Galaxy Ultra HD-LED® Mobile Models Instructions for Use
IU-GLED-MOBILE Issue B4 – 03rd December 2013
Galaxy Ultra HD-LED Mobile Models Instructions for Use
© Brandon Medical Company Ltd 2009 All rights are reserved. No part of the content may be reproduced, adapted, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without the prior permission of the Copyright owner. Whilst every effort is made to ensure the accuracy and completeness of this guide, we do not warrant that the content is error free. The brand names or product names mentioned or referred to throughout this guide are fully recognised as the trademark or registered trademark of their respective owners.
Page 2 of 21
IU-GLED-MOBILE
Contents
Contents 1
Before using the equipment ...4 1.1 1.2 1.3 1.4 1.5
2
Symbols used ...4 Models covered...4 Classifications ...4 Environmental conditions...5 Related guides ...6
Operating instructions...7 2.1 2.2 2.3 2.4 2.5 2.6 2.7
Using the Control Unit ...7 Removing/replacing the sterilisable handle cover ...8 Controlling the focus ...14 Storing the mains cable ...15 Moving the unit...15 Warnings...16 Emergency power system for EL models only ...17
3
Cleaning instructions...18
4
Disposal instructions ...19
5
Declaration of Conformity ...20
IU-GLED-MOBILE
Page 3 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
1
Before using the equipment For your full satisfaction with this equipment it is recommended that you read this section carefully before you start to use it.
1.1
Symbols used The symbols used in this guide and on the product are explained in the following table. Symbol
Description
!
WARNING/DANGER – Pay special attention. Degree of protection from electric shock.
1.2
Note this point. Refer to another section/guide.
Models covered Model
Approx. weight
GLED30M
93 Kg
GLED30MEL
135 Kg
500nm Cell Safety Light. This product is a special adaptation of a Galaxy Ultra HD-LED surgical light to reduce blue light toxicity in delicate cell tissue. Short wavelengths of visible light (visible violet and visible blue light) below 500 nanometers is known to damage delicate cells including embryonic cells, stem cells and human retinal cells. This product has been adapted to reduce this radiation to safe levels. A Cell Safety Light has the suffix F5. White light without the visible violet and blue appears yellow to the human eye. All other functions of the product described in this manual are standard.
Page 4 of 21
IU-GLED-MOBILE
Before using the equipment
1.3
Classifications The classifications are explained in the following table.
1.4
Classification
Description
Electrical classification
Class 1
Degree of protection against electric shock
Class B
Classification according to EU Medical Device Directive 93/42/EEC
Class 1
Lamp head degree of protection from ingress of dust and water
IPXO
Degree of protection against the presence of flammable anaesthetic mixtures
Not for use in a flammable atmosphere
Environmental conditions The environmental conditions relate to transportation/storage and operating.
1.4.1
Transportation and storage The environmental conditions relating to transportation and storage are explained in the following table.
1.4.2
Condition
Range
Temperature
-40oC to 70oC
Relative humidity
10% to 95% non-condensing
Pressure
500hPa to 1060hPa
Operating The environmental conditions relating to operating are explained in the following table.
IU-GLED-MOBILE
Condition
Range
Temperature
15oC to 30oC ambient temperature
Relative humidity
10% to 80% non-condensing
Page 5 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
1.5
Related guides The other guides that relate to the Galaxy Ultra lamps are:
Page 6 of 21
Galaxy Ultra Mobile & HD-LED Models Assembly Guide
Galaxy Ultra Mobile & HD-LED Models Routine Maintenance Guide.
IU-GLED-MOBILE
Operating instructions
2
Operating instructions This section covers:
2.1
Using the Control Unit
Removing/replacing the sterilisable handle cover
Controlling the focus
Storing the mains cable
Warnings.
Using the Control Unit The main features of the Control Unit are highlighted in the illustration. Fault/service indicator
ON/OFF switch
Decrease % visible red light
Camera zoom in
Increase % visible red light
Camera zoom out
Decrease light intensity
2.1.1
Increase light intensity
Initially, when the master control unit is switched on, all the LEDs are briefly illuminated while an automatic software test is performed.
Switching the lamp on/off You press the base unit power switch then the control unit switch to toggle the lamp on and off. When the lamp is:
2.1.2
On, the small LED immediately beneath and to the left of the switch glows green.
Off, the small LED immediately beneath and to the right of the switch glows red which indicates standby mode.
Increasing/decreasing the red balance The % visible red light, ie the red balance control, is by default set at the midpoint. To decrease the red balance, press the Up arrow until the required balance is achieved.
IU-GLED-MOBILE
Page 7 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
To increase the red balance, press the Down arrow until the required balance is achieved.
2.1.3
Increasing/decreasing the light intensity The light intensity control is by default set to maximum, ie all six LEDs are illuminated. To increase the light intensity, press the Plus button until the required intensity is achieved. To decrease the light intensity, press the Minus button until the required intensity is achieved.
2.1.4
Zooming the surgical camera in/out (if fitted) To zoom the camera in, press the Plus button until the required field of vision is achieved. To zoom the camera out, press the Minus button until the required field of vision is achieved.
2.2
Additional Controls
2.2.1
Entering the ‘default setting’ Mode To enter the ‘default setting’ mode press the ‘red balance up and down key’ with the ‘Increase intensity key’ simultaneously. The key pad display should have the lower Red Balance (RB-LED 1), On-LED and Stand-by LED illuminated when the ‘default mode’ has been entered.
To leave the default setting mode ensure the RB-led is at position 1 then press the ‘on/off button’. The only LED lit should be the red standby led. Table1 below lists the six default setting states, the first state allows entry and exit of the default setting mode. The next five are the parameters states.
Parameter list
Ct-Led’s position
Int-Led’s (Blue) position
Other Led’s On
Enter/Leave default mode
RB-led 1 On
No led’s are illuminated
Stand-By + ON-Led
On/Off-Behaviour
RB-led 2 On
Int-Led 1 or 2
Stand-By + ON-Led
Intensity after Power-On
RB-led 3 On
Refer to Table 2
Stand-By + ON-Led
RB after Power-On
RB-led 4 On
Refer to Table 3
Stand-By + ON-Led
Working Hours for service Led
RB-led 5 On
Refer to Table 4
Stand-By + ON-Led
Reset service hours
RB-led 1 and 2 On
Int-Led 1
Stand-By + ON-Led
Pairing Mode
RB-led 1 and 3 On
Int-Led 1
Stand-By + ON-Led
Table 1
Page 8 of 21
IU-GLED-MOBILE
Operating instructions
2.2.2
Changing Power-On behaviour This sets the lamp to power on or off when the main power is first applied to the lamp head. Press the Red Balance (RB) Up-button until RB-Led 2 is on. Refer to fig 1.
Stand-by LED RB-LED 2
Service LED
Int-LED 1
Fig 1 Int-LED 1(Blue) is on; the lamp will be off when it is first powered up. Int-LED 2(Blue) is on; the lamp will come on when it is first powered up. The Service LED will blink when a different setting is selected, to save the change to memory press the standby button. The service LED will stop flashing after the on/off button has been pressed. Note: Service light is lit in fig 1 because another setting has been chosen.
2.2.3
Changing the default intensity at power-up This sets the lamp intensity when the lamp is powered Press the Red Balance (RB) up button until RB-Led 3 is on. 1 of 6 Int-Leds will be on. To increase the light intensity, press the Plus button until the required intensity is achieved. To decrease the light intensity, press the Minus button until the required intensity is achieved. The Service LED will blink when a different setting is selected, to save the change to memory press the on/off button.
IU-GLED-MOBILE
Page 9 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
Int-Led
Intensity after PowerOn
1
10 %
2
40 %
3
55 %
4
70 %
5
85 %
6
100 % Table 2
The factory default level for light intensity is 3, i.e. three of the six Led’s are illuminated.
2.2.4
To leave the default setting mode ensure the (Red Balance) RB-led is at position 1 then press the ‘on/off button’. The only LED lit should be the red standby led.
Changing the default colour temperature at power-up This sets the colour temperature of the lamp when first powered up using the on/off button. Press the up button until RB-Led 4 is on. To increase the colour temperature, press the intensity Plus button until the required intensity is achieved. To decrease the colour temperature, press the intensity minus button until the required intensity is achieved. The Service LED will blink when a different setting is selected, to save the change to memory press the on/off button.
Int-Led
Colour Temperature after Power-On
1
3500 K
2
3700 K
3
4000 K
4
4300 K
5
5000 K Table 3
The factory default level for colour temperature is 3, i.e. three of the five LED’s will be illuminated.
To leave the default setting mode ensure the RB-led is at position 1 then press the ‘on/off button’. The only LED lit should be the red standby led.
Page 10 of 21
IU-GLED-MOBILE
Operating instructions
2.2.5
Setting the service hour counter Sets the service interval time, when the working hours are exceeded by the internal timer the service LED comes on. Press the up button until Ct-Led 5 is on. To adjust the working hours press the intensity plus or minus keys. 1 of the 6 Int-Leds will be on. The Service LED will blink when a different setting is selected, to save the change to memory press the on/off button.
Int-Led
Working Hours
1
32500
2
1000
3
2000
4
3000
5
4000
6
5000 Table 4
The factory default level for working hours is 1000Hours, i.e. two of the five Led’s will be illuminated.
2.2.6
To leave the default setting mode ensure the RB-led is at position 1 then press the ‘on/off button’. The only LED lit should be the red standby led.
Resetting the service hour counter Turns the service LED off and resets the service timer. Press the intensity up button until RB-Led 1 and RB-led2 are on. Step intensity led bar up until Int-led 2 goes on. The service LED will flash; now push the ON/OFF-button. Once this is done the service light and service timer will be reset
2.2.7
To leave the default setting mode ensure the RB-led is at position 1 then press the ‘on/off button’. The only LED lit should be the red standby led.
Responding to the fault/service indicator Should the fault/service indicator LED be lit, immediately contact Brandon Medical or Dealer/Supplier by telephoning +44 (0)113 277 7393.
IU-GLED-MOBILE
Page 11 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
2.2.8
Pairing the Lamp to Remote Control (if fitted) 1. Set the lamp to standby ’stand-by-led’ is on 2. Enter the ‘default mode’ by simultaneously pressing the ‘red balance down key’ with the ‘Increase and decrease intensity key. 3. Continualy Press the Red balance ‘Up key’ until the RB-Led 1 and RB-Led 3 are illuminated. 4. Find the pairing button on the bottom face of the remote control. Do not press
the button yet!
Pairing Button
5. Press the lamphead Increase intensity button until Int-Led 3 is illuminated, then the led’s will flash and initialise the pairing mode. Now promptly press the pairing button on the remote until the single green led comes on. There will be a delay of 3 seconds and then two Red led’s will light on the remote. The pairing of the light and remote should be complete. 6. Press On/OFF button on the lamp keypad. 7. Press the down button until only RB-led 1 is lit and then press On/OFF button on the lamp keypad 8. Press standby again and the lamp will come on at its default settings. The remote will reflect the settings on the lamp key pad. Pairing is now complete Notes: 1. The first lamp to be paired will be allocated to channel one on the remote. 2. To reset the remote press and hold the pairing button in until the display goes from a single green until the pair of red led’s come on. 3. Initialising the pairing mode from the lamp keypad will unpair the lamp from the remote control unless the paring button on the remote is pressed. 4. If you are having problems pairing the lamp and the remote control then reset the remote control.
Page 12 of 21
To leave the default setting mode ensure the RB-led is at poison 1 then press the ‘on/off button’. The only LED lit should be the red standby LED.
IU-GLED-MOBILE
Operating instructions
2.3
Replacing the sterilisable handle cover To fit a sterilisable handle cover you only need to push the handle cover over the Carrier until an audible click is heard and the handle has locked in position. As shown below. To release the sterilisable handle cover, simply push the blue release button and the cover can be removed.
Grub Screw
Release Button
Sterile Handle Cover (HM33562)
!
The sterilisable handle cover as supplied is not sterile. You must sterilise it by autoclave – typically steam sterilisation at 134°C for three minutes – or by using normal sterilising solutions in accordance with the manufacturer’s instructions.
Extra handle covers – Part No. HM33562C – are available from Brandon Medical.
Disposable handle cover systems are available from Brandon Medical.
IU-GLED-MOBILE
Page 13 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
2.4
Controlling the focus Galaxy Ultra lamps are fitted as standard with the Brandon Electronic Focus Control System. You adjust the focal point of the light beams, ie the point where they converge, by twisting the sterilisable handle until the desired focus pattern is achieved.
Page 14 of 21
IU-GLED-MOBILE
Operating instructions
2.5
Storing the mains cable When not in use, the mains cable can be stowed as shown in the illustration.
2.6
Moving the unit Before moving the unit ensure the brakes are released as illustrated below.
IU-GLED-MOBILE
Page 15 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
Only use the blue stem handle illustrated below to manoeuvre the unit and under no circumstances should you use the light head focus handle to move the unit.
2.7
Warnings The following warning applies:
!
Page 16 of 21
Disconnect from all electrical supplies before removing covers. Models operating from low voltage and emergency power systems may present a hazard due to the possibility of high discharge currents.
Refer to Galaxy Ultra Mobile Models Routine Maintenance Guide for more information about the general maintenance of lamp heads and spring arms.
IU-GLED-MOBILE
Operating instructions
2.8
Emergency power system for EL models only The EL model lamp is fitted with an emergency battery backup system so that in the event of mains power failure the lamp will still operate. On receipt of the lamp the batteries will require charging for 24 hours to ensure they are fully charged. The lamp should ALWAYS be connected to the mains supply whether in use or not in use to maintain the battery charge. If this is not adhered to then the batteries will become damaged. Indicator
Description Under normal mains supply the green indicator will be illuminated.
If there is a fault with the charger or the batteries are not fully charged, then the red indicator will illuminate. Ensure the lamp is connected to the mains supply, ie the green indicator is on, and leave to charge for a minimum of 12 hours and up to a maximum of 24 hours. If the indicator is still illuminated there is a fault with the charging system and you must consult your local service agent or contact Brandon Medical. When the mains supply is disconnected then the red indicator will illuminate, showing the batteries are in use. The lamp will continue to operate until the batteries become discharged at which point the batteries will become disconnected and the lamp will not illuminate. The mains supply should be re-established before the lamp will reilluminate. The batteries will need to be fully recharged.
IU-GLED-MOBILE
Page 17 of 21
Cleaning instructions
3
Cleaning instructions !
Do not clean the lamp heads with alkaline cleaning solutions or bleach, which may damage the plastic material.
Brandon Medical approves the use of the following cleaning solutions on its products:
HexaquartR plus – dilute to strength
MeliseptolR rapid – pre-prepared spray
MeliseptolR HBV – tissues.
These chemicals are manufactured by B Braun Ltd and are available from Brandon Medical. To order these products please contact your local stockist. Please ensure that the solutions are not applied to warm surfaces and that any excess solution is wiped from the surface and not allowed to dry as this may lead to a build-up of deposits. Make sure you check the manufacturer’s instructions regarding the dilution factors for a particular solution. You can clean the body of the lamp using proprietary non-alkaline, sterilising solutions as instructed by the manufacturer.
IU-GLED-MOBILE
Page 18 of 21
Galaxy Ultra HD-LED Mobile Models Instructions for Use
4
Disposal instructions
Do not dispose of the light with normal refuse. Depending on local regulations, dispose of the light at a recycling centre or return it to a dealer with a disposal service. Equipment must be de-contaminated before disposal or return to dealer.
Page 19 of 21
IU-GLED-MOBILE
Declaration of Conformity
5
Declaration of Conformity
Brandon Medical Co Ltd hereby declares that this product – Galaxy Ultra HD-LED Mobile GLED30M – is in compliance with the EC Directive 93/42/EEC. The Class 1 medical device specified above conforms to the essential requirements listed in the Annex 1 of the EC Directive 93/42/EEC of 14 June 1993. Despite any compliance with EMC standards, the device may emit radiation that interferes with adjacent equipment. If this becomes apparent then increase the distance between the two pieces of equipment.
IU-GLED-MOBILE
Page 20 of 21