CRONO PCA 50 Short Instructions July 2005
Short Instructions
2 Pages
Preview
Page 1
BUTTONS
SETTINGS
Operations feasible only with key-pad unlocked.
• Setting flow (from 0.05 to 35 ml/h with the following 1st depressing
SCREEN 10 20
10 20
increments:- from 0.05 to 10 with 0.02 ml increments; - from 10 to 35 with 0.05 ml increments).
• Setting bolus dose (from 0.10 to 9.98 ml with 0.02 ml 2nd depressing
BUTTONS
20
increments).
SCREEN
PUMP ACTIVATION
• The screen will show all symbols.
PUMP OFF
• Setting bolus time limitations (- from 5’ to 1 h with
INSERTING BATTERY PHASE
10
10 20
• Piston self-adjust.
3rd depressing
4th depressing
20
increments of 5’ - from 1 h to 24 h with increments of 15’).
• Setting boluses number limitations (0-12 per h).
• Pump switches OFF.
10
10 20
Bolus lock-out possibilities: combination of bolus time and bolus number limitations. The computer always chooses the programming that results in the smallest bolus volume.
BUTTONS and
REVERSING OF THE PUMP PISTON NUMBER OF INFUSIONS DELIVERED
SCREEN
• Setting partial volume (from 20 to 50 ml with 5th depressing
• Reversing of the pump piston to the start position.
10 20
increments of 1 ml - possible only at the begin of a new infusion).
depress simultaneously
/
• Reading the number of infusions delivered.
• Decrease/Increase above parameters.
depressed for 8 seconds
and
• Reset the number of bolus delivered (L0).
BUTTONS
depress simultaneously
PUMP OFF
Confirm the reset of bolus delivered.
20
• Access to PRIME phase.
button
PUMP ON
•Pump switches ON.
10
10
20
20
PRIME CONDITION
• Press for 7 seconds.
press without releasing
SCREEN
Priming possible: only at strat of a new infusion or start of a new partial infusion.
10
KEY-PAD LOCK-OUT
• Key-pad lock level selection.
PRIME
• PRIME (each depressing administers 0.5 ml - max 1.5 ml).
• PRIME phase finish, infusion start. and
• PRIME interruption (if in execution).
depress simultaneously
and depress simultaneously
• Device switch OFF (if PRIME carried out). In case there is a need to prime one or more times during an already started infusion, the clinician bolus can be used
10 20
10 20
10 20
10 20
BUTTONS
INFUSIONS
PUMP ON
• Bolus dose delivery.
SCREEN
BUTTONS
• Press for 7 seconds.
10 20
press without releasing button
and
2nd depressing
• After the screen will show the latest administered clinician bolus dose. SCREEN REVIEW OPTIONS
• Administered flow rate in ml since start of infusion. • Administered patient bolus in ml since start of infusion.
ON
• Administered number of bolus.
SCREEN
PUMP OFF
1st depressing
10 20
/
20
and
10
depress simultaneously
20
1st depressing
2nd depressing
SCREEN REVIEW OPTIONS
• Programmed flow rate. • Programmed bolus dose.
10 20
10 20
20
10 20
10 20
10 20
10 20
10 20
10 20
After the administration of one cllinician bolus, the lock level L 2c switches automatically back to the lock level programmed when the clinician bolus procedure started i. e. either L 0 or L 1 and no more clinician bolus can be delivered without repeating the clinician bolus procedure from the beginning. PUMP OFF
SCREEN
• Pump switches OFF.
depress simultaneously
BUTTONS
10
3rd depressing
• Programmed bolus number limit.
and
20
• Clinician bolus function interruption.
BUTTONS
SCREEN
END OF INFUSION
SCREEN
• End of infusion. and depress simultaneously
• Piston automatic reversing. • Automatic switch OFF.
4th depressing
• Programmed partial volume.
10 20
5th depressing
• Programmed clinician bolus dose. 6th depressing
10 20
AUG 01/USA/00 PCA50 07/05
ON
• Programmed bolus time limit.
20
PUMP ON
BUTTONS
• Total administered volume in ml since start of infusion.
10
END INFUSION
5th depressing
• Administered clinician bolus in ml since start of infusion.
10
20
ml with 0.02 ml increments).
• Clinician bolus delivery.
10
10
• Decrease/Increase clinician bolus (from 0.10 to 9.98 • After the screen will show that the clinician bolus is available.
3rd depressing
4th depressing
• Key-pad lock level selection. • Select clinician bolus level.
• Bolus interruption (if in execution).
depress simultaneously
BUTTONS
SCREEN
CLINICIAN BOLUS FUNCTION