CROnO 50: IDEnTIFYInG THE RESERVOIR COMPOnEnTS RESERVOIR
LUER LOCK CAP
The Crono 50 pump uses a 50 ml specific reservoir. The luer lock cap is connected to the stem of the reservoir from which it is detached by applying a little pressure. Functions of the luer lock cap:
Features of the reservoir: • Sterile. • Disposable. • Apyrogenous. • Do not use if the package is damaged or open.
• After filling the reservoir, it assists in unscrewing the stem from the piston; Stem
Reservoir body
• It assists in the correct connection between the pushing unit of the pump and the rubber piston of the syringe; • It protects the medication inside the reservoir should it not be immediately used.
Piston
Luer-Lock cap (detachable)
• For safety purposes the use of original CRN® Crono® Reservoir syringes is recommended. • The use of reservoirs of other types may damage the pump and harm its user. • CANè S.p.A. is exempt from any responsibility should the device be used with other syringes than those recommended.
CAnè S.p.A. Medical Technology Via Cuorgnè, 42/a 10098 Rivoli (TORInO) - Italy Tel. +39.011.9574872 Fax +39.011.9598880 www.canespa.it mailbox@canespa.it
MAN RS 01/USA/00 CRONO 50 07/15
Warning
1
CROnO 50: PREPARInG THE RESERVOIR 1. Refer to the Medication's Instructions for Use when selecting the appropriate device for transferring of the medication from the vial into the Crono Syringe.
3. Attach the syringe to the pump with a 90° rotation and continue the movement until the syringe clicks into its correct position. 4. Attach the butterfly needle cone onto the syringe
2. Attach the luer-lock cap and then unscrew the piston rod of the syringe counter-clock wise with a fairly swift movement;
AnTI FREE FLOW SAFETY SYSTEM
WARnInG The piston must not be removed from the syringe
Anti-Siphon Valve Performance Specification
1
2
3
Cracking pressure
3.5 PSI +/- 1
Back Pressure
> 45 PSI
Flow rate @ 5 PSI
203 ml/min
Static position
Closed
Applications
Liquids and air
The term “free flow” means possible uncontrolled administration of a medication caused by the effect of the force of gravity. To avoid this, the pushing unit is equipped with a special profile that is attached to the inside of the rubber piston of the reservoir and binds them together. The effectiveness of the safety system is guaranteed only by the correct attachment of the piston to the pushing unit as indicated in this figure:
4
nO ATTACHInG THE RESERVOIR TO THE DEVICE 1
2
3
4
YES
Attention The reservoir must be completely filled to allow the connection of the rubber piston to the pushing unit. The rubber piston of the reservoir must be connected to the pushing unit of the pump; if this does not occur the following conditions may happen: • The infusion incurs a delay and is imprecise. • The connection between the pushing unit and the piston is incorrect and therefore the safety of the anti free flow system is not guaranteed.