Cardinal Healthcare
CareFusion Avea Pulse Oximetry Option Operators Manual Appendix Rev B
Appendix
18 Pages
Preview
Page 1
Avea® Pulse Oximetry Option Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
This document is protected by United States and International Copyright laws. This document may not be copied, reproduced, translated, stored in a retrieval system, transmitted in any form, or reduced to any electronic medium or machine-readable form, in whole or in part, without the written permission of CareFusion. Information in this document is subject to change without notice. This document is for informational purposes only and should not be considered as replacing or supplementing the terms and conditions of the License Agreement. © 2010 CareFusion Corporation or one of its subsidiaries. All rights reserved. Avea is a registered trademark of CareFusion Corporation or one of its subsidiaries. All other trademarks are the property of their respective owners. USA CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887-4668
Authorized European Representative CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Hoechberg, Germany District Court Wuerzburg HRB7004
800.231.2466 tel +1.714.283.2228 tel +1.714.283.8493 fax
+49.931.4972.0 tel +49.931.4972.423 fax
carefusion.com
Literature number: L3252 Revision B
ii
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Notices EMC Notice This equipment generates, uses, and can radiate radio frequency (RF) energy. If this equipment is not installed and used in accordance with the instructions in this manual, electromagnetic interference may result. This equipment has been tested and found to comply with the limits of acceptance set forth in Standard EN 60601-1-2 for Medical Products. These limits provide reasonable protection against electromagnetic interference (EMC) when operated in the intended use environments described in this manual. This ventilator is also designed and manufactured to comply with the safety requirements of Standard EN 60601-1, EN/ISO 9919, IEC 60601-2-12, CAN/CSA-C22.2 No. 601.1-M90, and UL 2601-1. This ventilator can be affected by portable and mobile RF communications equipment. This ventilator should not be stacked with other equipment. The following cables were used in the evaluation of this ventilator. ● 15619 – Normally Open Patient Call Cable (Length – 1.7 meters) ● 15620 – Normally Closed Patient Call Cable (Length – 1.7 meters) ● 70600 – Cable, Communications (Length – 1 meter) ● 70693 – Cable, Communications (Length – 3 meters) ● Standard Centronix™ Printer Cable (Length – 2 meters) ● Standard SVGA Monitor Cable (Length – 2 meters) Use of other cables may result in increased emissions or decreased immunity. See Tables 201, 202, 203, and 205 (below) for further information regarding the Avea Ventilator and EMC.
L3252 Rev. B
iii
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Table 201. 60601-1-2 IEC:2001 (E) Guidance and manufacturer’s declaration – electromagnetic emissions The Avea Ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the Avea Ventilator should assure that it is used in such an environment. Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The Avea Ventilator uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The Avea Ventilator is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply that supplies buildings used for domestic purposes.
CISPR 11 RF emissions CISPR 11 Harmonic emissions
Class A
IEC 61000-3-3 Voltage Fluctuation/
Complies
Flicker emissions IEC 61000-3-3
iv
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Table 202. 60601-1-2 IEC:2001 (E) Guidance and manufacturer’s declaration - electromagnetic immunity The Avea Ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the Avea Ventilator should assure that it is used in such an environment. Immunity Test
IEC 60601
Compliance level
Electromagnetic environment - guidance
Test level Electrostatic discharge (ESD)
± 6 kV contact
± 6 kV contact
IEC 61000-4-2
± 8 kV air
± 8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
± 6 kV for power supply lines
± 6 kV for power supply lines
IEC 61000-4-4
± 1 kV for input/output lines
± 1 kV for input/output lines
Surge
± 1 kV differential mode
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
± 2 kV common mode
Voltage dips, short interruptions and voltage variations on power supply input lines
<5 % UT (>95% dip in UT)
<5 % UT (>95% dip in UT)
for 0,5 cycle
for 0,5 cycle
IEC 61000-4-11
40 % UT
40 % UT
for 5 cycles
for 5 cycles
70 % UT
70 % UT
(60 % dip in UT)
Power frequency (50/60 Hz) magnetic field
(60 % dip in UT)
(30 % dip in UT)
(30 % dip in UT)
for 25 cycle
for 25 cycle
<5 % UT
<5 % UT
(>95% dip in UT)
(>95% dip in UT)
for 5 seconds
for 5 seconds
3 A/m
3 A/m
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Compliance is dependent on the operator following recommended charging and maintenance of the installed battery backup.
Power frequency magnetic fields should be at level characteristic of a typical location in a typical commercial or hospital environment.
IEC 61000-4-8 Note: UT is the AC. mains voltage prior to application of the test level.
L3252 Rev. B
v
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Table 203. 60601-1-2 IEC:2001 (E) Guidance and manufacturer’s declaration – electromagnetic immunity The Avea Ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the Avea Ventilator should assure that it is used in such an environment. Immunity Test
IEC 60601 Test level
Compliance level
Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6
3 V rms 150 kHz to 80 MHz outside ISM bandsa
3V
10 V rms 150 kHz to 80 MHz In ISM bandsa
10 V
Portable and mobile RF communications equipment should be used no closer to any part of the Avea Ventilator, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
10 V/m 80 MHz to 2,5 GHz
10 V/m
Radiated RF IEC 61000-4-3
Recommended separation distance:
80 MHz to 800 MHz 800 MHz to 2.5 GHz Where is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and is the recommended separation distance in meters (m). b Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, c should be less than the compliance level in each frequency range. d Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a. The ISM (industrial, scientific, and medicinal) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27, 283 MHz; and 40,66 MHz to 40,70 MHz. b. The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges. c. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed FR transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Avea Ventilator is used exceeds the applicable RF compliance level above, the Avea Ventilator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Avea Ventilator. d. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
vi
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Table 205 60601-1-2 IEC:2001 (E) Recommended separation distance between portable and mobile RF communications equipment and the Avea Ventilator The Avea Ventilator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Avea Ventilator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Avea Ventilator as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter (W)
Separation distance according to frequency of transmitter (m) 150 kHz to 80 MHz outside ISM bands
150 kHz to 80 MHz in ISM bands
80 MHz to 800 MHz
80 MHz to 800 MHz
0.01
0.12
0.12
0.12
0.23
0.1
0.37
0.38
0.38
0.73
1
1.16
1.20
1.20
2.30
10
3.67
3.79
3.79
7.27
100
11.60
12.00
12.00
23.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer. Note1. At 80 MHz and 800 MHz, the separation distance of the higher frequency range applies. Note 2. The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. Note 3. An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. Note4. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
L3252 Rev. B
vii
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Regulatory Notice Federal law restricts the sale of this device, except by or on order of a physician.
Manufacturer CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887-4668 USA If you have a question regarding the Declaration of Conformity for this product, please contact CareFusion (see page ii for the contact information).
viii
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Safety Information Please review the following safety information before operating the ventilator. Attempting to operate the ventilator without fully understanding its features and functions may result in unsafe operating conditions. Warnings and Cautions, which are general to the use of the ventilator under all circumstances, are included in this section. Some Warnings and Cautions are also inserted within the manual where they are most meaningful. Notes are also located throughout the manual to provide additional information related to specific features. If you have a question regarding the installation, set up, operation, or maintenance of the ventilator, contact CareFusion Customer Care (see page ii for the contact information).
Definition of Terms The following list describes the use of notes, cautions, and warnings in this document. Notes provide supplemental information to help you understand how the ventilator works. Cautions identify conditions or practices that can cause damage to the ventilator or other equipment. Warnings identify conditions or practices that can cause a serious, adverse reaction or are potential safety hazards.
Warnings The Warnings and Cautions listed here apply generally anytime you operate the ventilator. The warnings contained in this addendum are in addition to the warnings in the Av ea Operator’s Manual . Excessive ambient light, poor perfusion states, patient movement, venous pulsations, a disconnected pulse oximeter probe and other clinical causes (significant levels of dysfunctional hemoglobin or low hemoglobin levels, intravascular dyes such as indocyanine or methylene blue, venous pulsations at the frequency of the arterial pulsations), may lead to incorrect SPO2 measurements. If any measurement seems questionable, first check the patient’s vital signs by alternate means and then check the equipment for proper functioning. Loss of the pulse signal may occur during periods of severe vasoconstriction, severe anemia, hypotension, hypothermia, cardiac arrest or when there is arterial occlusion proximal to the sensor. Place the sensor on patient site that has unrestricted blood flow, as specified in the direction for use provided in the packaging of the adhesive sensor. Do not constrict the monitoring site when securing the probe because inaccurate readings or tissue damage may occur. Do not position the cable in any manner that may cause entanglement, strangulation, or accidental self-extubation. The pulse oximeter can be used during defibrillation, but the readings may be inaccurate for a short time. Use only Masimo® pulse oximetry sensors.
L3252 Rev. B
ix
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Cautions The following cautions apply anytime you work with the ventilator: ● Do not use damaged sensors or cables. ● Do not immerse sensors or the oximeter module housing in any liquid. ● Do not attempt to sterilize sensors by irradiation, steam, or ethylene oxide. ● Do not apply excessive tension to any sensor cable. ● Do not open the oximeter module housing. There are no user-serviceable parts inside.
x
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Contents Notices
... i
Safety Information ... ix Theory of Operation ...1 Configuration and Setup ...2 Monitors
...4
Waveforms
...5
Alarms
...5
L3252 Rev. B
xi
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
xii
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Theory of Operation Masimo-SET™ This device is calibrated to display functional oxygen saturation. The device uses a two-wavelength system by passing red (660 nm wavelength) and infra-red (905 nm wavelength) light through the capillary bed and measuring changes in light absorbance through the cardiac cycle. The function principle of the pulse oximeter with Masimo-SET is based on (1) the fact that oxyhemoglobin and deoxyhemoglobin differ in their capability of absorbing red and infrared light, (2) that the volume of arterial blood in the tissues varies from pulse-to-pulse (thus also the light absorption by the blood), and (3) that arteriovenous shunting is very variable and the fluctuating absorption by the venous blood is the dominant noise component. Since oxyhemoglobin and deoxyhemoglobin absorption capabilities are different, the amount of red and infrared light absorbed by the blood is related to the oxygen saturation of the hemoglobin. Masimo-SET divides the red and infrared pulsating portion of the absorption signal into an arterial and a noise signal thereby calculating the ratio of the arterial signals without the noise portion. The ratio of the two arterial absorption values is used by an empirically-derived equation to determine the oxygen saturation. Calibration of this device is neither required nor possible, and a functional tester cannot be used to assess the accuracy of a pulse oximeter probe or system with a co-oximeter.
L3252 Rev. B
1
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Configuration and Setup The settings are found in the Monitoring tab in the Utility Screen. Depress the Screens button, touch the Utility button, and select the Monitoring tab.
Pulse Ox imeter – Enable / Disable This control activates and de-actives the Pulse Oximeter system. ● Range: Enable / Disable ● Default: Disable
Algorithm This control allows you to select the sensitivity mode of the pulse oximeter based on individual needs of the patient. Maximum
Normal
APOD (Adaptive Probe Off Detection)
This mode should be used for the sickest patients, where obtaining a reading is most difficult. Maximum Sensitivity is designed to interpret and display data for even the weakest of signals.
This mode provides the best combination of sensitivity, and probe-off detection performance. This mode is recommended for the majority of patients. It is also recommended during procedures and when clinician and patient contact is continuous.
This mode is the least sensitive in picking up a reading on patients with low perfusion but has the best detection for probe-off conditions. This mode is useful for patients that are at particular risk of the sensor becoming detached (pediatric, combative, etc.)
● Range: Max / Normal / APOD ● Default: Normal
Note: If using the AutoFiO2 option, the Algorithm is automatically set to Normal and cannot be changed as long as the AutoFiO2 option is enabled.
2
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Av eraging Determines the time period in which the SpO2 reading is averaged. ● Range: 2, 4, 8, 10, 12, 14, or 16 seconds ● Default: 8 seconds
Note: If using the AutoFiO2 option, Averaging is automatically set to eight (8) seconds and cannot be changed as long as the AutoFiO2 option is enabled.
SpO 2 Delay The SpO2 Alarm Delay determines the duration that a High or Low SpO2 Alarm or SpO2 Signal Quality alarm must persist before the alarm is activated. ● Delay Range: 10 to 120 seconds ● Resolution: 5 seconds ● Default: 60 seconds
L3252 Rev. B
3
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Monitors Pulse oximetry data may be displayed on any of the five monitors to the left of the graphic displays on the main screen. This data is also selectable on the trending screen. Monitored values are:
SpO 2 The patient’s SpO2 as measured and reported by the pulse oximeter. ● Range: 1 to 100% ● Resolution: 1% ● Accuracy: ± 3% from 70 – 100%. Unspecified < 70%.
Pulse Rate The patient’s pulse rate as measured and reported by the pulse oximeter. ● Range: 25 to 240 bpm ● Accuracy: ± 4 bpm ● Resolution: 1 bpm
Perf. Index The perfusion index (PI) indicates the percentage of pulsatile signal to non-pulsatile signal (pulse strength). The PI is useful in determining optimal probe placement and troubleshooting. ● Range: 0.02 to 20.0% ● Resolution: 0.01% or three significant digits, whichever is greater.
Note: The SpO2, Pulse Rate and Perfusion Index values may be subject to a delay of up to one breath period or ten seconds (whichever occurs first), because the update period is governed by the ventilatory monitoring functions.
4
L3252 Rev. B
Operator’s Manual Appendix
Avea® Pulse Oximetry Option
Wav eforms Pulse oximetry waveforms may be selected as one of the scalar waveforms on the Avea Ventilator.
Plethy smographic Wav eform The plethysmographic waveform is displayed as a scalar wave. The waveform is a continuous, autoscaled waveform. It is displayed to 90% of full scale for a signal strength ≥10%. Below 10%, the waveform is scaled monotonically with signal strength.
Signal IQ The height of the vertical line of the Signal IQ indicates the quality of the measured arterial pulse signal. A high vertical bar indicates that the SpO2 measurement is based on a good quality signal. A small vertical bar indicates that the SpO2 measurement is based on data with low signal quality. Low signal quality may compromise the accuracy of the SpO2 measurements.
Alarms Low SpO 2 Alarm The Low SpO2 Alarm is activated if the measured SpO2 is at, or lower than, the Low SpO2 Alarm setting for more than the SpO2 Alarm Delay setting. This is a high priority alarm. ● Range: 60 to 97% or Off ● Resolution: 1% ● Default: 87% (neonatal); 90% (pediatric and adult)
High SpO 2 Alarm The High SpO2 Alarm is activated if the measured SpO2 is at, or higher than, the High SpO2 Alarm setting for more than the SpO2 Alarm Delay setting. This is a high priority alarm. ● Range: 70 to 100% or Off ● Resolution: 1% ● Default: 96% (neonatal); Off (pediatric and adult)
Note: If you are using the AutoFiO2 Option, the High and Low SpO2 Alarms may only be disabled if the High and Low AutoFiO2 alarms are set (not disabled). The High SpO2 Alarm must be set at least 2% higher than the Low SpO2 Alarm.
High Pulse The device gives an alarm if the monitored pulse rate is greater than or equal to the High Pulse alarm setting. This is a high priority alarm. Range: 30 to 240 bpm, Off Resolution: 5 bpm Default: 180 bpm (neonatal); 140 bpm (pediatric and adult) L3252 Rev. B
5
Avea® Pulse Oximetry Option
Operator’s Manual Appendix
Low Pulse The device gives an alarm if the monitored pulse rate is less than or equal to the Low Pulse alarm setting. This is a high priority alarm. Range: Off, 25 to 235 bpm Resolution: 5 bpm Default: 100 bpm (neonatal); 50 bpm (pediatric and adult)
Note: The Low Pulse Alarm must be set at least 5 bpm below the High Pulse Alarm
Note: The Alarm signal generation delay for low and high SpO2 and for low and high pulse rate equal to the sum of the oximeter averaging time plus the SpO2 alarm delay time.
Poor SpO 2 Signal Quality Alarms Poor SpO2 Signal Quality alarms indicate a problem with the pulse oximetry signal. Only one alarm appears at a time. If more than one occurs simultaneously, the highest ranking alarm appears. These are low or high priority alarms. Alarm messages (ranked from highest to lowest priority): ● SpO2: Sensor off ● SpO2: Low SIQ ● SpO2: Low perfusion ● SpO2: Pulse search ● SpO2: Interference ● SpO2: Ambient Light
Note: The alarm starts as a low priority alarm and transitions to a high priority alarm when the alarm has been active for more than the SpO2 Alarm Delay period.
SpO 2 Inv alid Alarms These alarms indicate a problem with the sensor, cable, or interface box. These are high priority alarms. ● SpO2: No sensor ● SpO2: Defective sensor ● SpO2: Unrecognized sensor ● SpO2: Not connected
6
L3252 Rev. B