ConMed
Linvatec Arthroscopic Instruments Instructions for Use
12 Pages
Preview
Page 1
Linvatec® Arthroscopic Instruments ENGLISH FRANÇAIS ITALIANO DEUTSCH ESPAÑOL ΕΛΛΗΝΙΚΑ PORTUGUÊS NEDERLANDS DANSK SVENSKA
Linvatec® Arthroscopic Instruments General NOTE: Instruments are provided non-sterile. Sterilize prior to use. Prior to use and sterilization, remove all protective packaging and tip protector, if applicable. Follow universal precautions for protective apparel when handling and cleaning contaminated instruments.
Cleaning 1. Open and/or disassemble instrument immediately after use. Place in a basin with distilled water. NOTE: Saline solution is not recommended as it has a corrosive effect on certain metals. 2. Clean instruments using a neutral pH cleaning solution and a non-abrasive brush until all traces of blood, debris, and stains are removed. Repeat cleaning if necessary. 3. Rinse instruments with distilled water. 4. Dry with a clean, lint-free cloth.
Steam Sterilization NOTES: 1. Guidelines provided below do not guarantee sterility. Your institution is responsible for the normal sterility assurance validation. 2. Sterilization validation is based on AAMI guidelines (Association for the Advancement of Medical Instrumentation). 3. For sterilization of instruments in a tray, consult guidelines provided with the specific sterilization tray. Prepare instruments such that all surfaces have direct contact with steam. Disassemble the instruments whenever possible. Method
Cycle
Temperature
Exposure Time
Steam (unwrapped) Steam (wrapped) Steam (wrapped)
Pre-Vacuum Pre-Vacuum Gravity
270°F (132°C) 270°F (132°C) 270°F (132°C)
3 minutes 4 minutes 10 minutes
INFORMATION For more information or a product demonstration, contact your regional ConMed Linvatec sales representative, or call 1-800-237-0169 in the U.S.
EC REP MDSS GmbH Schiffgraben 41 D-30175 Hannover, Germany
ConMed Corporation 525 French Road Utica, NY 13502-4908 USA Phone: (727) 392-6464 Customer Service: 1-800-237-0169 USA FAX: (727) 399-5256 International FAX: +1 (727) 397-4540 email: [email protected] www.conmed.com ©2012, 2013, 2016 ConMed Corporation. All rights reserved. W55-000-741
Rev AC