InterVac Analogue Intermittent Low Suction Pump SUC 82643 220/240V 50Hz
User Manual
Manual No. SUC 91005 448 Issue 5
Safety Thank you for purchasing this Clements Analogue Intermittent Low Suction Pump.
InterVac
For your safety it is imperative that this unit be operated by authorised personnel only, in accordance with the instructions as described in this manual. Operated in this way, the Clements InterVac Analogue Intermittent Low Suction Pump will provide years of service. Due to continual improvements in product design, the Clements Analogue Intermittent Low Suction Pump may vary in detail from the descriptions in this manual. In the event of further questions please contact your local distributor or Clements direct.
User Manual Clements InterVac Analogue Intermittent Low Suction Pump Manual Number SUC 91005 448 Issue 5 Copyright © 1997-1999 by Clements Medical Equipment Pty Ltd The information in this manual was originated by, and is the exclusive property of Clements Medical Equipment Pty Ltd. It is furnished for customer information only, and is not an authorisation or licence to make this product or to furnish this information to others.
Clements Medical Equipment Pty Ltd 5 Cal Close, Somersby, NSW 2250 Tel: + 61 2 4340 2484 Fax: + 61 2 4340 5121 Email: sales@clements.net.au Website: www.clements.net.au
CLEMENTS
Contents
Specifications ...
4
Installation and Operation ...
5
Contents ...
6
Accessories and Spares ...
10
Maintenance ...
12
Troubleshooting ...
13
Disassembly and Repair ...
14
Warranty ...
17
Contents
3
CLEMENTS
Specifications Classification
Low Vacuum, Low Flow (per AS2120.3 or ISO 10079-1)
Power Supply
220-240 Volts 50 Hz
Power Consumption
36 watts
Gauge
Diaphragm Type with Dual Scale Calibrated in mm Hg and kPa
Scale Range
0 to -200 mm Hg and 0 to -25 kPa
Intermittent Cycle
2 Minutes On, 1 Minute Off
Regulator
Vacuum Regulating, Needle Valve
Flow Rate
5 Litres Free Air Per Minute
Vacuum
-20 kPa /-150 mm Hg Maximum
Collection Jars
Two Shatterproof Plastic 2 litre Jars with Calibration Marks at 100 ml Spacing
Dimensions
550 mm Wide by 270 mm High by 230 Deep
Weight
5 kg (Without Collection Jars)
4
Specifications
CLEMENTS
Installation and Operation The Clements InterVac Analogue Intermittent Low Suction Pump provides a time cycled operation which automatically switches the suction supply on and off, to help prevent blockage of the catheter and possible harm to the patient. The suction level is fully adjustable and regulated to give accurate and precise control of the strength of the suction applied to the patient.
8
7 6
5 4 3
2 9
1
10
Main Components of Unit 1
Main Control Unit
6
Bacteria Filter
2
Collection Jar
7
Main Unit Inlet
3
Collection Jar Outlet
8
Connecting Tubes
4
Collection Jar Inlet
9
Side Holder
5
Bung
10
Overflow Cut Off Valve
Installation and Operation
5
CLEMENTS
1 5
2 6 3 4
Unit Controls
1
Suction Gauge
4
Continuous / Intermittent Switch
2
Power On Indicator
5
Main Control Switch (rear)
3
Suction Control Knob
6
Intermittent Mode LED
Contents
6
Tubing Yellow
2.95m
Jar 2 Lt
2
Clip Jar “Fitted”
2
Tubing Yellow
2
Bung
2
Manual
1
Tech Info Sheet
1
Installation and Operation
CLEMENTS
Connecting The Unit Because the unit is fitted with two collection jars it can either be connected as a changeover arrangement or a series arrangement. With the changeover arrangement only one jar is in use at a time, with the second jar on standby. As one jar is filled it is disconnected, so it can be emptied, and the other jar is connected. The series arrangement allows both jars to be in use at the same time, thus providing longer periods between changeover.
To Patient
Overflow Cut Off Valve in Both Jars
Changeover Arrangement To connect the unit as a changeover arrangement, use the following procedure. 1 Place two clean jars in the side holders, ensuring that both jars are fitted with an overflow cut off valve. 2 Connect the suction outlet of the jar to be used first to the suction inlet of the main unit. 3 Connect the suction inlet of the jar to be used first to the tube leading to the patient. 4 When the first jar is full, disconnect it and use steps 2 and 3 to connect the second jar.
Installation and Operation
7
CLEMENTS To Patient
No Overflow Cut Off Valve in this Jar
Series Arrangement
To connect the unit as a series arrangement, use the following procedure. 1 Place two clean jars in the side holders, ensuring that only one jar is fitted with an overflow cut off valve. 2 Connect the suction outlet of the jar with the overflow cut off valve to the suction inlet of the main unit. 3 Connect the suction inlet of the jar with the overflow cut off valve to the suction outlet of the jar without the overflow cut off valve. 4 Connect the suction inlet of the jar without the overflow cut off valve to the tube leading to the patient.
8
Installation and Operation
CLEMENTS
Operation The Clements InterVac Analogue Intermittent Low Suction Pump has two operating modes. The pump can be operated in Continuous mode when suction is required all of the time or in Intermittent mode when the suction needs to be cycled on and off to minimise catheter blockage and injury to the patient. To operate the pump, follow the steps listed below.
1 Install the collection jars and tubing for either changeover or series arrangement as previously described. 2 Connect the pump to a 240 volt power outlet via the mains lead and 3 pin plug fitted to the pump. 3 Turn the power on at the power outlet and ensure that the Power On lamp is lit. 4 Using the Continuous / Intermittent mode switch select Continuous mode . 5 Occlude the suction tubing and adjust the regulator by turning the knob to obtain the required suction level, as indicated on the suction gauge. 6 Check that there are no leaks in the tubing and fittings. 7 Connect the catheter to the patient. 8 If intermittent mode is required, push the main control switch to the intermittent mode position and ensure that the intermittent mode light comes on. 9 Check the collection jars to ensure that suction is being applied correctly.
Installation and Operation
9
CLEMENTS
Accessories and Spares SUC 80297
Yellow Medical Suction Tubing
SUC 80301
Two (2) Litre Collection Jar, (Jar only)
SUC 80312
Two (2) Litre Collection Jar Complete. (Jar, Self Sealing Bung, Overflow Cut Off Valve)
SUC 80317
Plastic Clip for Attaching 2 Litre Collection Jar to 'V' Bracket
SUC 82640 002
Complete 2 Litre Jar Assembly Accessory Pack for Intermittent Pumps, (2 x 2 Litre Jar Assy, Suction Tubing, V Clips, Filters)
SUC 83201
Elevated Trolley with Collection Jar, Bung and Overflow Cut Off Valve
SUC 83203
Elevated Trolley (Trolley Only)
SUC 81000 081
Metal Ring Bracket for Attaching 1 Litre Collection Jar to 'V' Bracket
SUC 81003
1 Litre Collection Jar for Portable High Suction Pumps (Jar Only)
SUC 81004
Bung Assembly for 1 Litre Collection Jar
SUC 81009
1 Litre Collection Jar with Bung and Overflow Cut Off Valve for Portable High Suction Pumps
SUC 84200 519
Blue Male Nipple for Bacteria Filter (Pack of 10)
SUC 894200 520
Blue Female to Male Nipple for Bacteria Filter (Pack of 10)
SUC 84100 162
Blue Bacteria Filter (Set of 12)
SUC 84200 033
Two (2) Litre Jar Bung Assembly Complete with Cut Off Valve
SUC 89455
Plastic 'V' Bracket to Hold Suction Devices
10
Accessories and Spares
CLEMENTS
SUC 91005 448
User Manual for the Analogue Intermittent Suction Pump (240V 50Hz)
SUC 91005 454
Technical Information for the 2 Litre Collection Jar with Self-Sealing Bung and Overflow Cut Off Valve
SUC 91005 611
Technical Information for the 1 Litre Collection Jar with Self-Sealing Bung and Overflow Cut Off Valve
Spare Parts MUL 91000 576
Silicone Suction Tubing (6.4mm ID x 12.7mm OD)
MUL 91600 101
Black Rubber Feet for Machined Base Plates, (Set of 4)
MUL 94003 895
Neon Indicator Green 240V for Intermittent Suction Pumps
SUC 80253
Vacuum Gauge for Low Vacuum Products, Bottom Entry, Vacuum Range 0 to -25kPa
SUC 80285
Silicone Suction Tubing (5mm ID x 9mm OD)
SUC 82610 078
Electric Pump for Low Suction Pumps (240V 50Hz)
SUC 82660 055
Power and Timing Printed Circuit Board Assembly for Analogue Intermittent Suction Pump
SUC 89240 069
Regulator Knob Replacement Kit for Suction Pumps (Regulator Knob, Screw)
SUC 89107
Kit for Regulators (Needle Type)
SUC 94004 003
Rocker Switch On/Off Black Accessories and Spares
11
CLEMENTS
Maintenance To keep your Clements InterVac Analogue Intermittent Low Suction Pump in good working order and condition, the following maintenance is recommended.
After Each Use Clean and sterilise the collection jars, bungs, tubing and tubing nipples. These items may be safely autoclaved. Wipe the main unit with a damp cloth or sponge. Do not immerse the main unit or allow water to enter the cabinet.
After Each 100 Hours or 2 Months Replace the blue bacteria filter. Check all tubing and replace it if it is perished, soft or discoloured. Also check the tubing nipples. Check the seal rings on the bungs and replace them if they are hard, cracked or perished. Also check the fit of the seal in the collection jar. Check the float valve seals and replace them if they are perished or damaged.
The above maintenance should also be carried out immediately after use with infectious patients. Do not use phenolic solutions as disinfecting agents on the collection jars as they can cause the polycarbonate material to melt. The use of strong chemical agents on plastic parts will shorten their effective service life.
12
Maintenance
CLEMENTS
Troubleshooting Fault
Check
Rectify
No Suction but Motor is Working
Tubing Blocked Float Valve in Off Position Inlet Blocked Gauge Blocked Regulator Blocked Bung Not Sealing
Clear or Replace Clear, Replace or Empty Jar Clear Replace Gauge Clear or Replace Adjust or Fit New Seal Adjust Setting
Regulator at Zero No Suction Motor not Working
Suction On Continuously
Mains Fuse Blown Faulty Mains Power Lead Power Not Turned On Unit Fuses Blown
Replace Replace Turn On Replace
Unit Not Switched to Switch to Intermittent Mode Intermittent Mode Internal Connections and Tighten Connections PCB or Replace PCB
Not Enough Suction Regulator Setting Too Low Leaks In Tubing or Fittings Blockage In Tubing or Fittings
Adjust To Correct Setting Tighten Connections Clear or Replace
Excessive Noise Pump Worn Coming From Pump Pump Incorrectly Installed
Replace Pump Install Correctly
Troubleshooting 13
CLEMENTS
Disassembly and Repair Opening The Pump Casing To open the pump case in order to carry out servicing checks use the following procedure. 1 Remove the collection jars from the side holders. 2 Turn the unit over and remove the four screws in the base - two along the front edge and two along the rear edge. 3 Turn the unit upright again and remove the two screws located in the sides of the unit near the top edge. 4 Remove the top from the base taking care to prevent dislodging any electrical or tubing connections.
After completion of the service checks, reverse the above procedure to re-assemble the pump.
Filter Replacement The Clements InterVac Analogue Intermittent Low Suction Pump is fitted with a filter. On the outside of the unit, the blue bacteria filter is fitted to the pump inlet. To replace this filter, simply disconnect any tubing from the inlet, pull the filter out of the inlet and push a new filter into place. Replace the tubing.
14
Disassembly and Repair
CLEMENTS
Disassembly and Repair
15
CLEMENTS
Fuse Replacement The pump is fitted with two cartridge fuses located in a fuse holder that is accessible from outside the unit. To replace the fuses use the following procedure. 1 2 3 4
5
16
Turn off the power at the wall outlet and disconnect the power lead from the IEC power inlet. Turn the pump around and locate the fuse holder on the IEC power inlet. Release the fuse holder drawer. Remove the old fuses and inspect them. If they are blown, replace them with new fuses. Replace the fuse holder drawer. Plug the pump into the power socket, turn it on and ensure that it is operating correctly.
Disassembly and Repair
CLEMENTS
Warranty Clements Medical Equipment Pty Limited ("Clements") warrants that this product is free from defects in workmanship and materials for a period of 12 months from the date of shipment by Clements or its authorised agent to the purchaser. Subject to the conditions of this warranty, if the product fails to operate for any reason within the warranty period and the product is returned to the place of purchase at the purchaser's expense, Clements will repair or replace the product free of charge. If a valid warranty claim is made within 30 days from the date of shipment, then Clements will also reimburse the purchaser for reasonable freight costs in returning the product to the place of purchase.
Conditions of Warranty 1.
The product must be returned to the place of purchase with proof of purchase.
2.
This warranty is only available to the original purchaser of the product.
3.
The product must not have had its serial number removed, defaced or changed, its casing opened, its power supply altered or have been tampered with in any other way.
4.
This warranty does not cover : • inadequate or incorrect site preparation; • improper installation; • connection to the wrong voltage; • failure of the product due to misuse; • the use or operation of the product outside of the physical, electrical or environmental specifications of the product; • use in a manner or environment in which the product is not designed to be used; • improper adjustment, calibration or operation by the purchaser; • the use of accessories including consumables, hardware or software which were not manufactured or approved in writing by Clements; • any modifications of the product which were not authorised in writing by Clements;
Warranty
17
CLEMENTS • any contamination or leakages caused or induced by the purchaser; and • inadequate or improper maintenance of the product. 5.
This warranty does not cover normal wear and tear.
6.
Clements will not be responsible for damage or loss caused during shipping.
7.
In Australia, apart from any warranties implied by the Trade Practices Act 1974 all other warranties expressed or implied and whether arising by virtue of statute or otherwise are hereby excluded.
8.
Outside Australia, all other warranties expressed or implied and whether arising by virtue of statute or otherwise (including any warranties implied by the Vienna Convention) are hereby excluded.
9.
Clements' obligations under this warranty are limited to the repair or replacement of the product, within the terms of this warranty and the total liability of Clements for loss or damage of every kind whether arising pursuant to the terms of the sale of the product or otherwise in connection with the product is limited to the amount paid by the purchaser to Clements for the product.
10. Apart from any liability imposed by Part VA of the Trade Practices Act, Clements accepts no other liability for any loss or damage occasioned (including consequential loss or damages) in any way as a result of the use of the product. 11. The warranty does not extend to cover damage to the following parts as they are inherently prone to wear : • motor brushes 12. This warranty does not extend to cover corrosion due to any cause nor to any damage to painted or anodised surfaces. 13. Clements will give the purchaser the benefit of any manufacturer's warranty in respect of any components in the product which were not manufactured by Clements, if such a manufacturer's warranty is available.
18
Warranty
CLEMENTS
19
Clements is a wholly-owned Australian company supplying a wide range of quality products to the medical, scientific and wellness industries. Clements medical products embrace a range of : • portable suction pumps; • portable emergency suction pumps; and • suction and gas flow devices to complement and control supply systems. Clements scientific products include a comprehensive range of bench top centrifuges and mixers. Clements wellness products include a range of chiropractic adjusting tables and chiropractic accessories. Clements also makes a line of blood storage refrigerators to serve both the medical and scientific industries. Clements has the ability and the capacity to design and manufacture related products to satisfy specific customer requirements. Contact your local distributor as detailed on www.clements.net.au or Clements direct Clements Medical Equipment Pty Ltd 5 Cal Close, Somersby, NSW 2250 Tel: + 61 2 4340 2484 Fax: + 61 2 4340 5121 Email: sales@clements.net.au Website: www.clements.net.au