Brymill

Cryosurgical Equipment

Cry-Ac's Contact Probes Instructions for Use Issue 8 Feb 2021

Instructions for Use

3 Pages

105 Windermere Avenue Ellington. CT 06029 Tel: (860) 875 2460 Fax: (860) 872 2371  Instructions for Use – Contact Probes Intended Use Probes are used with Cry-Ac’s which is a Hand-Held Cryosurgical Device for the controlled dispensing of Liquid Nitrogen. 1) Scope This Instructions for Use applies to the following Brymill Cryogenic Systems products: • Mini probes, Model 214; • Conical probes, Model 203; • Ball probes, Model 201; • Flat probes, Model 205. 2) Warnings and Cautions: a. Extreme care must be taken to properly direct the silicone vent tube which is attached to the base of the probe. This tube will harden after approximately five (5) seconds and remain in a fixed position during the remainder of the procedure. b. It is imperative that they be pointed is a safe position away from the patient and the user at the onset of the procedure. c. In order to prevent freezing of underlying structures such as tendons, nerves, or blood vessels, pull the probe back slightly to raise the skin away from the underlying structures after the lubricant jelly has adhered to the skin.  February 2021 Issue 8 1
File Type: PDF
File Size: 1.24 MB
File Name: Brymill - Cry-Ac's Contact Probes Instructions for Use Issue 8 Feb 2021 - 2021-02.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions