Endoscope Storage Cabinet Troubleshooting Alarms Guide Rev B

Guide

2 Pages

ESC  Endoscope Storage Cabinet | Troubleshooting Alarms  Invalid ID Error indicating a “manual” user login failed. •  Retry your login or contact the administrator to amend your ID or PIN.  Invalid Scan Error indicating a “scanned” user login failed. •  Retry your login or contact the administrator to amend your ID or PIN.  Endoscope Not Found The barcode or serial number presented to the ESC Endoscope Storage Cabinet has not been assigned to the ESC cabinet endoscope library. Proceed as per “Add Scope” using the barcode and serial number that is on the scope. Cabinet Pressure Low Alarm (fault code 3039) The pressure inside the ESC cabinet has fallen below the minimum pressure setting in Config Limits •  Fan has failed  •  Fan speed control requires adjustment  •  The ADC Cal offset is incorrectly set  •  Pressure transducer is faulty  •  Blocked Pre-filter  •  Blocked high efficiency cabinet filter  •  Leaking door seal  •  Door lock not engaged  Cabinet Pressure High Alarm (fault code 3040) The pressure inside the ESC cabinet has risen above the maximum pressure setting in Config Limits •  ACD Cal Gain setting is incorrect  •  Fan speed control requires adjustment  •  The ADC Cal offset is incorrectly set  •  Pressure transducer is faulty  •  Cabinet exhaust is restricted  © 2017 Cantel 50098-1320-EN REV B