Blanketrol III Quick Reference Guide To turn on the Blanketrol II, depress power switch to “ON” position.
MANUAL CONTROL: Control of Water Temperature 1.
Set WATER temperature by depressing “Temp Set” button, then depress either the “Up” or “Down” arrow to raise or lower the SETPOINT temperature.
2.
Depress the “MANUAL CONTROL” button
AUTOMATIC CONTROL: Control of Patient Temperature 1.
Insert or attach 400 series probe in or onto patient.
2.
Plug probe into probe jack on side of unit. While in Auto, Gradient or Smart Mode, probe must be in place for unit to function.
3.
Set desired PATIENT temperature by depressing “Temp Set” button, then depress either the “Up” or “Down” arrow to raise or lower the SETPOINT.
4.
Press desired “AUTOMATIC CONTROL” buttons.
GRADIENT 10C MODE: 1.
Follow the Automatic Control steps above, but do not press “Auto Control”.
2.
Instead depress the “GRADIENT 10C” button.
*This feature will create a maximum of 10C difference between the patient’s current temperature and the current water temperature.
page 1
Medical Division Phone 513-772-8810 Toll Free: 800-989-7373 Fax: 513-772-9119 www.cszmedical.com
Blanketrol III Quick Reference Guide GRADIENT VARIABLE MODE: 1.
Follow the Automatic Control steps above, but do not press “Auto Control”.
2.
Instead depress the “GRADIENT VARIABLE” button.
3.
Set the desired offset between the Patient Temperature and the Water Temperature.
4.
Depress the “GRADIENT VARIABLE” button again to activate mode.
*This feature will create a difference between the patient’s current temperature and the current water temperature, the maximum difference will be the variable that you selected.
SMART MODE: 1.
Follow the Automatic Control steps above, but do not press “Auto Control”.
2.
Instead select either the “GRADIENT 10C” button or the “GRADIENT VARIABLE” button.
3.
Once a desired Gradient has been selected then depress the “SMART” button.
*This feature will compare the patient’s temperature to the target temperature every 30 minutes. If the temperatures are not the same then the Blanketrol will adjust the water + or - by 5.0˚C. Once the patient temperature and the target temperature are the same the SMART Mode will shut off and the Blanketrol will default to the orignial Gradient setting.
MONITOR ONLY: Patient Temperature Monitor 1.
Must be used with patient temperature probe to function.
2.
Depress the “MONITOR ONLY” button.
* While in the “MONITOR ONLY” mode, the unit does not circulate water.