Caesarea Medical Electronics Ltd

Syringe Driver

T34 and T34L (T60) Addendum to Directions for Use Appendix 1 March 2020

Addendum

1 Page

Appendix 1 – Addendum to the Directions for Use for the T34TM and T34LTM (T60TM) Ambulatory Syringe Pumps (T34™ Directions for Use document reference number: 100-090SM / 100090SS T34L™ (T60™) Directions for Use document reference number: 100-090SL / 100-090SLM)  Monitoring and Managing Infusions Pump and Infusion Safety Checks Inspect the lead screw prior to use (See 3.2 Pump Description, item 4). If there is white plastic debris on the lead screw, this is an indication of wear on the syringe mechanism. Therefore, discontinue use and send the pump for service.  March 2020
File Type: PDF
File Size: 111 KB
File Name: Caesarea Medical Electronics Ltd - T34L (T60) Addendum to Directions for Use Appendix 1 March 2020 - 2020-03.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions