Nippy 4+ Quick Reference Guide
Quick Reference Guide
2 Pages

Preview
Page 1
Symbols
Usage of Oxygen (Accessory) •
O2 must be turned off when the Nippy 4+ is not in
•
operating mode! It is recommended to monitor the oxygen concentrations.
SYMBOL
DESCRIPTION Internal battery Click-in battery level
1.
2.
Mount the oxygen tube to the connector.
Keypad lock activated Keypad lock deactivated
Connect the oxygen tube. A click sound is heard when the tube is properly attached.
Click-in Humidifier (accessory) Heated circuit (accessory)
Disconnect: 1. Press the release button to disconnect the oxygen tube. 2. Remove the tube. 3. Turn off the oxygen source.
Multiple pages
Replacing the Patient Air Inlet Filters The patient air filters are located in the filter cassette at the side of the ventilator. Washable Filter (grey): Replace at least once a year, wash at least once a week. Disposable Filter (white): Replace at least every 4th week, or more often in high pollution or pollenrich environments.
1 2
Quick Reference Guide Doc. 007768 M-2 2022-11-30
Quick Reference Guide This is not a complete guide for the Nippy 4+. See your manual for complete instructions.
Start/Stop Treatment
Connect the Click-in Battery (Accessory)
1.
To start treatment and enter operating mode first press and hold the Start/Stop button on the front panel.
1.
Release the side cover by pressing the button under the handle.
2.
Release the Start/Stop button when the progress bar is filled.
2.
Open and remove the side panel.
3.
To stop treatment and enter standby mode first press and hold the Start/Stop button on the front panel.
4.
Release the Start/Stop button when the progress bar is filled. Press the Audio Pause button to stop the treatment.
3.
Insert the click-in battery. Make sure the latch at the bottom of the click-in compartment is engaged.
4.
Close the side panel. Make sure there is a clicking sound to secure the side panel.
Switch On/Off Press the On/Off button on the side panel.
Connect the Patient Circuit Always perform a pre-use test when the patient circuit is replaced or modified. If the pre-use test is not prompted at startup, you can start it from the menu. The leakage from the mask leakage port should be at least 12 l/min at 4 cm H2O to prevent rebreathing of exhaled air. The recommended leakage is 20 to 50 l/min at 10 cmH2O pressure. 1.
2.
Connect the patient circuit to the patient air outlet on the ventilator.
Connect the other end of the patient circuit to a leakage port or to a patient interface with integrated leakage port.
5.
i-Button Use the i-button to get explanations about settings and alarms.
Connectors Make sure that the marking is facing up.
Pull the connector sleeve, not the cable itself to release the connector.