ConMed Linvatec
Linvatec Hall Surgery Advantage Drive System Instruction Manual Rev C Jan 2001
Instruction Manual
192 Pages
Preview
Page 1
The Advantage™ Drive System Instruction Manual
CONTROLLER OPERATION
A
E9000™ HANDPIECES
B
MICROCHOICE® HANDPIECES
C
SHAVER HANDPIECES
D
POWERPRO™ HANDPIECES
E
MINI-DRIVER™ HANDPIECES
F
Proprietary Information This manual contains information deemed proprietary to Linvatec Corporation. The information contained herein, including all of the designs and related materials, is the sole property of Linvatec and/or its licensors. Linvatec and/or its licensors reserve all patent, copyright and other proprietary rights to this document, including all design, manufacturing methodology and reproduction. This document, and any related materials, is confidential and is protected by copyright laws and shall not be duplicated, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language in any form or by any means, electronic, mechanical, magnetic, manual or otherwise, or disclosed to third parties, in whole or in part, without the prior express written consent of Linvatec. Linvatec reserves the right to revise this publication and to make changes from time to time in the contents hereof without obligation to notify any person of such revision or changes, unless otherwise required by law.
Linvatec, Hall, Advantage, Apex, Elite, MicroChoice, Mini-Driver, E9000 and Ultrapower are trademarks or registered trademarks of Linvatec Corporation.
© Linvatec Corporation 2001. All Rights Reserved. Printed in USA
Record the Model and Serial Numbers of the Controller and handpiece(s), and date received. Retain for future reference. Controller Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No.
Serial No Serial No Serial No Serial No Serial No Serial No Serial No Serial No Serial No Serial No
Date Date Date Date Date Date Date Date Date Date
Page
Table of Contents
1.0
INTRODUCTION 1.1
Intended Use... 1
1.2
General Warnings and Cautions... 2
1.3
Symbol Definitions... 4
1.4
System Indicators... 8
1.5
1.4.1
Controller Front Panel (Top Portion)... 8
1.4.2
Controller Front Panel (Bottom Portion)... 9
1.4.3
Controller Back Panel... 10
Handpieces... 11 1.5.1
1.5.2
1.5.3
GROUP 1 HANDPIECES... 11 1.5.1.1
Cranial Perforator Drive Handpiece (E9015)... 11
1.5.1.2
High Speed Shaver (E9005)... 11
1.5.1.3
High Speed Drill (E9010)... 12
GROUP 2 HANDPIECES... 13 1.5.2.1
MicroChoice® Drills and Saws... 13
1.5.2.2
MicroChoice Modular Handpiece (5020-027)... 14
GROUP 3 HANDPIECES... 15 1.5.3.1
1.5.4
1.5.5
Shaver Handpieces... 15
GROUP 4 HANDPIECES... 17 1.5.4.1
PowerPro® Handpieces... 17
1.5.4.2
PowerPro Electric Oscillator Handpiece (PRO2300E)... 18
1.5.4.3
PowerPro Modular (PRO2100E) and Two-Trigger (PRO2200E) Electric Handpieces... 19
GROUP 5 HANDPIECES... 20 1.5.5.1
Mini-Driver Electric Handpiece (K500)... 20
i
Page
Table of Contents
1.0
INTRODUCTION (Continued) 1.6
Footswitches... 21 1.6.1
1.6.2
1.6.3
Footswitch, 2-Pedal (5020-053)... 21 1.6.1.1
Operation with the E9005 High Speed Shaver and Group 3 Handpieces... 21
1.6.1.2
Operation with the E9010 High Speed Drill and Group 2 Drills . 21
1.6.1.3
Operation with Group 2 Saws and E9015 Cranial Perforator Drive... 21
Footswitch, 3-Pedal (C9863)... 22 1.6.2.1
Operation with the E9005 High Speed Shaver and Group 3 Handpieces... 22
1.6.2.2
Operation with the E9010 High Speed Drill and Group 2 Drills . 22
1.6.2.3
Operation with Group 2 Saws and E9015 Cranial Perforator Drive... 22
Tubing Sets... 23 1.6.3.1
Tubing Set without Bur Irrigation (E9414) for use with the E9010 High Speed Drill... 23
1.6.3.2
Tubing Set with Bur Irrigation (E9415/E9415A) for use with the E9010 High Speed Drill... 24
1.6.3.3
Tubing Sets (E9416 and E9417) for use with 4.2 mm and 3.7 mm Shaver Blades... 25
1.6.3.4
Tubing Set (E9418) for use with Group 2 MicroChoice Drills and Saws... 26
ii
Page
Table of Contents
2.0
SYSTEM INSTALLATION and OPERATION 2.1
Installation and Setup... 28 2.1.1
Handpiece Cord Installation... 28
2.1.2
Footswitch and Handpiece to Controller Installation... 29
2.1.3
Controller Options Switches and Settings... 30 2.1.3.1
Controller Options Switch Settings... 31
2.1.3.2
Switch 1 - “Day-to-Day” Memory Storage Operation... 32
2.1.3.3
Switch 2 (Top Portion Only) - Footswitch Mode Selection... 33
2.1.3.4
Switch 3 (Top Portion Only) - Audio Selection... 34
2.1.3.5
Switch 3 (Bottom Portion Only) Handpiece Operating Mode Selection (PowerPro Electric Two-Trigger Handpiece Only)... 34
2.1.3.6
Switch 4 - Oscillate Mode Selection (Shaver Handpieces Only) . 35
2.1.3.7
Switch 5 (Bottom Portion Only) Display Mode Selection... 36
2.1.4
Tubing Set Cassette Installation... 37
2.1.5
Controller Operation... 38
2.1.6
Footswitch Operation (Top Portion Only)... 39 2.1.6.1
Operation of the 2-Pedal Footswitch (REF 5020-053)... 40
2.1.6.2
Operation of the 3-Pedal Footswitch (REF C9863)... 42
iii
Page
Table of Contents
3.0
HANDPIECE OPERATION Group 1 Handpieces... 45 3.1
High Speed Shaver (E9005) Operation... 45
3.2
High Speed Drill (E9010) Operation... 47
3.3
3.2.1
Preoperative Functional Test for the E9010 High Speed Drill... 48
3.2.2
Bur Guard Attachment... 49
3.2.3
Neuro Guard Attachment... 50
3.2.4
Tubing Set Line Attachment... 51
3.2.5
Handpiece Adjustment... 55
Cranial Perforator Drive (E9015) Operation... 56
Group 2 Handpieces... 57 3.4
MicroChoice Drills... 57 3.4.1
Tubing Set Attachment for MicroChoice Handpieces... 57
3.4.2
Preoperative Functional Test for the MicroChoice Elite High Speed Drill . 58
3.4.3
Elite High Speed (5020-025) and Medium Speed Drill (5020-021) Attachments and Accessories... 59
3.4.4
3.4.3.1
Medium (5020-060), Long (5020-061) and Extra-Long (5020-062) Bur Guards... 60
3.4.3.2
20° Angle (5020-063) and Extra-Long 20° Angle (5020-064) Attachments... 61
3.4.3.3
70° Contra Angle with Surgical Head (5020-065), 70° Contra Angle with Dental Head (5020-069), and 90° Angle (5020-066) Attachments... 62
3.4.3.4
Tissue Retractor Bur Guard (5020-068)... 63
3.4.3.5
Laminectomy Bur Guard (5020-067)... 64
Low Speed Drill (5020-026) Operation... 65
iv
Page
Table of Contents
3.0
HANDPIECE OPERATION (Continued) 3.5
3.6
MicroChoice Saws... 67 3.5.1
Reciprocating Saw (5020-023) Operation... 67
3.5.2
Sagittal Saw (5022-022) Operation... 68
3.5.3
Oscillating Saw (5020-024) Operation... 70
3.5.4
Leverless Handpiece Operation... 72
MicroChoice Modular Handpiece (5020-027), Attachments and Accessories... 73 3.6.1
Connecting/Removing Attachments... 73
3.6.2
Pin and Wiredriver Attachments... 75
3.6.3
3.6.2.1
Wiredriver Attachment... 75
3.6.2.2
Pin Driver Attachment... 75
3.6.2.3
Pin and Wire Insertion... 75
Universal Drill Attachment... 76 3.6.3.1
Trinkle Shank Accessories for the Universal Drill Attachment . . 77
3.6.3.2
AO Drill Bits for the Universal Drill Attachment... 77
3.6.4
All Jacobs Chuck Attachments... 78
3.6.5
Group 2 Handpiece Cord Button Operation... 79 3.6.5.1
Button Operation Using the Elite High Speed Drill, Medium Speed Drill, Saws and Modular Handpiece... 79
3.6.5.2
Button Operation Using the Low Speed Drill (Port 2 Only)... 80
Group 3 Handpieces... 81 3.7
Preoperative Functional Test for Group 3 Shaver Handpieces... 82
3.8
Accessory Insertion and Removal... 83 3.8.1
Basic (C9820, MC9820), 2-Button (C9824), and Full-Function (C9828, MC9828) Large Shaver Handpieces... 83
3.8.2
Advantage™ Basic (D9820) and 2-Button (D9824) Shaver Handpieces . . . 83
3.8.3
Small Shaver Handpieces (MC9840, C9840)... 84
v
Page
Table of Contents
3.0
HANDPIECE OPERATION (Continued) 3.9
Group 3 Handpiece Operation... 85 3.9.1
3.9.2
3.9.3
Activating the Handpiece... 85 3.9.1.1
Using the Footswitch to Activate the Handpiece... 85
3.9.1.2
Using the Handpiece ON/OFF Button to Activate the Handpiece 85
Changing Handpiece Operating Speeds... 85 3.9.2.1
Using the Console Speed Buttons to Change Handpiece Speed . . 85
3.9.2.2
Using the Handpiece Direction Select and ON/OFF Buttons to Change Handpiece Speed... 86
3.9.2.3
Using the Handpiece Speed Buttons to Change Handpiece Speed 86
3.9.2.4
Using the Handpiece Cord Button to Change Handpiece Speed. . 86
Changing Handpiece Operating Directions... 87 3.9.3.1
Using the Footswitch to Change Handpiece Direction (Applies only to the 2-Pedal Footswitch - REF 5020-053)... 87
3.9.3.2
Using the Handpiece Direction Select Button... 87
3.10 Group 3 Handpiece Cord Button Operation... 88 3.10.1 Using APEX® and Advantage Shaver Handpieces in Aspirate Mode... 89 3.10.2 Using the MicroChoice Basic Shaver Handpiece (MC9820) in Aspirate Mode... 89 3.10.3 Using the Full-Function Large Shaver Handpieces (MC9828, C9828) in Aspirate Mode... 90 Group 4 Handpieces... 92 3.11 PowerPro Electric Modular Handpiece... 92 3.12 PowerPro Electric Two-Trigger Handpiece... 96 3.12.1 Default Mode of Operation... 96 3.12.2 Swiss Mode of Operation... 100 3.13 PowerPro Electric Oscillator Handpiece Operation... 104 3.13.1 Connecting Blades to the PowerPro Oscillator Handpiece... 105
vi
Page
Table of Contents
3.0
HANDPIECE OPERATION (Continued) 3.14 PowerPro Attachments... 106 3.14.1 Connecting and Removing Attachments... 108 3.14.2 Reaming Attachments... 109 3.14.2.1 AO Reamer (PRO2040)... 109 3.14.2.2 1/4” High Torque Jacobs Chuck (PRO2042)... 110 3.14.2.3 Zimmer/Hudson Reamer (PRO2047)... 110 3.14.2.4 Aesculap Reamer (PRO2060)... 111 3.14.3 Drilling Attachments... 111 3.14.3.1 Trinkle /AO Attachment (PRO2029)... 111 3.14.3.2 5/32” Jacobs Chuck (PRO2030)... 113 3.14.3.3 1/4” Jacobs Chuck (PRO2041)... 113 3.14.3.4 Zimmer/Hudson Drill (PRO2046)... 113 3.14.3.5 1/8” Keyless Chuck (PRO2050)... 114 3.14.3.6 AO Drill (PRO2070)... 114 3.14.4 Wire and Pin Driving Attachments... 115 3.14.4.1 Wire Driver (PRO2028)... 115 3.14.4.2 Pin Driver (PRO2032)... 115 3.14.4.3 Insertion of Pins and Wires... 116 3.14.5 Sawing Attachments... 117 3.14.5.1 Sagittal Saw Attachment (PRO2043)... 117 3.14.5.2 Reciprocating Saw Attachment (PRO2045)... 118 3.15 PowerPro Modular and Two-Trigger Electric Handpiece Attachment Matrix... 120
vii
Page
Table of Contents
3.0
HANDPIECE OPERATION (Continued) 3.16 Mini-Driver Handpiece, Attachments and Accessories... 122 3.16.1 Attachments and Accessories... 122 3.16.2 Connecting/Removing Attachments... 122 3.16.3 Pin and Wiredriver Attachments... 123 3.16.3.1 Pin Driver Attachment (K211)... 123 3.16.3.2 Wiredriver Attachment (K111A)... 124 3.16.3.3 Pin and Wire Insertion... 124 3.16.4 Sagittal Saw Attachment (K220)... 126 3.16.5 All Jacobs Chuck Attachments... 127 3.16.5.1 5/32” Jacobs Chuck (K110)... 127 3.16.5.2 1/4” Jacobs Chuck (K109)... 128 3.16.5.3 High-Torque Jacobs Chuck (K210)... 128 3.16.6 Trinkle Chuck Attachment (K112)... 129 3.16.6.1 Automatic Screwdrivers for the Trinkle Chuck Attachment (D520/D524)... 129 3.16.7 Hudson Chuck Attachment (K113)... 131 3.16.8 ASIF/AO Twist Drill Chuck Attachment (K114A)... 131 3.17 Handpiece Default Settings... 132
viii
Page
Table of Contents
4.0
MAINTENANCE 4.1
Cleaning and Sterilizing... 139 4.1.1
Cleaning Precautions... 139
4.1.2
Controller and Footswitch Cleaning Instructions... 139
4.1.3
Handpiece Cleaning Instructions... 140 4.1.3.1
MicroChoice, Advantage, and APEX Shaver Handpiece Cleaning Instructions... 140
4.1.3.2
E9000 Handpiece Cleaning Instructions... 141
4.1.3.3
MicroChoice Drills, Saws, Modular Handpiece, Mini-Driver Handpiece and PowerPro Handpiece Cleaning Instructions... 142
4.1.4
Attachment Lubricating Instructions... 143
4.1.5
Sterilization Information... 144 4.1.5.1
Sterilization Warnings, Precautions and Notes... 144
4.2
Troubleshooting... 147
4.3
Calibration... 151
4.4
Fuse Replacement and Voltage Selection Means... 151
4.5
Bur Guard Testing Procedure... 152
4.6
Ground Continuity Test... 153
4.7
Leakage Test... 153
4.8
Theory of Operation... 154
ix
Page
Table of Contents
5.0
6.0
TECHNICAL SPECIFICATIONS 5.1
Controller... 155
5.2
Power Cord Requirements... 156
5.3
System Environmental Requirements... 157
5.4
Footswitches (5020-053 and C9863)... 157
5.5
Handpiece Cords... 157
5.6
Tubing Sets... 158
5.7
Handpieces... 159
CUSTOMER SERVICE, REPAIRS and WARRANTY 6.1
Customer Service... 168
6.2
Handpieces, Attachments and Accessories... 170
6.3
Linvatec and Hall® Surgical Instrument Warranty... 176
x
1.0 INTRODUCTION
The fields of application for the Advantage Drive System include Arthroscopic, Foot, Hand, Medial Sternotomy, Neurosurgical, Orthopedic, Otolaryngological, Plastic/Reconstructive and Spinal surgical procedures.
t is recommended that personnel study this manual before attempting to operate, clean or sterilize the Advantage™ Drive System. The safe and effective use of this equipment requires the understanding of and compliance with all warnings, cautionary notices and instructions marked on the product and included in this manual.
I
1.1
Intended Use
he Advantage Drive System combines the functionality of the APEX Drive System, the E9000™ System and the PRO2000 PowerPro™ System into one unit. The top portion of the controller is a combination handpiece-drive and low-flow irrigation pump that operates a variety of handpieces consisting of drills, shavers, and saws, in conjunction with assorted attachments to perform bone cutting, reaming and drilling, and soft tissue resection.
T
The irrigation pump and tubing sets provide a sterile fluid supply at the surgical site to the blades, burs and drill bits, to prevent clogging and provide lavage and cooling. The bottom portion of the controller functions as a universal electric power source for an instrument system consisting of drills, shavers, saws and assorted modular attachments. The instruments perform bone cutting, reaming, drilling and driving screws, as well as soft tissue resection. This portion is designed for soft tissue resection, trauma procedures, total hip/knee arthroplasty, and ACLs.
1
•
1.2
1.
2.
General Warnings and Cautions This equipment is designed for use by medical professionals completely familiar with the required techniques and instructions for use of the equipment. Read and follow all warning and caution notices and instructions marked on the product and included in this manual. The Advantage System’s irrigation pump and tubing sets are contraindicated for distention.
3.
Do not open the controller chassis, as this may void the warranty. There are no user-serviceable parts inside. Removing the cover may introduce an electric shock hazard by exposing you to dangerously high voltages or other risks.
4.
Eye protection is recommended when operating equipment.
5.
Use only associated Hall® Surgical and Linvatec attachments and accessories (saw blades, bits, etc.) as defined within the descriptions of each attachment.
6.
Handle all equipment carefully. If any equipment is dropped or damaged in any way, return it immediately for service.
7.
Inspect all equipment for proper operation.
•
Ensure all attachments and accessories are correctly and completely attached to the handpiece.
8.
Handpieces are factory sealed. Do not disassemble or lubricate.
9.
Never operate the MicroChoice® Elite™ High Speed Drill (REF 5020-025) or Medium Speed (REF 5020-021) Drill without a bur and appropriate bur guard in place and the collet locked. Damage to the collet may result.
10. When operating shaver blades or burs in any Group 3 handpiece (reference “ Group 3 Handpieces” on page 81) the shaver handpiece suction port valve must be in the open or ON position, and the shaver blade or bur must be within the distention fluid of the joint. Otherwise, damage to the blade hub or bur hub will result. 11. Continually check all handpieces and attachments for overheating. If overheating is sensed, immediately discontinue use and return equipment for service. Overheating of the bur, bit or blade may cause damage to the bur, bit or blade and may cause thermal necrosis. 12. Do not attach, insert or remove accessories or attachments while the handpiece is operating. Place the handpiece safety switch, if equipped with one, to the appropriate safe position, prior to installation or removal of items.
Prior to each use, perform the following: •
Perform the required Performance Tests for each associated handpiece prior to each use.
13. Do not stall handpieces, damage can occur. 14. Do not operate the E9010 High Speed Drill or MicroChoice Elite™ High Speed Drill (REF 5020-025) with the collet open. Damage will occur. 2
15. Do not handle the controller, handpieces or footswitch by the cord. Do not pull on the cord to remove it from the handpiece or controller.
(b) While holding the bur, spin the guard. The guard should spin freely around the bur shaft without resistance.
16. Do not excessively bend or kink the instrument handpiece cord or power cord. Always inspect cords for signs of excessive wear or damage. If wear or damage is found, discontinue use and replace immediately. 17. Always inspect for bent, dull or damaged burs, blades or drill bits before each use. Do not attempt to straighten or sharpen. Do not use if damaged. After use, dispose of properly.
(c) Attach the guard and bur to the handpiece.
18. Always use a bur of the proper length for the particular guard.
(d) Operate the handpiece for approximately 30 seconds. Stop the handpiece and carefully feel the end of the guard where the guard encases the shaft of the bur for overheating. If overheating is noticed, return guard for service.
19. Do not use burs for plunge cutting. Damage or injury may occur. 20. Bur and neuro guards should be checked frequently. Overheating can occur if bur guard bearings are worn or not kept clean. Perform the following steps to verify guards are in good operating condition.
21. Tubing sets and E9000 Disposable Irrigation Tips are one-time use only. Do not resterilize or reuse. After use, dispose of properly. 22. All associated E9000 Tubing Sets are latex free.
NOTE: Bur and neuro guards are to be returned to the factory or a Linvatec authorized service facility for routine maintenance every six (6) months.
LATEX
23. Ensure the optical cassette sensor next to the pump cassette mechanism is kept clean at all times.
(a) Remove the guard from the handpiece and insert a bur into the nose of the guard.
24. Reaming attachments (red colored rings) for the PowerPro handpieces are not designed to be used in any screw, tap or oscillate drill modes.
3
25. When using the PowerPro Modular or Two-Trigger Handpiece with any saw attachment (Sagittal Saw - PRO2043, Reciprocating Saw - PRO2045), do not operate in the oscillate mode. These handpieces should only be operated in the drill/ream mode when used with these attachments.
1.3
Symbol Definitions Controller Markings Stand-by (ON/OFF)
Attention, consult accompanying documents
26. After each use, thoroughly clean the controller, handpieces, attachments and accessories (See “4.1 Cleaning and Sterilizing” on page 139).
Warning - For continued protection against risk of fire, replace only with the same type and rating fuse
27. Equipment grounding is vital to ensure safe operation. Plug the power cord into a properly earthed mains supply outlet whose voltage and frequency characteristics are compatible with those listed on the unit or in this manual. Do not use plug adapters or extension cords; such devices defeat the safety ground and could cause injury.
Mains Fuse Type B equipment
~
28. Do not use in the presence of flammable anesthetics, gases, disinfecting agents, cleaning solutions, or any material susceptible to ignition due to electrical sparking.
Alternating Current Protective earth ground
Footswitch Connection
29. The service intervals, as listed in “Table 4: Maintenance Schedule” on page 135, are required to keep the handpieces at their optimum operating performance.
E9000 Series Handpiece Connection MicroChoice Drills and Saws Connection MicroChoice, Advantage, and APEX Shaver Handpiece Connection PowerPro Handpiece Connection Speed (rpm) Selection
4
Irrigation (Flow) Control
Full-Function Handpiece Markings On/Off non-mains control
Adjustment to a maximum (Increase)
Increase Speed Button
Adjustment to a minimum (Decrease)
Decrease Speed Button
PRESET
Predetermined setting for irrigation flow using the E9010 Drill with the E9414 Tubing Set
Directional Selection Button (Forward/Reverse/ Oscillate)
RPM
Revolutions Per Minute
Max
Maximum Operating Speed
Min
PowerPro Handpiece Markings Indicates the safe or off position (only for a part of equipment) of the handpiece thumb lever
Minimum Operating Speed
Footswitch Markings
PowerPro Modular Handpiece Specific Markings
Shaver Handpieces
Indicates the forward position of the handpiece thumb lever
Directional Selection (Forward/Reverse/Oscillate) Variable Speed, ON/OFF
Indicates the reverse position of the handpiece thumb lever
MicroChoice or Hall Handpiece Forward Variable Speed
PowerPro Oscillator Handpiece Specific Markings
Reverse Variable Speed Oscillate
Indicates the on or run position (only for a part of equipment) of the handpiece thumb lever
5
PowerPro Two-Trigger Handpiece Specific Markings
1 min
Indicates the DRILL/ REAM position of the handpiece thumb lever
3 min
Indicates the SCREW position of the handpiece thumb lever
Not to be used for plunge cutting
Single Use Only
Indicates the reverse, or counterclockwise, direction of rotation of the handpiece trigger
Latex Free LATEX
Indicates oscillate direction of the handpiece trigger
Eye Protection Required No user service recommended. Refer servicing to qualified Linvatec service personnel.
Mini-Driver™ Handpiece Specific Markings
Indicates product component should not be sterilized.
Indicates the safe or off position (only for a part of equipment) of the handpiece safety switch
Indicates product component should not be immersed in any type of fluid.
Indicates the on or run position (only for a part of equipment) of the handpiece safety switch Indicates on in the forward, or clockwise, direction of rotation
R
Indicates on in the reverse, or counterclockwise, direction of rotation
Duty Cycle: 3 minutes off.
Miscellaneous Markings
Indicates the forward, or clockwise, direction of rotation of the handpiece trigger
F
Duty Cycle: 1 minute on.
Indicates handpiece should not be immersed in any type of fluid. Indicates product should not be oiled or lubricated. Lock Unlock
6
LOCK UN
Indicates the blade collet unlock position of the D9820 and D9824 Advantage Handpieces
Handpiece “Stalled” and “Maximum Torque” messages symbol
N ???? Graphical Display Symbols Drill - forward direction Drill - reverse direction OscDrill - oscillate function Tap - forward direction Tap - reverse direction Screw - forward direction Screw - reverse direction Irrigation message symbol Saw operating symbol Handpiece “Aspirate Mode” symbol Handpiece in “safe” position Handpiece “Motor Overspeed” message symbol
7
S
Handpiece “Magnetic Field” message symbol Handpiece Service Required message
❶
❷
❸
❹
❺
❻
❼
❽
❾
❿
PORT 1 PORT 2
PORT 3
11
1.4
13
12
(REF E9320) or the Mini-Driver Adapter (REF K501) - this receptacle will also accept and operate the remainder of the Group 3 Handpieces (REF C9820, C9824, C9828, C9840) or the Mini-Driver Handpiece (REF K500). This receptacle is NOT designed to operate any Group 4 PowerPro Handpieces (REF PRO2100E, PRO2200E, PRO2300E).
System Indicators
1.4.1
Controller Front Panel (Top Portion)
❶ Footswitch Receptacle - Accepts the footswitch cord connector.
❷ APEX® Handpiece Receptacle
NOTES:
(Port 1 - See page 27) - Directly accepts the handpiece cord connector of any APEX handpiece (REF C9820, C9824, C9828, C9840).
1.
All cord receptacles and connectors are keyed to prevent incorrect insertion. Likewise, receptacles and connectors are color-coded to easily recognize where to connect each cord (footswitch receptacle and connector - yellow, handpiece receptacles and connectors - red or grey).
2.
Only one handpiece can be connected at a time in either port of the top section. If a handpiece is connected while another handpiece is already connected, “HP SERVICE REQUIRED” will display. BOTH handpieces must be removed to clear the fault condition.
❸ Handpiece Receptacle (Port 2 - See page 27) - Directly accepts the handpiece cord connector of all Group 1 Handpieces (REF E9005, E9010, E9015), all Group 2 Handpieces (REF 5020-021, 022, -023, -024, -025, -026, -027), and most Group 3 Handpieces (REF D9820, MC9820, D9824, MC9828, and MC9840). However, with the addition of one of two handpiece adapters - the APEX Adapter 8