Cane Crono PCA Abbreviated Programming Guide Caution: These notes apply only to: Pumps labelled only with the model designation CRONO PCA.  Pumps produced after June 2004, with serial numbers greater than 1000.04 Caution: Users must familiarise themselves will full user instructions. These notes provide only a reminder of main functions. Pump Activation and Changing the Battery This is conducted when it is necessary to modify syringe size selection, or when replacing a depleted battery. Remove and replace battery. Screen will momentarily flash all symbols. Syringe size (10 or 20) will flash on screen. To change between values just press the (+) button for a few seconds until required value is displayed. Pump will re-set and [OFF] is displayed. Note: Partial Syringe Volume is reset to Full Syringe during this procedure. Unlocking the Keypad Prior to making any program changes the keyboard must be unlocked. With the pump displaying [OFF] press the (-) key for approximately 7 seconds. A number representing the number of completed infusions will appear on the screen. While still holding the (-) key depress the (+) key at the same time and the screen will display [Lo.0] and keyboard is unlocked. This will remain unlocked for 2 minutes following the last key press.