Cantel Medical

Endoscope Storage Cabinet Maintenance & Cleaning Procedure Guide Rev C

Guide

2 Pages

ESC  Endoscope Storage Cabinet | Maintenance & 								  Cleaning Procedure  Daily Maintenance • Check that the door alarms and locks are functioning properly. • Verify air flow to each hookup port using a hookup or QA Airflow Indicator.  • Check the internal cabinet pressure reading to ensure the system is maintaining pressure above 17 Pa when the cabinet doors are closed. • Check the Drip Mat for excessive moisture. Replace as needed.  Weekly Maintenance  Monthly Maintenance  • Lubricate the hookup O-rings with Silicone oil.  • Ensure the air vents are not obstructed and are dust free.  • Change the Drip Mats. • Wipe down the interior and exterior of the cabinet with a disinfectant wipe.  • Check the hookups for cracks, splits or tears in the tubing and replace as necessary. Ensure that all hookups are securely connected to the endoscope and cabinet hookup port. • If the hookups are not routinely used during High-Level Disinfection, wipe the surface of the hookups with a disinfectant wipe.  Cantel Medical Customer and Technical Support Toll Free: Phone: 1-800-444-4729 www.cantelmedical.com  © 2017 Cantel 50098-1323-EN REV C