Pacing System Analyzer (PSA) Application In-Service 1
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Connections
Patient Connections Inductive Telemetry Wand Model 6395 S-ICD Telemetry Wand Model 3203 Connection Port (for future use) Patient ECG Cable Model 3153 PSA Cable A / RV Model 6763 PSA Cable LV Model 6763
2
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Physician Connections Power (ON / OFF) Button is green when the Programmer is ON USB 2.0 Port (3)
USB 3.0 Port (1) Ethernet Port Verify proper adapter if video screen uses VGA or HDMI connection
DisplayPort Out AC Power Connection
Connectivity and Power a
b
Verify Desired Power and Connectivity
• Lightning bolt (a) appears inside the black battery symbol if the programmer is attached to AC power and battery charge is provided • When unplugged to AC power, battery icon changes color at different levels of charge: 25-100% charged 10-24%charged <10%charged
• Check for (b) Bluetooth connectivity on the Main Screen prior to selecting Quick Start or the PSA application to ensure Bluetooth printing capability, or connect USB cable to USB printer 3
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Note: When powering ON the LATITUDE™ Programming System, it may take up to one minute for the software to load and during this time the screen may be flashing or blank
Main Screen: Power Supply Indicator • If the Programmer is connected to external AC power, the internal battery (Model 6753) will be charging, whether the Programmer is ON or OFF • Depending on age of the battery, a full charge should last for approximately 2 hours of normal operation • Messages will display on the Programmer screen at different levels of battery depletion
Connected to AC Power
25%
4
5%
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Not Connected to AC Power
25%
5%
Startup Work Flow a
Device Interrogation
• • • •
5
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Interrogate the device via Quick Start button before starting the PSA application Device application button (a) will then be visible at upper right of startup screen next to PSA button Saved PSA data will be associated within the device being implanted During the implant case, toggle between the PSA and PG application as desired
Startup Work Flow
Data Management
c
b
6
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
• •
Select (b) Data tab Select method to Save, either (c) Hard Drive or USB Drive
Startup Work Flow f
•
Enter and Program (d) Patient Information and any preferred device settings
•
Check device (e) Battery and reform capacitor if a high-voltage device
•
Select (f) PSA application
d d
e
7
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
PSA application remains active until Programmer is powered OFF
Select PSA Setting and Lead Traces
h g
8
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
•
Select (g) PSA traces
•
Choose (h) Gain settings using the Up or Down arrows
PSA Main Screen Layouts (a) Lead Traces panel • 3-chamber pacing and sensing • Displays real-time EGMs and markers (b) Magnify and Additional Settings • Gain, surface filter, markers, pacing spikes a (c) Lead Trace button (d) Current of Injury (e) PSA Real-time Log (f) PSA Pacing and Amplitude panel • Enables / disables pacing • Mode adjusts accordingly (g) PSA Settings button (h) More Tests button (i) Test Results button (j) PSA Test Chamber Measurements panel
9
b c d e
j
f
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
g
h
i
Lead Trace Panel ECG / EGM Full Screen Display
From the PSA Main Screen the Magnify button enlarges the lead trace area
(a) Gain • Select the appropriate value to adjust the surface gain of the traces that are captured on printouts; range: AUTO, 0.5, 1, 2, 5, 10, 20 mm/mV
(b) Calibrate button • Transmits a 1 mV calibration pulse to the ECG display to establish a reference point to evaluate amplitudes (c) Baseline button • Forces the trace back to the baseline
a a a a b c
10
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Lead Trace Panel ECG / EGM Full Screen Display
(d) Trace Speed • Select the desired speed on the ECG display: 0 (stop), 25, or 50 mm/s (e) PSA Markers • When in a PSA application session, select the check box to enable the PSA markers
d
e
11
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
f
g
(f) Surface Filters • Select the check box to minimize noise on the surface ECG (g) Pacing Spikes • Select the check box to show detected pacing spikes, annotated by a marker on the top waveform
Lead Trace Panel
PSA Markers Parameter AS (AS) AP RVS RVP LVS LVP
12
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Measurement Atrial Sense after refractory Atrial Sense during refractory Atrial Pace Right Ventricular Sense after refractory Right Ventricular Pace Left Ventricular Sense after refractory Left Ventricular Pace
Dual Chamber Devices, PSA Cable Connections
Dual Chamber Pacemaker
Dual Chamber DF4 ICD Model 6763 PSA Cables
Model 7001EZ-4 Connector Tool RA
13
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
RV
BiV Quadripolar PSA Cable Connections
BiV Quadripolar CRT-D
Model 4625 Quadripolar Connector Tool White cables
V- / V+ or LV1 & LV2 Gray cables
RA LV RV
14
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
A- / A+ or LV3 & LV4
BiV Quadripolar PSA Cable Connections c
b
d a
For LV Unipolar Connection: (a) Move the A+ (red) clip of the A / RV cable to a temporary indifferent electrode as the can (b) Check the box “Use the A+ connection ...”
13 Bipolar 13 Bipolar LV Lead Vectors
(c) Check the desired Can electrode (d) Click Accept To end a unipolar configuration, select the
4 Unipolar desired bipolar LV configuration and the “Use the A+ connection ...” LV Lead Vectors deselect box; this results in normal operation of the atrial lead anode
15
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
LV Impedance Parameter Ranges
*The specified tolerance does not apply to LV lead impedance measurements using the LV PSA cable in combination with either the RV or RA cables. Clinical decisions using LV lead impedance values should be based on measurements using the LV PSA cable only.
Impedance
Voltage
100 - 3000 Ω 0.5 - 7.5 volts
16
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
Pulse Width Tolerance 0.4 to 2.0 ms ± 25%*
PSA Cable Connections (Model 6697)
Optional Remington Cables (Disposable) LV PSA cable RA / RV PSA cable A port
V port
RA LV3 & LV4 (A port)
RV
17
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
LV
LV1 & LV2 (V port)
PSA Settings
PSA Nominal Settings
a
NOTE: Verify PSA Settings prior to beginning lead testing, making modifications as necessary
Noise Filter
If noise is evident on electrogram trace, turn ON (a) Filter for (b) 50Hz or 60Hz frequencies b
18
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
This feature is nominally OFF.
Sensing
Typical Method for P/R Wave Amplitude Measurements
• Select desired (a) Surface ECG / PSA Lead a Traces based on lead connected • If testing (b) LV Lead, ensure Pace / Sense configuration is set from the PSA Test Chamber Measurements panel • Go to the (c) PSA Settings and Output (PW and amplitude) panel and turn OFF pacing to all channels where no leads are connected • Disable (c) Pacing and enable sensing in the channel you want to measure (alternatively, c decrease (d) pacing rate below the intrinsic rate of patient) • Sense measurement values are now displayed on the (e) EGM with each corresponding sensed beat • Repeat procedure for other channels
19
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA
d
e
b
Sensing
Slew Rate and RV-LV Timing
(a) Slew displayed with each sensed intrinsic activity (b) RV–LV interval displayed when RV and LV cables are connected; a negative value will be displayed if LV is sensed before RV
a
b
Sensed measurement values displayed beat-by-beat
Slew Rate displayed during sensed intrinsic activity on beat-by-beat basis RV–LV Interval displayed
20
© 2020 Boston Scientific Corporation or its affiliates. All rights reserved. CRM-925601-AA