ConMed
1.5 mm Percutaneous Drill Bit, 170 mm Length, SmartPin/SmartPin PDX
Instrument Disinfection Guide Reusable Medical Devices and Accessories
10 Pages
Preview
Page 1
Instrument Disinfection Guide Reusable Medical Devices and Accessories In accordance with ISO 17664
CONMED Corporation 525 French Road, Utica, NY 13502-5994 USA
US: 1-866-4CONMED (426-6633) (US) or [email protected] International: +1 727-219-3310 or [email protected] www.conmed.com
P000049261
A
05/2023
Table of Contents
1.0
IMPORTANT INFORMATION 1.1
2.0
Page
Introduction... 1
PRECAUTIONS, WARNINGS, and NOTES (Definitions) 2.1 2.2
Precautions... 2 Warnings... 2
3.0
EQUIPMENT REQUIRED
4.0
DISINFECTION INSTRUCTIONS
5.0
REFERENCE INFORMATION
6.0
EXAMPLES OF EACH FAMILY GROUP
i
Table of Contents
Page
ii
1.0
IMPORTANT INFORMATION
1.1 Introduction The purpose of this guide is to provide detailed, recommended practices for the disinfection instructions of CONMED manual reusable surgical instruments, battery operated equipment and associated accessories. All necessary pre-cleaning/cleaning steps for the instrument are expected to be completed according to the instruction for use (IFU) for that device before the use of this manual. This guide provides two methods of disinfecting CONMED re-usable instruments, a manual method and a method using an automatic washer disinfector. The two methods have been validated and the processes provided in the instructions have been proven to be effective for disinfection in accordance with the ISO 17664 standard for Sterilization of Medical Devices: Information To Be Provided By The Manufacturer For The Processing Of Resterilizable Medical Devices. Whichever method is used, staff should use suitable protective clothing and equipment at all times. CONMED, recommends using the automated disinfection method whenever possible. The automated method of disinfecting the devices provides staff with less exposure to contaminated devices and the disinfection agents used. This guide is intended to assist health care personnel in the proper and effective handling practices during disinfecting, and maintaining the equipment. This guide should be studied carefully. The user / processor should comply with local laws and ordinances in countries where reprocessing requirements are more stringent than those detailed in this guide. This guide is to be a supplement to the device’s Instructions For Use and / or User Manual. This guide is intended to be used with the Instructions For Use (IFU) and / or User Manual supplied with the devices and / or on CONMED's website. In addition, refer to the IFU and / or User Manual technical specifications or cleaning / sterilization recommendations related to the product / system. Keep all instructions where they are easily accessible for the personnel handling the product during the disinfection process. The Instructions for Use, Manuals and Instrument Disinfection Guide can be found on CONMED’s website (www.conmed.com). If you have any questions or need assistance during a procedure, contact CONMED Customer Service at 1-866-426-6633 or your CONMED Sales Representative. Outside the U.S. contact your local CONMED Representative. 1
2.0
PRECAUTIONS, WARNINGS, and NOTES (Definitions)
Precautions, Warnings and Notes The words PRECAUTION, WARNING and NOTE carry special meanings and should be carefully reviewed. PRECAUTION: Methods or procedures intended to prevent or avoid adverse outcomes. WARNING: The personal safety of the patient, user and / or surgical staff may be involved. Disregarding this information could result in injury, death or other serious adverse reactions associated with the use or misuse of the device to the patient, user and / or surgical staff. NOTE: This provides special instructions to make maintenance instructions clearer.
2.1 Precautions After use (post-operatively) do not allow soiled / contaminated devices to dry prior to pre-cleaning and disinfecting. Consider covering the re-usable device with a cloth saturated with distilled water to avoid drying of soil. NOTE: If the disinfection process is likely to be delayed and the devices are allowed to dry with blood, body fluid, bone, and tissue debris, it is highly recommended to spray the devices with distilled water to loosen the dried soil before proceeding to the disinfection process. All subsequent cleaning and sterilization steps are facilitated by not allowing blood, body fluid, bone and tissue debris, saline, or disinfectants to dry on devices. Immersion of manual instruments in distilled water immediately after use will facilitate cleaning and prevent proteinaceous material from drying on the instruments. Saline solutions may have corrosive effects on some metals and should not be used.
2.2 Warnings CaviCide® Spray Ready-to-use or a comparable disinfection solution of pH between 11.0 and 12.49 are highly recommended by CONMED. However, the use of alternative methods of disinfection or disinfecting agents outside the scope of this document may be suitable for disinfecting the devices. Alternative disinfection methods must be validated by the end user. The use of a washer/sanitizer or disinfection solution may affect the life expectancy of the device. CONMED does not define the maximum number of uses or number of reprocessing for re-usable devices. The useful life of these devices depends on many factors, including the handling between uses, methods and duration of each use. It is recommended to inspect and test functionality of the devices before each use and to determine the end of serviceable life.
2
3.0
EQUIPMENT REQUIRED
Automated Disinfection using washer disinfector: • Suitable protective clothing • Washer disinfector large enough to allow complete immersion of the re-usable instruments. (Do not exceed the temperatures provided in this instrument disinfection guide.) • Deionized water for rinsing purposes. • Washer disinfector with a thermal disinfection program with an approved rinse cycle at 194°F (90°C) for a minimum of five (5) minutes duration. • Sterile lint free wipes Manual Disinfection: • Suitable protective clothing • Disinfecting area /platform • Sterile lint free wipes • CaviCide® Spray Ready-to-use or a comparable disinfection solution of pH between 11.0 and 12.49
3
4.0
DISINFECTION INSTRUCTIONS
Device Families
Manual Disinfecting Instructions
Automated Disinfecting Instructions
• Fully immerse the device in CaviCide® Readyto-use disinfection solution for a minimum of three (3) minutes at room temperature.
• Place device in an automated washer disinfector filled with purified water.
• Manipulate the grabber mechanisms and/or hinges a minimum of three (3) times (when applicable). • Ensure valves are in open position (parallel with the openings) for exposure (when applicable). Scopes / Cameras
• Rinse under running deionized water for a minimum of 1 minute. • Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry.
• Run thermal rinse cycle at 194°F (90°C) for five (5) minutes • Let device cool before handling. • Remove device from washer and thoroughly dry the device with a sterile lint free wipe. • Proceed to cleaning and sterilization instructions per device Instructions for Use (IFU).
• Proceed to cleaning and sterilization instructions per device Instructions For Use (IFU). • Wet a sterile, lint free wipe with CaviCide® Spray Ready-to-use disinfectant until is it fully saturated. • After wetting the wipe, the excess liquid should be squeezed out. NOTE: The wipe should remain wet but not dripping. • Using the wipe, thoroughly wipe the device. Careful attention should be given to seams and recessed areas. Continue to wipe the surface areas for a minimum of thirty (30) seconds. Sterilization Trays
• Ensure the device remains wet for a minimum of three (3) minutes by spraying the device as needed with the CaviCide® Spray Ready-touse. • Rinse under running deionized water for a minimum of one (1) minute. • Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry. • Proceed to cleaning and sterilization instructions per device Instructions For Use (IFU).
4
• Place device in an automated washer disinfector filled with purified water. • Run thermal rinse cycle at 194°F (90°C) for five (5) minutes. • Let device cool before handling. • Remove device from washer and thoroughly dry the device with a sterile lint free wipe. • Proceed to cleaning and sterilization instructions per device Instructions for Use (IFU).
Device Families
Manual Disinfecting Instructions
Automated Disinfecting Instructions
• Wet a sterile, lint free wipe with CaviCide® Spray Ready-to-use disinfectant until is it fully saturated.
• Place device in an automated washer disinfector filled with purified water.
• After wetting the wipe, the excess liquid should be squeezed out. NOTE: The wipe should remain wet but not dripping. • Using the wipe, thoroughly wipe the device. Careful attention should be given to seams and recessed areas. Continue to wipe the surface areas for a minimum of thirty (30) seconds. Batteries
• Ensure the device remains wet for a minimum of three (3) minutes by spraying the device as needed with the CaviCide® Spray Ready-touse.
• Run thermal rinse cycle at 194°F (90°C) for five (5) minutes. • Let device cool before handling. • Remove device from washer and thoroughly dry the device with a sterile lint free wipe. • Proceed to cleaning and sterilization instructions per device Instructions for Use (IFU).
• Rinse under running deionized water for a minimum of ten (10) seconds. • Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry. • Proceed to cleaning and sterilization instructions per device Instructions For Use (IFU).
General Manual Instruments
• Spray the device thoroughly with CaviCide® Spray Ready-to-use disinfectant.
• Place device in an automated washer disinfector filled with purified water.
• Ensure the device remains wet for a minimum of ten (10) minutes by spraying the device as needed with the CaviCide® Spray Ready-to-use.
• Run thermal rinse cycle at 194°F (90°C) for five (5) minutes.
• Ensure valves are in open position (parallel with the openings) for exposure (when applicable).
• Remove device from washer and thoroughly dry the device with a sterile lint free wipe.
• Manipulate the grabber mechanisms and/or hinges a minimum of ten (10) times (when applicable) • Rinse under running deionized water for a minimum of thirty (30) seconds. • Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry. • Proceed to cleaning and sterilization instructions per device Instructions For Use (IFU).
5
• Let device cool before handling.
• Proceed to cleaning and sterilization instructions per device Instructions for Use (IFU).
Device Families
Manual Disinfecting Instructions
Automated Disinfecting Instructions
• Fully immerse the device in CaviCide® Spray Ready-to-use disinfection solution for a minimum of three (3) minutes at room temperature.
• Place device in an automated washer disinfector filled with purified water.
• Ensure valves are in open position (parallel with the openings) for exposure (when applicable). Arthroscopic Procedure Devices
• Manipulate the grabber mechanisms and/or hinges a minimum of ten (10) times (when applicable) • Rinse under running deionized water for a minimum of thirty (30) seconds.
• Run thermal rinse cycle at 194°F (90°C) for five (5) minutes. • Let device cool before handling. • Remove device from washer and thoroughly dry the device with a sterile lint free wipe. • Proceed to cleaning and sterilization instructions per device Instructions for Use (IFU).
• Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry. • Proceed to cleaning and sterilization instructions per device Instructions For Use (IFU). • Place device in an automated washer disinfector filled with purified water. • Spray the device thoroughly with CaviCide® spray Ready-to-use disinfectant. • Ensure the device remains wet for a minimum of ten (10) minutes by spraying the device as needed with the CaviCide® Spray Ready-to-use. Handpieces
• Rinse under running deionized water for a minimum of one (1) minute. • Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry. • Proceed to cleaning and sterilization instructions per device Instructions For Use (IFU).
5.0 1. 2. 3. 4.
• Do not immerse the device paddle connector. When applicable, disconnect the paddle connector from the handpiece. NOTE: Proceed to the individual device Instructions For Use (IFU) for cleaning and/or sterilization for the paddle connector. • Run thermal rinse cycle at 194°F (90°C) for five (5) minutes. • Let device cool before handling. • Remove device from washer and thoroughly dry the device with a sterile lint free wipe. • Proceed to cleaning and sterilization instructions per device Instructions for Use (IFU).
REFERENCE INFORMATION
Disinfection validations were performed by NAMSA and Nelson Laboratories. ISO 17664: Sterilization of Medical Devices Information To Be Provided By The Manufacturer For The Processing Of Re-sterilizable Medical Devices. ISO 15883-1 Washer-disinfector-Part-1 General requirements terms, and definitions and tests. Requirements Concerning Hygiene During the Reprocessing of Medical Devices, Federal Health Gazette, 2012. 55:1244-1310. 6
6.0
EXAMPLES OF EACH FAMILY GROUP The tables below contain types of devices for each family group: General Manual Instruments Forceps
Awls
Cannulas
Graspers
Guides
Handles
Clamps
Brushes
Drill Guides
Arthroscopic Procedure Devices Taps
Reamers
Punches
Suture Hooks
Scissors
Bur Guards
Obturators
Stop Cocks
Tendon Stripper
Badgers
Endura Blades/Burs
Suture Passers
Scopes/Cameras Eyecup Scopes
Cartridge Scopes
Laparoscopes
Camera Heads
Endoscopes
Arthroscopes
Sterilization Trays Arthroscopy Trays
Laparoscope Trays
Sterilization Cases
Endoscope Trays
Insert Trays
Rigid Containers
Charging Cases
Baskets
Aluminum Trays
Radel Trays
Handpieces Autoclavable Remote Controls
Drill Handpieces
Attachments
Batteries Lithium Batteries
Transfer Batteries
7