Operators Manual
264 Pages
Preview
Page 1
™
AVEA Ventilator Systems Operator’s Manual
ii
Operator's manual
This document is protected by United States and International Copyright laws. This document may not be copied, reproduced, translated, stored in a retrieval system, transmitted in any form, or reduced to any electronic medium or machine-readable form, in whole or in part, without the written permission of CareFusion. Information in this document is subject to change without notice. This document is for informational purposes only and should not be considered as replacing or supplementing the terms and conditions of the License Agreement. © 2006–2016 CareFusion Corporation or one of its affiliates. All rights reserved. AVEA, CareFusion, and the CareFusion logo are trademarks of CareFusion Corporation or one of its affiliates. All other trademarks are the property of their respective owners. CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887 U.S.A.
CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Hoechberg Germany
800.231.2466 toll-free 714.283.2228 tel 714.283.8493 fax
+49 931 4972.0 tel +49 931 4972.423 fax
carefusion.com
L2786 Version T
iii
AVEA ventilator systems
Version History Date
Version
Changes
September 2005
A
Release
May 2006
B
Removed references to the Plus model Removed “non-operational” from figure 2-19 Added a note regarding the setting of Peak Inspiratory Pressure Added Ppeak to the list of alert messages
December 2006
C
Updated the Primary Controls table Added NCPAP to the troubleshooting table Updated the Rate specification Added the chapter “Infant NCPAP”
February 2007
D
Updated the figure and added a description of the balloon size and type selection Added a note regarding the date and time Added Addendums Filter/Water Trap, Processing Instructions, Sterilization, L2860-101, L3004A, L3082A, L2889A, L3031A
September, 2008
E
Release of Volumetric Capnography
February 2010
F
Revised the manual to comply with the revised Medical Device Directive 2007/42/EC.
February 2010
G
Rebranded to the CareFusion style.
June 2010
H
Revised the Circuit Resistance Test regarding the use of ET tubes. Added a bullet under Warnings: Routine assessment of oxygenation and ventilation should be performed when a patient is receiving respiratory support.
January 2011
J
Changed the logo and company references to VIASYS.
January 2011
K
Changed the logo and company references to CareFusion.
June 2011
L
Added content for Volume Guarantee and Nasal Intermittent Mandatory Ventilation.
July 2011
M
Removed the first note from the section “Specific Controls” regarding the flow cycling of breaths.
February 2012
N
Updated Table E-2.
July 2012
P
Updated according to ECO 80620.
April 2014
R
Added the Manufacturer and EC REP symbols to the address information and added the Attention symbol to all caution and warning statements to meet the requirements of MDD 93/42/EEC as amended by 2007/47/EC Annex 1, Section 3.1. Updated references to standards. Changed all instances of “tracheal catheter” to “tracheal monitoring tube” and all instances of esophageal catheter to “esophageal balloon.”
February 2016
T
Updated according to ECO 84941.
L2786 Version T
iv
Operator's manual
Warranty THE AVEA™ ventilator systems are warranted to be free from defects in material and workmanship and to meet the published specifications for Two (2) years or 16,000 hours, whichever occurs first. The liability of CareFusion (referred to as the Company) under this warranty is limited to replacing, repairing or issuing credit, at the discretion of the Company, for parts that become defective or fail to meet published specifications during the warranty period; the Company will not be liable under this warranty unless (A) the Company is promptly notified in writing by Buyer upon discovery of defects or failure to meet published specifications; (B) the defective unit or part is returned to the Company, transportation charges prepaid by Buyer; (C) the defective unit or part is received by the Company for adjustment no later than four weeks following the last day of the warranty period; and (D) the Company’s examination of such unit or part shall disclose, to its satisfaction, that such defects or failures have not been caused by misuse, neglect, improper installation, unauthorized repair, alteration or accident. Any authorization of the Company for repair or alteration by the Buyer must be in writing to prevent voiding the warranty. In no event shall the Company be liable to the Buyer for loss of profits, loss of use, consequential damage or damages of any kind based upon a claim for breach of warranty, other than the purchase price of any defective product covered hereunder. The Company warranties as herein and above set forth shall not be enlarged, diminished or affected by, and no obligation or liability shall arise or grow out of the rendering of technical advice or service by the Company or its agents in connection with the Buyer's order of the products furnished hereunder.
Limitation of Liabilities This warranty does not cover normal maintenance such as cleaning, adjustment or lubrication and updating of equipment parts. This warranty shall be void and shall not apply if the equipment is used with accessories or parts not manufactured by the Company or authorized for use in writing by the Company or if the equipment is not maintained in accordance with the prescribed schedule of maintenance. The warranty stated above shall extend for a period of TWO (2) years from date of shipment or 16,000 hours of use, whichever occurs first, with the following exceptions: 1.
Components for monitoring of physical variables such as temperature, pressure, oxygen, or flow are warranted for ninety (90) days from date of receipt.
2.
Elastomeric components and other parts or components subject to deterioration, over which the Company has no control, are warranted for sixty (60) days from date of receipt.
3.
Internal batteries are warranted for ninety (90) days from the date of receipt.
The foregoing is in lieu of any warranty, expressed or implied, including, without limitation, any warranty of merchantability, except as to title, and can be amended only in writing by a duly authorized representative of the Company.
L2786 Version T
AVEA ventilator systems
Contents Revision History ...iii Warranty ... iv Limitation of Liabilities ... iv Notices ... viii EMC Notice ... viii MRI Notice ... viii Intended Use Notice ... viii Regulatory Notice ... viii Classification ... ix Declaration of Conformity Notice ... ix Safety Information ... x Terms ... x Warnings ... x Cautions ... xi Equipment Symbols ... xiii Chapter 1:
Introduction... 1
Chapter 2: Unpacking and Setup ... 7 Ventilator Assembly and Physical Setup ... 7 Setting Up the Front of the Ventilator ... 9 Front Panel Connections ... 17 Setting Up the Rear of the Ventilator ... 24 User Verification Test ... 39 AVEA User Verification Test Checklist ... 46 AVEA Troubleshooting ... 47 Chapter 3: Ventilator Operation ... 51 Membrane Buttons and LEDs ... 51 Patient Setup ... 59 Ventilation Setup ... 61 Setting the Ventilation Breath Type and Mode ... 66 Volume Guarantee (VG) ... 67 Primary Breath Controls ... 90 Advanced Settings... 96 Independent Lung Ventilation (ILV) ... 104
L2786 Version T
v
vi
Operator's manual
Chapter 4: Monitors, Displays and Maneuvers ...105 Graphic Displays ...105 Digital Displays ...124 Main Screen Displays ...130 Chapter 5: Volumetric Capnography ...133 Theory of Operation ...133 Setup ...134 Settings and Monitored Values ...137 Alarms ...141 Maneuvers ...142 Zeroing the CAPNOSTAT 5 ...144 Checking the Accuracy of the CAPNOSTAT 5 ...146 Chapter 6: Infant Non-invasive Ventilation ...149 Nasal CPAP (nCPAP) ...149 Nasal Intermittent Mandatory Ventilation (nIMV) ...155 Chapter 7: Alarms and Indicators ...167 Status Indicators ...167 Messages ...169 Alarms ...171 Alarm Controls ...172 Alarm Types ...173 Nasal CPAP / Nasal IMV Alarms ...182 Volume Guarantee Alarms ...185 Chapter 8: Maintenance and Cleaning ...187 Cleaning and Sterilization ...187 Recommended Periodic Maintenance ...190 Battery Care ...191 Fuses ...194 Appendix A: Contact and Ordering Information ...197 How to Call for Service ...197 Ordering Parts...198 Appendix B: Specifications ...201 Pneumatic Supply ...201 Electrical Supply ...201 Data Input / Output...202
L2786 Version T
vii
AVEA ventilator systems
Atmospheric and Environmental Specifications ... 209 Physical Dimensions ... 209 Accessories ... 209 Appendix C: Pneumatic Diagram ... 211 Appendix D: Monitor Ranges and Accuracies ... 213 Appendix E: Sensor Specifications and Circuit Resistance ... 217 VarFlex® Sensor Specifications ... 217 Hot Wire Flow Sensor Specifications... 218 Circuit Resistance Test... 219 Volumetric Capnography Specifications ... 220 Appendix F: AVEA Message Bar Text ... 223 Appendix G: Advanced Pulmonary Mechanics Monitored Parameters ... 225 Appendix H: Capnometry Troubleshooting ... 233 Appendix I: Volumetric CO2 Calculations... 235 Appendix J: Electromagnetic Declarations ... 239 Appendix K: Glossary ... 243 Index
L2786 Version T
... 245
viii
Operator's manual
Notices EMC Notice This equipment generates, uses, and can radiate radio frequency (RF) energy. If this equipment is not installed and used in accordance with the instructions in this manual, electromagnetic interference may result. This equipment has been tested and found to comply with the limits of acceptance set forth in Standard EN 60601-1-2 for Medical Products. These limits provide reasonable protection against electromagnetic interference (EMC) when operated in the intended use environments described in this manual. This ventilator is also designed and manufactured to comply with the safety requirements of Standard EN 60601-1, IEC 60601-2-12, CAN/CSA-C22.2 No. 601.1-M90, and UL 60601-1-1. This ventilator can be affected by portable and mobile RF communications equipment. This ventilator should not be stacked with other equipment. The following cables were used in the evaluation of this ventilator.
15619 – Normally Open Patient Call Cable (Length – 1.7 meters)
15620 – Normally Closed Patient Call Cable (Length – 1.7 meters)
70600 – Cable, Communications (Length – 1 meter)
70693 – Cable, Communications (Length – 3 meters)
Standard Centronics Printer Cable (Length – 2 meters)
Standard SVGA Monitor Cable (Length – 2 meters)
Use of other cables may result in increased emissions or decreased immunity. See Tables 201, 202, 203, and 205 for further information regarding the AVEA Ventilator and EMC.
MRI Notice This equipment contains electromagnetic components whose operation can be affected by intense electromagnetic fields. Do not operate the ventilator in a MRI environment or in the vicinity of high-frequency surgical diathermy equipment, defibrillators, or short-wave therapy equipment. Electromagnetic interference could disrupt the operation of the ventilator.
Intended Use Notice The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on neonatal through adult patients. It should only be operated by properly trained clinical personnel, under the direction of a physician.
Regulatory Notice Federal law restricts the sale of this device except by or on order of a physician. The benefit of treatment with medical respiratory support devices outweighs the remote possibility of exposure to phthalates.
L2786 Version T
ix
AVEA ventilator systems
Classification Type of Equipment:
Medical Equipment, Class 1 type B Adult/Pediatric/Infant Lung Ventilator
Declaration of Conformity Notice This medical equipment complies with the Medical Device Directive, 93/42/EEC, and the following Technical Standards, to which Conformity is declared: BS EN 60601-2-12:2006 EN 60601-1 EN 60601-1-1 EN 60601-1-2 ISO 13485 UL 60601-1 CAN / CSA C22.2 No. 601.1-M90 EU Notified Body: BSI (Notified Body No. 0086) Trade names: AVEA ventilator systems Manufactured by: CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887–4668 USA
If you have a question regarding the Declaration of Conformity for this product, please contact CareFusion at one of the numbers given in Appendix A.
L2786 Version T
x
Operator's manual
Safety Information Please review the following safety information prior to operating the ventilator. Attempting to operate the ventilator without fully understanding its features and functions may result in unsafe operating conditions. Warnings and Cautions, which are general to the use of the ventilator under all circumstances, are included in this section. Some Warnings and Cautions are also inserted within the manual where they are most meaningful. Notes are also located throughout the manual to provide additional information related to specific features. If you have a question regarding the installation, set up, operation, or maintenance of the ventilator, contact Customer Care, as shown in Appendix A Contact and Ordering Information.
Terms WARNINGS
identify conditions or practices that could result in serious adverse reactions or potential safety hazards.
CAUTIONS
identify conditions or practices that could result in damage to the ventilator or other equipment.
NOTES
identify supplemental information to help you better understand how the ventilator works.
Warnings Warnings and Cautions appear throughout this manual where they are relevant. The Warnings and Cautions listed here apply generally any time you operate the ventilator.
The AVEA Ventilator is intended for use by a trained practitioner, under the direction of a qualified physician.
When the ventilator is connected to a patient, a trained health care professional should be in attendance at all times to react to an alarm or other indications of a problem.
Alarm loudness must be set above ambient sound in order to be heard.
Always have an alternate means of ventilation available whenever the ventilator is in use.
The operator should not touch the electrical connectors of the ventilator or accessories, and the patient simultaneously.
Due to possible explosion hazard, the ventilator should not be used in the presence of flammable anesthetics.
An audible alarm indicates an anomalous condition and should never go unheeded.
Anti-static or electrically conductive hoses or tubing should not be used within the patient circuit.
If a mechanical or electrical problem is recognized while operating the ventilator, the ventilator must be removed from use and referred to qualified personnel for servicing. Using an inoperative ventilator may result in patient injury.
When a low gas supply alarm occurs, the oxygen concentration delivered to the patient will differ from that set on the O2 control setting.
A source gas failure will change the FIO2 and may result in patient injury.
The functioning of this equipment may be adversely affected by the operation of other equipment nearby, such as high frequency surgical (diathermy) equipment, defibrillators, short-wave therapy equipment, “walkie-talkies,” or cellular phones.
Water in the air supply can cause malfunction of this equipment.
L2786 Version T
AVEA ventilator systems
xi
Do not block or restrict the Oxygen bleed port located on the instrument back panel. Equipment malfunction may result.
Electric shock hazard - Do not remove any of the ventilator covers or panels. Refer all servicing to an authorized CareFusion service technician.
A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation. Upon loss of protective ground, all conductive parts including knobs and controls that may appear to be insulated can render an electric shock. To avoid electrical shock, plug the power cord into a properly wired receptacle, use only the power cord supplied with the ventilator, and make sure the power cord is in good condition.
The AVEA is designed to ensure that the user and patient are not exposed to excessive leakage current per applicable standards (UL 60601-1 and IEC60601-1). However, this cannot be guaranteed when external devices are attached to the ventilator. In order to prevent the risk of excessive enclosure leakage current from external equipment attached to the RS-232, printer and video ports, isolation of the protective earth paths must be provided to ensure proper connection. This isolation should ensure that the cable shields are isolated at the peripheral end of the cable. Only IEC 60950-1 or IEC 60601-1 certified external peripheral equipment shall be attached to the device.
Routine assessment of oxygenation and ventilation should be performed when a patient is receiving respiratory support.
Delivered and monitored flow as well as pressure and volume settings and values are subject to device accuracy specifications as described herein.
To avoid a potential safety hazard when operating two or more of the same or similar device in a single area, use the same audible alarm characteristics.
Cautions The following cautions apply any time you work with the ventilator.
Ensure that the voltage selection and installed fuses are set to match the voltage of the wall outlet, or damage may result.
A battery that is fully drained (i.e. void of any charge) may cause damage to the ventilator and should be replaced.
All accessory equipment that is connected to the ventilator should comply with CSA/IEC 60601-1/UL 60601-1.
To avoid damage to the equipment, clean the air filter regularly.
The following cautions apply when cleaning the ventilator or when sterilizing ventilator accessories.
Do not sterilize the ventilator. The internal components are not compatible with sterilization techniques.
Do not gas sterilize or steam autoclave tubing adapters or connectors in place. The tubing will, over time, take the shape of the adapter, causing poor connection and possible leaks.
DO NOT submerge the ventilator or pour cleaning liquids over or into the ventilator.
The patient-specific settings for the ventilator are stored in non-volatile memory every minute during use or each time a setting is changed. This allows indefinite recovery of the settings even in the event of total power loss. Exceptions would be any time the New Patient function is selected before ventilation or intentional clearing of the memory, such as installing new software.
L2786 Version T
xii
Operator's manual
When the recommended breathing system is in use, and normal ventilation is compromised, resulting in the opening of the safety valve, the inspiratory and expiratory gradient at the patient-connection port shall not be greater than shown below: 5 LPM
30 LPM
60 LPM
Inspiratory
0.7 cmH2O
2.1 cmH2O
5.0 cmH2O
Expiratory
0.7 cmH2O
3.8 cmH2O
5.3 cmH2O
The use of any attachments or accessories placed in the breathing systems (circuit) may increase the pressure gradient across the breathing system (resistance) for the patient.
Note: Maximum Circuit Pressure Limit: The ventilator has an independent mechanical pressure relief valve that limits the maximum pressure at the patient wye to 125 cmH2O. Safety Valve: Under certain conditions, the safety valve automatically opens, which gives the patient the ability to spontaneously breathe without having to draw a sub-ambient pressure to open the valve. Under this condition, the exhalation valve is also de-energized and allows one-way breathing through the circuit to prevent the re-breathing of exhaled gasses.
L2786 Version T
xiii
AVEA ventilator systems
Equipment Symbols
The following symbols may be referenced on the ventilator or in accompanying documentation Symbol
Source/Compliance ISO 7010-W001
General warning
ISO 7000-0434A
Caution
Symbol #5016 IEC 60417
This symbol indicates a FUSE.
Symbol #5034 IEC 60417 Symbol #01-36 IEC 60878 Symbol #5035 IEC 60417 Symbol #01-37 IEC 60878 Symbol #5031 IEC 60417 Symbol #5019 IEC 60417 Symbol #01-20 IEC 60878 Symbol #5021 IEC 60417 Symbol # 01-24 IEC 60878
Symbol # 5333 IEC 60417 Symbol #02-03 IEC 60878 Symbol #5032 IEC 60417
This symbol indicates INPUT. This symbol indicates OUTPUT This symbol indicates DIRECT CURRENT (DC) This symbol indicates protective EARTH (ground). This symbol indicates the EQUIPOTENTIAL connection used to connect various parts of the equipment or of a system to the same potential, not necessarily being the earth (ground) potential (e.g., for local bonding). This symbol indicates TYPE B equipment, which indicates equipment that provides a particular degree of protection against electric shock, particularly with regards to allowable leakage current and reliability of the protective earth connection.
Symbol #01-14 IEC 30878
This symbol is located on the rating plate. It indicates the equipment is suitable for alternating current.
Symbol #5007 IEC 60417 Symbol #01-01 IEC 60878
Indicates ON (Power)
Symbol #5008 IEC 60417 Symbol #01-02 IEC 60878
Indicates OFF (Power)
Symbol #0651 ISO 7000
Horizontal return with line feed. Indicates ACCEPT entered values for a specific field.
CareFusion Symbol
Indicates PATIENT EFFORT
CareFusion Respiratory Care symbol
Indicates MANUAL BREATH
ACCEPT
L2786 Version T
Meaning
xiv
Operator's manual
CANCEL
CareFusion Symbol
MAIN SCREEN
Symbol #417 IEC 5102
EVENT READY
CareFusion Symbol
MODE
CareFusion Symbol
ADVANCED SETTINGS
CareFusion Symbol
SET-UP for patient size selection
MDD Directive 93/42/EEC
CE Mark
Symbol #5307 IEC 60417
ALARM RESET
Symbol #5319 IEC 60417
ALARM SILENCE
CareFusion symbol
ADULT patient
CareFusion symbol
PEDIATRIC patient
CareFusion symbol
NEONATAL (Infant) patient
Graphical Symbol in general use internationally for “DO NOT”
CANCEL, i.e. do not accept entered values.
CareFusion symbol
Select DISPLAYED SCREEN function.
Symbol 5467 IEC 60417
FREEZE the current display.
L2786 Version T
xv
AVEA ventilator systems
L2786 Version T
CareFusion symbol
Enable the ALARM LIMITS screen
CareFusion symbol
This symbol indicates a CONTROL LOCK.
CareFusion symbol
NEBULIZER port
CareFusion symbol
Increase OXYGEN
CareFusion symbol
PRINT SCREEN
CareFusion symbol
SUCTION port
CareFusion symbol
VARIABLE ORIFICE FLOW SENSOR connection
CareFusion symbol
HOT WIRE FLOW SENSOR connection
CareFusion symbol
ANALOG IN/OUT connection
CareFusion symbol
Display the MAIN SCREEN
CareFusion symbol
DO NOT BLOCK PORT
CareFusion symbol
EXTERNAL BATTERY connection
CareFusion symbol
Indicates GAS ID port
CareFusion symbol
OXYGEN SENSOR connection
CareFusion symbol
OVERPRESSURE relief
xvi
Operator's manual
CareFusion symbol
REMOTE NURSE CALL connection
CareFusion symbol
USER INTERFACE MONITOR connection
CareFusion symbol
This symbol indicates an INTERNAL BATTERY FUSE
CareFusion symbol
This symbol indicates ALARM LOUDNESS
CareFusion symbol
This symbol indicates that the AVEA is being powered by the INTERNAL BATTERY only.
CareFusion symbol
This symbol indicates that the HELIOX configuration is in use.
CareFusion symbol
This symbol indicates the product contains phthalates.
Symbol #0434 IEC TR60878
Indicates ATTENTION or CAUTION
Manufacturer
Indicates the medical device manufacturer, as defined in EU Directives 90/385/EEC, 93/42/EEC and 98/79/EC
Authorized representative in the European Community
Indicates the authorized representative in the European Community
L2786 Version T
1
AVEA ventilator systems
Chapter 1: Introduction The AVEA is a fourth generation, servo-controlled, software-driven ventilator. It has a dynamic range of breathing gas delivery that provides for neonatal through adult patients. Its revolutionary user interface module (UIM) provides maximum flexibility with simple operator interaction. It has a flat panel color LCD with real time graphic displays and digital monitoring capabilities, a touch screen for easy interaction, membrane keys and a dial for changing settings and operating parameters. A precision gas delivery engine with servo controlled active inhalation and exhalation improves performance over previous generations. The AVEA has been designed to function using most commonly available accessories. It is easy to clean and its design does not allow liquids to pool on the casing, reducing the likelihood of fluid leakage into the body of the ventilator. There are two models of AVEA: Comprehensive and Standard. The following table shows the standard and optional functions available with each model.
Functions and Accessories Modes Proximal Hot Wire Flow Sensing Synchronized Nebulizer 24 Hour Trending Internal Battery Full Color Graphics Display Loops and Waveforms Standard Cart Proximal Variable Orifice flow sensing Proximal Airway Pressure Monitoring Tracheal Monitoring Tube Esophageal Balloon Internal Compressor Heliox Delivery
L2786 Version T
Standard
Comprehensive
All
All
2
Chapter 1: Introduction
Optional Functions and Accessories
Operator's manual
Standard
Comprehensive
Custom Cart
Option
Included
External Battery (on custom cart only)
Option
Option
Gas Tank Holder (on either cart)
Option
Option
Internal Compressor
Option
Included
Pflex Maneuver
Option
Included
Heliox Delivery
Option
Included
nIMV/ Volume Guarantee
Option
Included
Volumetric Capnography
Option
Option
Some AVEA Features Artificial Airway Compensation1 When Artificial Airway Compensation is turned on, the ventilator automatically calculates the pressure drop across the endotracheal tube. The AVEA then adjusts the airway pressure to deliver the set inspiratory pressure to the distal (carina) end of the endotracheal tube. This calculation takes into account flow, gas composition (Heliox or Nitrogen/Oxygen), Fraction of Inspired Oxygen (FIO2), tube diameter, length, and pharyngeal curvature based on patient size (Neonatal, Pediatric, Adult). This compensation only occurs during inspiration. Artificial Airway Compensation is active in all Pressure Support and Flow Cycled Pressure Control Breaths.
WARNING Activating of Artificial Airway Compensation while ventilating a patient will cause a sudden increase in the peak airway pressures and a resultant increase in tidal volume. If you choose to activate Artificial Airway Compensation while the patient is attached to the ventilator you will need to exercise caution to minimize the risk of excessive tidal volume delivery.
Note: Monitored airway pressures (inspiratory) will be higher than set values when Artificial Airway Compensation is active. With an inspiratory pressure setting of zero, Artificial Airway Compensation will still provide an elevated airway pressure, which will compensate for the resistance of the endotracheal tube. When turned on the Artificial Airway Compensation indicator will appear in all modes of ventilation even though the function may not be active (i.e.: Volume Controlled Breaths). This is to alert you to the fact that Artificial Airway Compensation will become active if a Pressure Support or combination mode (e.g. Volume Control SIMV) is selected. Range:
Off/On
Default:
Off
Available in all patient sizes
1
Estimation of Inspiratory Pressure Drop in Neonatal and Pediatric Endotracheal Tubes, by Perre-Henri Jarreau, American Physiological Society 1999
L2786 Version T
AVEA ventilator systems
Chapter 1: Introduction
3
Full range of Patient Size You can select a patient size of Adult, Pediatric, or Neonate. Once the selection is made, the ventilator offers only those parameters, which are available for your selected patient size. For a complete listing of compatible circuits and accessories, see “Appendix E: Sensor and Circuit Specifications” on page 217.
Non-Invasive Ventilation The ventilator can perform non-invasive ventilation with a standard dual limb circuit. Leak compensation should be turned on when using this feature. To turn leak compensation on, use the touch screen control displayed in the Ventilator Set-Up Screen.
Note: Noninvasive ventilation requires the use of a snug fitting mask with no bleed holes. Excessive leaks around the mask may result in false triggering of the ventilator or assertion of disconnect alarms.
Leak Compensation Leak Compensation is used to compensate for baseline leaks, which may occur at the patient mask interface or around the patient’s endotracheal tube. It only provides baseline leak compensation and is not active during breath delivery. During exhalation, PEEP is maintained by the cooperation of the Flow Control Valve (FCV) and the Exhalation Valve (ExV). The ExV pressure servo is set to a target pressure of PEEP and the FCV pressure servo is set to a pressure target of PEEP - 0.4 cmH2O. The ExV servo relieves when the pressure is above its target and the FCV supplies flow when the pressure drops below its target up to a maximum flow rate for the patient size Range:
Off/On
Default:
Off
Circuit Compliance Compensation When Circuit Compliance is active, the volume of gas delivered during a volume controlled or targeted breath is increased to include the set volume, plus the volume lost due to the compliance effect of the circuit. Circuit Compliance is active for the set Tidal Volume during volume control ventilation, the Target Tidal Volume in PRVC mode and for Machine Volume. It is only active in Adult and Pediatric applications. Exhaled volume monitors for all modes and breath types are also adjusted for the compliance compensation volume. Range:
0.0 to 7.5 ml/cmH2O
Default:
0.0 ml/cmH2O
The ventilator automatically measures Circuit Compliance during the Extended Systems Test (EST). The value cannot be entered manually.
L2786 Version T
4
Chapter 1: Introduction
Operator's manual
Note: Although circuit compliance is displayed on the Setup screen it is not active for neonatal patients. High circuit compliance with small tidal volumes may result in extended inspiratory times. This is a result of the delivery of the circuit compliance volume at the set flow rate. Setting extremely small delivered tidal volumes with Circuit Compliance Compensation not active and using a proximal flow sensor may result in assertion of Patient Circuit Disconnect Alarms.
Humidification You can select active or passive humidification (ON/active or OFF/passive). Active humidification assumes 99% RH; passive assumes 60% RH when using an HME. This feature adjusts the BTPS correction factor to correct exhaled tidal volumes. Range: Default:
Off/On Active (ON)
Note: Incorrect setting of the Humidification feature will affect monitored exhaled volume accuracy.
Heliox Delivery (Comprehensive only, option on Standard) Using patented “Smart” connector technology, the Comprehensive model AVEA can deliver Heliox blended gas instead of Medical air. By simply changing a connector on the back panel, the ventilator identifies the gas input and adjusts to accommodate the change. All volumes (numeric and graphic) are automatically compensated for accurate display. The clinical benefits of helium / oxygen gas are based on its significantly lower gas density when compared to nitrogen / oxygen gas. This lower gas density allows the same volumetric (tidal volume) of gas to be delivered to the patient at a significantly lower airway pressure. Additionally, the low-density properties of the gas allow it to diffuse past airway obstructions or restrictions much easier than nitrogen / oxygen gas mixtures.
Note: The Heliox “smart” connector is designed for use with an 80/20 Heliox tank only. Only a mixture of 20% oxygen and 80% Helium can be used as the Heliox gas supply. If Heliox gas is connected this green icon displays in bottom right of the touch screen. To set the Helium / Oxygen mixture during administration simply set the desired FIO2, the balance of the breathing gas is Helium. For example: A set FIO2 of 35% will deliver a 65/35 Heliox mixture to the patient.
L2786 Version T