59 Pages
Preview
Page 1
2B702-067EN*T
Guidelines for cleaning, disinfection, and sterilization of transducers Issued: 2021-04 This manual describes the cleaning, disinfection, and sterilization procedures for the ultrasound transducer. For the operating precautions and procedures for the transducers, refer to the operation manual for each transducer.
Trademarks This manual may include trademarks or registered trademarks of companies other than Canon Medical Systems. 2B702-067EN*T
1
1. Safety Precautions 1.1
Meaning of Signal Words In this manual, the signal words DANGER, WARNING, CAUTION, and are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual. Signal word
Meaning
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. Indicates a potentially hazardous situation which, if not avoided, may result in property damage.
1.2
Safety Precautions Observe the following precautions to ensure the safety of patients as well as operators when performing cleaning, disinfection, or sterilization of the transducer.
WARNING
CAUTION
Never immerse the transducer connector or any other nonwaterproof sections into liquids such as water or cleaning solution. Immersion may cause electric shock. Refer to the tables and figures in section 3 for the immersible range of each transducer model. 1. Precautions concerning cleaning, disinfection, and sterilization. 1) Observe the following precautions to prevent infection. • Wear protective gloves when performing cleaning.* • Wear sterile protective gloves when performing disinfection or sterilization.* • Wear new protective gloves each time cleaning, disinfection, and sterilization are performed. • Clean the transducer before and after examination. Disinfect or sterilize the transducer as required. • Note that local regulations may require that the transducer be disinfected before sterilization. • Sterilize the transducer and biopsy adaptor before and after an ultrasound-guided biopsy procedure is performed. Failure to do so may result in the transducer and biopsy adaptor becoming sources of infection. * Refer to the FDA's March 29, 1991 Medical Alert on Latex Products.
2
2B702-067EN*T
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
CAUTION
2) After cleaning, rinse the transducer thoroughly with purified water to remove all chemical residues. After disinfection, rinse the transducer thoroughly with sterile or deionized water to remove all chemical residues. Chemical residues on the transducer may be harmful to the human body. 3) After chemical cleaning or chemical disinfection, thoroughly dry the transducer surface. 4) After sterilization, degas the transducer. Gas residues on the transducer may be harmful to the human body. 5) The efficacy of the cleaning solutions, disinfectants, and sterilizing gases is not guaranteed by Canon Medical Systems. Contact the manufacturers for information on the activity of the products. 6) To ensure the prevention of infection, confirm the effectiveness of each chemical for cleaning, disinfection, or sterilization based on the criteria (such as effective period, number of times of use, discoloration, and results of using the effectiveness test kit) described in the documentation provided by the relevant manufacturer. 7) Do not clean, disinfect, or sterilize the transducer using chemicals or methods other than those specified in this guide. If chemicals or methods other than those specified in this guide are used, the transducer may not be properly cleaned, disinfected, or sterilized or may be damaged. 8) Observe the following precautions to prevent transducer malfunction. • Only soft materials such as soft cloth or soft gauze should be used when removing ultrasound gel from the transducer or wiping water or disinfectant from the transducer surface after cleaning or disinfection. Use of hard or abrasive cloth or gauze may damage the transducer. • The transducer must not be immersed in a chemical solution for more than three hours. • Do not permit the transducer to become overheated (more than 60°C (140°F)) during cleaning, disinfection, and sterilization. • The cleaning, disinfection, or sterilization conditions, such as the temperature and pressure, differ depending on the product. In addition, some products cannot be subjected to disinfection or sterilization procedures. Confirm the detailed conditions by referring to the tables and figures in section 3.
NOTE
2B702-067EN*T
The mouthpiece supplied with the TEE transducer can be high-pressure steam sterilized (temperature: 134°C (273.2°F), holding time: 18 min). Confirm that "AUTOCLAVABLE" is indicated on the mouthpiece.
3
1.3
Chemical Hazard (for USA only) Observe the following instruction in order to protect patients and operators from inflammation or poisoning by chemical substances.
WARNING:
1.4
This product can expose you to chemicals including lead and lead compounds, which are known to the State of California to cause cancer, and phthalates, which are known to the State of California to cause birth defects or other reproductive harm. For more information go to www.p65warnings.ca.gov.
Precautions Concerning Restriction of Sale or Use (for USA only) United States law restricts this device to sale or use by, or on the order of a physician.
4
2B702-067EN*T
This section describes the methods for cleaning, disinfection, and sterilization. Cleaning and disinfection between patients must be performed to prevent transmission of disease. All transducers must be thoroughly cleaned prior to disinfection. The level of disinfection required is based on patient contact.
2.1
Cleaning <<Items to be used: Protective gloves, cleaning solution or cleaning wipes, purified water, clean soft cloth or gauze, single-use sponge*>> * The single-use sponge must not include any abrasive parts or contain any abrasive cleanser. (1) Wear protective gloves to prevent infection. Wear new protective gloves each time cleaning is performed. (2) If an accessory that can be disassembled (e.g., biopsy adaptor) is provided, disassemble it. For details concerning such parts, refer to the operation manual for the transducer. (3) Wash off all organic materials (such as blood or other bodily fluids) from the transducer under purified water. A single-use sponge can be used for washing. Do not use a brush, because it may damage the transducer. (4) In accordance with the tables and figures in section 3, immerse the transducer in a cleaning solution or wipe the transducer using wipes to dissolve or remove all remaining organic materials. Use a single-use sponge if necessary. If dried organic materials are present on the transducer, immerse it in the cleaning solution for a prolonged period. (5) Remove all residual organic materials and cleaning solution from the transducer by rinsing it under purified water. Repeat steps (3) and (4) to visually confirm that all residual organic materials and cleaning solution have been completely removed from the transducer. Do not reuse the purified water. (6) Dry the surface of the transducer using clean soft cloth or gauze. Do not use heat to dry the transducer. Handle the cleaning solution or wipes as described in the documentation provided by the relevant manufacturer. To maintain the effectiveness of the cleaning solution or wipes, ensure that the concentration, temperature, and other conditions specified in the documentation provided by the manufacturer are met. To confirm the effectiveness of the cleaning solution or wipes, use the criteria (such as effective period, number of times of use, discoloration, and results of using the effectiveness test kit) described in the documentation provided by the manufacturer. (7) Confirm that the transducer shows no signs of damage, deformation, or peeling.
2B702-067EN*T
5
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
2. Cleaning, Disinfection, Sterilization
2.2
Disinfection Before disinfection, the transducer must be cleaned. Note that local regulations may require that the transducer be disinfected before sterilization. <<Items to be used: Sterile protective gloves, disinfectant, sterile water or deionized water, sterile soft cloth or gauze>> (1) Wear sterile protective gloves to prevent infection. Wear new sterile protective gloves each time disinfection is performed. (2) Disinfect the transducer using the chemicals listed in section 3 “List of Chemicals”. Ensure that the entire immersible range of the transducer is completely in contact with the disinfectant. (3) Rinse the transducer thoroughly with sterile or deionized water, as described in the documentation provided by the manufacturer, in order to remove all residual disinfectant. (Rinsing is not necessary when gas disinfection is performed using Trophon EPR or Trophon 2.) Do not reuse the sterile or deionized water. (4) Dry the surface of the transducer using sterile soft cloth or gauze. Do not use heat to dry the transducer. Handle the disinfectant as described in the documentation provided by the relevant manufacturer. To maintain the effectiveness of the disinfectant, ensure that the concentration, temperature, and other conditions specified in the documentation provided by the manufacturer are met. To confirm the effectiveness of the disinfectant, use the criteria (such as effective period, number of times of use, discoloration, and results of using the effectiveness test kit) described in the documentation provided by the manufacturer. (5) Confirm that the transducer shows no signs of damage, deformation, or peeling.
2.3
Sterilization Before sterilization, the transducer must be cleaned. Note that local regulations may require that the transducer be disinfected before sterilization. <<Items to be used: Sterile protective gloves, sterilant>> * Some types of transducers cannot be sterilized or the sterilization conditions may differ. (1) Wear sterile protective gloves to prevent infection. Wear new sterile protective gloves each time sterilization is performed. (2) Sterilize the transducer using the chemicals listed in section 3 “List of Chemicals”. * Place the transducer in a sterilization packing case and then place it in the sterilizer. (3) After gas sterilization, perform aeration to remove all gas residues on the transducer surface. (4) Confirm that the transducer shows no signs of damage, deformation, or peeling.
6
2B702-067EN*T
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
3. List of Chemicals This section lists the chemicals used when washing, disinfecting, or sterilizing the transducer. Refer to the lists for the available chemicals. In addition, refer to the manuals for the chemicals for the handling details. NOTE
1. When multiple model names included in the List of Chemicals are referred to, note that they are abbreviated. For example, "PLT-604AT and PLT-704AT" is abbreviated as "PLT-604AT/704AT". 2. For the transducers supported by each diagnostic ultrasound system, refer to the operation manual supplied with the diagnostic ultrasound system.
● Type A (PC-20M etc.)
● Type B (PET-508MA etc.) Immersible Immersible range range
Immersible Immersible range range
● Type C (PET-511BTM etc.)
● Type D (PLT-704AT etc.)
Immersible range
Immersible Immersible range range
Never immerse the non-waterproof sections of the transducer into liquids such as water or cleaning solution. Immersion may cause electric shock.
2B702-067EN*T
7
Cleaning (1/2) Country of origin :
Chemical name/type
Enzyme
Any : Any country FRA : France DEU : Germany GBR : United Kingdom USA : United States JPN : Japan AUS : Australia CAN : Canada
Model name PC-20M PC-50M PET-508MA PET-512MA/512MC PET-511BTM*1
Trade name
CIDEZYME®
3M™ Rapid Multi-Enzyme cleaner
Manufacturer
J&J
3M Company
Country of origin Concentration (Dilution ratio) Time Temperature Humidity Pressure Immersible range Type A Type A Type B Type B Type C
Any Any 0.8% 1% (125) (100) 1 min. 10 min. Room temperature Normal humidity Normal pressure
ANIOSYME DD1 *2
neodisher® mediclean *2
Laboratories DR. WEIGERT ANIOS GmbH & Co. KG FRA DEU 0.5% 1% (200) (100) 5 min. 10 min. Room temperature Normal humidity Normal pressure
OK
OK
OK
OK
OK OK OK OK
OK OK OK OK
OK – – OK
OK – – –
PET-512MD*1, PEI-512VX*1
Type C
OK
OK*2
OK
–
PET-609MA*1
Type C Type C
OK OK
OK*2 –
OK –
– –
Type D
OK
OK
OK
OK
Type D Type D Type D Type D Type D Type D Type D
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
Type D
OK
OK
OK
OK
Type D Type D
OK OK
OK OK
OK OK
OK OK
PET-805LA PLT-604AT/704AT/805AT/1202S/1204BT PVT-375BT/375SC/382BT/661VT/674BT PST-25BT/65AT PVT-745BTF/745BTH PLT-308P/308BTP, PVT-770RT PLT-705BTF/705BTH PVT-745BTV PVT-350BTP PLT-1204BX PLT-704SBT/705BT/1005BT PVT-781VT/781VTE PLT-704ST PLT-1204ST, PVT-375ST
OK : Use of the chemical is permitted.
–
: Use of the chemical is not permitted.
*1: If a waterproof cover is placed over the transducer connector, the entire transducer can be immersed in water or cleaning solution. *2: Not for use in the USA.
8
2B702-067EN*T
Tristel Endozime® ® Pre-Clean MetriZyme™ Klenzyme® Instru-Zyme Xtreme Gel Wipes *2 Power *2 Tristel Solutions Metrex STERIS Summit RUHOF Limited Research Corporation Medical Corporation GBR USA USA USA USA Working 0.77% 0.8% Working 0.2% solution (130) (125) solution (500) Wiping only 5 min. 5 min. Wiping only 2 min. Room temperature Room temperature Normal humidity Normal humidity Normal pressure Normal pressure
Alkyldimethylbenzylammonium chloride/2-ethanol/ Alkyl polysaccharide/ Ethylenediaminetetraacetic acid
CaviWipes™
Intercept® Wipes
Metrex Research USA
MEDIVATORS Inc. USA
Working solution
Working solution
Wiping only Room temperature Normal humidity Normal pressure
OK
OK
OK
OK
OK
OK
OK
OK OK OK
OK OK OK
OK – –
OK – –
OK OK OK
– – –
– – –
OK
OK
OK
–
OK
–
–
OK
OK*2
–
–
OK
–
–
OK
OK*2
–
–
OK
–
–
–
–
–
–
OK
–
–
OK
OK
OK
OK
OK
OK
OK
OK OK OK OK OK – OK
OK OK OK OK OK OK OK
OK OK OK OK OK – OK
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
OK OK OK OK OK – OK
OK OK OK OK OK – OK
OK
OK
OK
OK
OK
OK
OK
OK OK
OK OK
OK OK
OK OK
OK OK
OK OK
OK OK
2B702-067EN*T
9
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
Enzyme
Isopropyl alcohol/ Ethylene glycol monobutyl ether
Cleaning (2/2) Country of origin :
Chemical name/type
Enzyme
Any : Any country FRA : France DEU : Germany GBR : United Kingdom USA : United States JPN : Japan AUS : Australia CAN : Canada
Model name PST-25SX PST-25ST/50BT, PVT-712BT PST-30BT PLT-1204MV PVT-375MV/382MV/675MV/675MVL PVT-681MV PVT-681MVL PVL-715RST PVI-475BX/475BT/574BX/482BX/450BXP PVT-475BT/482BT/574BT PLI-1205BX/2004BX/705BX/605BX PSI-30BX, PST-28BT/65BT PSI-70BT PSI-30VX/40VX PSI-50VX PLI-2002BT PLI-3003BX PLT-1202BT PVT-675MVS
OK : Use of the chemical is permitted.
Trade name
CIDEZYME®
3M™ Rapid Multi-Enzyme cleaner
Manufacturer
J&J
3M Company
Country of origin Concentration (Dilution ratio) Time Temperature Humidity Pressure Immersible range Type D Type D Type D
Any Any 0.8% 1% (125) (100) 1 min. 10 min. Room temperature Normal humidity Normal pressure
ANIOSYME DD1 *2
neodisher® mediclean *2
Laboratories DR. WEIGERT ANIOS GmbH & Co. KG FRA DEU 0.5% 1% (200) (100) 5 min. 10 min. Room temperature Normal humidity Normal pressure
OK
OK
–
OK
OK OK
OK OK
OK OK
OK OK
Type D
OK
OK
OK
OK
Type D Type D Type D
OK OK OK
– OK OK
– OK OK
– OK OK
Type D
OK
OK
OK
OK
Type D Type D Type D Type D Type D Type D Type D
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
–
: Use of the chemical is not permitted.
*1: If a waterproof cover is placed over the transducer connector, the entire transducer can be immersed in water or cleaning solution. *2: Not for use in the USA.
10
2B702-067EN*T
Tristel Endozime® ® Pre-Clean MetriZyme™ Klenzyme® Instru-Zyme Xtreme Gel *2 Wipes Power *2 Tristel Solutions Metrex STERIS Summit RUHOF Limited Research Corporation Medical Corporation GBR USA USA USA USA Working 0.77% 0.8% Working 0.2% solution (130) (125) solution (500) Wiping only 5 min. 5 min. Wiping only 2 min. Room temperature Room temperature Normal humidity Normal humidity Normal pressure Normal pressure
Alkyldimethylbenzylammonium chloride/2-ethanol/ Alkyl polysaccharide/ Ethylenediaminetetraacetic acid
CaviWipes™
Intercept® Wipes
Metrex Research USA
MEDIVATORS Inc. USA
Working solution
Working solution
Wiping only Room temperature Normal humidity Normal pressure
–
OK
–
OK
OK
–
–
OK OK
OK OK
OK OK
OK OK
OK OK
OK OK
OK OK
–
OK
–
OK
OK
OK
OK
– OK OK
– OK OK
– – OK
– OK OK
OK OK OK
OK OK OK
OK OK OK
OK
OK
OK
–
OK
–
–
OK OK OK OK – OK OK
OK OK OK OK OK OK OK
OK OK OK OK OK OK OK
OK – – – – – –
OK OK OK OK OK OK OK
OK OK OK OK – OK OK
OK – – – – – –
2B702-067EN*T
11
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
Enzyme
Isopropyl alcohol/ Ethylene glycol monobutyl ether
Low/Middle-Level Disinfection -1 (1/2) (Except for USA) Note that local regulations may require that the transducer be disinfected before sterilization. Country of origin : Any : Any country
Chemical name/type
FRA : France DEU : Germany GBR : United Kingdom USA : United States JPN : Japan AUS : Australia CAN : Canada
Model name PC-20M PC-50M PET-508MA PET-512MA/512MC
Trade name Manufacturer Country of origin Concentration (Dilution ratio) Time Temperature Humidity Pressure Immersible range Type A Type A Type B Type B
Didecyldimethylammonium chlorides/ Quaternary ammonium compounds CLEANISEPT® WIPES Dr. Schumacher GmbH DEU
Didecyldimethylammonium chlorides
Glyoxal/Glutaral/ Didecyldimethylammonium chlorides
WIP'ANIOS Excel
Lysoformin® 3000
Laboratories ANIOS FRA
Working solution
Working solution
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Lysoform Disinfection Ltd. DEU 2% (50) 15 min. Room temperature Normal humidity Normal pressure
OK
OK
OK
OK – –
OK – –
– – –
PET-511BTM*1
Type C
–
–
OK
PET-512MD*1, PEI-512VX*1
Type C
–
–
–
PET-609MA*1
Type C Type C
– –
– –
– –
PET-805LA PLT-604AT/704AT/805AT/1204BT PVT-375BT/375SC/382BT/674BT PST-25BT/65AT PVT-745BTF/745BTH PLT-308P/308BTP PVT-770RT PLT-705BTF/705BTH
Type D
OK
OK
OK
Type D Type D Type D Type D Type D
OK – OK – OK*2
OK OK OK OK OK
OK – OK OK OK
PVT-745BTV
Type D
OK*2
OK
OK
PLT-1202S PVT-350BTP PLT-1204BX PLT-704SBT/705BT/1005BT PVT-781VT/781VTE PVT-661VT PLT-704ST PLT-1204ST, PVT-375ST
Type D Type D Type D Type D Type D Type D Type D Type D
OK*2 – OK OK – – OK OK
OK – OK OK OK – – –
OK OK – OK OK OK OK OK
OK : Use of the chemical is permitted.
–
: Use of the chemical is not permitted.
*1: If a waterproof cover is placed over the transducer connector, the entire transducer can be immersed in water or cleaning solution. *2: Only if the transducer is not used internally.
12
2B702-067EN*T
Propan-2-ol Ethanol
Didecyldimethylammonium chloride/Alkyl (C12-16) dimethylbenzylammonium chloride/Alkyl (C12-C14) ethylbenzylammonium chloride
Propan-1-ol Ethanol
GBR
mikrozid® universal wipes premium Schulke & Mayr GmbH DEU
Working solution
Working solution
Working solution
Working solution
Working solution
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
–
–
–
–
–
– – –
– – –
– – –
– – –
– – –
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
OK
OK
OK
OK
OK
OK – – – –
OK – – – –
OK OK OK – –
OK OK OK – –
OK OK OK – –
–
–
–
–
–
–
–
OK
OK
OK
– OK OK OK – – –
– OK OK OK – – –
– OK OK OK – – –
– OK OK OK – – –
– OK OK OK – – –
Clinell® Universal Wipes GAMA Healthcare Ltd.
2B702-067EN*T
mikrozid® sensitive wipes
mikrozid® AF liquid
mikrozid® AF wipes
Schulke & Mayr GmbH DEU
Schulke & Mayr GmbH DEU
Schulke & Mayr GmbH DEU
13
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
Benzalkonium chloride/ Didecyldimethylammonium chloride/ Polyhexamethylene biguanide (PHMB)
Low/Middle-Level Disinfection -1 (2/2) (Except for USA) Note that local regulations may require that the transducer be disinfected before sterilization. Country of origin : Any : Any country
Chemical name/type
FRA : France DEU : Germany GBR : United Kingdom USA : United States JPN : Japan AUS : Australia CAN : Canada
Model name PST-25SX PST-50BT, PVT-712BT PST-25ST PST-30BT PLT-1204MV PVT-375MV/382MV/675MV/675MVL PVT-681MVL PVT-681MV PVL-715RST PVI-475BX/574BX/482BX PVT-574BT, PLI-705BX PSI-30BX, PST-28BT/65BT PLI-1205BX/2004BX/605BX PVI-475BT, PVT-475BT/482BT PSI-70BT PSI-30VX/40VX PSI-50VX PLI-2002BT PVI-450BXP PLI-3003BX PLT-1202BT PVT-675MVS
OK : Use of the chemical is permitted.
Trade name
Didecyldimethylammonium chlorides/ Quaternary ammonium compounds CLEANISEPT® WIPES Dr. Schumacher GmbH DEU
Didecyldimethylammonium chlorides
Glyoxal/Glutaral/ Didecyldimethylammonium chlorides
WIP'ANIOS Excel
Lysoformin® 3000
Laboratories ANIOS FRA
Working solution
Working solution
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Lysoform Disinfection Ltd. DEU 2% (50) 15 min. Room temperature Normal humidity Normal pressure
–
–
–
OK OK OK
OK – OK
OK OK OK
Type D
OK
OK
OK
Type D Type D Type D
– – –
OK OK OK
OK – –
Type D
OK
–
–
Type D
OK
–
–
Type D Type D Type D Type D Type D Type D Type D Type D
OK OK OK OK OK OK OK OK
OK – – – – – OK OK
OK – – – – OK OK –
Manufacturer Country of origin Concentration (Dilution ratio) Time Temperature Humidity Pressure Immersible range Type D Type D Type D Type D
–
: Use of the chemical is not permitted.
*1: If a waterproof cover is placed over the transducer connector, the entire transducer can be immersed in water or cleaning solution. *2: Not for use in the USA.
14
2B702-067EN*T
Propan-2-ol Ethanol
Didecyldimethylammonium chloride/Alkyl (C12-16) dimethylbenzylammonium chloride/Alkyl (C12-C14) ethylbenzylammonium chloride
Propan-1-ol Ethanol
GBR
mikrozid® universal wipes premium Schulke & Mayr GmbH DEU
Working solution
Working solution
Working solution
Working solution
Working solution
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
–
–
–
–
–
OK – OK
OK – OK
OK – OK
OK – OK
OK – OK
–
–
–
–
–
OK – OK
OK – OK
OK – OK
OK – OK
OK – OK
OK
OK
OK
OK
OK
OK
–
OK
–
–
OK OK OK – – – – OK
OK OK OK – – – – –
OK OK OK OK OK – OK –
OK OK OK – OK – OK –
OK OK OK – OK – OK –
Clinell® Universal Wipes GAMA Healthcare Ltd.
2B702-067EN*T
mikrozid® sensitive wipes
mikrozid® AF liquid
mikrozid® AF wipes
Schulke & Mayr GmbH DEU
Schulke & Mayr GmbH DEU
Schulke & Mayr GmbH DEU
15
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
Benzalkonium chloride/ Didecyldimethylammonium chloride/ Polyhexamethylene biguanide (PHMB)
Low/Middle-Level Disinfection -2 (1/2) Note that local regulations may require that the transducer be disinfected before sterilization.
Country of origin :
Chemical name/type
Any : Any country FRA : France
Ethanol
Isopropyl alcohol
Isopropyl alcohol/ Ethylene glycol monobutyl ether
Sodium hypochlorite
DEU : Germany GBR : United Kingdom USA : United States
Trade name
Ethanol
Isopropyl alcohol
CaviWipes™
Dispatch® Hospital Cleaner Disinfectant Towels with Bleach
Manufacturer
Any manufacturer
Any manufacturer
Metrex Research
The Clorox Company
Country of origin
Any
Any
USA
USA
Concentration (Dilution ratio)
80% (original solution)
70% (original solution)
Working solution
0.65% (working solution)
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature
Wiping only Room temperature
Normal humidity
Normal humidity
Normal pressure
Normal pressure
–
OK
OK
OK
OK OK
OK – –
– – –
– – –
–
–
–
JPN : Japan AUS : Australia CAN : Canada
Time Temperature Humidity Model name PC-20M PC-50M PET-508MA PET-512MA/512MC
Pressure Immersible range Type A Type A Type B Type B
PET-511BTM*1
Type C
OK*2 OK
PET-512MD*1, PEI-512VX*1
Type C
OK
OK
–
–
PET-609MA*1
Type C Type C
OK –
– –
– –
– –
Type D
OK
OK
OK
OK
Type D Type D Type D Type D Type D Type D Type D Type D
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK – OK
OK OK OK OK – – – OK
Type D
OK
OK
OK
OK
Type D Type D
OK OK
OK OK
OK OK
OK OK
PET-805LA PLT-604AT/704AT/805AT/1202S/1204BT PVT-375BT/375SC/382BT/674BT PST-25BT/65AT PVT-745BTF/745BTH PLT-308P/308BTP PVT-770RT PLT-705BTF/705BTH PVT-745BTV PVT-350BTP PLT-1204BX PLT-704SBT/705BT/1005BT PVT-781VT/781VTE PLT-704ST, PVT-661VT PLT-1204ST, PVT-375ST
OK : Use of the chemical is permitted.
–
: Use of the chemical is not permitted.
*1: If a waterproof cover is placed over the transducer connector, the entire transducer can be immersed in water or cleaning solution. *2: Not for use in the USA.
16
2B702-067EN*T
Quaternary ammonium chlorides/ Isopropyl alcohol
Hydrogen peroxide
Hydrogen peroxide
Isopropanol/ Ethanol/ Ethylene glycol monobutyl ether (2Butoxyethanol)/ Didecyldimethylammonium chloride/ Water
Octyl decyl dimethyl ammonium chloride
Sani-Cloth® Trophon® SONO ® Super AF3, Oxivir 1 PREempt Sani-Cloth CaviWipes1™ Companion Ultrasound wipes Wipes Sani-Cloth® Plus Cleaning Sani-Cloth® Wipes Wipes AF Professional Parker Advanced Virox Diversey, Metrex Nanosonics Disposables Professional Disposables Laboratories, Ultrasound Technologies International, Inc. Inc. Research Limited International, Inc. Solutions Inc. Inc. Inc. USA USA USA USA USA CAN USA AUS 0.25% / 0.232% 0.28% 0.5% / 55% 0.5% 0.5% 14.85% Working Working Working (working (working (working (working (working (working solution solution solution solution) solution) solution) solution) solution) solution) Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Room Room Room Room Room Room Room Room Room temperature temperature temperature temperature temperature temperature temperature temperature temperature Normal Normal Normal Normal Normal Normal Normal Normal Normal humidity humidity humidity humidity humidity humidity humidity humidity humidity Normal Normal Normal Normal Normal Normal Normal Normal Normal pressure pressure pressure pressure pressure pressure pressure pressure pressure Protex™ Ultra Disinfectant Wipes
OK
OK
OK
OK
–
OK
OK
–
–
– –
OK OK
OK –
OK –
– –
– –
– –
– –
– –
–
OK
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
OK
–
–
–
–
–
–
–
–
OK
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK OK OK OK – – – OK
OK OK OK OK OK OK – OK
OK OK OK OK OK OK – OK
OK OK OK OK OK OK – OK
OK OK OK – – – – OK
OK OK OK OK – – – OK
OK OK OK OK – – – OK
OK OK OK – – – – OK
OK OK OK – – – – OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK OK
OK OK
OK OK
OK OK
– –
OK OK
– –
– –
– –
2B702-067EN*T
17
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
Quaternary ammonium chlorides
Decyldimethyloctylammonium chloride/ Dimethyldioctylammonium chloride/ Didecyldimethylammonium chloride/ Quaternary ammonium compounds, benzyl-C12-16alkyldimethyl, chlorides
Low/Middle-Level Disinfection -2 (2/2) Note that local regulations may require that the transducer be disinfected before sterilization.
Country of origin :
Isopropyl alcohol/ Ethylene glycol monobutyl ether
Sodium hypochlorite
Ethanol
Isopropyl alcohol
CaviWipes™
Dispatch® Hospital Cleaner Disinfectant Towels with Bleach
Manufacturer
Any manufacturer
Any manufacturer
Metrex Research
The Clorox Company
Country of origin
Any
Any
USA
USA
Concentration (Dilution ratio)
80% (original solution)
70% (original solution)
Working solution
0.65% (working solution)
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature Normal humidity Normal pressure
Wiping only Room temperature
Wiping only Room temperature
Normal humidity
Normal humidity
Normal pressure
Normal pressure
Chemical name/type
Any : Any country FRA : France
Ethanol
Isopropyl alcohol
Trade name
DEU : Germany GBR : United Kingdom USA : United States JPN : Japan AUS : Australia CAN : Canada
Time Temperature Humidity Model name PST-25SX PST-50BT, PVT-712BT PST-25ST PST-30BT PLT-1204MV, PVT-375MV/382MV PVT-675MV/675MVL PVT-681MVL PVT-681MV PVL-715RST PVI-475BX/475BT/574BX/482BX PVT-475BT/482BT/574BT PLI-1205BX/2004BX/705BX/605BX PSI-30BX, PST-28BT/65BT PVI-450BXP PSI-70BT PSI-30VX/40VX PSI-50VX PLI-2002BT PLI-3003BX PLT-1202BT PVT-675MVS
Pressure Immersible range Type D Type D Type D Type D
OK
OK
–
–
OK OK OK
OK OK OK
OK OK OK
OK OK OK
Type D
OK
OK
OK
OK
Type D Type D Type D
OK OK OK
OK OK OK
OK OK OK
OK OK OK
Type D
OK
OK
OK
OK
Type D Type D Type D Type D Type D Type D Type D Type D
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK
OK OK OK OK OK – OK OK
OK : Use of the chemical is permitted.
–
: Use of the chemical is not permitted.
*1: If a waterproof cover is placed over the transducer connector, the entire transducer can be immersed in water or cleaning solution. *2: Not for use in the USA.
18
2B702-067EN*T
Quaternary ammonium chlorides/ Isopropyl alcohol
Hydrogen peroxide
Hydrogen peroxide
Isopropanol/ Ethanol/ Ethylene glycol monobutyl ether (2Butoxyethanol)/ Didecyldimethylammonium chloride/ Water
Octyl decyl dimethyl ammonium chloride
Sani-Cloth® Trophon® SONO ® Super AF3, Oxivir 1 PREempt Sani-Cloth CaviWipes1™ Companion Ultrasound wipes Wipes Sani-Cloth® Plus Cleaning Sani-Cloth® Wipes Wipes AF Professional Parker Advanced Virox Diversey, Metrex Nanosonics Disposables Professional Disposables Laboratories, Ultrasound Technologies International, Inc. Inc. Research Limited International, Inc. Solutions Inc. Inc. Inc. USA USA USA USA USA CAN USA AUS 0.25% / 0.232% 0.28% 0.5% / 55% 0.5% 0.5% 14.85% Working Working Working (working (working (working (working (working (working solution solution solution solution) solution) solution) solution) solution) solution) Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Wiping only Room Room Room Room Room Room Room Room Room temperature temperature temperature temperature temperature temperature temperature temperature temperature Normal Normal Normal Normal Normal Normal Normal Normal Normal humidity humidity humidity humidity humidity humidity humidity humidity humidity Normal Normal Normal Normal Normal Normal Normal Normal Normal pressure pressure pressure pressure pressure pressure pressure pressure pressure Protex™ Ultra Disinfectant Wipes
–
OK
OK
OK
–
–
–
–
–
OK OK OK
OK OK OK
OK OK OK
OK OK OK
OK – OK
OK OK OK
OK – OK
OK – OK
OK – OK
OK
OK
OK
OK
OK
OK
–
–
–
OK OK OK
OK OK OK
OK OK OK
OK OK OK
OK OK OK
OK OK OK
OK – OK
OK – OK
OK – OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK
– OK OK OK OK OK OK OK
OK OK OK OK OK – OK OK
OK OK OK OK OK – OK OK
OK OK OK OK OK – OK OK
OK OK OK OK OK – OK –
2B702-067EN*T
19
Except for Japan and China / ⽇本・中国以外向け / 日本及中国以外的国家或地区适用
Quaternary ammonium chlorides
Decyldimethyloctylammonium chloride/ Dimethyldioctylammonium chloride/ Didecyldimethylammonium chloride/ Quaternary ammonium compounds, benzyl-C12-16alkyldimethyl, chlorides