User’s Guide Before connecting, operating or adjusting this product, please read this instruction booklet carefully and completely.
Table of Contents Introduction... 3 Symbol Definitions... 4 Safety Instructions...6 Caution...8 FCC Information... 10 Parts, VP4826 / VP4832... 14 Connectors, VP4826 / VP4832... 15 Mechanical Drawing, VP4826 / VP4832... 16 Controls... 18 Power Management... 19 On Screen Display (OSD)... 20 Remote Control (VP4832 only)... 30 Standard Signal Tables... 31 Signal Connector Pin Assignments... 32 General Specification VP4826... 34 VP4832... 35 Cleaning Instructions... 36 Contacts... 38
The specifications and information in this document are subject to change without notice. 2 - English
Overview
VP4826 / VP4832, from ConMed Corporation is a high-end surgical display monitor designed for advanced digital OR applications. This medical display is uniquely equipped to handle tasks in the demanding operating room environment. VP4826 / VP4832 has the latest advanced features built in. The input/output options have been expanded, including dual DVI and on-board DC out power for smaller component needs. This unit features LED backlight technology. We have implemented methods to fine-tune the properties of ConMed displays. Distinct color space settings have been calibrated to well-known surgical color preferences, providing the user with the ideal surgical visualization system. Features in VP4826 / VP4832 include: • • • • •
Rapid signal detection, robust mode tables Artifact-free HD images Fanless - sterile field compatible Calibrated to clinical color Image pan, zoom, freeze, picture-in-picture
ConMed Corporation has solutions for managing your operating room video signals. Our products are engineered and built for compatibility with other highly specialized surgical and diagnostic equipment used in surgical suites, operating rooms, emergency rooms, and procedural facilities.
English - 3
Symbol Definitions The following symbols appear on the product, its labeling, or the product packing. Each symbol carries a special definition, as defined below:
4 - English
Dangerous : High Voltage
Consult accompanying documents
Direct Current
Power adapter
Indicates protective earth ground
lndicates equipotential earth ground
DC Power control switch
Indicates top-bottom direction
Do not get wet
Fragile
Consult the operating instructions.
Maximum Stacking
Indicates the manufacturing date
Indicates the manufacturer
Representative in the USA
Authorized representative in the European community
Serial Number
Humidity limitation
Temperature limitation
Atmospheric pressure limitation
Indicates proof of conformity to applicable European Economic Community Council directives and to harmonized standards published in the official journal of the European Communities. Medical Equipment is in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1 in regards to electric shock, fire hazards, and mechanical hazard. Tested to comply with FCC Class B standard. Indicates the display is approved according to the CCC regulations. China RoHS labels. This symbol indicates that the waste of electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your equipment. Language: English Note: A printed copy of the manual in English is provided with the product. Users within EU member states, please contact local distributor for other languages or refer to the CD manual enclosed with the product. This applies to EU member states where the product has been purchased through authorized channels.
English - 5
Safety Instructions On Safety 1. Before connecting the AC power cord to the DC adapter outlet make sure the voltage designation of the DC adapter corresponds to the local electrical supply. 2. Never insert anything metallic into the cabinet openings of the medical LCD monitor. Doing so may create the danger of electric shock. 3. To reduce the risk of electric shock, do not remove cover. No user-serviceable parts inside. Only a qualified technician should open the case of the medical LCD monitor. 4. Never use your medical LCD monitor if the power cord has been damaged. Do not allow anything to rest on the power cord, and keep the cord away from areas where people can trip over it. 5. Be sure to hold the plug, not the cord,when disconnecting the medical LCD monitor power cord from an electric socket. 6. Unplug your medical LCD monitor power cord when it is going to be left unused for an extended period of time. 7. Unplug your medical LCD monitor power cord from the AC outlet before any service. 8. If your medical LCD monitor does not operate normally, in particular, if there are any unusual sounds or smells coming from it, unplug it immediately and contact an authorized dealer or service center. 9. Please contact the manufacturer if the set should be installed in an inaccessible area. Warning: Do not touch input or output connectors and the patient simultaneously. Warning: This medical LCD monitor is intended for connection to input/output signals and other connectors that comply with relevant IEC standard (e.g.,IEC60950 for IT equipment and IEC60601 series for medical electrical equipment). In addition, all such combination-system shall comply with the standard IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-1, respectively, safety requirements for medical electrical systems. Any person who has formed a combination-system is responsible for the system to comply with the requirements of IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-1, respectively. If in doubt, contact qualified technician or your local representative. Warning: To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. Power supply (AC/DC Adapter) is specified as a part of the LCD Color Display. Do not to position equipment so that it is difficult to disconnect the power cord plug from the appliance inlet. Warning: Do not modify this equipment without authorization of the manufacturer.
6 - English
On installation 1. Openings in the medical LCD monitor cabinet are provided for ventilation. To prevent overheating, these openings should not be blocked or covered. If you put the medical LCD monitor in a bookcase or some other enclosed space, be sure to provide adequate ventilation. 2. Put your medical LCD monitor in a location with low humidity and a minimum of dust. 3. Do not expose the medical LCD monitor to rain or use it near water (in kitchens, near swimming pools, etc.). If the medical LCD monitor accidentally gets wet, unplug it and contact an authorized dealer immediately. You can clean the medical LCD monitor with a damp cloth if necessary, but be sure to unplug the medical LCD monitor first. 4. Place your medical LCD monitor near an easily accessible AC outlet. 5. High temperature can cause problems. Don’t use your medical LCD monitor in direct sunlight and keep it away from heaters, stoves, fireplaces,and sources of heat. 6. Don’t place your medical LCD Monitor on an unstable stand, Medical LCD monitor may malfunction or fall. 7. This medical LCD monitor should not topple over when tilted at a 5° angle, in any position, during NORMAL USE, excluding transport. 8. In the position specified for transport, medical LCD monitor shall not overbalance when tilted at a 10 degree angle. 9. When carrying this product, please use both handles (if included) on the left and right side of the product, and carry using two people. If you want the product to be installed in another place, please call ConMed. 10. Do not use other cables or accessories that are not provided. 11. Do not lay this monitor on the other equipment.
Environmental Conditions for Operation and Storage
Temperature range within 0°C to 40°C(operation), -20°C to 60°C (storage) Relative humidity range 10% to 85% Atmospheric pressure range within 500 to 1060hPa.
Intended Use
This Medical LCD Monitor is an accessory intended for use with Medical Equipment to display alphabetical, numerical and graphical data.
English - 7
Cautions Caution CAUTION RISK OF ELECTRICAL SHOCK DO NOT OPEN
This symbol alerts the user that important literature concerning the operation of this unit has been included. Therefore, it should be read carefully in order to avoid potential problems. This symbol warns users that un-insulated voltage within the unit may have sufficient magnitude to cause electrical shock. Therefore, it is dangerous to make contact with any part inside the unit. To reduce the risk of electrical shock, DO NOT remove cover (or back). There are no user-serviceable parts inside. Refer servicing to qualified service personnel. To prevent fire or shock hazards, do not expose this unit to rain or moisture. Do not use this unit’s polarized plug with an extension cord receptacle or other outlets unless the prongs can be fully inserted. This display is designed to meet the medical safety requirements for a patient vicinity device. This device may not be used in connection with life support equipment. Underwriters Laboratories (UL) Classification: UL safety Compliance: This medical LCD monitor is U.L. Classified WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1/CAN/CSA C22.2 NO. 601.1 EEC Safety Compliance: This medical LCD monitor unit meets the requirements of EN-60601-1 so as to conform to the Medical Device Directive 93/42/EEC (general safety information). Use 120V rating 5-15P type plug only in the U.S. This medical LCD monitor complies to the above standards only when used with the supplied medical grade power supply. 26” & 32” LCD Monitor (VP4826 / VP4832) - BPM150S24F10(BRIDGE POWER CORP.) Caution: Make sure the power cord is the correct type that is required in your area. This medical LCD monitor has a universal power supply that allows operation in either 100-120V AC or 200240V AC voltage areas (no user adjustment is required). Use the proper power cord with correct attachment plug type. If the power source is 120 V AC, use a power cord which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for 125 volts AC with UL and C-UL approvals. If the power source is a 240 V AC supply, use the tandem (T blade) type attachment plug with ground conductor power cord that meets the respective European country’s safety regulations. 8 - English
The hospital-grade plug for medical products intended for use in Denmark has DEMKO approval and is rated 13 amps at 250Vac. Plug is recommended for use in medical applications and specifications are being added to the standard SB 107-2-D1. Plug mates with maker’s Danish hospital-grade socket. Hospital sockets have slightly different shaped openings allowing only the hospital plug, not the standard Danish plug, to be inserted, to protect the ac circuit in specific medical settings. A ground post, located on the back of the display, may be used for the purpose of grounding the display’s chassis. Any such ground must be installed in accordance with applicable electrical codes. The ground post is shown on the mechanical drawing found in this user’s guide. Recycling Follow local governing ordinances and recycling plans regarding the recycling or disposal of this equipment.
Cleaning Instructions Follow your hospital protocol for the handling of blood and body fluids. Clean the display with a diluted mixture of mild detergent and water. Use a soft towel or swab. Use of certain detergents may cause degradation to the labels and plastic components of the product. Consult cleanser manufacturer to see if agent is compatible. Do not allow liquid to enter the display.
Servicing
Do not attempt to service the medical LCD monitor yourself, as opening or removing covers may expose you to dangerous voltages or other hazards, and will void the warranty. Refer all servicing to qualified service personnel. Unplug the medical LCD monitor from its power source and refer servicing to qualified personnel under the following conditions: - If the power cord or plug is damaged or frayed. - If liquid has been spilled into the medical LCD monitor. - If objects have fallen into the medical LCD monitor. - If the medical LCD monitor has been exposed to rain or moisture. - If the medical LCD monitor has been subjected to excessive shock by being dropped. - If the cabinet has been damaged. - If the medical LCD monitor seems to be overheated. - If the medical LCD monitor emits smoke or abnormal odor. - If the medical LCD monitor fails to operate in accordance with the operating instructions.
Accessories
Use only accessories specified by the manufacturer, or sold with the medical LCD monitor.
Classification
- Protection against electric shock : Class I including AC/DC Adapter - Applied Parts : No Applied Parts - Degree of safety in the presence of flammable anesthetics mixture with air or with oxygen or with nitrous oxide. Not suitable for use in the presence of a flammable anesthetics mixture with oxygen or with nitrous oxide. - Mode of operation : Continuous. English - 9
FCC Information This medical LCD monitor unit has been tested and found to comply with the limits of a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against interference. This monitor can radiate radio frequency energy and, if not installed and used in accordance with the instructions, it may interfere with other radio communications equipment. There is no guarantee that interference will not occur in a particular installation. If this equipment is found to cause harmful interference to radio or television reception, the user is encouraged to try to correct the interference by carrying out one or more of the following measures: 1. Reorient or relocate the receiving antenna. 2. Increase the distance between the medical LCD monitor and the subject of interference. 3. Plug the monitor into an outlet on a different electrical circuit than that to which the subject of interference is connected. 4. Consult the dealer or an experienced radio/TV technician for help.
NOTICES TO USER This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
FCC WARNING This medical LCD monitor generates or uses radio frequency energy. Changes or modifications to this medical LCD monitor may cause harmful interference unless the modifications are expressly approved in the instruction manual. The user could lose authority to operate this equipment if an unauthorized change or modification is made.
10 - English
1. Guidance and manufacturer’s declaration - electromagnetic emission The medical LCD monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the medical LCD monitor should assure that it is used in such an environment. Emission test
Compliance
Electromagnetic environment -guidance
RF Emissions CISPR 11
Group 1
The medical LCD monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
D
Voltage fluctuations IEC 61000-3-3
Complies
The medical LCD monitor is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
2. Guidance and manufacturer’s declaration - electromagnetic immunity This medical LCD monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the medical LCD monitor should assure that it is used in such an environment. Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment-guidance
Electrostatic discharge(ESD) IEC 61000-4-2
6 kV contact 8 kV air
6 kV contact 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%
Electrical fast transient/burst IEC 61000-4-4
2 kV for power supply lines 1 kV for input/ output lines
2 kV for power supply lines 1 kV for input/ output lines
Main power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
1 kV differential mode 2 kV common mode
1 kV differential mode 2 kV common mode
Main power quality should be that of a typical commercial or hospital environment.
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80MHz
3 Vrms 150 kHz to 80MHz
Portable and mobile RF communications equipment should be used no closer to any part of the LCD monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance : d
[
]√
Where P is the maximum output power rating of the transmitter in watts (W)
English - 11
3. Guidance and manufacturer’s declaration - electromagnetic immunity This medical LCD monitor is intended for use in the electromagnetic environment specified below. The customer or the user of monitor should assure that it is used in such an environment. Immunity test Power frequency (50/60Hz) Magnetic field IEC 61000-4-8 Voltage dips, short Interruptions and voltage variations on power supply input lines IEC 61000-4-l 1
IEC 60601 Test level
Compliance level
Electromagnetic environment-guidance
3.0 A/m
3.0 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
<5% UT (>95% dip in UT) for 0.5 cycle
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycle
40% UT (60% dip in UT) for 5 cycle
70% UT (30% dip in UT) for 25 cycle
70% UT (30% dip in UT) for 25 cycle
Main power quality should be that of a typical commercial or hospital environment. If the user of monitor requires continued operation during power mains interruptions, it is recommended that monitor be powered from an uninterruptible power supply or a battery.
<5% UT (<95% dip in UT) for 5 sec.
<5% UT (<95% dip in UT) for 5 sec.
Note: UT is the A.C. mains voltage prior to application of the test level. Recommended separation distance
Radiated RF IEC 61000-4-3
12 - English
3 V/m 80.0 MHz to 2.5 GHz
3 V/m 80.0 MHz to 2.5 GHz
[
]√
80MHz to 800MHz
[
]√
80MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
4. Recommended separation distances between portable and mobile RF communications equipment and this medical LCD monitor. The medical LCD monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the medical LCD monitor as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter[m] Rated maximum output power [W] of transmitter
150kHz to 80MHz
[
]√
V1=3Vrms
80MHz to 800MHz
[
]√
E1=3V/m
800MHz to 2.5GHz
[
]√
E1=3V/m
0.01
0.116
0.116
0.2333
0.1
0.368
0.3687
0.7378
1
1.166
1.1660
0.2333
10
3.687
3.6872
0.7375
100
11.660
11.6600
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
English - 13
Parts Monitor
26” VP4826 32” VP4832
26” VP4826 or
32” VP4832
Accessories
User Manual
AC-DC Adaptor (BPM150S24F10)
AC Power Cord (6ft/1.8/m US,UK,EU, China) (Hospital Grade)
DVI-D Cable (6ft/1.8m)
BNC Cable (6ft/1.8m)
DC Cable Terminal Male / female
Remote Control (VP4832 only)
Screw BH M4 x 14
14 - English
Connectors 26” VP4826 Monitor Connector
32” VP4832 Monitor Connector
English - 15
Mechanical Product Drawing 26” VP4826 Dimension
88.8 / 3.5 668 mm / 26.3 in.
39.5 1.56
421 16.57
OSD membrane switch 88.8 / 3.5
.5 56
100 / 3.94
16 - English
Mechanical Product Drawing 32” VP4832 Dimension
80.5 / 3.17 35.5 1.40
770 mm / 30.31 in. 702 mm / 27.64 in.
396.4 15.60
471.5 18.56
141.5 5.57
OSD membrane switch
/ 3.17
400 / 15.75 300 / 11.81 200 / 7.87 100 / 3.94
12-M4 VESA MOUNT HOLES
4-M6 VESA MOUNT HOLES
100 3.94
200 7.87
141.5 5.57
English - 17
Controls
On Screen Display (OSD) Buttons An 8 button keypad, located in bottom right corner on the front of the display, allows the user to make adjustments to various display parameters using the On Screen Display (OSD) system. QUICK ADJUST. Press to decrease the display contrast. No need to enter the OSD main menu. When the OSD menu is activated, press to decrease the adjustment of the selected function.
QUICK ADJUST. Press to decrease the display brightness. No need to enter the OSD main menu. When the OSD menu is activated, press to move the menu selection downward.
Press to enable PIP (picture in Picture) function. Select from: PIP PBP1 PBP2
No Light - Normal, display has power and an active signal. Blinking Light - Standby Mode, no active signal is present. Constant Light - Off Mode, power to the LCD screen has been turned off using the OSD power button, however the display unit is still connected to a power source. Note 1 : Normal and Off Mode behavior of the Power Indicator Light can be changed according to the requirements of the customer. Constant light can be changed to indicate Normal if desired. Note 2 : The main AC power switch on the display’s back panel should be in the ON position. The front OSD power button is used to turn on the LCD screen on and off.
Press to show the input selection menu and to change the display signal source. Input choices:
DVI 1, DVI 2, SDI 1, SDI 2, SOG 1, SOG 2, VGA, RGBS, YPbPr, S-VIDEO, C-VIDEO
18 - English
QUICK ADJUST. Press to increase the display contrast. No need to enter the OSD main menu.
QUICK ADJUST. Press to increase the display brightness. No need to enter the OSD main menu.
When the OSD menu is activated, press to increase the adjustment of the selected function.
When the OSD menu is activated, press to move the menu selection upward.
Press to activate the OSD menu. When the OSD menu is active, press to exit from the main menu or submenu.
Press to turn power on/off to the display’s front LCD screen.
GPIO There are four pins on the RJ9 GPIO connector. Each pin has a pre programmed function assigned to it. The function is initiated when the pin is grounded. Pin 1. Primary and Secondary Swap. Grounding this pin will swap the primary and secondary image.
Pin 2. PIP, PBP1, PBP2, Single. Continuously grounding this pin causes the position and size choices to cycle.
Pin 3. Record Indicator. The record indicator is displayed in the top left corner when the pin is ground to pin 4. The indicator will vanish when the contact is opened.
Pin 4. Connector Ground. This is the common ground location. Monitor RJ9
Power Management This monitor does not adhere to the VESA DPMS standard when no signal is present on the video input. Model
Status
LED sign
Power Consumption
VP4826 (LCD)
Normal mode Standby mode OSD power off DC power off
Off Blinking Constant Off
<70W (83VA) <16W (38VA) <15W (37VA) <0.5W (17VA)
VP4832 (LCD)
Normal mode Standby mode OSD power off DC power off
Off Blinking Constant Off
<75W (86VA) <16W (38VA) <15W (37VA) <0.5W (17VA)
English - 19
On Screen Display (OSD) Menus ConMed display monitors come equipped with a rich set of features for system set-up, image adjustments, and screen layout control. These features are managed through the On Screen Display, or OSD. Some options presented in the OSD are contextual and vary depending on the active input signal. See the Controls section for a complete description of each OSD button.
1. Enter the OSD
To activate the OSD menu, press the MENU button on the front of the display monitor. To close the OSD menu, press the menu button to exit from the main menu or a sub menu.
2. Pick a Main Menu Category After entering the OSD, use the UP and Down to navigate to a main menu category.
or
buttons on the front of the display monitor
The ADJUST menu controls brightness, contrast, and more.
The COLOR SETTING menu controls preset or customized color settings.
The IMAGE menu controls horizontal / vertical positioning and sharpness.
The SETUP menu controls language, OSD behavior, and more.
The PIP menu controls picture in picture layouts.
3. Pick a Submenu Category After using the UP and Down buttons to navigate to the desired main menu category, press the button to enter the submenus associated with the selected main menu.
+
20 - English