Cantel Medical

Endoscope Storage Cabinet Daily Use Procedure Guide Rev C

Guide

1 Page

ESC  Endoscope Storage Cabinet | Daily Use Procedure  Loading an Endoscope NOTE: choose an appropriate method of hand decontamination according to your local policy before handling endoscopes.  • Once the endoscope has been reprocessed it should be placed into the cabinet without delay. • Scan USER ID or log in manually. If screen is dark, touch screen to animate it.  • Place endoscope onto chosen hook. • Once the endoscope is in place, ensure all channels are connected and, if applicable, the channel separator is in place.  • Choose hook number  • Connect the DSD-hookup to the cabinet by inserting the male CPC connector into the cabinet manifold until a “click” is heard.  • Scan ENDOSCOPE ID or select OK to unlock door.  • Close door and press OK, a prompt will ask if you want a printout.  • Select ENDOSCOPE IN on screen  Removing an Endoscope • Scan USER ID or or log in manually. • Select ENDOSCOPE OUT on screen • Open door. • Disconnect the DSD hookup from the cabinet and remove endoscope. • Scan the ENDOSCOPE ID, close the door and press OK to confirm what scope has been removed. Cantel Medical Customer and Technical Support Toll Free: Phone: 1-800-444-4729 www.cantelmedical.com © 2017 Cantel 50098-1321-EN REV C  • A printout will then be produced with endoscope status, storage details, and user load/unload information.