CLEMENTS
SUC 89540 093 Direct Suction Nozzles User Manual Issue 5
User Manual
11 Pages
Preview
Page 1
Manual No. SUC 89540 093 Issue 5
3 # !
2 Vacuum Range
0 to -100 kPa, 0 to -750 mmHg (Subject to available suction)
Flow Rate
0 to 35 Litres of Free Air Per Minute
Gauge
Bourdon Tube type Dual scale, graduated in kPa and mmHg
Gauge Range
0 to -100 kPa, graduated at 5 kPa 0 to -760 mmHg graduated at 20 mmHg
Control
Lever styled tap O-Off 906rotation
Australian Standard
AS 2103.3 - 1992 Gauge
No Gauge
Height
210 mm
135mm
Width
70 mm
60 mm
Depth
50 mm
50 mm
Weight
325 g
220 g
Duty Cycle
Continuous
Ambient Temperature
+5°C to 35° C
Standard Conditions
25°C, Sea Level, 101 kPa
GMDN Code
14325
ARTG Number )
141960 2
2
7
SUC 89771
Ring Index Connector Kit for Wall Suction Products ( Hand wheel, 1/8 BSP Threaded Nipple, O-Ring)
SUC 89240 019
Puritan Bennett Connector with 1/8 BSP Thread
SUC 89240 074
BS MKIV Connector with 1/8 BSP Thread
SUC 89250 006
Drager Connector with 1/8 BSP Thread
SUC 80970 001
Plastic Hose Tail with O-Ring (Pack of 20)
SUC 89310 001
Porous Bronze Filter (Pack of 20)
SUC 89660
Duplex Suction Adaptor with Lever
SUC 89661
Duplex Suction Adaptor Without Lever
SUC 80971
Metal Suction Nozzle Assembly with O-Ring
SUC 89540 093
User Manual for Direct Suction Nozzles
'
% 9
!$
'
% 2
! !
-
5
+
1
8
9
No Vacuum / Flow
On / Off Lever
Ensure turned on
Wall outlet blocked
Refer to engineering
Body blocked
Strip and clean
Filter blocked
Clean or replace
Gauge faulty
Replace
Gauge Faulty
Check against master gauge
Replace
Bronze Filter
Damage or corrosion
Autoclave, wash
-