairs into a mains socket whenever one is available. Is not necessary to wait for the battery to be fully discharged. If the equipment is left out of use for a long period we recommend charging the battery every month. If the battery is stored for more than 50 days without being charged its performance may be affected, or even reduced irreversibly.
The light in the on/off switch indicates the battery condition
If the light is green, the wheel is charged. When the light is orange, the wheel is at half charge and we strongly recommend charging the battery. To charge the battery, connect the plug situated under the document holder into the mains. If the light is red it is essential to charge the battery. When the light flashes red the wheel is completely discharged and it is impossible to use it.
Once the battery is connected to the mains the light flashes. For a complete charge the wheel must be connected to the mains for 1 to 2 hours. When the wheel is charged the light switches to green. MU - I-MOVE –V5
If the light does not flash although the battery is connected, check the connection and the mains plug. Before using the chair make sure that it is not connected to the mains. A flashing light shows that the battery is connected Connection conditions to the mains, and electromagnetic compatibility: - Connection to the mains must comply with the following standards: NF C 15-100 and NF C 15211 (230 V supply socket). The circuit used should be protected by a 30mA (maximum) differential circuit breaker. - Work on the electrical parts must only be performed by qualified and authorised personnel. - Although the chair satisfies the requirements of the EN 60 601-1-2 standard, electromagnetic interference caused by the chair is not impossible. In this case we recommend inhibiting the chair functions and keeping it well clear of the appliances affected by the interference.
2.2 – MOTORISED DRIVE Motorised drive by an autonomous wheel facilitates movement of the chair. 1) Put the brake pedal in the top position to release the chair brake.
2) Activate the on/off switch and turn the rotating grip. Rotating grip for progressive acceleration
.
2
1
On/off switch
Check regularly that no foreign bodies on the floor could hinder rotation of the wheel.
MU - I-MOVE –V5
2.3 – BRAKING -
To release the chair, lift the red brake pedal with your foot. To immobilise the chair press the red brake pedal with your foot.
When the brake pedal is activated, make sure that the transfer chair is properly immobilised by trying to move it manually. Braking the transfer chair on a slope is not recommended.
2.4 – ARM RESTS To raise or lower the armrests, simply raise them or lover them manually.
2.5 – FOOTREST The opening at the front of the tray allows a foot to be inserted to advance or push back the footrest. -
To deploy the footrest, extended as far as the stop to lock it.
-
To retract the footrest, lift it and push it.
- When moving a patient make sure that the footrest is deployed. MU - I-MOVE –V5
- Do not step on the foot rest to sit in the transfer chair. - Do not stand up on the footrest.
2.6 – OPTIONS
2.6.1 – Leg rest To install the leg rest, insert the bar in the jaws provided for the purpose under the chair. It is possible to incline the leg rest by placing its bar in the 2nd notch.
Leg rest fitting
Straight position
Inclined position
2.6.2 – Removable drip stand / Anti-theft bar In the rear chassis of the transfer chair there are 2 locations where the drip stand or the anti-theft bar can be inserted.
MU - I-MOVE –V5
2 locations for the drip stand or the anti-theft bar
Maximum safe working load of the drip stand: 10 kg. This value for information corresponds to safe use of the product, but it does not indicate normal use for which the load is much less. When moving the transfer chair fitted with the drip stand, make sure that the drip stand does not hit any obstacles (e.g.: doorways, lighting, etc.) or persons. Never use the drip stand or anti-theft bar to move the chair.
MU - I-MOVE –V5
III.
CLEANING AND DISINFECTION
3.1 – INTRODUCTION Regular and proper cleaning ensures the service life of the appliance (cf. cleaning programme drawn up at the time of entry into service of the appliance with the identities of the personnel responsible). Failure to perform cleaning operations may cause a risk of infection due to the presence of secretions and/or residues on the transfer chair. The cleaning advice below does not replace the internal protocol in your establishment.
3.2 – INTERVALS Cleaning should be performed: - for each patient, i.e. between 2 patients - after transporting an infected patient - completely every week
3.3 – RECOMMENDED EQUIPMENT AND PRODUCTS - Only use authorised and standardised cleaning and disinfectant products. -
Be aware of the products strictly prohibited by the manufacturer (Cf. list below)
List of products strictly prohibited: Solvents: Gasoline/higher aliphatic and aromatic alcohols / ketones-esters / trichloroethylene-perchlorethylene Base: Soda 10% / Ammonia 10% Acids: Nitric 10% / Hydrochloric 30% / Citric 10% Miscellaneous: Distilled water/ Liquid bleach / Abrasive powder/ All abrasive products -
Do not use abrasive utensils or strong liquid bleach solution. Do not mix products. Follow the recommended proportions and dilution. Use clean disposable wipes. Wear protective clothing and gloves when cleaning and disinfecting.
MU - I-MOVE –V5
3.4 – CLEANING AND DISINFECTION ADVICE 3.4.1 Cleaning operation according to the elements ELEMENTS
OPERATIONS TO BE PERFORMED Washing with mains water and soap or dilute alcohol
METAL PARTS (90% epoxy covered)
Dry the transfer chair after cleaning Check operation Wash with soap and water
UPHOLSTERY
Dry the transfer chair after cleaning Check operation
3.4.2 Advice to be followed - All surfaces of the furniture must be cleaned - When cleaning or disinfecting we recommend: o Spraying the product or applying it with a dish cloth. o Starting by the furniture the closest to the patient and giving particular attention to parts in contact with users. o Proceeding from the top downwards o Allow the equipment to dry. Staining products (e.g.: "blue" or eosin) should be cleaned off as quickly as possible to prevent penetration.
IV.
PRECAUTIONS
4.1 – PRECAUTIONS WHEN USING THIS APPLIANCE 4.1.1 Precautions when using the transfer chair
When moving a patient on a downhill slope (maximum 10°) the patient must be positioned with his/her back to the direction of movement. When moving a patient the footrest should be deployed and the 2 armrests lowered to prevent falling. When the chair is not in use it should be immobilised with its brakes. Check that the wheels are locked by trying to move it. Failure to follow this recommendation may result in a fall by a person trying to use the chair as a support. When transferring a patient (from a bed to the chair or from the chair to a bed) the chair should be positioned as close as possible to the bed to provide maximum safety for the patient. Adjust the bed height so that is the same as that of the transfer chair.
MU - I-MOVE –V5
Make sure that the brakes are active and that the footrest and armrests are retracted. All the actions relating to the precautions above are described in paragraph II – Instructions for use We do not recommend action without suitable training.
4.1.2 Precautions applying to servicing, and maintenance of the appliance in good condition The appliance must not be modified in any way without prior authorisation from the manufacturer: a modification could cause unexpected operation and consequent injury to the patient and/or the medical personnel. The appliance requires regular inspection. The recommended maintenance is essential to maintain the appliance in good condition and ensure safe and durable operation. If malfunctions or damage are observed that could compromise the operation and safety of the appliance it must be withdrawn from use immediately. It is essential to use the appliance in accordance with the instructions in this user's manual at all times. The appliance must not be exposed to, or be in contact with, combustion heat sources or flammable elements. BMB MEDICAL accepts no responsibility in the event of incidents or accidents resulting from inappropriate use or modification of the product by the purchaser or user.
4.2 – ABNORMAL USE This transfer chair is designed and adapted to medical practices and procedures; hence it has a specific use. Abnormal use may be the cause of accidents for the user or even irreversible damage. The following list of abnormal uses of the transfer chair is not exhaustive and BMB MEDICAL expects users to apply their common sense. - Use by any person who is not familiar with the contents of this user's manual. - Use by any person other than medical personnel having received authorisation and explanations from a qualified person. - Handling by a child. - Use of the transfer chair for installing a child aged less than 4 years. - Use of the transfer chair with a load greater than the safe working load. - Use of the chassis for holding or storing objects. - Use in the presence of a flammable anaesthetic mixture. - Use outside the establishment or on an unprepared floor. - Use on a floor with a slope greater than 10°. - Washing down, high-pressure water or steam jets or passage through a washing the tunnel. - Any use other than use within the scope of operation of a hospital transfer chair.
MU - I-MOVE –V5
4.3 – TRANSPORT AND STORAGE 4.3.1 Transport When transporting the transfer chair with a view to storage, we recommend, to avoid risks of impact or falling: securing it to a support on the means of transport covering it with a tarpaulin applying the brake putting it in its bottom position Damage caused during transport is not covered by warranty. Repairs and replacement of damaged parts will be for the customer's account. 4.3.2 Storage When the transfer chair is stored, make sure that it is: braked covered with a tarpaulin in a temperate room free from damp (temperature : +10°C to +40°C). No heavy weight should rest on the appliance during storage. The transfer chair must not be considered as a support for any type of equipment. The appliance must not be exposed, to or be in contact with, combustion heat sources or flammable elements.
4.4 – WARRANTY The equipment is covered by one year's warranty, parts and labour However, the warranty is cancelled in the event of:
abnormal use (Paragraph 4.2)
intervention without having first obtained advice from our technical or maintenance departments on: - a mechanical part of the chair - a cylinder or motor
failure to observe any other warranty condition
inappropriate cleaning
use of accessories not supplied by the manufacturer
Failure to implement preventive maintenance operations recommended in the maintenance logbook.
4.5 – AFTER SALES SERVICE For any information concerning the installation, use or maintenance of the product, or any other question contact our Service Department at the following address: [email protected] or your local distributor for export.
MU - I-MOVE –V5
Using your on-line customer centre on www.bmb-medical.com, access your equipment and make your requests for intervention, and follow them up directly on this secure interface. The operator should read the information in this user's manual carefully, especially that concerning safety precautions and the methods of installation and use. If in doubt over the proper interpretation of the instructions, do not hesitate to contact our Service Department which is always ready to help.
4.6 – ENVIRONMENT 4.6.1 Worn out equipment
Recycling electrical equipment contributes to preserving natural resources and preventing risks of pollution. To this end BMB MEDICAL fulfils its obligations in relation to end-of-life I-MOVE EZ-GO transfer chairs which it puts on the market by financing the Recylum recycling system dedicated to DEEE Pro, which takes them back free of charge (More information on www.recylum.com)".
MU - I-MOVE –V5