Caesarea Medical Electronics Ltd
Syringe Driver
T34 Syringe Pump 2nd and 3rd Edition Updates Issue 1 2019
Updates
1 Page
Preview
Page 1
T34 Syringe Pump 2nd Edition-3rd Edition updates
T34 changes include: • Compliance with standard IEC 60601-1 Edition 3.1 • IP22 • Bug fixes
• PCBA update to alarms • Update to DFU, QRG & Patient guide • Bolus capability (new)
Alarms Product changes T34 Syringe - alarm tones and priorities Description
Alarm Type
Audio Signal as per 60601-1-8
Visual Signal as per 60601-1-8
High Priority Alarm Requires immediate user response
High Priority Volume = ~ 58 dBA
RED flashing visual Operation LED flashes RED
Medium Priority Alarm Requires prompt user response
3 tones Volume = ~ 58 dBA
YELLOW flashing visual Operation LED flashes YELLOW
Low Priority Alarm Requires user awareness
Low Priority 2 or 3 pulses Volume = ~ 54 dBA
YELLOW solid visual Operation LED solid YELLOW
Information Signal Provides information that may or may not require actions from clinicians
1 or 2 pulses
No visual
Down occlusion End battery End of infusion Syringe displaced during infusion Restart pump Switch off and on ERROR XX End of infusion (keep vein open) Pump paused too long Low battery Near end Bolus started/completed Keypad lock/unlock Syringe plunger hit the limit Syringe loaded Purge started/end Power on/off Infusion started/resumed/ stopped by user KVO stopped by user Service interval alert
Visual changes
2E
International
3E
UK
T34
Text
1030 Eskdale Rd, Winnersh Triangle, Wokingham, Berkshire, RG41 5TS
bd.com/en-uk BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company. © 2019 BD. 0000CF04374 Issue 1.
IEC Symbols