CANÈ S.p.A. Medical Technology

CRONO CRN PUMP Syringe Preparation Guide Sept 2006

Guide

1 Page

02/I/00 CRN SYRINGE PREPARATION ISCRN Date of issue: 05/09/2006 FOR CRONO PUMP  1 - Before syringe filling-up  Begin by unscrewing the piston rod half a turn and then back again before moving the plunger. This way it will be easier to remove the piston rod after the syringe has been filled.  Slide the plunger twice along the syringe to spread the lubricant along the walls.  2 - Syringe filling-up Draw up the drug slowly. The syringe is not to be filled up beyond the maximum volume allowed of 10 or 20 ml. The piston rod is to be unscrewed with a fairly swift movement.  3 - Connecting the syringe onto the pump Connect the infusion set (or the luer cap) to avoid any possible drug leakage during syringe connection onto the pump.  Hold the syringe from the syringe body upper top during syringe connection onto the pump and not along the syringe body wall.  During connection avoid applying pressure onto the syringe body walls since this could cause drug leakage from piston rings.  During syringe filling up and connection onto the pump a light drug leakage may occur between the first and second rubber piston retaining ring: such condition does not affect either the correct syringe working or drug delivery. Canè S.r.l. Medical Technology Via Pavia 105/i 10090 Rivoli Cascine Vica (TO) Italy Tel.+39 011 9574872 - Fax +39 011 9598880 www.microjet.it - mailbox@microjet.it