ConMed Corp
ClearView Operators Manual Rev B
Operators Manual
24 Pages
Preview
Page 1
ClearView™
Smoke Evacuation System Operator’s Manual
For a period of one (1) year following the date of delivery, CONMED warrants the ClearView™ against any defects in material or workmanship. CONMED will repair or replace (at CONMED’S option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by CONMED. This warranty is void if the product is used in a manner or for purposes other than intended.
ConMed Corporation 525 French Road, Utica, NY 13502 USA (800) 448-6506 (United States) or +1 (315) 797-8375 (International) www.conmed.com Made in USA The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages (if any). Manual Number 904568 REVA Unit Serial Number ______________________
Emergo Europe Molenstraat 15 2513 BH, The Hague The Netherlands
MEDICAL – GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1, ANSI/AAMI ES60601-1 (2005, 3rd ed.), CAN/CSA C22.2 NO. 601.1, AND CAN/CSA-C22.2 No. 60601-1 (2008) 9D93 This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
Table of Contents & List of Illustrations Section Title Page 1.0 SYSTEM DESCRIPTION 3 1.1 Introduction 1.2 Inspection 1.3 Operational Information 1.4 Cautions and Warnings 2.0 OPERATING INSTRUCTIONS 10 2.1 System Controls 2.2 Filter Instructions 2.3 Set-up and Operation 2.4 Specifications 3.0 MAINTENANCE 21 3.1 General Maintenance Information 3.2 Cleaning 3.3 Periodic Inspection 3.4 Troubleshooting 4.0 CUSTOMER SERVICE 22 4.1 Equipment Return 4.2 Ordering Information 5.0 TERMS & WARRANTY 23 Figure Title Page 1 Control Panel 14
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 3
System Description Section 1.0 1.1
Introduction
CONMED ClearView™ Smoke Evacuation Systems are intended to evacuate and filter surgical smoke plume and aerosols created by the interface surgical tools with tissue, examples being lasers, electrosurgery systems, and ultrasonic devices. The ClearView™ Smoke Evacuation Systems have been designed with a high suction, high flow rate vacuum motor. The ultra-quiet motor is used to draw the surgical smoke from the surgical site through the vacuum tubing and into the filter where the surgical smoke is processed by a series of filters. A single disposable filter is used to simplify the installation and removal during filter changes. The filter is completely enclosed to protect the healthcare personnel from potential contamination during filter changes. One ClearView™ filter contains four different stages within to capture the smoke plume. The first stage filtration is a prefilter whose function is to trap and remove gross particulate and casual fluid. The second stage filtration is ULPA grade (Ultra Low Penetration Air) filter whose high-tech patented (U.S. Patent #5874052) design captures particulates and micro-organisms from .1 to .2 microns at an efficiency of 99.999%. The third stage filtration uses the highest grade virgin activated carbon, especially designed for CONMED for the removal and adsorption of odors and toxic gases produced by burning tissues. These harmful gases may constitute a health hazard to healthcare professionals who are subjected to prolonged exposure. The activated carbon used in the ClearView™ Smoke Evacuation Systems preferentially removes toxic organic gases rather than water vapor and provides optimal odor removal. The fourth stage filtration is an expanded foam used to trap activated carbon fines from migrating out of the filter. The electronic controls on the face panel of the ClearView™ Smoke Evacuation System has been designed “user friendly” and facilitate unit set up and operation. Please refer to Section 2.0 for Operating Instructions.
1.2 Inspection The ClearView™ Smoke Evacuation System has been thoroughly tested and inspected before shipment from the factory. Please check the unit before using it to insure that no damage has occurred in transit. If damage is evident, please contact CONMED Customer Service at (800) 448-6506 (United States) or +1 (315) 797-8375 (International).
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 4
In addition, please compare the accessories you receive with the standard accessories list below. If an item is missing, please notify CONMED Customer Service. Standard Accessories: • Operator’s Manual • Power Cord • Pneumatic Footswitch • Automatic Activation Device Please contact CONMED Customer Service to purchase the following accessories: • Replacement Filters • Tubing Accessories • Fluid Trap
1.3 Operational Information The operational information contained in this section is intended for the customer review of regulatory issues. The information pertains to the use of the products both domestically and internationally: 1. CONMED ClearView™ Smoke Evacuation System(s) complies with IEC60601.1 electrical specifications in the following systems: 100/120 VAC 50/60 Hz, 220/240 VAC 50/60 Hz 2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I 3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part 4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1 5. Method of sterilization or disinfection recommended by CONMED (UL 60601-1, Clause 5.4): Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water. Wipe dry with a clean cloth. Do not steam sterilize. 6. Degree of safety of application in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable 7. Mode of operation (UL 60601-1, Clause 5.6): Continuous 8. Upon request, CONMED will provide the following: Service and Repair Instructions, including Circuit Diagrams and Parts List 9. The fuses on the circuit board are to be serviced by an authorized CONMED technician as follows: 100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo) 220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo)
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 5
10. The fuses on the motor curcuit are to be serviced by an authorized CONMED technician as follows: 220/240 VAC, 50/60 Hz use 3,15 Amp 250 Volt Fuse (Fast-Acting), (F3) 11. This equipment needs special precautions regarding ElectroMagnetic Compatibility and needs to be installed according to EMC information found in this manual. 12. This equipment utilizes mobile RF communications equipment that can affect medical electrical equipment. 13. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their expense. 14. This equipment operates in the following radio frequency specifications: RX modulation: Pulse-width coded, AM 100% modulation TX Frequencies: Manchester encoded, A = fc =/- 423.75kHz, B = fc +/- 484.29kHz Low bit: transition A to B High bit: transition B to A 15. To isolate equipment from supply mains, unplug the power cord from the appliance inlet on the unit or receptacle in the wall. Position the equipment to allow for ease of unplugging power cord. 16. Potential Equalization Conductor: Terminal located on back panel for connection of potential equalization. Conductor complies with requirements per IEC 60601-1 (2005). The ClearView™ Smoke Evacuation System(s) and all filters are not intended for contact with patients.
1.4 Cautions and Warnings Please note that all Cautions and Warnings should be read and understood before any use of this equipment.
Please note that all Cautions and Warnings should be read and understood before any use of this equipment.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 6
1.4.1 WARNINGS: • Read this manual thoroughly, and be familiar with its contents prior to using this equipment. • Test this equipment prior to a surgical procedure. This product was thoroughly tested at the factory before shipment. • Disconnect the unit from the electrical outlet prior to inspecting system components. • The ClearView™ Smoke Evacuation System is only intended and suitable for the applications that are mentioned in the operating instructions. • The smoke evacuator produces a strong vacuum. Adjust the airflow and the position of the inlet end of the wand or tubing to prevent patient injury and to prevent suction of surgical materials and surgical specimens. • If the smoke evacuator is activated while the airflow is set to a high speed, it may produce a sudden, strong suction action. Check the airflow setting before activating the smoke evacuator to prevent patient injury and to prevent suction of surgical materials and surgical specimens. • To maximize patient safety, the tubing or wand should not come into direct contact with tissue. Otherwise, patient injury may result. • The CONMED filters and single-use accessories are completely disposable. Please dispose of according to your local codes or regulations and hospital policy. These filters may be disposed of or incinerated, whichever is appropriate for your institution. • Care should be taken to route the power cord, foot pedal, smoke evacuation tubing, and Automatic Activation Device cable as to not cause a tripping hazard or crimping of cords. • Do not operate this device in the presence of flammable or explosive gases. • To avoid risk of Electric Shock, this equipment must only be connected to a supply mains with protective earth. • This equipment is intended for use by healthcare professionals only. This equipment may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the ClearView™ or shielding the location. • The use of ACCESSORIES other than those specified by CONMED, or sold by CONMED as replacement parts for internal components, may result in increased emissions or decreased immunity of the ClearView™ Smoke Evacuation System. • This equipment should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use in necessary, the ClearView™ Smoke Evacuation System should be observed to verify normal operation in the configuration in which it will be used.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 7
• Refer routine servicing to qualified biomedical technical personnel. • Changes or modifications not expressly approved by CONMED could void the user’s authority to operate the equipment. The warranty on this product is void if any of these warnings are disregarded.
1.4.2 CAUTIONS: • Federal law (United States of America) restricts this device to be used by, or on the order of a physician. • Do not block either the tubing or the filter. If either becomes occluded or significantly restricted, the motor/blower may overheat and cause the unit to fail. • Using any other filter or accessory not supplied by CONMED may cause damage and/ or cause the system to be inoperable and may void the warranty. • Care must be exercised in the installation of hoses, adapters and suction canisters. Failure to follow the procedures outlined in this manual may result in overheating of the motor and may void the unit warranty. • This device is not intended for evacuation of fluid. If fluid is expected to be aspirated to the ClearView™ Filter, fluid collection devices must be installed with the vacuum hose assembly. Failure to install a fluid collection device could cause filter blockage and electrical damage. • The Filter should be changed according to the life of the filter. The Filter, used with the ClearView™ Smoke Evacuation System(s), should not be used for more the time specified for each filter. Failure to change the filter may result in decreased efficiency and contamination of the electric motor, vacuum pump, and sound absorbing media within the unit. • Do not block either the tubing or the filter during operation. An occlusion or significant restriction may cause the motor to overheat and the unit to stop working. • The installation of this equipment must be performed such that the intake and exhaust vents located on the bottom of the system are not obstructed. Failure to properly install the unit may cause reduced performance, damage and/or cause the system to be inoperable and may void the warranty. • The ambient temperature during operation must be kept between 50°F to 104°F (10°C to 40°C) • The relative humidity during operation must be kept between 10% to 75%. • An atmospheric pressure range of 700 hPA to 1,060 hPa.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 8
• Storage environmental ambient temperature 14°F to140°F (-10°C to 60°C). • Storage environmental relative humidity 10% to 75%. There are no user serviceable components in the ClearView™ Smoke Evacuation System(s). Refer service to qualified service personnel. Use only with the power cord provided and always plug into a grounded outlet.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 9
Symbol
Description / Meaning DANGER HIGH VOLTAGE CAUTION - ELECTRICAL SHOCK HAZARD. DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. DANGER CAUTION - RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS. WARNING
CAUTION
TYPE CF APPLIED PART
IPX1
PROTECTION AGAINST INGRESS OF WATER AS DETAILED IN IEC 60529
ALTERNATING CURRENT
PROTECTIVE EARTH, (GROUND)
EQUIPOTENTIALITY
DENOTES THE DATE THE EQUIPMENT WAS MANUFACTURED
DENOTES THE MANUFACTURER OF THE DEVICE
NON-IONIZING RADIATION
CONSULT INSTRUCTIONS
REMOTE ACTIVATOR
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 10
Operating Instructions Section 2.0 2.1 System Controls The electronic system controls on the ClearView™ Smoke Evacuation System(s) are easy to understand and simple to use. The membrane control panel contains the suction on/off switch, suction power adjustment, filter life indicator, and service indicator light. See Figure 1. Note: Please be sure to read all instructions before installing accessories or operating this equipment. Failure to do so may result in damage to the unit and/or personal injury. • SUCTION ON & Standby There is one ON/OFF button on the ClearView™ Smoke Evacuation System(s). The suction ON switch on the electronic membrane control panel is located in the upper right hand corner of the membrane panel. To power up the machine, (|), connect the supplied power cord to a grounded outlet and the appliance inlet on the back of the smoke evacuation system. Once power has been applied, the yellow standby LED will illuminate on the keypad. Press the ON/OFF button on the membrane panel to illuminate the “fan running” green LED indicating active suction. Place the unit in “standby” by pressing the suction ON membrane switch to illuminate yellow STANDBY LED. Turn the system main power off, (O), by unplugging the power cord from the appliance inlet on the unit or the receptacle in the wall. • SUCTION CONTROL (Membrane Control Panel) The amount of suction may be adjusted by pressing the suction control button. Each time the suction control button is depressed, the motor speed in increased. Once the suction has reached the maximum level, depressing the button again will return suction level to lowest setting. The suction control should be set at the lowest practical setting to completely remove the surgical smoke from the operative site. Each time the arrow button is pressed, the suction will change to a different flow setting (Low/Lap, Medium, High/Turbo). • FOOTSWITCH / AUTOMATIC ACTIVATION DEVICE (Membrane Control Panel) The ClearView™ Smoke Evacuation System also comes equipped with a pneumatic footswitch. A footswitch or a Automatic Activation Device may be added to any system by simply plugging in a CONMED activation accessory into the appropriate jack on the front or back of the unit. When the footswitch is plugged in, the unit may be turned on or off by depressing the footswitch pedal once for each operation. For directions on using the Automatic Activation Device, please see instructions that accompany that product.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 11
• FILTER LIFE INDICATOR (Membrane Control Panel) The filter life indicator on the membrane control panel provides a visual indication of the status of the life of the filter in use. The filter life indicator for the ClearView™ Smoke Evacuation System will automatically adjust according to the flow setting selected.
Low/Lap Flow Setting = 35 hours of filter life Medium Flow Setting = 24 hours of filter life High/Turbo Flow Setting = 18 hours of filter life
The ClearView™ Filter Life Indicator is factory set. All filter life timing is automatic. Reading the Filter Life Indicator: Install an unused ClearView™ Filter into the system per the installation instructions contained in this operator’s manual. When the system is turned on, the filter life indicator with light up the leftmost GREEN LED, indicating 100% filter life. The indicator will progress thru subsequent GREEN LEDs, to an AMBER LED as time elapses and begin flashing RED to indicate that the filter has expended its useful life and requires replacement. When the maximum filter life is consumed and the smoke evacuator is not powered off for greater than six (6) hours or the main power is disconnected – a new filter is required to activate the smoke evacuator and make operational. • FUSES (circuit board) Two 10 AMP fuses (8 AMP for 220/240 ClearView™ Smoke Evacuation Systems) are located on the circuit board within the housing of the system. It electrically protects both the system and the operator from damage or injury. If the system is overheated or if there is an electrical surge in the electrical system, fuses will break and the system will not operate. When the Service light illuminates, please contact CONMED Customer Service for system service instructions.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 12
2.2 Filter Instructions CONMED P/N: Configuration: Filter - Multi Port Filter: Filter(s): Particle Size, mm: Filter Life: Filter Life Indicator: Filter P/N:
60-8585-120 and 60-8585-230 Portable or tabletop 4-Stage Filtration in One Casing, (Pre-Filter, ULPA, Carbon, Post-Filter) ULPA 0.1 to 0.2 Microns at 99.999% Efficiency Automatic Factory Set Filter Sensor Time Replacement 60-8590-001
Note: Before installing or removing any filter, be sure that the system is turned off. Filter Installation Instructions: The installation of the Filter into the CONMED ClearView™ Smoke Evacuation System(s) is quick and simple. 1. Remove the Filter from the shipping box and discard any protective wrapping. Examine all filters for damage during shipping and storage. Do not install any filter with visible signs of structural damage. 2. Insert the Filter into the filter receptacle. Be sure that the filter is seated completely against the bottom of the filter chamber and clip is fully engaged. WARNING: This device is not intended for evacuation of fluid. If fluid is expected to be aspirated using the Filter or the CONMED ClearView™ Smoke Evacuation System, fluid collection devices must be installed with the vacuum hose assembly. Failure to install a fluid collection device may cause filter blockage and/or electrical damage. Filter Removal Instructions: 1. After the Filter has been exhausted and requires changing, turn the smoke evacuation system off and disconnect any accessory tubing attached to the filter. 2. Depress the filter tab and pull the Filter from the smoke evacuation system and dispose of in accordance with hospital policy. The Filter may be disposed of or incinerated. 3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions for maintenance and installation of a new Filter. CAUTION: Using any other filter or accessory not supplied by CONMED may cause damage to the system and/or cause the system to be inoperable and may void the warranty. WARNING: The Filter should be changed when the Filter Life Indicator shows red flashing LED (replace). Failure to change this filter may result in decreased efficiency and contamination of the electric motor, vacuum pump, and sound absorbing media within the system, or non-operation of smoke evacuator.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 13
Figure 1 Unit Control Panel
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 14
2.4
Performance References*
Performance Model Number
60-8585-120 and 60-8585-230
Maximum Flow Setting (CFM-U.S.) Standard Hose I.D. 7/8” 25 CFM ** 3/8” 4.5 CFM 1/4” 2 CFM Standard Hose I.D. 22 mm 708 LPM ** 9.5 mm 130 LPM 6.4 mm 57 LPM Dimensions (H x W x D )
inches 6 x 11 x 15.5
Dimensions (H x W x D )
centimeters 15.2 x 27.9 x 39.4
Weight
lbs 4.3
Noise Level, dBA
(nominal) < 55.0 dBA
Footswitch Pneumatic
Standard
Automatic Control Activation
YES (optional)
Safety Features
UL Classified Fuse protection
Display
LED Filter Status Flow Rate Service Required
Voltage Available
100/120 VAC, 220/240 VAC
Frequency, auto sensed
50/60 Hz
Variable Flow Control
Yes Watts 1000 ± 10%
Motor Motor Static Suction
kPa (6.5 mm orifice) 21.04
*For reference purposes only **Using a new 7/8 in tubing.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 15
2.5
Electromagnetic Compatibility Information per IEC60601-1-2
Table 1 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The Smoke Evacuation System model ClearView™ is intended for use in the electromagnetic environment specified below. The customer or user of the ClearView™ should assure that it is used in such an environment. Emissions Test RF Emissions
Compliance
Electromagnetic Environment - Guidance
Group 1
The ClearView™ uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class A
The model ClearView™ is suitable for use in all establishments, other than domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Class A
Not applicable.
CISPR 11 RF Emissions CISPR 11
Harmonic Emissions IEC 61000-3-2
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 16
Table 2 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The model ClearView™ is intended for use in the electromagnetic environment specified below. The customer or user of the model ClearView™ should assure that it is used in such an environment. Immunity Test Electromagnetic discharge (ESD)
Compliance Level
+6 kV contact
+6 kV contact
+8 kV air
+8 kV air
Electrical fast transient/burst
+2 kV for power supply lines
+2 kV for power supply lines
IEC 61000-4-4
+1 kV for input/ output lines
+1 kV for input/ output lines
Surge
+1 kV differential mode
+1 kV differential mode
+2 kV common mode
+2 kV common mode
<5 % UT (>95 % dip in UT) for 0.5 cycle
<5 % UT (>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
40 % UT (60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
<5 % UT (>95 % dip in UT) for 5 sec
3 A/m
3 A/m
IEC 61000-4-2
IEC 61000-4-5
Voltage dips, short interruptions, and voltage variations on power supply input lines. IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
IEC 60601 Test Level
Electromagnetic Environment Guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Model ClearView™ requires continued operation during power mains interruptions, it is recommended that the Model ClearView™ be powered from an uniterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 17
Table 3
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The model ClearView™ is intended for use in the electromagnetic environment specified below. The customer or user of the model ClearView™ should assure that it is used in such an environment. Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance Portable and mobile RF communications equipment should be used no closer to any part of the Model EVL including cables, than the Recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Radiated RF
3 V/m
d = 1.7 √P 80 MHz to 800 MHz
IEC 61000-4-3
80 MHz to 2.5 GHz 3 V/m
d = 2.3 √P 800 MHz to 2.5 GHz
3 Vrms Conducted RF IEC 61000-4-6
150 kHz to 80 MHz
d = [3.5/V1} √P Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structure, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Model is used exceeds the applicable RF compliance level above, the model ClearView™ should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model ClearView™. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 18
Table 4 Recommended Separation Distance Between Portable and Mobile RF Communications Equipment and the Model @ 3 Vrms The model ClearView™ is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the model ClearView™ can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model ClearView™ as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter m 150 kHz to 80 MHz Rated maximum output power of transmitter W
d=
[] 3.5
v1
√P
80 kHz to 800 MHz
[]
d = 3.5 v1
√P
800 kHz to 2.5 GHz
[]
d = 3.5 v1
√P
0.01
0.12
0.12
0.23
0.1
0.34
0.34
0.74
1
1.7
1.7
2.3
10
3.7
3.7
7.4
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance (D) in meters (M) can be estimated using the equation applicable to the frequency of the transmitter, where (P) is the maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structure, objects, and people.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 19
Maintenance Section 3.0 3.1 General Maintenance Information This section contains information for ordinary upkeep of the CONMED ClearView™ Smoke Evacuation System. While the system has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed by a qualified Biomedical Technician to ensure continued safe and effective operation. 3.2 Cleaning Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water. Wipe dry with a clean cloth. Do not steam sterilize. 3.3 Periodic Inspection The ClearView™ Smoke Evacuation System should be visually inspected at least every year. This inspection should include checks for: • Damage to the power cord. • Damage to the power plug or power inlet module. • Proper mating, cleanliness and absence of damage to the filter inlet. • Obvious external or internal damage to the system. 3.4
Troubleshooting the System – see below.
PROBLEM
POTENTIAL CAUSE
CORRECTIVE ACTION
1. Smoke Evacuation System is ON but suction is minimal or none.
1. Filter is not seated completely.
1. Re-install Filter, press firmly into place and fully engage clip.
2. Filter is clogged.
2. Replace filter with a genuine CONMED Filter.
3. Vacuum hose or tube is clogged.
3. Replace vacuum hose or tube with genuine CONMED products.
4. Motor/blower is obstructed. 4. Call BioMed or CONMED Customer Service at 1.800.448.6506 or +1.315.797.8375. 2. Smoke Evacuation System does not function even though suction ON button is depressed.
1. Not plugged into an electrical outlet.
1. Check power outlet and connection to rear or side panel of the machine.
2. Fuses are blown.
2./3. Call BioMed or CONMED Customer Service at 1.800.448.6506 or +1.315.797.8375.
3. Electronic system failure. 4. Filter life has expired or invalid filter inserted.
4. Replace filter with a genuine CONMED Filter.
© 2016 All rights reserved. 904568 Rev B CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 20