Radio Frequency Ablation RF 3000™ Generator Error Codes HARDWARE SELF-TEST ERRORS Error Code
Malfunction (Unrecoverable)
Description
H01
CPU Configuration Error
√
H02
COP Timeout Error
√
H03
ROM CRC Error
√
H04
RAM Error
√
H05
Illegal CPU Instruction
√
H06
Duplicate RAM Data Error
√
H07
POST Error: Power Generation/Measurement
√
H08
POST Error: Impedance Generation/Measurement
√
H09
POST Error: Voltage Generation/Measurement
√
H10
POST Error: Current Generation/Measurement
√
H11
POST Error: Pad Current Generation/Measurement
√
H12
POST Error: Pad Current Offset Measurement
√
H13
POST Error: Watchdog Error
√
Error (Recoverable)
MEASUREMENT ERRORS E01
Low Impedance Measured
√
E02
High Initial Impedance Measured
√
E03
High Impedance Measured
√
P-1
High Current in Pad 1
√
P-2
High Current in Pad 2
√
P-3
High Current in Pad 3
√
P-4
High Current in Pad 4
√
E80
Pad Current Sum Error
√
E81
Hardware Shutdown: Over Voltage
√
E82
Hardware Shutdown: Over Current
√
E83
Hardware Shutdown: Over Power
√
E84
Hardware Shutdown: Multiple
√
E85
Reference Voltage Error
√
E86
Power Offset Error
√
E87
High Voltage Measured
√
E88
High Current Measured
√
E89
High Power Measured
√
E90
RESERVED (RF-2000 POST Measurement Error)
√
E91
Software Shutdown
√
E92
PAD Current Offset Error
√
E93
Vrms Offset Error
√
E94
Irms Offset Error
√
Customer Service, Ordering Information 1.888.272.1001
Radio Frequency Ablation
RF 3000™ Generator Error Codes
HARDWARE SELF-TEST ERRORS Error Code
Description
Malfunction
Explanation
Action
H01
CPU Configuration Error
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H02
COP Timeout Error
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H03
ROM CRC Error
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H04
RAM Error
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H05
Illegal CPU Instruction
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H06
Duplicate RAM Data Error
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H07
POST Error: Power Generation/Measurement
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
H08
POST Error: Impedance Generation/Measurement
Unrecoverable Hardware
May occur when generator is very cold.
Cycle AC Power Off/On. If problem persists, contact BSC service.
H09
POST Error: Voltage Generation/Measurement
Unrecoverable Hardware
May occur when generator is very cold.
Cycle AC Power Off/On. If problem persists, contact BSC service.
H10
POST Error: Current Generation/Measurement
Unrecoverable Hardware
May occur when generator is very cold.
Cycle AC Power Off/On. If problem persists, contact BSC service.
H11
POST Error: Pad Current Generation/Measurement
Unrecoverable Hardware
May occur when generator is very cold.
Cycle AC Power Off/On. If problem persists, contact BSC service.
H12
POST Error: Pad Current Offset Measurement
Unrecoverable Hardware
May occur when generator is very cold.
Cycle AC Power Off/On. If problem persists, contact BSC service.
H13
POST Error: Watchdog Error
Unrecoverable Hardware
Cycle AC Power Off/On. If problem persists, contact BSC service.
MEASUREMENT ERRORS E01
Low Impedance Measured
Recoverable
Starting Impedance is < 15 ohms.
Check electrode, pads and connections for a short circuit. Correct problem and press RESET.
E02
High Initial Impedance Measured
Recoverable
Starting Impedance is > 400 ohms.
Check electrode, pads and connections for open circuit. Correct problem and press RESET.
E03
High Impedance Measured
Recoverable
After START, high impedance observed with RF power on.
Check electrode, pads and connections for disconnection or complete desiccation. Correct problem and press RESET.
Customer Service, Ordering Information 1.888.272.1001
Radio Frequency Ablation
RF 3000™ Generator Error Codes
MEASUREMENT ERRORS (cont.) Error Code
Description
E80
Pad Current Sum Error
E81
Hardware Shutdown: Over Voltage
E82
Hardware Shutdown: Over Current
Malfunction
Explanation
Action
Unrecoverable
Current flow to the four return pads differs. Or External RF source (e.g. ESU in use)
Discontinue use of ESU generator. Check pad position. Cycle AC Power Off/ On. If problem persists, contact BSC service.
Unrecoverable
RF voltage in excess of the hardware limit was detected. This may possibly result from a very rapid impedance increase (e.g. and open circuit).
Check pads and electrode. Discontinue use of ESU generator. Cycle AC Power Off/On. If problem persists, contact BSC service.
Unrecoverable
RF current in excess of the hardware limit was detected. This may possibly result from Check for possible short circuit. Cycle a very rapid impedance decrease (e.g. a AC Power Off/On. If problem persists, short circuit). External RF source (e.g. ESU in contact BSC service. use) or low impedance. Cycle AC Power Off/On. If problem persists, contact BSC service.
E83
Hardware Shutdown: Over Power
Unrecoverable
RF power in excess of the hardware limit was detected. This may possibly result from a very rapid impedance decrease. External RF source (e.g. ESU in use) or impedance decreased rapidly and power exceeded limit.
E84
Hardware Shutdown: Multiple
Unrecoverable
More than one of the conditions for E81-E83 occurred. Impedance may have decreased enough to allow Power to rise above limit.
Cycle AC Power Off/On. If problem persists, contact BSC service.
E85
Reference Voltage Error
Unrecoverable
May occur when generator is very cold.
Allow generator to warm to room temperature. Cycle AC Power Off/On. If problem persists, contact BSC service.
E86
Power Offset Error
Unrecoverable
May occur when generator is very cold.
Allow generator to warm to room temperature. Cycle AC Power Off/On. If problem persists, contact BSC service.
E87
High Voltage Measured
Unrecoverable
Sudden open circuit. E01, E02 or E03 usually occur first and suppresses the display of E87.
Check for possible open circuit. Cycle AC Power Off/On. If problem persists, contact BSC service.
E88
High Current Measured
Unrecoverable
Sudden decrease in impedance.
Check electrode, pads and connections. Cycle AC Power Off/On. If problem persists, contact BSC service.
Sudden decrease in impedance after setting a high power level. Or: While impedance is rapidly rising, the operator is pressing the DECR PWR button.
Cycle AC Power Off/On. If problem persists, contact BSC service.
E89
High Power Measured
Unrecoverable
(E90)
RESERVED (POST Measurement Error)
Unrecoverable
E91
Software Shutdown
Unrecoverable
Multiple possible causes.
Cycle AC Power Off/On. If problem persists, contact BSC service.
E92
PAD Current Offset Error
Unrecoverable
May occur when generator is very cold.
Allow generator to warm to room temperature. Cycle AC Power off/On. If problem persists.
E93
Vrms Offset Error
Unrecoverable
May occur when generator is very cold.
Allow generator to warm to room temperature. Cycle AC Power off/On. If problem persists.
E94
Irms Offset Error
Unrecoverable
May occur when generator is very cold.
Allow generator to warm to room temperature. Cycle AC Power off/On. If problem persists.
Customer Service, Ordering Information 1.888.272.1001
RF 3000™ Generator Error Codes
Radio Frequency Ablation PAD ERRORS Error Code
P-1
P-2
P-3
P-4
Description
Malfunction
Explanation
Action
High Current Pad 1
Recoverable / RESETABLE
Current flow to pad 1 exceeds .78 amperes. There is uneven current distribution amongst the four return pads.
Check pad placement relative to the other pads. Check cable routing or “coiling”. Correct problem. Press RESET band continue.
High Current Pad 2
Recoverable / RESETABLE
Current flow to pad 2 exceeds .78 amperes. There is uneven current distribution amongst the four return pads.
Check pad placement relative to the other pads. Check cable routing or “coiling”. Correct problem. Press RESET band continue.
High Current Pad 3
Recoverable / RESETABLE
Current flow to pad 3 exceeds .78 amperes. There is uneven current distribution amongst the four return pads.
Check pad placement relative to the other pads. Check cable routing or “coiling”. Correct problem. Press RESET band continue.
High Current Pad 4
Recoverable / RESETABLE
Current flow to pad 4 exceeds .78 amperes. There is uneven current distribution amongst the four return pads.
Check pad placement relative to the other pads. Check cable routing or “coiling”. Correct problem. Press RESET band continue.
Peripheral Interventions 300 Boston Scientific Way Marlborough, MA 01752-1234 www.bostonscientific.com To order product or for more information contact customer service at 1.888.272.1001. © 2016 Boston Scientific Corporation or its affiliates. All rights reserved.
RF 3000 is a registered trademark of Boston Scientific Corporation or its affiliates.
ENDO-149637-AA PI-376902-AA FEB2016 FEB2016