Codman
INTEGRA Codman Electrosurgical Generator Instruction Manual REF901001ESUO
Instruction Manual
269 Pages
Preview
Page 1
Codman®
Electrosurgical Generator Instruction Manual 901001ESUO EN – ENGLISH... 3 DA – DANSK... 27 NL – NEDERLANDS... 49 FI – SUOMI... 71 FR – FRANÇAIS... 93 DE – DEUTSCH... 115 EL – ΕΛΛΗΝΙΚΑ... 137 IT – ITALIANO... 159 NO – NORSK... 181 PT (EU) – PORTUGUÊS... 203 ES – ESPAÑOL... 225 SV – SVENSKA... 247
THIS PAGE IS INTENTIONALLY LEFT BLANK.
2
EN - ENGLISH
Codman®
Electrosurgical Generator Instruction Manual
SYSTEM OVERVIEW Important Information: Please Read Before Use Codman® Electrosurgical Generator Description The Codman® Electrosurgical Generator (the Generator) is a unit designed for use with monopolar and bipolar instruments to cut and coagulate tissue during surgical procedures. The Generator provides an audible indication of the operating mode each time cutting or coagulation is performed. The Generator can be used with an electrosurgical irrigator, such as the Codman® Electrosurgical Irrigator, and an interconnecting cable, such as the Codman® Electrosurgical Interconnecting Cable, where irrigation is available to the user while performing electrosurgical procedures. Indications The Generator is indicated for use for cutting and coagulation of soft tissue in neurosurgical procedures. Contraindications There are no known contraindications relating to safe application of the Generator. WARNINGS • Read entire instruction manual before attempting to operate the Generator. • Use of monopolar operating modes should be restricted to extracranial cutting/coagulation during neurosurgical procedures (i.e., the opening and closing of neurosurgical cases) and to tissues contained in the extradural compartment. The monopolar operating modes should NOT be used in the cutting/coagulation of tissues in the subdural compartment or the neural parenchyma. • Do not remove the Generator cover. There are no user serviceable components in this system. • Replacement fuses must meet the specifications outlined in this manual. See Technical Specifications. • This Generator must not be modified in any way. Unauthorized modifications could cause the Generator to malfunction or fail and will void the warranty. • The operator of the Generator must be a healthcare professional. The operator must be experienced in the use of electrosurgical devices and generators in surgical procedures. • Failure of the Generator could result in unintentional increase of output power. • The Generator performs a self-test upon initial start-up. An indicator error in any of the self-test sequence indicates a device defect. See the Error Diagnosis/Troubleshooting section to determine if the error can be resolved. If not, discontinue using the device and contact your Integra sales representative. 3
EN - ENGLISH
SYSTEM OVERVIEW Electrical Shock Hazard • To avoid risk of electric shock, the Generator must only be connected to supply mains with protective ground. • Disconnect the power cord to isolate the Generator from the supply mains. Serious shock hazards may be present even with the power switch off. • Do not bypass the grounding prong on the Generator by using an adapter. The Generator must be properly grounded to ensure user and patient safety. Grounding reliability can only be achieved when connected to a receptacle marked “Hospital Only” or “Hospital Grade.” • The output from the Generator must not be applied directly to the heart. • Do not use in patients who have electronic implants such as cardiac pacemakers without first consulting a qualified professional (e.g., cardiologist). A possible hazard exists because interference with the action of the electronic implant may occur, or the implant may be damaged. • Do not connect multiple portable socket outlets (MPSO) or extension cords to the Generator. • Connect adapters and accessories to the Generator only when the energy is off. Failure to do so may result in an injury or electrical shock to the patient or operating room personnel. • Never simultaneously touch the patient and the external fuse. • Inspect all cords, monopolar and bipolar accessory devices, and Foot Pedal for proper function and intact insulation prior to each use. • Use in typical commercial or hospital environment, except in RF (radiofrequency) shielded room for magnetic resonance imaging, where the intensity of EM (electromagnetic) disturbances is high. • Avoid HF (high-frequency) output settings where maximum output voltage may exceed rated accessory voltage. • Activating the RF current is not permitted while inserting the electrode and during electrode replacement due to risk of burns. • Do not use in the presence of flammable anesthetics or oxidizing gases (such as nitrous oxide (N2O) and oxygen) or in close proximity to volatile solvents (such as ether or alcohol), as explosion may occur. • Do not immerse the Generator in any type of liquid. • Non-flammable agents should be used for cleaning and disinfection whenever possible. • Flammable agents, including disinfectants, cleaning fluids, and solvents, must fully evaporate prior to HF surgical procedures. There is a risk of pooling of flammable solutions under the patient or in body depressions such as the umbilicus, and in body cavities such as the vagina. Any fluid pooled in these areas should be mopped up before HF surgical equipment is used. Attention should be called to the danger of ignition of endogenous gases. Some materials, for example cotton and gauze, when saturated with oxygen may be ignited by sparks produced in normal use of the HF surgical equipment. • Unless a compatible monitoring neutral electrode is used with a contact quality monitor, loss of safe contact between the neutral electrode and the patient will not result in an auditory alarm. • Other equipment may be adversely influenced by the interference produced by the operation of HF surgical equipment. Portable and mobile RF communications equipment can affect medical electrical equipment. • Prevent patient contact with metal parts that are grounded or that have an appreciable capacitance to ground; e.g., operating table supports. Use antistatic sheeting. • Prevent patient skin-to-skin contact (e.g., between the arm and the body) when the Generator is in use. Use dry gauze between body parts that might touch during surgery. • When HF surgical equipment and physiological monitoring equipment are used simultaneously on a patient, any monitoring electrodes must be placed as far as possible from the surgical electrodes. Needle monitoring electrodes are not recommended. If needle electrodes are used, do not place in direct contact with forceps tips. Monitoring systems incorporating high frequency current-limiting devices are recommended. • During surgical procedures where the HF current could flow through parts of the body having a relatively small area, bipolar techniques can prevent unwanted tissue damage. • To prevent injury to the patient and/or user, do not connect the Generator to any item not specified as part of the system or identified as a recommended accessory within this manual. Use of unspecified items with the system could result in excessive leakage current. Refer to individual Instructions for Use for additional information regarding accessories or use with other devices.
4
EN - ENGLISH
SYSTEM OVERVIEW • The system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the system should be observed to verify normal operation in the configuration in which it is used. • Refer to the Guidance and Manufacturer’s Declaration section for EMC (Electromagnetic Compatibility) information. • In order for the Generator to be able to fulfill its intended clinical function effectively, a maximum cord and accessory length of 4.5 meters between the Generator and accessories is defined for Bipolar Accessories, 3.2 meters is defined for Monopolar Accessories and 3.0 meters for Neutral Electrodes. • The use of accessories, transducers and/or cords other than those specified, except for those sold by the manufacturer as replacement parts for internal components, could result in increased emissions or decreased immunity of the equipment or system. • The application of a current with high voltage, and in particular of a high voltage monopolar coagulation current, may cause neuromuscular stimulation in the patient. • When vaporizing tissue, do not inhale the combustion fumes that form by design over a longer period in concentrated form. During regular use of the Generator, no pollutants other than these combustion products will be formed. For removal of such combustion products, a fume suction system can be used.
Precautions • The Generator must be separated from other electrosurgical devices and their cords. Close proximity to the Generator and its cords to electrosurgical units that produce excessive RF current radiation could cause the Generator to change output power. Surgical electrode cords must be positioned to avoid contact with the patient or other leads. Temporarily unused active electrodes should be stored in a location that is isolated from the patient. • Do not operate the Generator at temperatures below 50°F (10°C). Allow the Generator to warm up to at least 50°F (10°C) before use. • Visually inspect the power cord, bipolar cord and Foot Pedal for damage prior to each use. Connect Foot Pedal, adjust Generator to desired power setting, and activate Foot Pedal to verify proper function. Replace accessories if necessary. • Non-insulated bipolar forceps are not recommended for use with higher power settings. When higher power settings are employed, insulated bipolar forceps are recommended. • Keep bipolar instrument tips clean. Build-up of eschar may reduce the instrument's effectiveness. Do not activate the instrument while cleaning, as injury to operating personnel may result. • The Generator is designed for use without a neutral electrode while operated in bipolar modes. • The Generator does not generate waste. Disposable products used with the Generator must be disposed of in accordance with hospital procedures and regulations. • Always position Generator so that the power cord can be easily unplugged from the rear panel. • The output power should be set as low as is necessary to achieve the desired effect. • Anyone who connects additional units to the input or output section thereby configures the system and is thus responsible for compliance with section IEC 60601-1:16. • In order to be able to completely separate the unit from the grid in the event of danger, either the power socket of the unit or the outlet into which the mains cable is plugged should remain accessible.
Wiping • Wipe the exterior surfaces of the Generator as necessary with a soft, damp cloth or sponge. Use 70% Isopropyl Alcohol wipes or a mild cleaning solution to remove stains or adhesives that stick to the surface. Subjecting the Generator to excessive moisture can damage the electronic components and nullify the warranty. • Never sterilize the Generator.
5
EN - ENGLISH
CONTROLS, INDICATORS & CONNECTIONS Front Panel
1
3
2
4
5
7
6
8
Codman® Electrosurgical Generator CUT
COAG
Watt
Watt
max.
MONOCUT
BICUT
max.
MONOCOAG
BICOAG
Monopolar
9
10
Bipolar
11
12
1. Power Button This button controls the primary AC power to the Generator. When the Generator is ON, the Display Screen, Operating Mode Buttons and Power Button illuminate. 2. CUT Output Power Setting Knob Turning the output power setting knob clockwise increases the power setting and turning it counterclockwise decreases the power setting of the Generator. The setting is visible on the CUT Display Screen. 3. CUT Display Screen The screen shows the output power setting for CUT (MONOCUT or BICUT, depending on which Operating Mode Button is activated). 4. CUT Indicator Light The light is illuminated when the user activates the CUT mode, either by finger switch or Foot Pedal. 5. Operating Mode Buttons These buttons control the operating mode of the Generator. The Display Screen will show the output power setting for the selected Operating Mode. Only one (1) CUT and one (1) COAG mode can be selected simultaneously. MONOCUT BICUT MONOCOAG BICOAG 6. COAG Indicator Light The light is illuminated when the user activates the COAG mode, either by finger switch or Foot Pedal.
6
EN - ENGLISH
CONTROLS, INDICATORS & CONNECTIONS
7. COAG Display Screen The screen shows the output power setting for COAG (MONOCOAG or BICOAG, depending on which Operating Mode Button is activated). 8. COAG Output Power Setting Knob Turning the output power setting knob clockwise increases the power setting and turning it counterclockwise decreases the power setting of the Generator. The setting is visible on the COAG Display Screen. 9. Neutral Electrode Receptacle The neutral electrode receptacle accepts the international rectangular connection. 10. “Neutral Electrode not connected” Indicator Light If the light is illuminated and flashing red, the Neutral Electrode is not properly connected to either the Generator or the patient. 11. Bovie and 3-Pronged Monopolar Receptacle The Monopolar receptacle accepts either a bovie connection or 3-pronged connection cord for use with monopolar instruments. 12. Bipolar Receptacle The Bipolar receptacle accepts a Codman bipolar cord for use with bipolar instruments. (See Recommended Accessories section for product codes.)
7
EN - ENGLISH
CONTROLS, INDICATORS & CONNECTIONS Rear Panel
13
14
15 16
17
18 19
100-240 V-50/60 Hz T3.15 AH 250V
13. Foot Pedal Connection Receptacle The connection receptacle accepts the four-pin end of the Foot Pedal cord. 14. Irrigator Connection Receptacle The connection receptacle accepts the three-pin end of the Interconnecting Cable. 15. Speaker The speaker produces audible warnings and indications. 16. Volume Control Knob Allows the user to adjust the volume of the speaker. 17. Device Type Plate Provides information to the user about the device. 18. Power Receptacle The power receptacle accepts the three-prong power cord plug. 19. Equipotential Connector Connect a potential equalization conductor to a bus bar in rooms requiring potential equalization. A bus bar is a grounded metal bar in a breaker box to which all neutral and grounding wires are connected.
8
EN - ENGLISH
CONTROLS, INDICATORS & CONNECTIONS Foot Pedal 20 21
22
23
20. Adjustment Bar When raised, the adjustment bar allows the user to reposition the Foot Pedal with his/her foot mid-procedure. 21. Cord management feature Allows the user to wrap the Foot Pedal cord for storage. 22. CUT pedal (yellow) Depressing this pedal activates cutting power to the instrument. 23. COAG pedal (blue) Depressing this pedal activates coagulation power to the instrument.
9
EN - ENGLISH
INSTRUCTIONS FOR USE Generator Setup
1
Place Generator
Place the Generator on a stand, shelf, or other flat surface capable of securely supporting the device.
Codman® Electrosurgical Generator CUT
COAG
W MONOCUT
BICUT
Monopolar
2
W MONOCOAG
BICOAG
Bipolar
Plug In Generator
a
Plug the power cord into the rear panel of the Generator.
b
Plug the Generator into a grounded electrical outlet of the appropriate voltage rating.
REAR PANEL
Foot Pedal & Electrosurgical Irrigator Setup
1
Plug In Foot Pedal
REAR PANEL
2
If you are using the Foot Pedal with the Generator: Insert the plug of the Foot Pedal into the Foot Pedal Receptacle on the rear of the Generator. Place the Foot Pedal in an area accessible to the user.
b
Twist the cable collar 1/4 turn to lock the Foot Pedal cable in place.
Note: Verify activation of the Foot Pedal before each use.
FOOT PEDAL
Codman Electrosurgical Irrigator Setup (Optional)
IRRIGATOR: REAR PANEL
a
If you are using the Electrosurgical Irrigator with the Generator:
GENERATOR: REAR PANEL
10
a
Follow the Instructions for Use provided with the Irrigator.
b
Using the Electrosurgical Interconnecting Cable, connect one end of the cable to the Generator receptacle on the rear of the Irrigator. This cable allows you to control irrigation while cutting or coagulating when using the Foot Pedal to activate the system.
c
Connect other end of the cable to the receptacle on the rear of the generator.
EN - ENGLISH
INSTRUCTIONS FOR USE Power On Power on the Generator by pushing the power button. The power button will illuminate blue. An error in the start-up indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
Self-Test The Generator performs a self-test after it is powered on. All display elements on the front panel are turned on during this process. The following light up briefly in the order they are listed: The MONOCUT and BICUT mode buttons, then MONOCOAG and BICOAG mode buttons, followed by the red indicator light above the neutral electrode receptacle. The two Output Power displays light up with “88” and the white CUT and COAG Indicator lights illuminate sequentially; the respective activation tone must sound simultaneously. The default start-up of the device is BICUT and BICOAG with both displays lit up with “00”. An error in the self-test sequence indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
11
EN - ENGLISH
INSTRUCTIONS FOR USE: CONNECTING ACCESSORIES Connecting the Neutral Electrode
1 Connect the Neutral Electrode to the receptacle on the front of the device. NOTE: • If the neutral electrode is NOT connected during monopolar mode, the “Neutral electrode not connected” indicator above the receptacle will light up red and flash. • If the finger switch or Foot Pedal is activated while in monopolar mode without the neutral electrode connected, an audible warning sounds. RF current cannot be activated. This has no influence on the bipolar current. • When using a multi-surface electrode, the “Neutral electrode not connected” indicator will turn off after the safe application state is reached. • The entire area of the neutral electrode should be reliably attached to a suitably prepared and appropriate area of the patient’s body as defined by the manufacturer.
Connecting Monopolar Accessories
1 Connect a 3-pronged connection monopolar cord or bovie connection monopolar cord to the appropriate location in the monopolar receptacle. Ensure the 3-pronged connection is in the correct orientation (see illustration below).
2 Connect the desired Monopolar Instruments to this cord. NOTE:
a1 Connect a Monopolar device with a finger switch for CUT and COAG activation. OR Monopolar
a Monopolar device without a finger switch, in combination with the Foot Pedal for CUT and a2 ConnectBipolar COAG activation.
Connecting Bipolar Accessories
1 Connect a Codman Bipolar cord to the receptacle. See Recommended Accessories section. 2 Connect the desired bipolar instrument to the cord. The Foot Pedal is used for CUT and COAG activation. NOTE: • The neutral electrode is NOT required for bipolar applications.
Bipolar
12
EN - ENGLISH
RECOMMENDED ACCESSORIES AND ORDERING INFORMATION Description
Catalog No.
Codman Integrated Bipolar Cord and Tubing Set, Rotary Pump, Fixed Leads
9190001RP
Codman Bipolar Cord Sets, Reusable, Unitized Leads
801186
Codman Bipolar Cord Sets, Disposable, Unitized Leads
801188
Codman Electrosurgical Interconnecting Cable
901001CBL1
Codman Electrosurgical Foot Pedal
901001FPED
Codman Electrosurgical Irrigator
901001IRRO
Codman VersaTru [Standard/Slim] Disposable Non-Stick Bipolar Forceps
9007050ST, 9007100ST, 9007150ST, 9008050ST, 9008100ST, 9008150ST, 9009050ST, 9009100ST, 9009150ST, 9008050SL, 9008100SL, 9008150SL, 9009050SL, 9009100SL, 9009150SL
13
EN - ENGLISH
ERROR DIAGNOSIS/TROUBLESHOOTING The Generator conducts extensive functional tests during the power-on self-test. All safety-related functions are also checked continuously during operation. In case of errors, the RF energy output is stopped and the error is indicated by an error message. Table 1 below describes an example of an error diagnosis, possible cause, and solutions. The messages where resolution by the user is possible are listed in the Possible Solutions column. Note: After an error message is displayed, the device can only be reset by turning it off and back on again. Note: Contact your Integra sales representative if a displayed message is not described below. Table 1 - Error Diagnosis Error Messages
Possible Causes
Solutions
Apparent low output or failure of the HF surgical equipment to function correctly at the normal operating settings.
Faulty application of the neutral electrode or poor contact in its connections.
Application of the neutral electrode and its connections should be checked before selecting a higher output power.
Power cord not plugged into the outlet.
Plug power cord into the outlet.
Power cord is not fully connected to Generator.
Assure that the power cord is completely plugged into both the device and the outlet.
Device is not turned ON.
Switch ON the device.
Defective fuse.
Device defect; service required.
Defective internal power supply.
Device defect; service required.
Visual elements on the front work normally; no RF output power.
Defective accessory.
Replace the accessory.
Defective device.
Turn the device OFF and ON again. If any error message is displayed, contact your Integra sales representative.
Err 41, Err 42
Monopolar hand piece switch activated before the self-test was complete.
Release the finger switch.
Defective instrument or connection cord.
Disconnect the hand piece and turn the device OFF and ON.
Defective device.
Contact your Integra sales representative.
Defective instrument or connection cord.
Disconnect the hand piece and turn the device OFF and ON.
Defective device.
Contact your Integra sales representative.
Activation of the Foot Pedal before the self-test was complete.
Release the Foot Pedal.
Short circuit in the Foot Pedal or its connection cable.
Disconnect the Foot Pedal and turn the device OFF and ON.
Defective device.
Contact your Integra sales representative.
Operator panel button was activated before the self-test was completed.
Turn the device OFF and ON again. If any error message is displayed, contact your Integra sales representative.
Defective device.
Contact your Integra sales representative.
Visual elements on the front are dark; device does not function.
CUT
COAG
Err 45, Err 46 CUT
COAG
Err 47, Err 48 CUT
COAG
Err 51 CUT
COAG
Err 89 CUT
COAG
Overheating caused by continuous RF activation Turn OFF the Generator and let it cool down for for an extended time outside the specification. a few minutes before turning it ON again. If the error is still displayed, contact your Integra sales representative. 14
EN - ENGLISH
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS Current Mode MONOPOLAR BIPOLAR
Current Type
RF Output Power
RF Output Current
RF Output Voltage
Nominal
Modulation
MONOCUT (Cutting) max. 80 W ±20 % on 600 Ω
0.51 A on 600 Ω max 800 Vp
4.00 MHz
33 kHz
MONOCOAG (Coag.) max. 80 W ±20 % on 400 Ω
0.51 A on 400 Ω max 560 Vp
4.00 MHz
33 kHz
BICUT (cutting)
max. 80 W ±20 % on 300 Ω
0.6 A on 300 Ω
max 400 Vp
4.00 MHz
33 kHz
BICOAG (coag.)
max. 60 W ±20 % on 50 Ω
1.2 A on 50 Ω
max 440 Vp
4.00 MHz
33 kHz
Classification Mains Supply
100-240 VAC ; 50/60 Hz
Power Consumption
Without HF output: approximately 50 VA At maximum output power: approximately 500 VA
Protection Class
I
LF and RF Leakage Currents
According to IEC 60601-2-2
Application Type
CF; defibrillator safe
Operating Mode
Intermittent 10 seconds ON /30 seconds OFF
Mains Fuse
2 x T 3.15 AH 250 V 5x20 mm
Signal Level
RF indicator: 40-65 dB(A) adjustable. Alarm: 65 dB(A)
Weight
Approximately 5.0 kg
Dimensions
Width 403 mm Height 170 mm Depth 313 mm
Standards
IEC 60601-1:2005 + A1:2012, Ed. 3.1 CAN/CSA-C22.2 No. 60601-1:14 ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 IEC 60601-1-2:2007, Ed. 3.0 IEC 60601-1-2:2014, Ed. 4.0 IEC 60601-2-2:2017, Ed. 6.0
Environmental
Operational
Transport
Storage
Temperature
10° to 40°C / 50° to 104°F
-40° to 60°C / -40° to 140°F
-40° to 60°C / -40° to 140°F
Relative Humidity (non-condensing)
30% to 75%
15% to 85%
15% to 85%
Atmospheric Pressure
700 hPa to 1050 hPa
500 hPa to 1050 hPa
500 hPa to 1050 hPa
15
EN - ENGLISH
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS Service and Repair The manufacturer will provide schematics, parts list, and other technical information to authorized service centers. For service, repair or replacements, contact your local Integra representative. Equipment under warranty should be returned to Integra LifeSciences Corporation. It will be repaired or replaced without charge to the purchaser.
End of Useful Life The Generator contains electrical components. Consider recycling at the end of service life. The product can be returned to Integra for recycling. Dispose of the equipment in accordance with local ordinances. PRODUCT INFORMATION DISCLOSURE INTEGRA LIFESCIENCES CORPORATION (“INTEGRA”) HAS EXERCISED REASONABLE CARE IN THE SELECTION OF MATERIALS AND THE MANUFACTURE OF THESE PRODUCTS. INTEGRA WARRANTS THAT THESE PRODUCTS SHALL CONFORM TO THE PRODUCT LIMITED WARRANTY AS PROVIDED IN THE PRODUCT LABELING OR APPLICABLE PRODUCT CATALOG. THIS WARRANTY IS EXCLUSIVE, AND INTEGRA DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. INTEGRA SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL LOSS, DAMAGE, OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING FROM USE OF THESE PRODUCTS. INTEGRA NEITHER ASSUMES NOR AUTHORIZES ANY PERSON TO ASSUME ANY OTHER OR ADDITIONAL LIABILITY OR RESPONSIBILITY IN CONNECTION WITH THESE PRODUCTS.
Field Maintenance and Repair Equipment under warranty should be returned to Integra LifeSciences Corporation. It will be repaired or replaced without charge to the purchaser. Annual safety and functional checks are recommended.
Guidance and Manufacturer’s Declaration Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the tables below. NOTE: Portable and mobile RF communications equipment can affect medical electrical equipment.
Table 2 -Guidance and Manufacturer’s Declaration-Emissions All ME Equipment and ME Systems Guidance and Manufacturer’s Declaration-Emissions The Generator is intended for use in the electromagnetic environment specified below. The customer or user of the Generator should ensure that it is used in such an environment. Emissions Test
IEC 60601-1-2, 3rd edition Compliance Level
IEC 60601-1-2, 4th edition Compliance Level
Electromagnetic Environment-Guidance
Radiated RF Emissions CISPR 11
Group 2
Group 2
The Generator must emit electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.
Conducted RF Emissions CISPR 11
Class A
Class A
Harmonics IEC 61000-3-2
Class A
Class A
Emissions of voltage fluctuations/flicker IEC 61000-3-3
Complies
Complies
The Generator is suitable for use in all establishments, other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
NOTE: The Generator must be installed and operated according to the EMC regulations laid out in this document. NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 class B is normally required), this equipment might not offer adequate protection to radiofrequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. 16
EN - ENGLISH
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS
Table 3 -Guidance and Manufacturer’s Declaration-Immunity All ME Equipment and ME Systems Guidance and Manufacturer’s Declaration-Immunity The Generator is intended for use in the electromagnetic environment specified below. The customer or user of the Generator should ensure that it is used in such an environment. Immunity Test
IEC 60601-1-2, 3rd edition Compliance Level
IEC 60601-1-2, 4th Electromagnetic Environment-Guidance edition Compliance Level
ESD IEC 61000-4-2
±6 kV Contact ±8 kV Air
±8 kV Contact ±15 kV Air
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the relative humidity should be at least 30%.
EFT IEC 61000-4-4
±2 kV Mains ±1 kV I/Os
±2 kV Mains ±1 kV I/Os
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV Differential ±2 kV Common
±1 kV Differential ±2 kV Common
Mains power quality should be that of a typical commercial or hospital environment.
Voltage Dips/ Dropout IEC 61000-4-11
<5% UT for 0.5 Cycle
0% UT for 0.5 Cycle
40% UT for 5 Cycles
0% UT for 1 Cycle
70% UT for 25 Cycles
70% UT for 25/30 Cycles
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Generator requires continued operation during power mains interruptions, it is recommended that the Generator be powered from an uninterruptible power supply or battery.
<5% UT for 5 Seconds
0% UT for 250/300 Cycles
Power Frequency 3 A/m, 50/60 Hz Magnetic Field IEC 61000-4-8
30 A/m, 50/60 Hz
17
Power frequency magnetic fields should be that of a typical commercial or hospital environment.
EN - ENGLISH
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS Table 4-Guidance and Manufacturer’s Declaration – Electromagnetic Immunity Reference Table 204 IEC 60601-1-2 3rd edition Guidance and Manufacturer’s Declaration-Immunity The Generator is intended for use in the electromagnetic environment specified below. The customer or the user of the Generator should ensure that it is used in such an environment. Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic Environment – Guidance Portable and mobile RF communications equipment should be used no closer to any part of the Generator, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance:
Conducted RF (Mains and I/Os) IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
3 Vrms
d = 1.2 √ P
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3 V/m
d = 1.2 √ P 80 MHz to 800 MHz d = 2.3 √ P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmittersa, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency rangeb. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Generator is used exceeds the applicable RF compliance level above, the Generator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Generator. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
18
EN - ENGLISH
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS Table 5-Recommended Separation Distances for the Generator, per IEC 60601-1-2, 3rd edition The Generator is intended for use in the electromagnetic environment specified below. The customer or user of the Generator should ensure that it is used in such an environment. Maximum Output Separation (m) Power (Watts) 150 kHz to 80 MHz
Separation (m) 80 to 800 MHz
Separation (m) 800 MHz to 2.5 GHz
d = 1.2√P
d = 1.2√P
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
WARNING: Per IEC 60601-1-2, 4th edition, portable RF communications should be used no closer than 30 cm (12 inches) to any part of the Generator, including cords specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
19
EN - ENGLISH
POWER CURVES RF Output The below power charts show typical output characteristics of the Generator for all available modes. For each mode, the power output at the maximum (1) and half (2) power level as a function of the resistive load is depicted. Actual measurements of the output power may vary depending on the chosen test accessory and the measurement setup. According to IEC 60601-2-2:2017 paragraph 201.12.1.101, the power output shall not deviate by more than ±20 %.
Monopolar Curve 1: Output setting “maximum output” Curve 2: Output setting “half output”
Curve 1: Output setting “maximum output” Curve 2: Output setting “half output”
MONOCOAG MONOCOAG 90
90
80
80
80
80
60
90
MONOCUT MONOCUT
70
1
60
90
50
80 70
50 40
60 50
20
10
10
40
0 0
30
0
20
20
10
10
0 400
200600
400800
6001000
8001200
1000 1400
1200 1600
1400 1800
1600 2000 1800
BICUT BICUT 800 1200 1000 1400 1200 1600 1400 1800 1600 2000 1800 Curve 1: Output setting “maximum output” 90 90 Resistance (Ω) Curve 2: Output Resistance setting “half output” (Ω) 200
Output Power (Was) Output Power (Was)
80 70 60
90 80 70 60 50
50 40 30 20 10
40
0
0 400
30 20
400800
600 1000
70
10
BICUT
0 0
20
20
10
10
BICUT
20
1
10
40
0
100 30
0200
100300
200 400
300 500
400 600
500 700
600 800
700 900
800 1000
1000
60
50
40
50
20
40
10
30
0
0 400
1000 1400
12001600
14001800
100 300
200 400
300 500
400 600
500 700
600 800
700 900
800 1000
900
800
700
700
500
900
400 300
700
200 100
800
600
600
MONOCUT MONOCUT
600 1000
60
BICOAG BICOAG
50
30 100
0 0200
100 300
200 400
300 500
400 600
500 700
600 800
700 900
Resistance Resistance (Ω) (Ω)
10
500
500
400
500
300 200 100
20
400
600
300
300
500
200
400 100
800 1000
900
1000
1 2
0 100
600
400
2000
40 10
600
600
500
( Vp)
600
2000
50 20
0
1000
Voltage ( Vp)
800
Voltage ( Vp)
900
1600 2000 1800
60 30
Voltage ( Vp)
0200
900
Vp)
Voltage ( Vp) Vp)
600
400800
20
( Vp)
100
MONOCUT MONOCUT Resistance Resistance (Ω) (Ω)
700
200600
0
0 0
800
8001200
10
0
900
6001000
70 40
10 10
400800
Resistance Resistance (Ω) (Ω)
0
20
2
Resistance Resistance (Ω) (Ω)
20
900
200600
0
60 30
30
50
0 400
BICOAG BICOAG 800 1200 1000 1400 1200 1600 1400 1800 1600 2000 1800 Curve 1: Output setting “maximum output” 70 70 Resistance Resistance (Ω) (Ω) Curve 2: Output setting “half output”
40
60
0
200
70
70
10
200 30
0
50
80
2
20
40
0
2000
60
90
0
200600
80
Output Power (Was) Output Power (Was)
0
30
50
30
Resistance Resistance (Ω) (Ω)
Bipolar 0
0
2000
1
40
60
20
40
200 30
50
70
30
Output Power (Was) Output Power (Was)
40
40
50
50
60
80
60
60
20
50
70
2
30
MONOCOAG MONOCOAG
70
90
80
70
30
60
90
80
40
70
Output Power (Was) Output Power (Was)
70
Output Power (Was) Output Power (Was)
90
Output Power (Was) Output Power (Was)
90
Output Power (Was) Output Power (Was)
Output Power (Was) Output Power (Was)
MONOCUT MONOCUT
200
400 100
0200
100 300
200 400
300 500
400 600
500 700
600 800
700 900
MONOCOAG MONOCOAG Resistance Resistance (Ω) (Ω)
MONOCOAG MONOCOAG
800 1000
900
1000