Appendix 1 – Addendum to the Directions for Use for the T34TM and T34LTM (T60TM) Ambulatory Syringe Pumps (T34™ Directions for Use document reference number: 100-090SM / 100090SS T34L™ (T60™) Directions for Use document reference number: 100-090SL / 100-090SLM)
Monitoring and Managing Infusions Pump and Infusion Safety Checks Inspect the lead screw prior to use (See 3.2 Pump Description, item 4). If there is white plastic debris on the lead screw, this is an indication of wear on the syringe mechanism. Therefore, discontinue use and send the pump for service.