Endosafe nexgen-pts Assay Guide with User Management

Guide

2 Pages

the endosafe  ®  nexgen-pts  ™  01  Turn on the reader by pressing the button at the rear righthand side of the back panel. As the instrument boots up, the Charles River logo will display on the touch screen.  02  At the Enter Operator ID prompt, enter your username and click Next.  03  At the Enter Operator Password prompt, enter your password and click Next.  ASSAY GUIDE WITH USER MANAGEMENT  04  As the reader warms in preparation for use, the Initializing message appears. When the unit has reached 37.0 °C, the screen will display the Insert Cartridge message.  05  ‌‌ Allow cartridges to reach room temperature in pouch before using. Remove cartridge from pouch and avoid touching the sample wells.  06  ‌‌ Insert the cartridge firmly into the slot at the front of the reader with sample wells facing up.  07  ‌‌ The reader will then verify cartridge integrity; used or damaged cartridges will signal an alert.  Bubbles  Splash out  08  Prior to testing, be sure you have access to the Certificate of Analysis (COA), which can be found within your Endosafe®-PTS™ cartridge package or on the Endosafe® customer portal (www.criver.com/ endosafeportal). This information includes the cartridge lot number and calibration code.  09  ‌ series of A prompts will appear for you to enter the information necessary to run the assay, including Cartridge Lot #, Cartridge CAL Code, Sample ID, Sample Lot, Dilution and Endotoxin Limit. Enter values for each field, clicking the forward arrow until all fields are complete.  10  Once the necessary values are entered, confirm all inputted values on the next screen. The reader will display a prompt to Add Sample to Cartridge.  11  Using the included pippetor, dispense exactly 25 µL of sample into all 4 wells of the cartridge. Be sure to hold the pipette so that the tip is at an angle, and not completely at the bottom of the sample reservoir.* Click the Done – Start Test icon on the home screen to initiate testing.  12  Test results will appear on screen when the assay is complete. You can send this data to a network printer, Zebra printer or download to a computer via USB.  *IMPR OPER TECHNIQU E  13  Holding the pipette straight up and down at the bottom of the sample well can force some sample into the channel early and disrupt the reagent mixing stations. Avoid any technique which is prone to splash out or bubbles.  14  ‌ plash out S will render the volume inaccurate, which can lead to invalid test results. Bubbles can cause false onset times due to the disturbance of light transmittance and can interfere with reconstitution of reagents within the cartridge.
File Type: PDF
File Size: 901 KB
File Name: Charles River - Endosafe nexgen-pts Assay Guide with User Management.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions