ASSISTANCE
CareFusion Respiratory Systems 22745 Savi Ranch Parkway Yorba Linda, California 92887-4645, USA Tel: 763.398.8500 800.754.1914 Fax: 763.398.8403 Website: www.carefusion.com Email:
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
ii
TABLE OF CONTENTS Table of Contents: Front and Side Panel Reference ... 1 Front Panel Display and Description... 1 Side Panel Descriptions... 3 Turning the Ventilator On and Off ... 5 Turning the Ventilator On ... 5 Turning the Ventilator Off ... 6 LTV® 1200/1150 Preset values Table... 7 LTV® 1200/1150 Preset Values Table (cont.) ... 8 Variable Controls... 9 SETTING UP MODES OF VENTILATION ... 11 Setting Up Assist/Control Mode ... 11 Setting Up SIMV Mode ... 13 Setting Up CPAP Mode ... 15 Setting Up NPPV Mode ... 17 Monitored Data ... 19 P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
iii
Extended Features ... 21 SBT (Spontaneous Breathing Trial) ... 23 Using AC/DC Power ... 27 Using the AC Adapter ... 27 Using an External DC Power Source ... 28 Power Displays and Indicators ... 29 Attaching a Breathing Circuit ... 32 Oxygen Computer Chart ... 33 Alarms ... 35
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
iv
FRONT AND SIDE PANEL REFERENCE Front Panel Display and Description (LTV® 1200 shown)
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
1
A - Mode and Breath Selection – Selects ventilation modes, and selects breath types. B - On/Standby Button – Turns the ventilator “On” or to “Standby”. C - Variable Control Settings – Sets and displays each ventilation characteristic. D - Display Window – Displays Alarm Messages, Monitored Data, and Extended Features menus. E - Airway Pressure Display – Displays real-time airway circuit pressure. F - Patient Effort Indicator – LED is lit briefly each time a patient trigger is detected. G - Power Source – Displays power source and charge levels. H - Variable Alarm Settings – Sets and displays variable alarm levels. I-
Alarm Silence/Reset – Silences audible alarms. Clears visual alarms.
J - Set Value Knob – Changes variable control settings. Navigates Extended Features. K - Special Controls – Activates special controls such as Manual Breath, Low ® Pressure O2 Source (LTV 1200 only), Insp/Exp Hold and Control Lock feature. L - PEEP – PEEP control setting and display.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
2
FRONT AND SIDE PANEL REFERENCE Side Panel Descriptions
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
3
A - 22mm Outlet Port – Patient Breathing Circuit outlet port. B - Flow Xducer – Flow Transducer high pressure sensing port. C - Flow Xducer – Flow Transducer low pressure sensing port. D - Exh Valve – Exhalation Valve drive line port. E - Alarm Sounder Port F - Cooling Fan G - DC Input – DC power port pigtail connector. H - Patient Assist – Patient Assist Call jack. I-
Comm Port – Communications port.
J - O2 Inlet – Oxygen Inlet fitting. K - Filter – Air Inlet.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
4
TURNING THE VENTILATOR ON AND OFF Turning the Ventilator On 1) Push the On/Standby button. If the Patient Query feature is enabled/on when the ventilator is powered up, ventilation and alarm activation are suspended and the message SAME PATIENT is displayed.
To enable the suspended alarms and begin ventilation with the settings in use during the last power cycle, press the Select button while SAME PATIENT is displayed.
To enable the suspended alarms and begin ventilation with Preset values appropriate for a new patient, turn the Set Value knob until NEW PATIENT is displayed and press the Select button. Then turn the Set Value knob until the desired patient type is displayed (INFANT, PEDIATRIC or ADULT) and press the ® ® Select button (see the LTV 1200 or LTV 1150 Operator’s Manual, Chapter 10, for detailed settings and information).
If the Patient Query feature is disabled/off when the ventilator is powered up and passes POST, it will begin ventilation (appropriate alarms enabled) using the settings in use during the last power cycle.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
5
Turning the Ventilator Off To turn the ventilator off: 1)
Disconnect the patient from the ventilator.
2)
Press and hold the On/Standby button for 3 seconds. The ventilator ceases operating, the audible alarm sounds continuously and the Vent Inop LED is lit.
3)
Press the Silence/Reset button to silence the audible alarm.
4)
Verify a confirming audible chirp is activated immediately after the alarm is silenced.
The ventilator continues to charge the internal battery as long as it is connected to an external power source. Note: The Vent Inop LED will remain lit for a minimum of 5 minutes and does not impact battery life.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
6
LTV® 1200/1150 PRESET VALUES TABLE Feature/Function
Infant (>5kg < 10kg)
Pediatric (10-40 kg)
Adult (> 40 kg)
Alarm volume HP delay LPP alarm High f alarm
85 NO DELAY ALL BREATHS 80 bpm, 30 sec
85 NO DELAY ALL BREATHS 60 bpm, 30 sec
High PEEP alarm
PEEP +5 cmH2O
PEEP +5 cmH2O
Low PEEP alarm Rise Time FLOW TERM TIME TERM PC FLOW TERM Leak Comp Breath Rate Breath type Tidal Volume
PEEP -3 cmH2O 5 35% TERM 0.5 sec Off
PEEP -3 cmH2O 4 30% TERM 1.0 sec Off
85 NO DELAY ALL BREATHS 40 bpm, 30 sec PEEP +5 cmH2O PEEP -3 cmH2O 4 25% TERM 2.0 sec Off
P/N 19822-001 Rev. B
LEAK COMP ON
LEAK COMP ON
LEAK COMP ON
20 bpm Pressure 50 ml
15 bpm Pressure 250 ml
12 bpm Volume 500 ml
®
LTV 1200/1150 Ventilator
7
LTV® 1200/1150 Preset Values Table (cont.) Feature/Function
Infant (>5kg < 10kg)
Pediatric (10-40 kg)
Adult (> 40 kg)
Insp. Time Pres. Control Pres. Support Sensitivity High Pres. Limit Low Pressure Low Min. Vol. PEEP Mode
0.3 sec 15 cmH2O 10 cmH2O 2 Lpm 30 cmH2O 10 cmH2O 0.5 ml 0 cmH2O Assist/Ctrl
0.7 sec 15 cmH2O 10 cmH2O 3 Lpm 30 cmH2O 10 cmH2O 1.0 ml 0 cmH2O Assist/Ctrl
1.0 sec 15 cmH2O 10 cmH2O 3 Lpm 40 cmH2O 10 cmH2O 3.0 ml 0 cmH2O Assist/Ctrl
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
8
VARIABLE CONTROLS
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
9
To set a variable control: 1) Select the control by pressing the associated button. The display for the selected control will be displayed at normal brightness and all other control displays will be dimmed. 2) Change the control value by rotating the Set Value Knob. Rotate clockwise to increase and counter-clockwise to decrease the value. 3) The new control value goes into effect when the operator: Presses the selected button again, or Selects another control, or Presses the Control Lock button, or Waits 5 seconds All controls will then return to their normal brightness.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
10
Setting Up the Ventilator in Assist/Control Mode: Press the Select button twice to toggle the modes between Assist/Control and SIMV/CPAP. Select the Assist/Control mode. 2) Press the Select button twice to toggle between Volume and Pressure ventilation. Select Volume or Pressure, as desired. 3) Set the Breath Rate. 4) If Volume ventilation is selected, set the Tidal Volume. The calculated peak flow Vcalc is displayed in the window while Tidal Volume is being changed. 5) If Pressure ventilation is selected, set the Pressure Control. 6) Set the Inspiratory Time. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation. ® 7) Set O2% (LTV 1200 only). 8) Set the Sensitivity to a setting from 1 to 9. 9) Set the High Pres. Limit alarm. 10) Set the Low Pressure alarm. 11) Set the Low Min. Vol. alarm. 12) Adjust the PEEP control. 1)
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
12
To set the Ventilator up in SIMV mode: Press the Select button twice to toggle the modes between Assist/Control and SIMV/CPAP. Select the SIMV/CPAP mode. 2) Press the Select button to toggle between Volume and Pressure ventilation. Select Volume or Pressure, as desired. 3) Set the Breath Rate. 4) If Volume ventilation is selected, set the Tidal Volume. The calculated peak flow Vcalc is displayed in the window while Tidal Volume is being changed. 5) If Pressure ventilation is selected, set the Pressure Control. 6) Set the Inspiratory Time. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation. 7) Set the Pressure Support, if desired. ® 8) Set O2% (LTV 1200 only). 9) Set the Sensitivity to a setting from 1 to 9. 10) Set the High Pres. Limit alarm. 11) Set the Low Pressure alarm. 12) Set the Low Min. Vol. alarm. 13) Adjust the PEEP control. 1)
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
14
To set the ventilator up in CPAP mode: Press the Select button twice to toggle the modes between Assist/Control and SIMV/CPAP. Select the SIMV/CPAP mode. 2) Press the Select button twice to toggle between Volume and Pressure ventilation for Apnea backup. Select Volume or Pressure for Apnea backup. 3) Set the Breath Rate to Off (dashes “- -“). 4) If Volume ventilation is selected, set the Tidal Volume for Apnea backup. The calculated peak flow Vcalc is displayed in the window while Tidal Volume is being changed. 5) If Pressure ventilation is selected, set the Pressure Control for Apnea backup. 6) Set the Inspiratory Time for Apnea backup. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation. 7) Set the Pressure Support, if desired. ® 8) Set O2% (LTV 1200 only). 9) Set the Sensitivity to a setting from 1 to 9. 10) Set the High Pres. Limit alarm. 11) Set the Low Pressure alarm for Apnea backup. 12) Set the Low Min. Vol. alarm. 13) Adjust the PEEP control. 1)
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
16
To set the Ventilator up in NPPV mode: Set any desired Extended Features options and: 1) Push the Assist/Control, SIMV/CPAP mode button until the NPPV LED flashes. Press the button once more to confirm. The NPPV LED continues to flash and SET IPAP displays. The Pres. Support control display is bright and all other controls dim. 2) Turn the Set Value knob to adjust the IPAP value (shown in Pres. Support LED window). Press the Pres. Support button to confirm, SET EPAP will display. The PEEP control display is bright and all other controls are dim. 3) Turn the Set Value knob to adjust the EPAP value (shown in the PEEP LED window). Press the PEEP button to confirm. 4) The PEEP button push confirms NPPV operation and LED then turns solid. ® 5) Set O2% (LTV 1200 only). 6) Set the High Pres. Limit alarm.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
18
MONITORED DATA The monitored data displays may be automatically scrolled or manually scrolled. To cycle through the available monitored data automatically from a halted scan, press the Monitor Select button (left of display window) twice. Pressing the Select button once while scan is active shall halt scanning and the currently display monitor shall remain in the display window. Each time you press the button once; the next data item in the list will be displayed. To resume scan, press the Select button twice within 0.3 seconds. The monitored data is displayed in the following order: Display PIP MAP PEEP f Vte VE
Description Displays the Peak Inspiratory Pressure measured during the inspiratory phase. PIP is not updated for spontaneous breaths. Displays a running average of the airway pressure for the last 60 seconds. Displays the pressure in the airway circuit at the end of exhalation. Displays the breaths per minute and includes all breath types. Displays the exhaled tidal volume as measured at the patient wye. Displays the exhaled tidal volume for the last 60 seconds as calculated from the last 8 breaths.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
19
Display I:E I:Ecalc
Vcalc
SBT min
f/Vt f
Description Displays the ratio between measured inspiratory time and measured exhalation time. Both normal and inverse I:E Ratios are displayed. Displays the ratio between the set Breath Rate and Inspiratory Time. The display is updated in real-time while the Breath Rate setting is being changed. Is based on the Tidal Volume and Inspiratory Time settings. Displayed when selected and whenever Tidal Volume or Inspiratory Time is selected for change. Displays the time remaining until the number of minutes preset in the SBT OP, MINUTES menu have elapsed. (Only displayed in the SBT mode of ventilation.) f/Vt is computed every time the Total Breath Rate (f) or Total Minute Volume (VE) is calculated. (Only displayed when SBT mode is selected.)
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
20
EXTENDED FEATURES Navigating the Extended Features Menus: To enter the Extended Features menu (in normal ventilation mode), press and hold the Monitor Select button for three seconds. To view the next item in a menu, turn the Set Value knob clockwise. To view the previous item, turn the Set Value knob counterclockwise. To enter a menu item or select a setting, press the Select button. To exit a menu, turn the Set Value knob until the EXIT option is displayed, then press the Select button or press Control Lock.
P/N 19822-001 Rev. B
®
LTV 1200/1150 Ventilator
21