Operator Manual
46 Pages
Preview
Page 1
T60
TM
Syringe Pump System Operator Manual
0344
Caesarea Medical Electronics Ltd. 16 Shacham street Industrial Park Caesarea North P.O.Box 4294 Caesarea 38900, Israel E: [email protected] W: www.cme-infusion.com
Software Version LCAT / LCATBOL
MedNet GmbH. Borkstrasse 10 48163 Muenster Germany T: +49 251 32266 0
100-090SLM Rev. 03
Contents SECTION 1: GENERAL INFORMATION ... 3 1.1 Preface ... 3 1.2 About this manual ... 3 1.3 Advisory terms and operating precautions and warnings ... 3 1.4 Intended use ... 5 1.5 Contraindications ... 5 1.6 System Symbols ... 5 1.7 Syringe pump specifications ... 6 1.8 Limited warranty ... 7 1.9 Pump inspection and unpacking ... 7 SECTION 2: DISPOSABLES AND ACCESSORIES ... 8 2.1 Syringe brands and sizes... 8 2.2 Administration sets ... 9 2.3 Lockbox ... 10 2.4 Carry Pouches ... 10 SECTION 3: PUMP FEATURES AND DESCRIPTION ... 12 3.1 Overview ... 12 3.2 Pump description ... 13 3.3 Event log ... 15 3.4 Post occlusion bolus reduction system (POBRS) ... 16 3.5 Mains power supply and internal battery ... 17 SECTION 4: MODES OF OPERATION ... 19 4.1 Modes of operation ... 19 SECTION 5: PUMP CONFIGURATION ... 22 5.1 Pump configuration ... 22 5.2 Pump access codes ... 22 5.3 Pump INFO and configuration menu`s ... 23 5.4 Pump Configurable settings for modes of operation ... 25 5.5 Optional configurable settings ... 26 5.6 Practise scenarios for changing pump configuration ... 28 SECTION 6: STARTING A NEW INFUSION ... 31 6.1 Sequence for starting an infusion ... 31 SECTION 7: MONITORING AND MANAGING INFUSIONS... 38 7.1. Pump and infusion safety checks ... 38 7.2 Keypad lock... 39 7.3 Program protection and “Resume” ... 40 7.4 Stopping/pausing the infusion and powering off ... 41 7.5 Alerts, alarms and troubleshooting ... 42 7.6 Changing syringes/administration sets ... 43 7.7 Servicing and maintenance ... 44 Software Version LCAT / LCATBOL
100-090SLM Rev. 03 2
SECTION 1: GENERAL INFORMATION 1.1 Preface Duplication or distribution of this manual and any information contained within is strictly prohibited without the express written permission of CME Ltd. This manual and any information contained within, may not be reproduced, distributed, or transmitted in any form, or by any means, for any purpose, without the express written permission of CME Ltd. To order additional copies of this manual, or other related manuals, contact CME Ltd. Head Office. The information in this document has been carefully compiled, and is believed to be entirely reliable. However, no responsibility is assumed for inaccuracies. Furthermore, CME Ltd. reserves the right to make changes to any products herein to improve readability, function, or design. CME Ltd. does not assume any liability arising out of the application or use of any product described herein; neither does it cover any license under its patent rights nor the rights of others.
1.2 About this manual The operator must be thoroughly familiar with the T60™ syringe pump described in this manual prior to use, and in particular must read and understand any warnings and precautions stated herein. If a software change occurs and the operation/specification for the pump changes, new or additional operating instructions will be issued, if needed. All illustrations used in this manual show typical settings and values that may be used in setting up the functions of the pump. These settings and values are for illustrative use only. The complete range of settings and values are listed in the specifications section of this manual. This operation manual document has been developed with consideration to the requirements in relevant Harmonised Standards. Data presented in the technical specifications reflect specific test conditions defined in this standard. Other external factors such as varying back pressure, temperature, head height, set usage, fluid restrictions, solution viscosity or combinations of these factors, may result in deviations from the performance data enclosed.
1.3 Advisory terms and operating precautions and warnings Warnings, precautions and notes Warnings and notes will be seen throughout this manual. These are described as: NOTE: Indicate that the information that follows is additional important information, a tip that will help you recover from an error or point you to related information within the manual. PRECAUTION: Indicates that the information is a precaution. Precautions advise you of circumstances that could result in damage to the device. Read and understand this manual and all precautions completely before operating T60™ ambulatory syringe pump. WARNING: Warnings advise of circumstances that could result in injury or death to the user/operator or circumstances that could result in damage to the device. Read and understand this manual and all warnings before operating the T60™ syringe pump.
Operating precautions and warnings o
o
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference (2) this device must accept any interference received, including interference that may cause undesired operation. TM Although the T60 syringe pump has been designed and manufactured to exact specifications, it is not intended to replace trained personnel in the supervision of infusions.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 3
o
o
o
o
o
o o o
CME Ltd will assume no responsibility for incidents which may occur if the product is not used, stored or transported in accordance with the environmental conditions stipulated in this document or on the package labelling. This syringe pump is designed for ambulatory use and should withstand everyday handling. If the pump is dropped onto a hard surface, or is suspected of being dropped, the operation and calibration should be checked by a qualified technician. The specified accuracy of the pump can only be maintained if the pump is used, maintained and serviced in accordance with the instructions given in this manual. If the pump has failed to calibrate during the servicing procedure, it must be returned for repair or disposal. Adjustments, maintenance, or repair made by un-certified service personnel may impair the operation of the pump and/or the accuracy of the infusion. Make sure any adjustments, maintenance, or repair of the device are carried out only by authorised and skilled technicians. Refer all service, repair and adjustments only to qualified and certified technical personnel. Unauthorised modifications or the use of any spare parts, other than those supplied by the manufacturer or their distributor, will void any warranty. If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to have been damaged, remove it from service for inspection by qualified personnel. The pump has been designed to be as safe as possible to handle; however, care should be exercised to avoid trapping of fingers or other body parts in the mechanism. TM o o o The T60 syringe pump system should be operated within a temperature range of +15 C to +45 C (+59 F to + 113 o F). Operation at temperatures outside this range may affect accuracy.
Infusion precautions and warnings Carefully read and follow accompanying set instructions for priming the set and the recommended set change interval. o The syringe and administration set should be disposed of in an appropriate manner, considering the nature of the residual fluid that may be contained within and in accordance with the hospital/homecare provider’s disposal practices. o Drugs for infusion to be used with the pump may only be prescribed by a qualified medical practitioner. Caution must be exercised in the selection of drugs and the amount and rate intended to be delivered via any infusion pump. o If the drug contained in the syringe will be exposed to extreme environmental conditions for prolonged time periods, it is important to select drugs that will not change pharmacologically upon such exposure. o As with all automatic infusion devices, whenever a toxic or dangerous level of drug is stored in the reservoir, constant/frequent monitoring of the infusion is required. o In all applications, time to alarm under occlusion or other fault conditions will depend on the infusion rate and levels of alarm settings. It is recommended to consider these parameters when using drugs requiring infusion stability or low flow rates and therefore a quick time to alarm.
General precautions and warnings o
The T60™ syringe pump must always be used in a CME Ltd. pouch or similar receptacle if used in direct sunlight. If the pump is exposed to direct sunlight, it may affect functionality.
o o
Do not use hard or sharp objects on the keypad. Do not bath or shower whilst using the pump. The pump is resistant to a limited amount of splashing, but its construction does not make it resistant to large amounts of spraying or immersion in liquids. Damage to the internal components may result. If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to have been damaged, remove it from service for inspection by qualified personnel. Do not operate the pump near high-energy radio-frequency emitting equipment, such as electro-surgical cauterising equipment as this may result in false alarm signals or error messages being displayed. The pump has been designed to be as safe as possible to handle; however, care should be exercised to avoid trapping of fingers or other body parts in the mechanism.
o o o
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 4
1.4 Intended use TM
The T60 syringe pump is designed for infusion of medications or fluids requiring continuous or intermittent delivery at precisely controlled infusion rates through all clinically acceptable routes of administration including intravenous, subcutaneous, percutaneous, epidural in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site). The system is intended for patients who require maintenance medications, analgesics, chemotherapeutic agents and general fluids therapy in hospital and homecare environments.
1.5 Contraindications
Infusion of blood and blood products Infusion of insulin Infusion of critical medications whose stoppage or interruption could cause serious injury or death Use in ambulatory regimens by patients who do not possess the mental, physical or emotional capability to self-administer their therapy; or who are not under the care of a responsible individual
1.6 System Symbols The following symbols are used on the T60™ syringe pump and components. Labels on the pump or statements in this manual preceded by any of the following words and/or symbols are of special significance and/or are intended to help you to operate the pump in a safe and successful manner.
System symbols Symbols
IPX1
Description Degree of protection against ingress of water.
Type of protection against electric shock - Class 11 equipment.
Attention, consult the accompanying instructions.
CE mark indicates conformance to Medical Device Directive 93/42/EEC. Do not dispose of in municipal waste. Symbol indicates separate collection for electrical and electronic equipment. (WEEE Directive 2002/96/EEC) NOTE: Does not apply to the battery.
Do not dispose of battery in municipal waste. Symbol indicates separate collection for battery is required. The use of single-use disposable components on more than one patient is a biological hazard. Do not reuse single-use disposable components. Type CF applied part.
Date of manufacture.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 5
SN
Serial number. Expiry date (consumables).
LOT
Lot number (consumables).
STERILE EO
Sterilized with Ethylene Oxide.
1.7 Syringe pump specifications TM
T60
Syringe Pump Specifications
Type:
Syringe pump with motor driven linear actuator, pulsed motion, (560 pulses per mm)
Flow rate:
0.05- 10mL/h in 0.01mL increments, 10 to 650mL/h in 1 mL increments
Actuator travel: Syringe sizes: Accuracy:
c.106 mm available 2 mL to 50/60 mL (most commonly used manufacturers, see section T) ± 5 % system accuracy (pump and set combined)
Occlusion pressure: 100-1500mmHg configurable (10mmHG increments) Maximum actuator force 50N (5 Kgf) Battery:
Integral rechargeable Li-Polymer 1800 mAh
Battery operation:
100 full deliveries depending on infusion and set up parameters
Indicators:
4 Line LCD display (128 x 64 pixels). Dual colour operation LED with backlight When a problem is detected, the T60™ displays the following alarm messages, sounds an audible alarm and the LED lights red Occlusion or Syringe Empty Pump Paused Too Long End Program Syringe Displaced End Battery Syringe Empty 38 x 55 x 190mm Type CF Equipment, degree of protection against electrical shock, Class II equipment. IPX1 protection against ingress of water. ABS (fire retardant) 400 gr. with battery Complies with IEC 60601-1 (Medical Electrical Equipment Safety), IEC 601-2-24 (Infusion pumps and controllers), IEC 601-1-4 (Programmable Electrical Medical Systems) Manufactured in accordance to ISO 9001:2008 and ISO 13485:2003. CE marked in accordance with the Medical Devices Directive 93/42/EEC Medical T60™ syringe pump is designed to be in compliance with EN60601-1 (Safety) and IEC 601-1-2 (EMC) Non-Operating Conditions (transportation and storage): o o o o Temperature: -25 C to 55 C (-13 F to + 131 F) Humidity: 5 % to 100% R.H., non-condensing Air pressure: 48kPa to 110kPa Operating Conditions: (the system may not meet all performance specifications if operated outside of the following conditions: o o o o Temperature: -15 C to 45 C (-59 F to 113 F) o Humidity: 20 % to 85% R.H. at +40 C, non-condensing Air pressure: 70kPa to 110kPa
Alarms:
dimensions: Classification: Housing: Weight: Electrical safety: Standards: EMC:
Environmental Specifications:
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 6
1.8 Limited warranty The T60™ syringe pump has been carefully manufactured from the highest quality components. CME Ltd guarantees the pump against defects in material and workmanship for twelve (12) months from date of purchase. CME Ltd.’s obligation, or that of its designated representative under this Limited Warranty, shall be limited, at CME Ltd.’s option, or that of its designated representative, to repairing or replacing pumps, which upon examination, are found to be defective in material or workmanship. The repair or replacement of any product under this Limited Warranty shall not extend the above-mentioned Warranty period. All repairs under this Limited Warranty should be undertaken only by qualified, trained service personnel. In the event that a pump is found to be defective during the warranty period, the purchaser shall notify CME Ltd. or its designated representative within thirty (30) days after such defect is discovered. The defective pump should be sent immediately to CME Ltd. or its designated representative for inspection, repair or replacement. Shipping costs are the purchaser’s responsibility. Material returned to CME Ltd. or its designated representative should be properly packaged using CME Ltd. shipping cartons and inserts. Inadequate packaging may result in damage to the pump. This Limited Warranty shall not apply to defects or damage caused, wholly or in part, by negligence, spilt fluids, dropping of the pump, misuse, abuse, improper installation or alteration by anyone other than qualified, trained personnel; or to damage resulting from inadequate packaging in shipping the pump to CME Ltd. or its designated representative. If, after inspection, CME Ltd. or its designated representative is unable to identify a problem, CME Ltd. or its designated representative reserves the right to invoice purchaser for said inspection. This Warranty is the sole and entire warranty pertaining to CME Ltd.’s products and is in lieu of and excludes all other warranties of any nature whatsoever, whether stated, or implied or arising by operation of law, trade, usage or course of dealing, including but not limited to, warranties of merchantability and warranties of fitness for a particular purpose. Purchaser expressly agrees that the remedies granted to it under this limited warranty are purchaser’s sole and exclusive remedies with respect to any claim of purchaser arising under this Warranty.
Caesarea Medical Electronics Ltd 16 Shacham street Industrial Park Caesarea North P.O.Box 4294 Caesarea 38900, Israel Email: [email protected] www.cme-infusion.com
MedNet GmbH Borkstrasse 10 48163 Muenster Germany T: +49 251 32266 0
0344
1.9 Pump inspection and unpacking TM
Remove the T60 syringe pump and accessories from the packaging and inspect for damage during shipment or TM storage. Do not use the T60 syringe pump and accessories if there are any obvious signs of damage. Return for inspection by authorised service personnel. Make sure you have the following items: TM T60 syringe pump Pole clamp External charger (mains power lead) WARNING: Visually inspect packaging and contents before each use. WARNING: Only use accessories designed for the system. Unsafe operation may result from using improper TM accessories. Use only accessories and options designed for this system and supplied or recommended by the T60 syringe pump distributor. Software Version LCAT / LCATBOL
100-090SLM Rev. 03 7
SECTION 2: DISPOSABLES AND ACCESSORIES 2.1 Syringe brands and sizes The T60™ syringe pump is programmed to recognize most commonly used syringes from 2 mL to 50 mL. Luer lock syringes should always be used to ensure secure connection of the infusion set and syringe. For homecare applications it is recommended that all but the brand of syringe in regular use are disabled per the technician menu to prevent accidental selection of the incorrect brand during set up . WARNING: DO NOT USE Slip-tip syringes. Luer Lock syringes must always be used to ensure secure connection of the infusion set and the pump is calibrated to Luer Lock. Failure to do so may result in under or over infusion as the dimension of some manufacturers slip-tip syringes differ from their own Luer Lock variants. It is recommended that all but the brand of syringe in regular use are disabled via the BodyComm™ Communication Software to prevent accidental selection of the incorrect brand during set up. Should the need arise to program the T60™ syringe pump to operate with a manufacturer and/or brand other than one of those listed above you should consult either your local medical engineering department or CME Ltd. Technical Services.
Default syringe brands configured for use
Braun Omnifix 2, 5, 10, 20, 30 & 50mL BD Plastipak 3, 5,10, 20, 30l & 50mL Monoject 6, 12, 20, 35 & 50mL Codan/Once 2.5, 5, 10, 20, 30 & 50mL Terumo 5, 10, 20, 30 & 50mL
It is possible to disable default syringes from the memory or replace them with one not listed above. This procedure is not detailed in this manual as it should only be undertaken by trained, certified service centres or biomedical engineers. Please consult a biomed engineer or your local T60™ syringe pump distributor should this need arise.
Syringe volumes Due to the physical length of the screw that drives the syringe plunger forward there are limits to the maximum amount that can be delivered from larger syringes and on some smaller syringes there is an undeliverable volume that will be left in the syringe once the actuator has driven to the zero position. NOTE: Braun 2mL, Codan 2.5mL and Monoject 6mL syringes exhibit a slight undeliverable volume of 0.1mL due to their design. The T60 will display the volume as “VOLUME 2.4 (of 2.5mL)”. In this example the pump can only deliver 2.4 of the 2.5mL in the syringe and when the pump has driven as far forward as possible 0.1mL will be left in the syringe. WARNING: Some manufacturers have several brand names within their ranges (e.g. Braun Omnifix and Braun Perfusor). Only use the brands named above with the T60 as failing to do so could result in an under or over infusion.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 8
2.2 Administration sets Introduction The T60™ syringe pump can be operated with any extension set with a Luer connection. However, it is recommended, to optimise system accuracy and performance, that proprietary sets from CME Ltd. are used. All CME Ltd. sets have siphon/free flow protection.
Features and characteristics Feature Materials Tubing
Slide Clamp Check valve (one-way valve)
Description The administration sets are manufactured using PVC materials that do not contain latex or Di (2-ethylhexyl) phthalate (DEHP). Micro-bore: require small priming volumes. Anti-kink: to prevent kinking or occlusion particularly in ambulatory configuration. Various lengths available. Clamps: to prevent fluid flow to patient (optional on some sets). All CME Ltd. syringe extension sets contain a combination check valve to prevent uncontrolled flow of fluid either towards or from the patient. The syringe extension set with pressure activated anti-siphon/anti-reflux valve reduces the potential for gravity flow and backflow (backtracking). The pressure-activated anti-siphon valve requires pressure to open. The pump occlusion pressure setting may require adjustment to prevent occlusion alarms. The combination check valve enhances pump functioning by: Preventing siphoning (free-flow) in the event the set is detached from the pump or mechanical malfunction Preventing reflux (back-flow) in the event several infusion pumps are connected simultaneously to the same patient.
Luer Lock end connector
The administration set is designed to be connected to Luer Lock syringes. Luer Lock syringe allows a needle to be twisted onto the tip and then locked in place. This provides a secure connection and prevents accidental removal.
Color-coded sets
Color coded sets are available as a visual aid to assist users in identifying a drug delivery route: o Clear sets may be used for any drug delivery route o Blue sets normally indicate intravenous drug delivery route o Yellow sets indicate epidural or intrathecal drug delivery route
WARNING: Ensure the administration set is NOT connected to the patient during priming. WARNING: A kinked/occluded infusion line may impair the operation and accuracy of the pump. Before operation, verify that the infusion line is not kinked or occluded. NOTE: Manufacturing data for set durability and accuracy of proprietary sets is up to 72 hours. Follow local guidelines for set use.
Advisory warnings and notes for syringes and administration sets: WARNING: Never prime or purge a set with the extension set attached to the patient. WARNING: Component damage may occur if the set is not correctly attached to the syringe. Assure all connections are secure: DO NOT over-tighten. This will help minimise leaks, disconnection and component damage. Software Version LCAT / LCATBOL
100-090SLM Rev. 03 9
WARNING: Disposables (as with any infusion) used with the syringe pump must be compatible with the drug/fluid being delivered. Check with the manufacturer of the disposables before use. Consult the fluid or drug manufacturer’s information for precautions, guidelines, and instructions for preparation and use of disposables. WARNING: Replace the syringe and/or set in accordance with local guidelines. WARNING: Use good aseptic technique when filling the syringe and priming the administration set. Patient infection may result from the use of non-sterile components. Maintain sterility of all disposable components and do not re-use singe use sets. WARNING: Syringes and administration sets should be disposed of in an appropriate manner, considering the nature of the residual fluid that may be contained within, in accordance with the hospital/homecare provider’s disposal practices.
2.3 Lockbox Uses and features Lockboxes are designed to help protect the syringe from displacement and/or tampering. Lock boxes are made from Polycarbonate due to its high impact, temperature resistance and optical properties, durability tests confirm that the overall design and construction of the T60™ syringe pump lock box ensures that they are fit for their intended purpose of protecting the T60™ syringe pump from damage caused through normal daily use and drops within the accepted normal range of one meter.
Types and sizes Lockbox for commonly used syringes up to 50/60mL o Lockboxes are available in clear polycarbonate o A clear lockbox can be used with any drug delivery route o The lockbox can be placed in a disposable (single patient use) and re-useable (washable) pouch o The pump and lockbox can be attached to a pole mount if required. Refer to your local sales representative or CME Ltd. website for information on lockbox types, codes and costs.
Using the lockbox When ready, place the pump into the lockbox +/- carry pouch: 1) Open the lockbox using the standard key that operates all T60™ lockboxes. 2) Place the pump into the lockbox so that the LCD display and keypad line up with the cut out opening. 3) Close the lockbox, guiding the administration set out of the slot at the side of the front section of the box. 4) Place the lockbox (or syringe pump if the security of the lockbox is not required for your practice) into the carry pouch and secure to the patient.
2.4 Carry Pouches Pouch use and types Disposable (single patient use) and re-useable (washable) pouches are available. Refer to your local sales representative or CME Ltd. website for information on types, codes and costs. Refer to Section 7 for re-usable pouch cleaning instructions.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 10
Use of pouches for protection of pumps during transportation WARNING: The T60™ syringe pump must always be used in a CME Ltd pouch or similar receptacle if used in direct sunlight. If the pump is exposed to direct sunlight, it may affect functionality. Using the pump with a CME Ltd. pouch or similar receptacle during transportation or patient ambulation whilst the pump is infusing protects the pump functionality and the medication in the syringe from exposure to direct sunlight. The pouch will also protect the pump from damage or syringe displacement. When using the CME Ltd re-useable (washable) pouch, it is possible to access the screen and keypad of the device during infusion by lifting the velcro flap of the pouch whilst the pump remains in the pouch. When using either a CME Ltd re-useable (washable) or disposable (single patient use) pouch it is possible to remove the forward part of the device during infusion from the pouch to inspect the syringe without removing the rear section of the device. CME Ltd pouches can be carried on the shoulder or around the waist for convenience.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 11
SECTION 3: PUMP FEATURES AND DESCRIPTION 3.1 Overview The T60™ syringe pump is a small, lightweight, robust and compact mains and battery powered infusion pump. The size and weight of the T60™ makes it ideal for ambulatory use in a hospital or alternate site / homecare setting. It is equally suitable for adults or paediatric use. The T60™ syringe pump may be used to deliver indicated medications via any common infusion routes including intravenous and subcutaneous.
Pump features and benefits Other Features Adjustable pole mount Mains powered and integral rechargeable Li-polymer back up battery Integral protective syringe cover Syringe library
Three syringe sensors
Access code protected areas Locked infusion program
User friendly and simple display screen and keypad that are easy to read and use
Comprehensive range of alerts and alarms Lockable keypad Event Log
Benefit Allows the syringe pump to be positioned according to the syringe contents being delivered. Provides added flexibility to the user Battery status can be viewed at any time Low battery alert and end battery alarm Helps to protect against syringe displacement Provides syringe recognition of commonly used syringe brands with sizes ranging from 2mL to 50/60mL. Custom brands and sizes can be added to the library. The sensors measure the syringe dimensions to identify all commonly used (or programmed) syringe brands and sizes. This feature also enables the pump to calculate the syringe volume and, depending on the mode of operation, deliver the contents over a specified duration or as mLs per hour infusion. If the syringe is removed or partially displaced during infusion an alarm will activate Code protected programs that are configured and locked into the pump memory and can only be accessed and changed by authorised personnel. Can be programmed and locked to deliver the contents over a specified duration or as mLs per hour infusion, (adjustable increments) therefore minimising the risk of user programming errors. (Refer to specifications). This makes the pump user friendly and simple to operate: Constant display for monitoring during infusion (e.g. flow rate and infusion time remaining information) provides user confidence that the infusion is continuing Key press during infusion to display volume infused, volume to be infused and battery status) Screen information and user prompts when managing or troubleshooting Call attention to situations requiring action and the display screen provides descriptive display of alerts, alarm state and resolution To prevent tampering or inadvertent key activation Date and time stamped key presses, pump events (e.g. infusion details) and self-tests. Accessible via the pump or downloaded to PC through BodyComm™ Communication Software.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 12
3.2 Pump description Top of pump: syringe fitting
No. 1. 2. 3.
Area Barrel clamp arm sensor Collar sensor Plunger sensor and actuator
4. 5.
Lead screw Guide rails
Function Detects and secures syringe barrel. Detects correct loading of the syringe collar. Plunger sensor detects and secures syringe plunger Actuator drives the plunger sensor to deliver syringe contents. Moves actuator forward and backwards. The two guide rails support the actuator position.
Front of pump: keys and display screen
No. 1. 2.
Area ON/OFF key Operation LED light
3. 4. 5. 6. 7. 8. 9.
FF (Forward) key BACK (Reverse) key UP (+) key DOWN (-) key STOP/NO key START/YES key INFO key
10. 11.
6-line LCD screen Infra-red port
Function Powers the pump on and off A green light indicates infusion in progress. A red lights indicates an alarm state or pump is in stand-by mode. Moves actuator forward or accesses purge function (if enabled). Moves actuator backward. Scrolls between options. Scrolls between options. Stops infusion or takes user back to previous screen. Confirms selections or starts infusion. Repeated presses during infusion gives infusion information when pump paused, accesses the Info Menu or activates/deactivates keypad lock. Displays pump and delivery status, programming choices and instructions. TM For connecting dongle to Bodycomm communications software.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 13
Back of pump: battery area and pump markings
No. 1. Pump label 2. 3.
Area
DC mains power lead connector Pump locator pins
Function Label includes universal symbols, device identity, manufacturer, and communicates information on using the pump safely. Power lead connection (diagram shows power lead in position). For fixing pump pins to pole mount.
Pole mounting unit
No. 1. 2. 3. 4.
Area Locator holes Pole mount locking catch Adjusting screw Adjusting screw location area
Function To fix pole mount to pump pins. To lock and release the pole mounting unit to the pump. For fixing the pole mount unit to the drip stand (vertical position). The adjusting screw can be moved here for horizontal pump positioning.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 14
Attaching the pole mounting unit to the drip stand a. Position the pole mounting unit onto the pole as shown. b. Turn the adjusting screw to fix the mount securely.
Adjusting screw
Attaching the pole mounting unit to the pump A. Pull up the locking catch (1) on the pole mounting unit and keep it held up. B. Insert the pump locator pins (2) into the pole mount locator holes (3) and position the pump into place. c. Release the locking catch (1).
3
1
3 2
2
Attaching the mains power lead to the pump a. Insert the mains power lead into the pump as shown.
Equipment safety checks It is good practice to inspect medical equipment and related accessories between patients and certainly immediately before use (e.g. whilst setting it up on the patient). One of the issues with damage to cables or internal electrical components is that visual inspection alone will not always be able to identify this type of failure.
3.3 Event log The event log shows a complete time and date stamped record of the last 512 pump events along with a record of pump status (volume infused, rate, etc) at the time of the event. Event log data cannot be deleted or altered and it is not patient specific i.e. the 512 events are likely to span multiple patients treated with that particular pump. Events recorded include hourly self-testing when an infusion is running and key presses. When the pump is infusing, the pump will record pump status every hour irrespective of any key presses. NOTE: Each event is assigned a new number and the pump stores the last 512 in memory. For event s over 512, the pump deletes the oldest event in the log each time a new event occurs. For example, after some time, the first event to appear when you enter the events history may be number 754. This means there have been 754 events in this pumps life and events 242-754 are stored in this history. The pump deletes the oldest event in the log each time a new event occurs. Software Version LCAT / LCATBOL
100-090SLM Rev. 03 15
The event log can be viewed via the pump or BodyComm™ Communications Software and a print out of events can be obtained using BodyComm™.
Event log access and navigation The event log cannot be accessed whilst an infusion is running, if necessary, stop the infusion and remove keypad lock. 1. Press the INFO key: Info Menu
Battery Level
Info Menu 2. Use +/- to scroll to “Event Log”
Select +/-, Press YES
Event Log
3. Press YES
Select +/-, Press YES
The most recent event displays: Event No: 1854 27.07.2011 16:01 Start Infusion Press INFO - Details
Line 1: Event Number Line 2: Date and time of event Line 3: Operating state Line 4: View details on this event
VI 1.03ml VTBI 14.35ml Rate 0.64ml/h 30ml BD Plastipak
VI = Volume infused VTBI = Volume to be infused Rate = mL/hour rate 30mL BD Plastipak = Syringe brand and size confirmed
Use +/- to scroll up or down events of interest or, Press INFO to display detailed information for the event
Use + or – key to display further information
Travel 2.73mm/ml Travel = Distance of actuator travel (mm) to deliver 1mL (for the syringe size and brand confirmed) 1N (0mmHg) =18mA 1N = Minimum actuator travel force (related to start up motor movement) 28N (540mmHg) =83mA 28N = Maximum actuator travel force (related to start up motor movement) Motor Current 0 mA Motor Current = motor current level in mA Occlusion 720mmHg Occlusion = pump occlusion alarm setting Battery 7.8V Battery = battery voltage
To return to the previous screen, press NO. NOTE: The information that displays when INFO is pressed will vary depending on the operational status of the pump. Some events may only record one or two parameters, other events record numerous parameters. Other event examples are: Switched ON
Syringe Removed
Anti-bolus Reverse
Keypad Lock ON
Switched OFF
Occlusion / End
Stop Infusion
Keypad Lock OFF
Volume Change
Purge
Pump Operating
System Error
NOTE: Other events may be recorded relating to technical information, refer to the technical service Manual. NOTE: The pump does not automatically change for daylight saving, the date and time can be updated manually via the pump Change Set Up Menu or via the BodyComm™ Communications Software. When INFO is pressed:
3.4 Post occlusion bolus reduction system (POBRS) What is the Post Occlusion Bolus Reduction system? During an occlusion, the pressure in the downstream section of the line and/or inside the syringe can increase above the occlusion pressure defined in pump settings. When the pump alarms the user must check the line and attempt to clear the occlusion. During an occlusion the pump’s Post Occlusion Bolus Reduction System feature will reverse the Software Version LCAT / LCATBOL
100-090SLM Rev. 03 16
operation of the motor and drive the actuator backwards otherwise the pressure build up could cause a surge of fluid into the patient’ on release of the occlusion. When the infusion is resumed, the user will notice that the volume to be infused (VTBI) increases and the volume infused (VI) decreases to indicate the pump back off feature and the infusion time remaining increases; this protects the mL/hour infusion rate. Following activation of the POBRS and if the user presses “YES to Resume” the infusion VTBI increases and the VI decreases to indicate the pump back off feature.
Occlusion pressure The occlusion pressure of a pump is the pressure in the system, registered at the pump, when the pump is still operating but cannot sustain the flow rate. The resultant build-up of pressure sets off the occlusion alarm. (Device Bulletin, 2010).
Occlusion and response An occlusion alarm can be activated by: A blockage in the delivery tubing - often inadvertently caused by kinking or leaving a roller clamp or a tap closed. A clotted-off cannula. A partially occluded cannula - if it causes the required driving pressure to rise above the occlusion alarm level. A very long or narrow bore cannula or/with extension line. Occlusion response is characterised in terms of three measurable parameters: 1. Pressure to alarm 2. Time to alarm 3. Bolus release when occlusion is resolved 1. Pressure to alarm If an occlusion occurs the pump attempts to maintain sufficient pressure on the fluid to cause it to flow through all restrictions and overcome any additional resistance. Although fluid is incompressible, the administration set and other components of the system have some ‘give’ (compliance) and the tubing can expand under the increasing pressure. Other components of the system, such as the bung of the syringe, become compressed. This expansion and compression takes some time to occur. 2. Time to alarm If the occlusion is present from the beginning of the infusion, the time to alarm will increase. The pressure in the pumping chamber increases from zero at the start of the infusion up to the alarm level. This is the most likely situation, as leaving clamps closed is the most usual cause of occlusions. If the occlusion occurs after the pump has stabilised at its set flow rate, the alarm time will not be unusually long as the pressure in the pumping chamber increases from the already high running pressure up to the alarm level. For example, a time to alarm with an infusion rate of 1mL/h is 16 minutes 7 seconds (PASA, 2006) Generally, shorter time to occlusion alarm occur with high flow rates, small syringes and good quality syringes. 3. Bolus release In the case of a complete occlusion, there is no flow to the patient whilst pressure in the system is increasing. When the occlusion is released, the build-up of fluid in the tubing can result in a bolus being delivered to the patient.
3.5 Mains power supply and internal battery Mains Power Pump operates on AC (mains) or backup battery power. Connection to wall power is recommended whenever possible.
Battery use When fully charged, batteries can provide up to 100 full deliveries depending on infusion and set up parameters. Battery life may vary depending on settings, infusion rate, and the number of times the information options/screen Software Version LCAT / LCATBOL
100-090SLM Rev. 03 17
backlight are used. Using the pump on mains power will extend the battery life. Leaving the battery in an uncharged state for long periods will shorten its life. After the End Battery signal appears, or following long periods of storage, wait two minutes after the pump has been connected to a mains power supply before operating.
Safety checks It is good practice to inspect medical equipment and accessories between patients and certainly immediately before use (e.g. whilst setting it up on the patient). One of the issues with damage to cables or internal electrical components is that visual inspection will not always be able to identify this type of failure.
Battery and charging precautions
Replace the battery once every two years. After the “End Battery” signal has been activated or following long periods of storage, wait 2 minutes after the pump has been connected to an AC power supply before operating. Whenever possible, use the pump connected to an AC power. This preserves the battery power supply for emergency use or for situations where the AC power is not available. The pump is protected against overcharging. In order to keep your battery fully charged connect the pump to mains power whenever possible. Blown fuses could cause a fire hazard. Replace blown fuses on the charger only with fuses of the same type and rating (see fuse values on the charger PCB). Battery charging is enabled as long as the charger cord is connected to the mains and the pump. Switching the pump off does not disconnect it from the mains. To disconnect from the mains, remove the charger cord from the mains.
WARNING: Unsafe operation may result from using improper accessories. Only use accessories designed for the system. Unsafe operation may result from using improper accessories. Use only accessories and options designed for this system and supplied or recommended by the T60™ syringe pump distributor.
Checking battery level a)
Checking Battery Level with no Infusion Running (LED light is red)
Load Syringe
Info Menu
Battery Level
Press INFO key:
Press YES:
Select +/-, Press YES Battery Level 80% Empty
Full
Load Syringe
Wait a few seconds for the screen to display:
b) Checking Battery Level with Infusion Running LED light is green) Time Remaining 24:00
Rate 0.50 ml/h <<< Pump Delivering 20ml BD Plastipak
Battery Level Press INFO key twice:
80% Empty
Full
Wait a few seconds for the screen to display:
Time Remaining 24:00
Rate 0.50 ml/h <<< Pump Delivering 20ml BD Plastipak
NOTE: Repeated key presses on the INFO key cycles through infusion running screen, volume history, battery level and infusion running screen.
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 18
SECTION 4: MODES OF OPERATION 4.1 Modes of operation Introduction This section provides describes the common infusion mode of operation in which the T60™ syringe pump can be configured to operate. The pump can be configured for a continuous infusion with either duration or mL/hour as the primary setting.
Duration infusion The primary setting is duration (volume over time infusion) which can be configured with a locked or changeable duration time. Once the duration time is confirmed, the pump will calculate the mL/hour rate.
Rate infusion (mL/hour) The primary setting is mL/hour rate (rate over time infusion) which can be configured with a locked or changeable mL/hour infusion rate. Once the mL/hour rate is confirmed, the pump will calculate the duration time.
Common pump configurations Four common pump configurations are: Lock On mode - fixed duration Lock Off mode - adjustable duration Rate mode (Lock ON) - fixed mL/hour rate Rate mode (Lock OFF) - adjustable mL/hour rate Each of these modes can have additional functions enabled or disabled, to suit local requirements. The pump default mode of operation is Lock On 24-hour duration. WARNING: For the correct pump configuration, mode of operation and start up sequence, you must refer to your local policy.
Lock on mode (fixed duration) The pump will deliver the syringe volume over the fixed (locked) duration. Once a syringe is detected and confirmed, the pump calculates the mL/hour infusion rate: Syringe volume Fixed duration o o
= mL/hour infusion rate
With the Program Lock On, rate change (titration) during infusion cannot be enabled. KVO and purge can be enabled if required.
NOTE: During programming, the user must check and confirm the infusion summary screen. This includes checking the syringe volume, duration, and the calculated rate matches the prescription and what is required for that infusion before it is commenced.
Lock off mode (adjustable duration) The pump will deliver the syringe volume over the confirmed default duration or the duration inputted and confirmed by the user during programming, the pump then calculates the mL/hr infusion rate: Syringe volume = mL/hour infusion rate Default or duration entered
Software Version LCAT / LCATBOL
100-090SLM Rev. 03 19
o o
With Program Lock Off, rate change (titration) during infusion can be enabled if required. There is no availability to re-program the infusion duration during delivery KVO, purge and rate titration can be enabled if required
NOTE: During programming, the user must check, change and/or confirm infusion information and programming screens. This includes checking that the program summary screen (syringe volume, duration, and the calculated rate) matches the prescription and what is required for that infusion before it is commenced. Comparison of programming screens between Lock On and Lock Off (Duration) modes Lock On (Duration) Lock off Mode (Duration)
20ml BD Plastipak
20ml BD Plastipak
Select +/-, Press YES
Select +/-, Press YES
Volume 12.0ml Duration 24:00 Rate 0.50ml/h Confirm, Press YES
Occlusion xx x mmHg Max Rate 5ml/h Program Lock OFF Battery Status xx% 20ml BD Plastipak
Volume
Start Infusion?
12 ml
Change +/-, Press YES 20ml BD Plastipak
Duration
24:00
Change +/-, Press YES 20ml BD Plastipak
Rate
0.5 ml/h
Confirm, Press YES Volume 12.0ml Duration 24:00 Rate 0.50ml/h Confirm, Press YES
Start Infusion? If the default duration is changed and/or titration is enabled additional screen prompts will display
Rate mode (lock on) - fixed mL/hour rate The pump will deliver the syringe volume over the fixed (locked) mL/hour rate. Once a syringe is detected and confirmed, the pump calculates infusion delivery duration: Syringe volume Fixed mL/hour rate o o
= Duration
With the Program Lock On, rate change (titration) during infusion cannot be enabled. KVO and purge can be enabled if required.
NOTE: During programming, the user must check and confirm the infusion summary screen. This includes checking the syringe volume, duration, and the rate matches the prescription and what is required for that infusion before it is commenced.
Rate mode (lock off) - adjustable mL/hour rate The pump will deliver the syringe volume over the mL/hour rate inputted and confirmed by the user during programming the pump calculates the infusion delivery duration: Software Version LCAT / LCATBOL
100-090SLM Rev. 03 20